Hantek DSO3062L, DSO3062AL User Manual


USER’S MANUAL
http://www.hantek.com
  

Content
General Safety Summary .............................................................................. 1
Chapter 1 Getting Start .................................................................................. 2
1.1 System Requirement ........................................................................................... 3
1.2 Install Software ..................................................................................................... 4
1.3 Install Driver ......................................................................................................... 7
1.4 General Features ............................................................................................... 10
1.5 General Check .................................................................................................... 11
1.6 Function Check .................................................................................................. 13
1.7 Self Calibration ................................................................................................... 14
1.8 Accessories ........................................................................................................ 15
Chapter 2 Operating Basics ........................................................................ 16
2.1 The User’s Interface ........................................................................................... 17
2.2 The Menu System .............................................................................................. 19
1.3 The Vertical System ........................................................................................... 21
2.4 The Horizontal System ....................................................................................... 22
2.5 The Trigger System ............................................................................................ 23
2.6 Input Connector ................................................................................................. 25
Chapter 3 Oscilloscope Functions ............................................................. 26
3.1 Setup the Oscilloscope ...................................................................................... 27
3.2 Set Vertical System ............................................................................................ 28
3.3 Setup Horizontal System .................................................................................... 34
3.4 Set Trigger System ............................................................................................ 35
3.5 Save/Load .......................................................................................................... 42
3.6 Utility Function ................................................................................................... 43
3.7 Measure Signal .................................................................................................. 48
3.8 The Display System ........................................................................................... 55
3.9 Zoom In/Out and Drag Waveforms .................................................................... 59
3.10 Interpolation ..................................................................................................... 61
3.11 Acquisition Modes ............................................................................................ 63
3.12 Print And Print Preview .................................................................................... 63
Chapter 4 Application Example .................................................................. 65
4.1 Simple Measurement ......................................................................................... 66
4.2 Capturing a Single-Shot Signal .......................................................................... 67
4.3 The Application of the X-Y Operation ................................................................. 67
4.5 Taking Cursor Measurements ............................................................................ 69
4.6 Logic Analyzer .................................................................................................... 72
4.7 Waveform Generator .......................................................................................... 75
Chapter 5 Appendix ..................................................................................... 83
Appendix A: Specifications ....................................................................................... 84
Appendix B: Accessories ......................................................................................... 86
Appendix C: General Maintenance .......................................................................... 87
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 1
General Safety Summary
Review the following safety precautions carefully before operate the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards use the device as specified by this user’s guide only.
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified
for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Connect and Disconnect Properly. Connect the probe output to the measurement device before connecting the probe to the circuit under test. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement device.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
Use Proper Probe. To avoid shock hazard, use a properly rated probe for your measurement.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the device, have it inspected by qualified service personnel before further operations.
Provide Proper Ventilation. Refer to the installation instructions for proper ventilation of the device.
Do not operate in Wet/Damp Conditions.
Do not operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 2
Chapter 1 Getting Start
The oscilloscope is small, lightweight, no external power required, portable oscilloscopes! The oscilloscopes is ideal for production test, research and design and all of the applications involving analog circuits test and troubleshooting, as well as education and training.
In addition to the list of general features on the next page, this chapter describes how to do the following tasks:
System Requirements
Install Software
Install Driver
General Features
General Check
Function Check
Self Calibration
Accessories
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 3
1.1 System Requirement
To run the oscilloscope software, the needs of computer configuration are as follows:
Minimum System Requirements
Operating System
Window XP/Vista/Win7/Win 8
Processor
Upwards of 1.00G processor
Memory
256M byte
Disk Space
500M disk free space
Screen resolution
800 x 600
Recommended Configuration
Operating System Windows XP SP3 System
Processor
2.4G Processor
Memory
1G Byte Memory
Disk Space
80G Disk Space
Screen resolution
1024 x 768 or 1280 x 1024 resolution
DPI Setting
Normal Size (96DPI)
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 4
1.2 Install Software
Caution: You must install the software before using the oscilloscope.
1. While in Windows, insert the installation CD into the CD-ROM drive.
2. The installation should start up automatically. Otherwise in Windows Explorer, switch to the CD-ROM drive and run Setup.exe.
3. The software Installation is started. Click 'Next' to continue.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 5
4. Choose a destination directory. Click 'Next' to continue.
5. Check the setup information. Click Next to start copying of files.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 6
6. This Status dialog is displayed during copying of files.
7. Updating Your System Configuration.
8. The installation is complete.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 7 
1.3 Install Driver
Example: DSO3062AL
1.Connect the A-Type Plug of USB cable to your PC’S USB port.
2.Connect the other A -Type Plug of USB cable to the device’s USB port.
3.New hardware is found.
4.New hardware search wizard starts.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 8
5.Select the specific location.
6.New hardware search wizard starts to search the driver.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 9
7.The wizard has finished installing for the device.
8.The new hardware is installed and ready to use.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 10
1.4 General Features
Product Features:
Two Channels, Bandwidth:
60MHz
Maximum real-time sample rate:
200MSa/s
Memory depth:
10K-16M /CH
Automatic setup for ease of use (AUTOSET);
Pass/Fail;
Built-in Fast Fourier Transform function(FFT);
20 Automatic measurements;
Automatic cursor tracking measurements;
Waveform storage, record and replay dynamic waveforms;
User selectable fast offset calibration;
Add, Subtract and Multiply Mathematic Functions;
Selectable 20 MHz bandwidth limit;
External trigger;
Waveform average;
Adjustable waveform intensity, more effective waveform view;
User interface in several user-selectable languages;
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 11
1.5 General Check
Please check the instrument as following steps after receiving an oscilloscope:
Check the shipping container for damage:
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
Check the accessories:
Accessories supplied with the instrument are listed in "Accessories" in this guide. If the contents are incomplete or damaged, please notify the franchiser.
Check the instrument:
In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, please notify the franchiser.
Probe Compensation
Perform this function to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel at the first time.
From the “Probe” menu, select attenuation to 1:10. Set the switch to “X10” on the probe
and connect it to CH1 of the oscilloscope. When using the probe hook-tip, insert the tip onto the probe firmly to ensure a proper connection.
Attach the probe tip to the Probe Compensator and the reference lead to the ground
connector, select CH1, and then press the “AUTOSET“ button into the menu or the toolbar.
Check the shape of the displayed waveform.
Correctly Compensated
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 12
Over Compensated
Under Compensated
1.
I
f necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the
fattest square wave being displayed on the oscilloscope.
2. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 13
1.6 Function Check
Perform this functional check to verify that your oscilloscope is operating correctly
.
Connect the oscilloscope
You should connect the A-Type Plug of USB cable to your PC USB port and connect the other A-Type Plug of USB cable to oscilloscope USB port.
Input a signal to a channel of the oscilloscope
The oscilloscope is equipped with two channels plus external trigger.
Please input signal in the following steps:
1. Set the attenuation switch on the probe as 10X and connect the probe on the
oscilloscope with CH1. Aim the slot in the probe connector at the faucet on BNC of CH1 and insert, then, turn right to lock the probe. Finally, attach the tip of probe and ground nip to the Connector of Probe compensator.
2. Set the CH1 probe attenuation of the oscilloscope to X10. (The default is X1).
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 14
3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click the
button. A square wave will be displayed within a several seconds. (Approximately 1
kHz, 2V, peak- to- peak).
4. Inspect CH2 with the same method. Repeat steps 2 and 3.
1.7 Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but you should always run the routine if the ambient temperature changes by 5v or more. For accurate calibration, power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To compensate the signal path, disconnect any probes or cables from the input connectors. Then, access the “Utility -> Calibration” option and follow the directions on the screen. The self calibration routine takes about several minutes.
DSO3062L/DSO3062AL
Digital Oscilloscope

USER’S MANUAL 15
1.8 Accessories
All the accessories listed below are standard accessories for the oscilloscope:
DSO3062L:
Probe x 2 (1.5m), 1:1(10:1), Passive Probes
A power adapter special for this product
A USB cable
A logic analyzer cable
Eighteen little test hook
A software installation CD
DSO3062AL:
Probe x 2 (1.5m), 1:1(10:1), Passive Probes
A power adapter special for this product
A USB cable
A BNC to BNC
A logic analyzer cable
Eighteen little test hook
A software installation CD
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 16
Chapter 2 Operating Basics
The User’s Interface
The Menu System
The Vertical System
The Horizontal System
The Trigger System
Input Connectors
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 17
2.1 The User’s Interface
Click the software icon on the desk after you finished the software setting and equipment connecting. Then a user interface will be showed as follows:
In addition to displaying waveforms, the display area is filled with many details about the waveform and the oscilloscope control settings.
1.The Main Menu
All settings can be found in the main menu.
2.The Toolbar
3.Displays the trigger information
Displays the edge trigger slope, source and level.
4.The DDS Panel (DSO3062AL)
5.The Horizontal Panel
The user can change Time/Div, format in the panel.
6.The Vertical Panel
The user can turn on/off the CH1/CH2. Also the user can change the CH1/ CH2 volt/div,
coupling and probe attenuation.
7.The LA Panel
In this panel, the user can set the LA voltage value.
8.The Trigger Panel
In this panel, the user can change the trigger mode, sweep, source and slope.
9.Displays the system time.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 18
10.Marker shows Edge trigger level.
11.Displays the main time base setting.
12.Displays the LA channel setting on-off status.
13.Displays the CH2 information
Readouts show the coupling of the channels. Readouts show the vertical scale factors of the channels. A “B” icon indicates that the channel is bandwidth limited
14.Displays the CH1 information
Readouts show the coupling of the channels. Readouts show the vertical scale factors of the channels. A “B” icon indicates that the channel is bandwidth limited
15.The markers show the CH2 reference points of the displayed waveforms. If there
is no marker, the channel is not displayed.
16. The markers show the CH1reference points of the displayed waveforms. If there
is no marker, the channel is not displayed.
17. A window that shows the display waveform in buffer position.
18. Marker shows horizontal trigger position.
19. Trigger status indicates the following:
AUTO: The oscilloscope is in auto mode and is acquiring waveforms in the absence of
triggers.
Trig’D: The oscilloscope has seen a trigger and is acquiring the post trigger data. WAIT: All pretrigger data has been acquired and the oscilloscope is ready to accept a
trigger.
STOP: The oscilloscope has stopped acquiring waveform data. RUN: The oscilloscope is running. PLAY: The oscilloscope is displaying the record waveforms.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 19
2.2 The Menu System
The Main Menu:
1.File: Load or Save data, setup
2.View: Change the user interface
3.Setup: Setup setting
4.Display: Change wave display type
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 20
5.Cursor: Set Cursor measure type
6. Measure: Set measurement parameters
7.Acquire: Run ,Stop or other operation setting
8. Utility: Utility setting
9.Vehicle:
10. Window: Window setting
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 21
11.Help: Turn on help file
1.3 The Vertical System
Click “Setup->Vertical”
The following figure shows the vertical Setup window. It shows the vertical parameters
setting.
1.Select channel: User can select the channel by clicking the Combo box.
2.Select channel: User can select the channel by clicking the Combo box.
3.VOLTS/DIV: Set the selected channel voltage range.
4.Coupling: Set the selected channel to DC/AC.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 22
5.Probe: Set the Select one according to the probe attenuation factor to ensure correct
vertical scale reading
6.BW Limit: Reject the frequency component higher than 20MHz.
7.Invert: Invert the selected wave.
2.4 The Horizontal System
Click “Setup->Horizontal”
The following figure shows the Horizontal System window. It shows the horizontal parameters settings.
1. Time/DIV: leads the setting of the time base parameters
2. Format: leads the setting of the horizontal format parameters
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 23
2.5 The Trigger System
Click “Setup-> Trigger”
The following figure shows the trigger system control.
Single Trigger:
1.Trigger Mode: Sets the trigger mode
2.Trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE
3.Trigger Source: Selects the trigger source to CH1, CH2, D0-D15, EXT.
4.Trigger Slope: Selects the edge trigger slope to Positive or Negative slope
Bus Trigger:
The Bus mode includes three trigger modes:
Trigger value: Trigger at the appearance of a specified value on the selected bus. Duration: Trigger at x time after the appearance of a specified value. Edge: Trigger at the appearance of a specified value on the selected bus and rising,
falling, or rising edge of the selected signal.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 24
Advance Trigger:
There are four parts in each condition, Data, Duration, Edge and Relative trigger value.
DSO3062L/DSO3062AL
Digital Oscilloscope
USER’S MANUAL 25
2.6 Input Connector
CH1/CH2: Input connectors for waveform display.
E-TRIG/S:
1. Input connector for an external trigger source. Use the Trigger Menu to select the Ext.
source.
2. Output Synchronized signal(DSO3062AL).
Other Connector:
OUTPUT: Signal output(DSO3062AL).
GND.: A ground terminal
USB PORT: Connect the USB cable to this port.
I0-I15: The LA input data channels.
CAL.: Probe compensation output.
GPower: Power Input port.
Loading...
+ 62 hidden pages