TDS1
TDS Meter
with
Automatic
Temperature
Compensation
OPERATING:
• Remove the protective cap.
solution is displayed.
• Calibration is now complete.
• Turn TDS1
on.
• Immerse the tester into solution up to the maximum
immersion level without
touching the bottom of the
beaker.
• Stir gently and wait until
the display stabilizes.
TDS1 automatically compensates for temperature
variations.
CALIBRATION:
• Immerse the tester into HI
70080 (800 ppm @25ºC/
77ºF) calibration solution,
without exceeding the maximum level.
• Allow the reading to stabilize and using a
small screwdriver adjust the calibration trimmer until the nominal ppm value of the
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
When the
TDS1 TDS1
TDS1 can not be switched on or
TDS1 TDS1
::
:
::
the display fades, batteries have to be replaced.
Pull out the batteries compartment and replace all
four 1.5V batteries, while
paying attention to their polarity.
Batteries should only be replaced in a non hazardous
area using the battery types
specified in this instruction manual.
RECOMMENDATIONS FOR USERS:
Before using this product, make sure that it is entirely suitable
for the environment in which it is used. Operation of this
instrument in residential areas could cause unacceptable interferences to radio and TV equipment. Any variation introduced by the user to the supplied equipment may degrade
the instrument's EMC performance. To avoid electrical shocks,
do not use this instrument when voltage at the measurement
surface exceeds 24 VAC or 60 VDC. To avoid damages or burns,
do not perform any measurement in microwave ovens.