Harris e-Core, e-Core 12, e-Core 20, e-Core 30 User Manual

User Guide 2009
For Laboratory Use Only
TM
“Cuts, retrieves and stores cored samples in one operation”
For Your Safety Read Before Operating
Read the following safety precautions in their entirety before using this equipment. Failure to observe these precautions will void the warranty. Keep these safety instructions where all those who use the product will read them.
WARNINGS - READ
Read all instructions before operating the Harris e-Core
• The Harris e-Core is to be used only in a laboratory by trained professionals.
• The Harris e-Core is not a medical device, do not use to sample living tissue.
• Do not operate the Harris e-Core near sinks, running water, spills, or wet surfaces.
• Do not immerse the Harris e-Core in water.
• Unplug the Harris e-Core in the event of malfunction.
Observe the following precautions when handling the power supply
• Do not place the Harris e-Core plug or cord near or in water.
• Do not put strain on the cord. Ensure the Harris e-Core is operated with adequate extra cord length. The use of grounded extension cords is acceptable.
• Keep dry. Failure to observe this precaution could result in re or electric shock.
• Do not handle the plug with wet hands. Failure to observe this precaution could result in electric shock.
• Do not modify the plug in any way.
• Dust on or near metal parts of the plug should be removed with a dry cloth. Continued use could result in re. Unplug the device before dusting.
• Unplug the power supply during thunderstorms. Do not handle the power cable or go near the adapter during thunderstorms. Failure to observe this precaution could result in electric shock.
• Do not damage, modify, forcibly tug or bend the power cable. Do not place the power cable under heavy objects, or expose it to heat or ame. Should the insulation be damaged and the wires become exposed, contact your distributor. Failure to observe these precautions could result in re or electric shock.
• Ensure that the optional plug sets are completely seated in the main plug base.
• Use only plug adapters designed for use with this product.
• Do not allow the cord to hang over the side of a bench or other working surface.
• Do not operate the Harris e-Core with a damaged cord or plug.
• Use care when changing plugs on power supply. Removing a plug from power supply may require additional effort. To assist this removal it is suggested that you cover the prongs on the
plug to be removed with a cloth or use gloves when prying plug from power supply sleeve.
(See Section 6.0)g
i
Do not operate the Harris e-Core with a damaged cord or plug or if the Harris e-Core malfunctions, or is dropped or damaged in any manner. Turn off immediately and contact the distributor for assistance.
• Should you notice smoke or an unusual smell coming from the Harris e-Core, unplug the power supply immediately. Continued operation could result in injury. Contact the distributor.
• Always unplug the Harris e-Core before replacing with dedicated Harris Micro-Punch Tips, when
it is not in use and before cleaning.
• Keep free hand away from coring zone.o
Unplug the unit to change coring tips (See Section 5.0). Make sure the tip
cover cap is on at all times when the Harris e-Core is not in use.
Do not disassemble the Harris e-Core.
• Touching the internal parts of the Harris e-Core or power supply could result in injury.
If repairs or servicing is required contact the distributor.
• Never keep the Harris e-Core with the Harris Micro-Punch Tip rotating in contact with the sub
strate for more than 3 seconds as it can cause permanent damage to the motor. If the motor
stops working, unplug the power supply and let the motor cool for a few hours before attempting
to use it again. If the motor continues to be inoperable then contact the distributor for assistance.
• To reduce the risk of injury, never activate the coring tip unless the Harris e-Core is to be used to
core a sample from a substrate.
• Always use the Harris e-Core on a clean, at, dry Harris Cutting Mat.
• The use of Harris Micro-Punch Tips not designed for the respective model of Harris e-Core or the use of attachments not recommended by the manufacturer may damage the Harris e-Core,
may cause injury to the user and will void the warranty.
• Be certain that the Harris Micro-Punch Tip is completely and rmly attached onto the Harris e-Core drive shaft thread.
• Do not depress Harris e-Core ejection rod/plunger when preparing to core a sample. The Harris e-Core is not a punch and the plunger is not a punching mechanism.
ii
NOTICES
Required Guidelines For Use of the Harris e-Core and Related Products
The information presented here is accurate and reliable to the best of our knowledge and belief, but is not guaranteed to be so. The manufacturer recommends that the buyer and other persons using the Harris e-Core follow the guidelines in this instruction booklet. Nothing herein is to be construed as recommending any practice or in violation of any law or regulation. It is the user’s responsibility to determine for himself or herself the suitability of any materials and/or procedure for a specic purpose and to adopt such safety precautions as may be necessary.
Warranty
Shunderson Communication Inc. (SCInc.) products are designed for R&D and general laboratory use. In using these products, the customer agrees that SCInc. shall not be held liable for any direct or indirect damages, including, but not limited to, personal injury, property damage or lost prots (or other economic loss) resulting from the use or inability to use the products. The manufacturer warrants that the Harris e-Core will meet the specications stated in Section 8.0. The manufacturer offers a 90 day limited warranty against any defects.
Research Use Only
This product is intended for research purposes by qualied persons for use in laboratories.
Trademarks
Harris e-Core™, Harris Micro-Punch™, Harris Uni-Core™ and Harris Cutting Mat™ are trademarks of Shunderson Communications Inc., FTA® is a registered trademark of Whatman® Part of GE Healthcare.
Copyright
Reproduction, transfer, distribution or storage of part or all of the contents in this document in any form without prior written permission from SCInc. is prohibited.
iii
Table of Contents
For Your Safety Read Before Operating .................................................................i
Notices....................................................................................................................iii
Table of Contents....................................................................................................iv
1.0 Introduction ......................................................................................................1
2.0 Harris e-Core Kit Content and Description ......................................................2
2.1 Harris e-Core Body .....................................................................................3
2.2 Harris Micro-Punch Tips ..............................................................................4
2.3 Optional Cord Wrist Strap ...........................................................................4
2.4 International Multi-Plug Set ......................................................................... 5
2.5 Power Supply .............................................................................................. 5
2.6 Harris Cutting Mat .......................................................................................6
2.7 Harris e-Core Carrying Case.......................................................................6
3.0 Harris e-Core Operating Instructions ............................................................... 6
3.1 For Your Safety Read Before Operating .....................................................6
3.2 Harris e-Core and Sample Preparation ....................................................... 8
3.3 Harris e-Core Operation .............................................................................. 9
3.4 Sample Ejection and Clean Up ...................................................................11
4.0 Cleaning ..........................................................................................................11
5.0 Tip Replacement..............................................................................................12
6.0 Plug Replacement ...........................................................................................12
7.0 User Notes.......................................................................................................13
8.0 Specications ..................................................................................................14
iv
1.0 Introduction
The Harris e-Core is a compact, versatile, electric coring device designed for unrestricted sampling of a wide range of DNA storage products or other sample substrates. The Harris e-Core was created to bridge the gap between the lower cost, manual Harris Micro-Punch and Harris Uni-Core and the automated bench top punching platforms. The ergonomic “joy stick” grip and the e-Core’s simple
operation provide comfort and efciency, integrating smoothly into all levels of laboratory sample
throughput. The user has complete management of the disc recovery process, the coring speed and disc delivery.
The Harris e-Core, like the Harris Micro-Punch and Harris Uni-Core, continues to offer the same unique combination of patented simultaneous sampling features: coring, retrieving and storing in a single operation, but with dramatically increased processing times. The Harris e-Core ejection system, like its predecessors, still offers the user the ability to control both the speed and delivery of the ejected disc. Most importantly the Harris e-Core ejection system does not generate static, affecting both accuracy and precision. The manually controlled ejection system and lack of static generated ensures accurate disc delivery.
The Harris e-Core is a coring tool and not a punch. The Harris e-Core has been designed to utilize dedicated replaceable Harris Micro-Punch Tips. The razor sharp cutting edge on the Harris Micro­Punch Tip cleanly and precisely cores discs from substrates, all but eliminating the creation of artefacts and potential cross contamination between sampling.
The Harris e-Core is currently available in three models identied by their respective industry standard coring diameters: 1.2 mm, 2.0 mm and 3.0 mm. Each model has a xed diameter plunger ejection rod ensuring efcient sample ejection. The Harris e-Core utilizes Harris Micro-Punch Tips which
conveniently thread to the Harris e-Core drive shaft. Each tip is made from 304 stainless steel with a nitride coating and Rockwell Hardness of Rc60. The ejection rod is made from 304 stainless steel.
1