Hantek DSO2090 Software Manual

Page 1
DSO-2090 USB Manual

Certificate of Product Warranty

This product’s warranty, provided by our company, covers a period of 1 year form the date of purchase. All faulty parts and/or functions, resulting despite user’s normal use, will be repaired and/or replaced at no charge during the warranty period. However, if any of the malfunctions are caused by user carelessness, inadequate maintenance, or natural disaster, we will the provide repair and/or replacement services for a fee regardless of warranty period. In the event our product does not prove suitable for your application you can return the product for an exchange or refund. To claim, the product must be returned in good condition within 14 days. Before returning a product, please contact and provide us a serial number. The customer is responsible for all shipping costs. Products covered by this warranty are limited to those that are registered at the We headquarter or branch offices and repairs, replacements, or refunds will be issued according to manufacturer’s discretion. When in need of warranty services, the purchaser may visit our headquarter or one of our sales offices with the product or send the product with a detailed description of services required via postal service to any of our sales offices (the purchaser will be responsible for packaging, postage, and any postal insurance).Once repair services are completed, the supplier will return the product to the purchaser (return postage will be paid by the supplier). However, if we determined required services to be outside the boundaries of the warranty coverage, we will first contact the purchaser and provide an estimate of repair costs and obtain authorization to conduct services before any work is done. In such a case, return postage along with the repair costs will be billed to the purchaser when the repaired product is returned. We do not warrant that the software and the product with which it was supplied are completely error free or that will function correctly in all operating environments. It is essential that you, the users, should verify that the software
1
Page 2
DSO-2090 USB Manual
and product are functioning to your requirements before relying caused on them or the data that they generate. We do not accept responsibility for any loss or injury caused by the use of the product or software. It is the user’s responsibility to ensure that the product is suitable for the user’s application.
We will not be held legally for any malfunction resulting from user carelessness, abnormal use, and/or natural disaster. Even if we are in receipt of prior notification of such dangers, we will not be held legally liable.

Product Warranty Guide

Please read through the “Precautions for Safety” carefully to obtain a long and safe use of this product.
This product has been processed through careful quality control and testing procedures. Any malfunctions occurring during normal use are covered under the guidelines of the “Certificate of Product Warranty”, which is included in this Manual.
If you experience product malfunction, please contact our headquarters or any of our sales offices.
2
Page 3
DSO-2090 USB Manual

Precautions for Safety

The following instructions are provided for safe and correct use of the product and to prevent a potentially dangerous situation or damage in advance. Be sure to read through them before you use the product.

Recommended Usage Environment

The

Power Source

The USB terminal
Warning: Be careful not to input e
The product’s input power capacity is fixed as shown below to prevent
ambient temperature and relative humidity should be 0-40 and
10-80% respective for the normal operation of the product.
and DSO-2090 USB are connected via a USB cable for
DC +5V power from the PC. The product does not require an outside power source.
xcessive power voltage
electric shock and/or fire. Please DO NOT use voltage higher than that prescribed below.
(Warning) Be sure to check the probe attenuation before measuring the
voltage. The voltage that is higher than the maximum input voltage can cause damage to the device and you. In particular, there is the risk of an electric shock when measuring high voltage. So you need to pay extra attention not to have your hand touch the terminal.
Please eliminate all unused probes or tester leads so that they do not come in
contact with surrounding high voltage parts.
Make sure that the PC’s power source is grounded.
3
Page 4
DSO-2090 USB Manual

Handing Precautions

The probe g
lead to an elevated voltage. Connect the ground lead of the probe to earth ground only. If measuring the floating potential, we will recommend the measurement by the differential motion method using CH1 and CH2.
Do not connect or disconnect the product with the probe or test lead which
is connected to a voltage source.
Do not attempt to operate the product at the situation that a risk of
dysfunction or shock expected to happen.
Stop using the product immediately when you notice something unusual
about the smell, smoke or sound. And then remove the USB connection form your computer.
Keep the product out of the reach of children or those who cannot be trusted
with the use.
Do not use the product with your hands wet. It can cause an electric shock. Do not try the product for the purposes other than those stated.

Usage Location

round lead is at ground potential. Do not connect the ground
If th
DO NOT use this product near gas as well as other flammable and/or
Store the product where direct sunlight can’t reach. DO NOT use the product where it’s dirty, other machines are operated, or
DO NOT place the product on an unstable cart, stand, or tripod. Such usage
e products crevice is exposed to electric conductors (solids or liquids), the product will short circuit and possibly create dangers of electric shock or fire. Please keep this product away form humidity, water, or dust.
explosive materials.
there is a high level of electromagnetic waves.
creates the dangers of accident and/or damage to product.
4
Page 5
DSO-2090 USB Manual

Maintenance and Storage

If not using The recommended temperature and humidity for equipment storage is 25
at 50% humidity.

After Service

There are no parts that you can replace inside the product. Do not attempt
replacing a part yourself. It can give you an electric shock.
Do not attempt to open the case or repair yourself. It can cause an electric
shock or other safety accidents.
Once you open the case, you cannot have the privilege of an after-sales
service.
it for a long term, remove the USB cable from the product.

Minimum System Requirements

To install and run DSO-2090 USB, you should have the following.

Operating System

Windows 98/ME/2000/XP/VISTA

Memory

128Mb
yte

Graphic Card

Microsoft DirectX supported Screen r
esolution: 800x600
Color depth: 16bit
5
Page 6
DSO-2090 USB Manual

Hardware Specification

Input Max. sample rate Realtime sampling: 100MS/s using one channel,
50MS/s using two channels Channels 2 Bandwidth 40MHz analog bandwidth Vertical resolution Gain range 10mV ~ 5V/div @ x1 probe(10mV, 20mV, 50mV,
Range 8 divisions Offset level +/-4 divisions Coupling AC, DC Offset increments Impedance 1M ohm DC accuracy +/-3% Input protection 35Vpk (DC + peak AC < 10 kHz, without external
Display Mode Y-T, X-Y Timebase Timebase range 4ns/div ~ 1h/div(4ns, 10ns, 20ns, 40ns, 100ns, 200ns,
8 bits/channel
100mV, 200mV, 500mV, 1V, 2V, 5V/div1,2,5 sequence)
100mV ~ 50V/div @ x10 probe
1V ~ 500V/div @ x100 probe
10V ~ 5KV/div @ x1000 probe
0.02 div
attenuation)
400ns, 1us, 2us, 4us, 10us, 20us, 40us, 100us, 200us, 400us, 1ms, 2ms, 4ms, 10ms, 20ms, 40ms, 100ms, 200ms, 400ms, 1s,2s, 4s, 10s,20s,40s,10m,20m,40m,1h /div
6
Page 7
DSO-2090 USB Manual
1-2-4 sequence)
Acquisition mode
Range 10 divisions Buffer size 10K ~ 64K samples Trigger Mode Auto, Normal and Single Type Edge trigger: Rising edge, falling edge. Autoset Yes Range 10 divisions Trigger level +/-4 divisions Settabillity 0.02 div increments Math Measurements Vp-p, Vmax, Vmin, Vmean, Vrms, Vamp, Vhigh, Vlow,
Cursor Time/frequency difference, voltage difference Frequency
Math Addition, Subtraction, Multiplication, Division FFT Rectangular, Hanning, Hamming, Blackman Window Physical Interface Universal Serial Bus Power No external power source required.
Realtime sampling: 4ns /div ~ 400ms/div Roll mode: 1s/div ~ 1h/div
positive overshoot, negative overshoot, cycle mean, cycle rms, period, frequency, positive pulse width, negative pulse width, rise time (10%~90%), fall time (10%~90%), positive duty cycle, negative duty cycle
only in FFT mode
Bus-powered from USB
7
Page 8
DSO-2090 USB Manual
Dimensions 187 x100 x 33(mm)

Introduction

DSO-2090 USB developed by our company. is a portable PC-based Digital Oscilloscope.
High performance
DSO-2090 USB has the following features: 40MHz analog bandwidth, 100MHz real-time sampling.
USB connected
DSO-2090 USB uses USB that supports plug’n play, with 12Mbp communication speed.
Best performance for your dollar
DSO-2090 USB has many features that are comparable to the high speed stand-alone DSOs. But it costs a fraction of the price.
No external power required
DSO-2090 USB does not need an external power source, because it is bus-powered from USB
Easy to use
DSO-2090 USB is easy to use. It is intuitive and easy to understand.
8
Page 9
DSO-2090 USB Manual
Big screen
You can change DSO-2090 USB screen size as you need.
Various data format processing
DSO-2090 USB can save waveform in the following formats: text file, jpg/bmp graphic file, MS excel/word file.

Software Installation

1. While in Windows, insert the installation CD into the CD-ROM drive.
2. The installation should start up automatically. Otherwise in Windows Explorer, switch to the CD-ROM drive and run “ Setup.exe”.
3. The DSO-2090 USB Installation is started. Click 'Next' to continue.
4. Choose a destination directory. Click 'Next' to continue.
9
Page 10
DSO-2090 USB Manual
5. Check the setup information. Click Next to start copying of files.
6. This Status dialog is displayed during copying of files.
7. Updating Your System Configuration.
8. The installation is complete.
10
Page 11
DSO-2090 USB Manual

Hardware Setup

1. Connect the A-Type Plug of USB cable to your PC’S USB port.
2. Connect the B-Type Plug of USB cable to DSO-2090 USB’S USB port.
3. New hardware is found.
11
Page 12
DSO-2090 USB Manual
4. New hardware search wizard starts.
5. New hardware search wizard starts to search
6. New hardware wizard installs software
12
Page 13
DSO-2090 USB Manual
13
Page 14
DSO-2090 USB Manual
7. Finish new hardware search wizard.
8. New hardware is found.
9. New hardware search wizard starts to search
10. New hardware search wizard starts to search.
14
Page 15
DSO-2090 USB Manual
11. New hardware wizard installs software and finish new hardware search
wizard.
15
Page 16
DSO-2090 USB Manual

Setup DSO-2090 USB.inf under windows 2000

When DSO-209 following dialog box. In this case you must install DSO-2090 USB.inf file manually.
1. Choose 'Search for a suitable driver for my device' button. Click Next to continue.
2. Specify a location. Click Next to continue.
3. The DSO-2090 USB.inf file is located in DSO-2090 USB CD-ROM root
directory. Specify the location by either entering or browsing.
4. DSO-2090 USB.inf file is detected automatically. Click Next to continue.
5. Click Finish.

Probe Calibration

DSO-2090 USB Calibration
1. When manufactured, DSO-2090 USB is calibrated manually to obtain
maximum performance.
2. You may calibrate DSO-2090 USB 5 or 6 months after purchase.
0USB.inf file is not detected automatically, you see the
Probe Compensation
1. You must use a probe with more than 40MHz bandwidth to get undistorted signal.
2. Probe should be compensated whenever it is connected for the first time.
3. Connect calibration signal to channel 1, then push AUTOSET.
4. Check the shape of the displayed waveform.
5. Adjust the probe until the displayed waveform is compensated.
16
Page 17
DSO-2090 USB Manual
17
Loading...