Hausted Horizon User manual

OPERATING MANUAL
Hausted® Horizon® Series
Multi-Purpose Stretchers – Electric Powered
Models: 4E1 & 4E2 (120 VAC), 4D2 (230 VAC)
(08/30/06) 069389
A WORD FROM STERIS CORPORATION
Indications for Use
This manual contains important information on proper use and maintenance of the Hausted® Horizon® Series Multi-Purpose Stretchers – Electric Powered. All personnel involved in the use and maintenance of this equipment mustAll personnel involved in the use and maintenance of this equipment must
All personnel involved in the use and maintenance of this equipment must
All personnel involved in the use and maintenance of this equipment mustAll personnel involved in the use and maintenance of this equipment must carefully review and comply with the warnings, cautions and instructionscarefully review and comply with the warnings, cautions and instructions
carefully review and comply with the warnings, cautions and instructions
carefully review and comply with the warnings, cautions and instructionscarefully review and comply with the warnings, cautions and instructions contained in this manual. contained in this manual.
contained in this manual. These instructions are important to protect the
contained in this manual. contained in this manual. health and safety of personnel operating a Hausted Horizon Series Electric Powered Multi-Purpose Stretcher and should be retained in a conveniently accessible area for quick reference.
Complete instructions for uncrating and connecting utilities, as well as equipment drawings, have been furnished. If missing, contact STERIS for replacement copies, giving the serial number and model numbers of the unit.
STERIS carries a complete line of accessories for use with these stretchers. A STERIS representative will gladly review these with you.
The Hausted Horizon Series Multi-Purpose Stretchers – Electric Powered are intended for use in patient treatment, transport or recovery.
Service Information
Advisory
A thorough preventive maintenance program is essential to safe and proper unit operation. This manual contains maintenance schedules and procedures which should be followed for satisfactory equipment performance.
You are encouraged to contact STERIS concerning our comprehensive Annual Maintenance Agreements. Under the terms of these agreements, preventive maintenance, adjustments, and replacement of worn parts are done on a scheduled basis to assure equipment performance at peak capability and to help avoid untimely or costly interruptions. STERIS maintains a global staff of well equipped, factory-trained technicians to provide this service, as well as expert repair services. Please contact your STERIS representative for details.
A listing of the safety precautions to be observed when operating and servicing this equipment can be found in operate or service the equipment until you have become familiar with this information.
Any alteration of this equipment not authorized or performed by STERIS Engineering Service which could affect its operation will void the warranty, could adversely affect sterilization efficacy, could violate national, state, and local regulations, and could jeopardize your insurance coverage.
S
ECTION
1
of this manual. Do not
©2006, STERIS Corporation. All rights reserved. Printed in U.S.A.
Table of Contents Operator Manual 069389
EC Authorized Representative
STERIS Limited STERIS House Jays Close Viables Basingstoke Hampshire RG22 4AX UNITED KINGDOM
Manufactured by: STERIS Corporation 2720 Gunter Park East Montgomery, AL 36109 • USA
www.steris.com
TEL: 334 277 6660 FAX: 334 271 5450
Class 1 Equipment Type B Equipment Equipment not suitable for use
in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide.
IPX4 (Splash-proof Equipment) Not suitable for continuous opera-
tion (Duty Cycle: 10% six minutes in one hour)
STERIS Corporation,
Montgomery, Alabama is an
ISO 13485 certified facility.
The base language of this document is
ENGLISH. Any translations must be
made from the base language docu-
ment.
ii
069389 Operator Manual Table of Contents

TABLE OF CONTENTS

Section Title Page
1 LISTING OF WARNINGS AND CAUTIONS ....................................... 1-1
2 UNPACKING INSTRUCTIONS ......................................................... 2-1
3 OPERATING INSTRUCTIONS .......................................................... 3-1
3.1 Stretcher Specifications ............................................................................................... 3-1
3.2 Features, Warnings and Proper Operation ................................................................... 3-3
3.3 Braking and Steering Operation ................................................................................... 3-5
3.3.1 Applying the Brakes ............................................................................................. 3-5
3.3.2 Releasing the Brakes............................................................................................ 3-5
3.3.3 Applying the Steering Lock/Fifth Wheel ................................................................ 3-5
3.3.4 Releasing the Steering Lock/Fifth Wheel .............................................................. 3-5
3.4 Litter Top Height Adjustment (4E1 Models Only) ......................................................... 3-6
3.4.1 Height Adjustment ................................................................................................ 3-6
3.4.2 Lowering Litter Top ................................................................................................ 3-6
3.4.3 Trendelenburg Adjustment .................................................................................... 3-6
3.4.4 Reverse Trendelenburg Adjustment ...................................................................... 3-7
3.5 Electric Control Locations ............................................................................................. 3-7
3.5.1 Nurses Control Panel ............................................................................................ 3-7
3.5.2 Patient Pendant ..................................................................................................... 3-7
3.5.3 Plug Locations ...................................................................................................... 3-7
3.5.4 Emergency Electric Backrest Overide .................................................................. 3-8
3.5.5 Low Battery Light (L.E.D.) ..................................................................................... 3-8
3.6 Litter Top Adjustment ................................................................................................... 3-8
3.6.1 Height Adjustment (4E2 & 4D2 Models Only) ....................................................... 3-8

3.6.2 Fowler Adjustment ................................................................................................ 3-9

3.6.3 Quick Drop Activation (Fowler Backrest) .............................................................. 3-9
3.6.4 Knee Flex Adjustment .......................................................................................... 3-9
3.6.5 Trendelenburg Adjustment (4E2 & 4D2 Models only) ........................................... 3-10
3.7 Airglide Rail Operation ................................................................................................3-10
3.7.1. Raising the Rail ................................................................................................... 3-10
3.7.2 Lowering the Rail.................................................................................................. 3-10
3.8 Retracto Rail Operation ............................................................................................... 3-11
3.8.1 Raising the Rail .................................................................................................... 3-11
3.8.2 Half Height ........................................................................................................... 3-11
3.8.3 Lowering the Rail.................................................................................................. 3-11

3.9 Push Handle Operation (Optional Accessory) .............................................................3-12

3.9.1 Raising the Push Handles .................................................................................... 3-12
3.9.2 Lowering the Push Handles ................................................................................. 3-12
3.10 Permanently Mounted IV Rod Operation ................................................................... 3-12
3.10.1 Putting I.V. Rod in Up Position ........................................................................... 3-12
3.10.2 Extending I.V. Rod ............................................................................................. 3-13
3.10.3 Retracting I.V. Rod .............................................................................................3-13
4 TROUBLE SHOOTING GUIDE .......................................................... 4-1
4.1 Electric Powered Stretchers ......................................................................................... 4-1
4.2 Battery Replacement Instructions ................................................................................ 4-2
Table of Contents Operator Manual 069389
iii
Section Title Page
5 RECOMMENDED PREVENTIVE MAINTENANCE............................ 5-1
5.1 Recommended Cleaning Instructions .......................................................................... 5-1
6 OPTIONAL ACCESSORIES.............................................................. 6-1
7 RECOMMENDED REPLACEMENT PARTS ...................................... 7-1
iv
069389 Operator Manual Table of Contents

LISTING OF WARNINGS AND CAUTIONS

1
The following Stretchers. WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain
Safety Precautions
Strictly following these customer avoid improper maintenance methods which may damage the unit or render it unsafe. It is importantto understand that these policies and procedures to enhance and compliment these
Safety Precautions
before operating or servicing the unit.
Safety Precautions
Safety Precautions
must be observed when operating or servicing this Hausted® Horizon® Multi-Purpose
Safety Precautions
enhances your ability to safely and effectively utilize the unit and helpsthe
are not exhaustive; customers are encouraged to develop their ownsafety
are repeated throughout the manual. It is important toreview ALL
Safety Precautions.
WARNING–LACERATION HAZARD:
When cutting bands always use a tool specifically designed for that purpose. This will help to avoid personal injuries frequently incurred when bands are cut and tension released.
WARNINGS – PERSONAL INJURY:
Do not sit on end – tipping may occur.
Be sure I.V. rod is inserted completely into socket up to the arrow before applying load.
To reduce the risk of electric shock, do not remove the cover. Unit is to be service by qualified personnel only.
See
S
ECTION
3
for the Warnings below:
WARNINGS – CAUTIONS AND PROPER OPERATION:
The stretchers have a warning label located at the head end stating: Maximum patient weight 226 Kilograms (500 lbs.) This is true for all models except the 4E2DPA or 4D2DPA which states: Maximum patient weight 283 Kilograms (625 lbs.).
Patient entry, egress and transfer should always be done with the brakes locked.
The brakes should always be on and patient side rails up when the patient is not in transport.
The patient pendant has a warning label on it stating: Clip Pendant to rail when not in use - Keep cord clear of moving parts.
It is important to remember that all electric stretchers are equipped with a built in battery back- up system but: it is recommended that the unit remain plugged in wall receptacle during normal use. The battery backup is recommended for transport and emergency only.
The Fowler backrest quick drop handle is intended to be used during emergency situations only.
Be sure rail is locked before leaving patient.
The lock out controls on the nurse control box controls both the pendant and the control panel.
To turn electric controls on, plug into wall receptacle. To turn off, remove plug from wall receptacle. The electric powered stretcher do no have a separate on/off switch.
Continued. . . .
Listing of Warnings and Cautions Operator Manual 069389
1-1
The stretcher has a warning label located on the control box cover stating: To reduce the risk of electrical shock do not remove the cover. Service by qualified personnel only.
Always disconnect the power source whenever servicing any electric powered stretcher.
Do not use a sharp instrument to remove fuse, as it may scratch the circuit board.
The batteries are wired in series, failure to install or rewire the same way may cause the batteries to explode.
WARNING – CAUTIONS AND PROPER OPERATION (Continued):
Floors must be smooth and level to maintain optimum fifth wheel steering. Fifth wheel steering functions can be influenced by floor irregularities (bumps or dips) greater than 1/2" (13 mm) across the span of the stretcher.
1-2
069389 Operator Manual Listing of Warnings and Cautions

UNPACKING INSTRUCTIONS

IMPORTANT: Report any shipping damage immediately! Inform shipper of
any damages, and leave carton intact. Leave the stretcher in receiving area until inspection is complete.
Your Hausted® Horizon® Multi-Purpose Stretcher has been carefully packaged at our manufacturing plant to ensure safe shipment to you medical facility. There are several procedures you must follow to put your new stretcher into service. These procedures only take a few minutes to complete and are required to ensure proper operation of the stretcher.
2
WARNING – LACERATION HAZARD: When cutting bands always use a tool spe­cifically designed for that purpose. This will help to avoid personal injuries fre­quently incurred when bands are cut and tension released.
Step 1 – Cut the two bands around the shipping carton. Step 2 – Remove the top half of the carton and cut one side of the bottom half. Step 3 – Remove the stretcher from the carton. Cut the bands from around the
stretcher.
Step 4 – Check to see if all features of the stretcher work properly.
NOTE: You must first plug the unit into a wall socket to check any electric features, as the battery has not been charged (see Step 6).
If all features work properly, go to step 5. If any of the features of the stretcher do not work properly, call STERIS for service.
Step 5 – Clean the stretcher using a mild detergent to remove any dirt
accumulated during shipment.
Step 6 – Plug the unit into a wall socket to charge for approximately 8 hours
to charge the battery backup.
Step 7 – Place the unit into service.
2-1
Unpacking Instructions Operator Manual 069389

OPERATING INSTRUCTIONS

3.1 Stretcher
Specifications
WARNING –PERSONAL INJURY: The stretchers have a warning label located at the head end stating: maximum patient weight 226 Kilograms (500 lbs.). This is true for all models except for the 4E2DPA or 4D2DPA which states: Maxi­mum patient weight 283 Kilograms (625 lbs.).
NOTE: All dimensions are specified
in inches. All dimension are ±.375 inches.
3
STERIS reserves the right to change specifications without notice.
Stretcher weight: 315 lbs.
KNEE FLEX
Electrical Specifications
(Applies only to 4E1 & 4E2 Models)
Product Classifiation: 1 Input Voltage: 120V ~, +/-5%, 60Hz Amperage: Max. 3.5A Duty cycle: 10% 6 Min in 1 Hr. IP Rating: IPX 4 Grounding Protection: Type B
KNEE FLEX
90° MAx
DPA Models Specifications Shown
Electrical Specifications
(Applies only to 4D2 Model)
Product Classifiation: 1 Input Voltage: 230V ~, +/-10%, 50Hz Amperage: Max. 1.9A Duty cycle: 10% 6 Min in 1 Hr. IP Rating: IPX 4 Grounding Protection: Type B
3-1
Operating Instructions Operator Manual 069389
WARNING–PERSONAL INJURY: The stretchers have a warning label located at the head end stating: maximum patient weight 226 Kilograms (500 lbs.). This is true for all models except for the 4E2DPA or 4D2DPA which state: Maxi­mum patient weight 283 Kilograms (625 lbs.).
Model 4E1HMC (hydraulic base) maximum Patient weight: 500 lbs.
Model 4E2HMC (electric base) maximum Patient weight: 500 lbs.
NOTE: All dimensions are specified in inches.
All dimension are ±.375 inches. STERIS reserves the right to change
specifications without notice.
Stretcher weight: 315 lbs.
KNEE FLEX
Electrical Specifications
(Applies only to 4E1 & 4E2 Models)
Product Classifiation: 1 Input Voltage: 120V ~, +/-5%, 60Hz Amperage: Max. 3.5A Duty cycle: 10% 6 Min in 1 Hr. IP Rating: IPX 4 Grounding Protection: Type B
KNEE FLEX
90° MAx
HMC Models Specifications Shown
Electrical Specifications
(Applies only to 4D2 Model)
Product Classifiation: 1 Input Voltage: 230V ~, +/-10%, 50Hz Amperage: Max. 1.9A Duty cycle: 10% 6 Min in 1 Hr. IP Rating: IPX 4 Grounding Protection: Type B
3-2
069389 Operator Manual Operating Instructions
3.2 Features,
Warnings and Proper
Operation
WARNINGS –CAUTIONS AND PROPER OPERATION:
A– The stretchers have a warning label located at
the head end stating: maximum patient weight 226 Kilograms (500 lbs.). This is true for all models except the 4E2DPAor 4D2DPA which states: Maximum patient weight 283 Kilograms (625 lbs.).
B– Patient entery, egress and transfer should
always be done with the brakes locked.
C– The brakes should always be on and patient
side rails up when the patient is not in trasnport.
D– The patient pendant has a warning label on it
stating: Clip Pendant to rail when not in use – Keep cord clear of moving parts.
E –It is important to remember that all electric
stretchers are equipped with a built in battery backup system but: it is recommended that the unit remain plugged in wall receptacle during normal use. The battery backup is recommended for transport and emergency only.
I– Floors must be smooth and level to maintain
optimum fifth wheel steering. Fifth wheel steer­ing functions can be influenced by floor ir­regularities (bumps or dips) greater than 1/2" (13 mm) across the span of the stretcher.
WARNINGS RELATED TO SERVICE & REPAIR:
J– To turn electric controls on, plug into wall
receptacle, to turn off, remove plug from wall receptacle, the electric powered stretcher do no have a separate on/off switch.
K– The stretcher has a warning label located on
the control box cover stating: To reduce the risk of electrical shock don not remove the cover. Service by qualified personnel only.
L– Always disconnect the power source when-
ever servicing any electric powered stretcher.
M–Do not use a sharp instrument to remove fuse,
as it may scratch the circuit board.
N– The batteries are wired in series, failure to
install or rewire the same way may cause the batteries to explode.
F– The Fowler backrest quick drop handle is
intended to be used during emergency situa-
tions only. G–Be sure rail is locked before leaving patient. H– The lock out controls on the nurse control box
controls both the pendant and the control
panel.
Important Note: The picture shown is of the “DPA” models, all cautions and warnings also apply to the “HMC” models.
3-3
Operating Instructions Operator Manual 069389
See Warnings C & G
See Warning H
See Warnings E & J
See Warnings D & H
See Warnings J thru N
See Warnings B & C
See Warning F
3-4
069389 Operator Manual Operating Instructions
3.3 Braking and
Steering Operation

3.3.1 Applying the Brakes

Figure 3-1.

3.3.2 Releasing the Brakes

Figure 3-2.
The four wheel central braking system is activated by depressing the red pedal at any of the four corners of the unit (Figure 3-1). To fully engage the brakes, the pedal should be pressed to approximately 45º. All four caster wheels should be locked from swiveling and rotating.
Depress the green pedal at any of the four corners of the unit, until the pedal is in a horizontal position (Figure 3-2). All four wheels should rotate and swivel freely.
NOTE: Brakes should always be activated when unit is not in transport.
3.3.3 Applying the
Steering Lock/Fifth Wheel
WARNING – PROPER OP­ERATION: Floors must be smooth and level to main­tain optimum fifth wheel steering. Fifth wheel steer­ing functions can be influ­enced by floor irregularities (bumps or dips) greater than 1/2" (13 mm) across the span of the stretcher.
3.3.4 Releasing the
Steering Lock/Fifth Wheel
From any corner of the stretcher, depress the green pedal downward into locked position (Figure 3-3). Push the stretcher forward. Either one caster at the foot end will lock into a non-swivel mode, or the optional fifth wheel will lower and apply pressure to the floor. This allows straight steering by the attendant.
Figure 3-3.
Depress the red pedal at any of the four corners of the unit, until the pedal is in a horizontal position (Figure 3-2). All four wheels should rotate and swivel freely, and/or the optional fifth wheel will retract.
3-5
Operating Instructions Operator Manual 069389
3.4 Litter Top Height Adjustment
(4E1 Models Only)

3.4.1 Height Adjustment

Figure 3-4

3.4.2 Lowering Litter Top

Press the pump pedal to the floor (Figure 3-4), then release. Repeat this process until desired height is obtained. Use smooth strokes on the pedal to ensure patient comfort.
Press down on the two release pedals at the same time (Figure 3-5) until desired height is obtained.
NOTE: Do not stand on release pedals.
Figure 3-5

3.4.3 Trendelenburg Adjustment

Place the unit at maximum height, see release pedal nearest the head end (Figure 3-6) until desired position is obtained. Then remove pressure.
Height Adjustment
. Press down on the
Figure 3-6
3-6
069389 Operator Manual Operating Instructions
3.4.4 Reverse
Trendelenburg Adjustment
Figure 3-7

3.5 Electric Control Locations

Place the unit at maximum height, see release pedal nearest the foot end (Figure 3-7) until desired position is obtained. Then remove pressure.
Height Adjustment
. Press down on the

3.5.1 Nurses Control Panel

Figure 3-8.

3.5.2 Patient Pendant

CAUTION: Clip pendant to rail when not in use. Keep cord clear of moving parts.
Figure 3-9.
To access the nurses control panel, pull out on the handle located at the foot end of the unit, under the top (Figure 3-8).
The patient Pendant has an adjustable location and may be located on either side of the rail (Figure 3-9).

3.5.3 Plug Locations The hospital grade plug is located under the head end of top section

(Figure 3-10). Unwrap cord from bracket and plug into the nearest wall receptacle.
Important: It is recommended that unit remain plugged in wall receptacle during normal use. Battery back-up is recommended for transport and emergency only.
Figure 3-10.
3-7
Operating Instructions Operator Manual 069389

3.5.4 Emergency Electric Backrest Override

The unit is equipped with a Fowler backrest quick drop handle located under the litter top at stretcher mid section. (See the stretcher specification page). This option is only to be used in an emergency situation. The label states:
Drop Backrest Pull Handle to Activate
.
Emergency

3.5.5 Low Battery Light (L.E.D.)

Figure 3-11.

3.6 Litter Top Adjustment

CAUTION: The lock-out con­trols both the pendant and
the control panel.
3.6.1 Height Adjustment
(4E2 & 4D2 Models Only)
The stretchers are equipped with a low battery indictor light, located on the controller cover under the head end of the stretcher (Figure 3-11). When a button on the pendant is pressed, and the L.E.D. light comes on, it indicates that the backup batteries require charging. Plug the unit into a wall receptacle.
Lockout – control panel, clockwise approximately 90º (Figure 3-12). under the third row of buttons, counter-clockwise approximately 90º.
Unlocking
: Turn the knob located under the third row of buttons on the
Locking
: Turn the knob
Figure 3-12.
NOTE: This lockout knob also controls the trendelenburg functions.
Nurse: Press either of the up or down arrows located on the third row of buttons on the control panel (Figure 3-13). Hold in the button until the desired height is achieved.
Patient: Press either of the up or down arrows located on the third row of buttons on the patient pendant (Figure 3-14). Hold the button in until desired height is achieved.
Figure 3-13.
Figure 3-14.
3-8
069389 Operator Manual Operating Instructions

3.6.2 Fowler Adjustment

Lockout – control panel, clockwise approximately 90º (Figure 3-15). under the first row of buttons, counter-clockwise approximately 90º.
Nurse: Press either of the up or down arrows located on the first row of buttons on the control panel (Figure 3-16). Hold in the button until desired incline is achieved.
NOTE: This unit is equipped with a Fowler backrest quick drop handle located on either side of the stretcher near the foot section (see stretcher specification page). This option is only to be used during an emergency situation.
Patient: Press either of the up or down arrows located on the first row of buttons on the patient pendant (Figure 3-17). Hold in the button until desired incline is achieved.
Unlocking
: Turn the knob located under the first row of buttons on the
Locking
: Turn the knob
Figure 3-15. Figure 3-16.
3.6.3 Quick Drop
Activation (Fowler
Backrest)
The stretcher is equipped with manual override function for the Fowler backrest. This option should only be used in an emergency situation. To activate the quick drop, pull out on the red “T” handle (Figure 3-18) located under the litter top on either side of the patient, near the middle of the unit.
Figure 3-18.

3.6.4 Knee Flex Adjustment

Lockout – control panel, clockwise approximately 90º (Figure 3-19). Turn the knob under the second row of buttons, counter-clockwise approxi­mately 90º.
Nurse: Press either of the up or down arrows located on the second row of buttons on the control panel (Figure 3-20). Hold in the button until desired height is achieved.
Patient: Press either of the up or down arrows located on the first row of buttons on the patient pendant (Figure 3-21). Hold in the button until desired height is achieved.
Unlocking
: Turn the knob located under the second row of buttons on
Figure 3-17.
Locking
:
Figure 3-19. Figure 3-20. Figure 3-21.
3-9
Operating Instructions Operator Manual 069389
3.6.5 Trendelenburg
Adjustment (4E2 & 4D2
Models only)
Lockout: See
Nurse: Trendelenburg: Press the top button on the fourth row of buttons on the control panel (Figure 3-22). Hold in the button until desired angle is achieved.
Rev-Trendelenburg: Press the bottom button on the fourth row of buttons on the control panel (Figure 3-23). Hold in the button until the desired angle is achieved.
Patient: There is no Trendelenburg functions on the pendant.
Lockout
in the Section 3.6.1
Height Adjustment Instructions
.
Figure 3-22.

3.7 Airglide Rail Operation

WARNING–PERSONAL INJURY HAZARD: Be sure rail is locked before leaving patient.

3.7.1. Raising the Rail

Figure 3-24.
Figure 3-23.
Grasp rail top tube from under litter top firmly, then pull up (Figure 3-24) until rail locks into full up position.

3.7.2 Lowering the Rail

Firmly grasp top of rail, pull out on red tab, located under litter top at the foot or head end (Figure 3-25). Lower side rail completely under litter top.
Figure 3-25.
3-10
069389 Operator Manual Operating Instructions

3.8 Retracto Rail Operation

WARNING –PERSONAL INJURY HAZARD: Be sure rail is locked before leaving patient.

3.8.1 Raising the Rail Grasp the rail top cap in the middle of the rail (Figure 3-27) and lift.

Figure 3-27.

3.8.2 Half Height Grasp the rail, lift up on the red trigger under the liter top (Figure 3-28) while

lowering the rail. When rail starts to move down release trigger. Lower rail until it locks into half height position.
Figure 3-28.

3.8.3 Lowering the Rail Grasp the rail, lift up on the red trigger under the top (Figure 3-28). While lowering

the rail, hold up on the trigger until the rail is all the way down.
3-11
Operating Instructions Operator Manual 069389
3.9 Push Handle
Operation (Optional
Accessory)

3.9.1 Raising the Push Handles

Figure 3-29.

3.9.2 Lowering the Push Handles

3.10 Permanently Mounted IV Rod
Operation
Hinge the push handle up (Figure 3-30) until it stops. Slide the push handle down into the socket until it stops. Repeat process for handle on the other side. (Figure 3-31).
Figure 3-30.
Lift up on push handle and rotate (Figure 3-30) until it is in the down position (Figure 3-29). Repeat process for the other side.
Figure 3-31.
WARNING–PERSONAL INJURY HAZARD: Be sure I.V. rod is inserted completely into socket up to the arrow before applying and load.

3.10.1 Putting I.V. Rod in Up Position

Figure 3-32.
Grasp I.V. rod and hinge upward until it stops (Figure 3-32). Push down on I.V. rod until it slides firmly into the socket.
3-12
069389 Operator Manual Operating Instructions

3.10.2 Extending I.V. Rod Lift up on top of I.V. rod (Figure 3-33) until desired length is achieved.

Figure 3-33.

3.10.3 Retracting I.V. Rod

Figure 3-34.
Press button located in the gray housing (Figure 3-34) and lower I.V. Rod until desired height is achieved. Repeat process with second housing as required.
3-13
Operating Instructions Operator Manual 069389

TROUBLE SHOOTING GUIDE

4.1 Electric Powered Stretchers

WARNING: To reduce the risk of electric shock, do not remove the cover. Unit is to be serviced by qualified per­sonnel only.
If Then
4
One motor or one column does not move but all others are operating correctly.
Nothing moves
The unit runs when plugged into wall receptacle, but does not run on the battery backup.
Step1: Check all motor and column plug connections at the controller box.
Step 2: If a column does not move: Check the the connection at the column.
Step 3: Plug a connector from the faulty component into a different socket:
If the component does not run:
– Replace that component.
If the component does run:
– Test pendant by plugging a functioning component into the non-
functioning socket on the controller. If this component does not run, replace the pendant. If replacing the pendant does not fix the problem, then the controller must be replaced.
Step 1: Plug unit into a mains supply wall receptacle, observe the pilot light
on the controller:
If the pilot light is off:
- Replace the controller
If the pilot light is on:
1. Check the nurse control plug connection at the controller.
2. Check the pendant control plug connection at the controller.
Step 1: Plug unit into a wall receptacle overnight. If the batteries do not hold
a charge, replace the batteries.
4-1
Troubleshooting Guide Operator Manual 069389
4.2 Battery
Replacement
Instructions
WARNING– SHOCK HAZ­ARD: To reduce the risk of electric shock, do not remove the cover. Unit is to be ser­viced by qualified personnel only.
WARNING –PERSONAL IN­JURY HAZARD: The batter­ies are wired in series, fail­ure to connect the same way can cause the batteries to explode.
The batteries are the only field serviceable components - do not attempt to repair the circuit boards.
Battery replacement Step 1: Remove the controller.
Step 2: Remove the 4 screws from the right half, located on top of the
controller.
Step 3: Replace (both) batteries with ‘YUSHA’ #NP 1.2-12, 12 volt, 1.2
Ah batteries. (Steris P/N 075759 – 2 required.
Step 4: Be sure battery connector in in place on left side of batteries.
Step 5: Replace cover.
4-2
069389 Operator Manual Troubleshooting Guide
8
8
8
8
8
8
8
8
8

RECOMMENDED PREVENTIVE MAINTENANCE

5
Procedure
Lubricate all moving and sliding parts and hinge points.
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
23456789012345678901234567890121234567890123456789012345678901212345678901234567890123456789012123456789012345678901234567890121234567
>!< Never Lubricate Gas Spring, Motor and Mech-Lock Shafts >!<
position and tightness (including nuts,
Schedule
3 Months
3 MonthsInspect all fasteners to ensure proper fit,
Lubricating oil, light-duty grease, wax stick lubricant.
Proper size wrench and screw­driver.
Material
!!
bolts, etc.).
3 MonthsInspect all surfaces and remove any
user developed sharp or burred areas.
Metal File, proper color paint (specify color when ordering).
Apply touch-up where required.
5.1 Recommended Cleaning
Instructions
NOTE: Steam cleaning and pressure washing of stretcher is not recommended and can void warranty.
Component
Pads
Electric Components
Mechanical Stretcher Components
Mechanical Accessories
Cleaning Procedure
Wipe with Damp Cloth To remove any foreign materials
Wipe with damp cloth to remove any foreign materials
Wipe with damp cloth to remove any foreign materials
Wipe with damp cloth to remove any foreign materials
Schedule
After Each Use
After each use
After each use
After each use
For more details information, please contact:
Cleaning Agent
Routine Hospital Grade Disinfectants, Soap and Water
Routine hospital grade disinfectants, soap and water
Routine hospital grade disinfectants, soap and water
Routine hospital grade disinfectants, soap and water
STERIS Corporation at:
1-800-333-8828
Special Notes
• Use only medium strength cleaners
• Do not steam clean or pressure wash
• Use only medium strength cleaners
• Do not steam clean or pressure wash
• Be sure any wet surface on electic components is dry before pulgging into wall receptacle
• Lubricate pivot points after cleaning
• Lubricate pivot points after cleaning
5-1
Recommended Preventive Maintenance Operator Manual 069389

OPTIONAL ACCESSORIES

Universal
000E17-00 Stainless steel 42" long fixed height I.V. rod - removable 000018-00 Stainless steel 27" to 54" long telescoping I. V. rod 128770-00 000018 - telescoping I.V. rod with cable attachment 126740-00 Infusion pump I.V. rod 1 1/8" dia - 42" long - removable IVSTOW-00 Mobile I.V. stand attachment
Airglide Rail (D) Style Tops
066AGR-00 Mounted stainless telescoping I.V. rod - patient right IVSTOW-00 Mobile I.V. stand attachment
Retracto Rail (H) Style Tops
066RR0-00 Mounted stainless telescoping I.V. rod - head end 065897-00 Mobile I.V. stand attachment
Airglide Rail (D) Style Tops
068965-60 Pressurecare - 4" x 32" x 77" conductive foam 069383-60 Enhanced pressurecare - 5" x 32" x 77" conductive foam 069230-00 Slick patient transfer mattress system
Retracto Rail (H) Style Tops
031661-60 3" x 29 1/2 “ x 76” conductive foam 061894-60 Pressurecare - 4" x 29 1/2" x 76" cond. foam - E & G tops 068922-60 Enhanced pressurecare - 5" x 29 1/2" x 76" -E & G tops 069206-00 Slick patient transfer mattress system
6
I.V. Poles
Pads and Matresses
Head and Foot Boards
Airglide Rail (D) Style Tops
000W6B-00 Head and foot board - stationary 0CRW6B-00 Head and foot board - stationary with chart rack 000W7B-00 Extension head and foot board 0CRW7B-00 Extension head and foot board - with chart rack
X-ray Accessories
Retracto Rail (H) Style Tops
000054-00 Lateral x-ray cassette holder- holds up to 14" x 17"
Monitor Shelves
Airglide Rail (D) Style Tops
069299-00 Extension footboard/monitor shelf w/chart rack
Retracto Rail (H) Style tops
00WN45-00 Monitor shelf - removable - folding 068846-00 Extension footboard/monitor shelf w/chart rack
6-1
Optional Accessories Operator Manual 069389
Miscellaneous
Universal
000012-00 Restraint straps with buckle 000014-00 Restraint straps with velcro 128450-00 Vertical oxygen tank holder 065366-00 Universal patient tray 069257-00 Paper roll holder
Airglide Rail (D) Style Tops
075665-00 Self storing push handles 065180-00 Armboard - height adjustable - 1" pad included 00C16H-00 Heel stirrups 00C16L-00 Knee crutch stirrups 00L16M-00 Mounting adaptor clamps for heel stirrups/leg holders 067741-00 Side rail pads 068577-00 Burgundy pin striping for departmental identification 068578-00 Teal pin striping for departmental identification 068585-00 Plum pin striping for departmental identification
Retracto Rail (H) Style Tops
075656-00 Self storing push handles 133171-00 Under fowler utility tray 076334-00 Surgical armboard – pad & adapter included 00N16A-00 Heel stirrups 00N16C-00 Knee crutch stirrups 00N16E-00 Heel stirrups and leg holders with mounting adapter clamps 075626-00 Side rail pads 128840-00 Drainage bag hooks 131934-00 Burgundy bumpers and pin striping for department identification 131935-00 Teal bumpers and pin striping for department identification 131936-00 Plum bumper and pin striping for department identification
Fifth Wheels
4E2 Base (Electric)
075798-00 Fifth wheel, electric horizon auto track
4E1 (Hydraulic)
075797-00 Fifth wheel, horizon auto track
It is recommended that only STERIS approved accessories be used with this device.
To order accessories or for more detailed information on accessories please contact STERIS At: 1-800-428-7833
6-2
069389 Operator Manual Optional Accessories
Loading...
+ 60 hidden pages