Hardy HI 6500 User Manual

5 (1)
Hardy HI 6500 User Manual

HI 6500 series Quick Start Guide

The HI 6500 series are single channel instruments designed to process signals from analog load cells and output stable gross or net weight readings to a display, a PLC or PC.

This Quick Start Guide is intended for users that are already familiar with setting up Hardy Process Solutions weighing instruments.

A complete User’s Guide can be found under Documents and Programs online: www.hardysolutions.com Click the product menu> Weighing Instruments > Weight Processors > HI 6500. Then click on the Docs & Programs tab to download the User’s Guide.

Panel Installation:

1.Use template provided to drill the hole pattern into panel.

2.Install the display onto the panel using 4mm nuts and washers, tighten until gasket is compressed and display bezel comes in contact with the panel. Do not overtighten or the instrument’s case may crack.

3.Connect the display cable to display and then to the instrument.

4.Slide the instrument onto threaded rods and secure in place with 8mm x 4mm retaining nuts using an 8mm nut wrench (do not over tighten instrument retaining nuts, overtighting could cause the instrument to crack).

DIN Rail Installation:

1.Use template provided to drill the hole pattern into panel

2.Install the display panel using 4mm nuts and washers, tighten until gasket is compressed (4mm screws tightened with an 8 mm nut wrench) with a length length appropriate to panel thickness can be used in place of the 4mm rods sent with the unit).

3.Place the instrument onto the DIN rail, slide DIN rail clip up until it snaps into place to secure.

4.Connect the display cable to the display panel and instrument (legthen the cable as needed by using a thin screwdriver to release and replace wires between the two connectors).

Wiring:

1.Remove connectors from the instrument and wire as per markings found on the connector labels.

2.When using 4 wire sensors place a jumper between Excitation(+) and Sense(+) and a jumper between Excitation( ) and Sense( ).

3.Grounding of the instrument is recommended by connecting to either of the 2 screws located left of power connector or right of the RS 485 connector.

 

Analog

 

(HI6510 only)

USB

 

 

4. Connect to 12 27 VDC power source