Hanna Instruments Primo 1 User Manual

Instruction Manual
Primo Series
EC and TDS Testers with
Dear Customer, Thank you for choosing a Hanna product. Please read carefully this instruction manual before using the
meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
FUNCTIONAL DESCRIPTION
Automatic Temperature
Compensation
www.hannainst.com
WARRANTY
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam­ine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.
Each meter is supplied with batteries and instructions.
Note: Conserve all packing material until the instrument has
been observed to function correctly. Any defective item must be returned in its original packing.
ACCESSORIES
HI 70031P 1413 µS/cm solution, 20 mL sachet, 25 pcs HI 70032P 1382 ppm solution, 20 mL sachet, 25 pcs HI 70039P 5.00 mS/cm solution, 20 mL sachet, 25 pcs HI 7031L 1413 µS/cm solution, 500 mL bottle HI 7032L 1382 ppm solution, 500 mL bottle HI 7039L 5.00 mS/cm solution, 500 mL bottle
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
ISTPRIMOR1 05/05
SPECIFICATIONS
OPERATION
ERROR CODES
Primo, Primo 2, Primo 5
Range
Primo 1999 ppm (mg/L) Primo 2 1999 ppm (mg/L) / 0.0 to 60°C Primo 5 1999 µS/cm
Resolution
Primo 1 ppm (mg/L) Primo 2 1 ppm (mg/L) / 0.1°C Primo 5 1 µS/cm
Accuracy (@20°C/68°F)
Primo ±2 % F.S. Primo 2 ±2 % F.S. / ±0.5°C Primo 5 ±2 % F.S.
Calibration solution
Primo HI 70032 (1382 ppm) Primo 2 HI 70032 (1382 ppm) Primo 5 HI 70031 (1413 µS/cm)
Calibration
Primo Automatic, at 1382 ppm Primo 2 Automatic, at 1382 ppm Primo 5 Automatic, at 1413 µS/cm
Temperature Compensation
Automatic, 0 to 60°C (32 to 122°F); ß = 2%/°C
Probe
Direct, fixed
Environment
0 to 50°C (32 to 122°F); RH max 95%
Battery Type / Life
2 x 1.5V / approx. 200 hours of continuous use auto-off after 5 minutes of non-use
Dimensions / Weight
180 x 50 x 25 mm (7.1x2.0x1.0”) / 50 g (1.8 oz.)
Primo 1, Primo 3, Primo 4
Range
Primo 1 1999 ppm (mg/L) Primo 3 1999 ppm (mg/L) / 0.0 to 60°C Primo 4 10.00 mS/cm
Resolution
Primo 1 1 ppm (mg/L) Primo 3 1 ppm (mg/L) / 0.1°C Primo 4 0.01 mS/cm
Accuracy (@20°C/68°F)
Primo 1 ±2 % F.S. Primo 3 ±2 % F.S. / ±0.5°C Primo 4 ±2 % F.S.
Calibration solution
Primo 1 HI 70032 (1382 ppm) Primo 3 HI 70032 (1382 ppm) Primo 4 HI 70039 (5.00 mS/cm)
Calibration
Primo 1 Automatic, at 1382 ppm Primo 3 Automatic, at 1382 ppm Primo 4 Automatic, at 5.00 mS/cm
Temperature Compensation
Automatic, 0 to 60°C (32 to 122°F); ß = 2%/°C
Probe
Fixed, with 1 m (3.3’) cable
Environment
0 to 50°C (32 to 122°F); RH max 95%
Battery Type / Life
2 x 1.5V / approx. 200 hours of continuous use auto-off after 5 minutes of non-use
Dimensioni/Peso
66 x 50 x 25 mm/ 50 g
• Turn the meter on by pressing the ON/OFF button.
• Immerse the probe in the sample to be measured, without exceeding the maximum immersion level.
• Stir gently and wait for stable reading. The meter auto­matically compensates for temperature variations.
• To extend the batteries life, after 5 minutes of non-use the meter provides for auto-off.
CALIBRATION
• Switch the meter on.
• Press and hold the ON/OFF button for about 3 seconds. The required calibration solution value will start blinking on the display to indicate that the calibration mode has been entered.
“1382” for Primo, Primo 1, Primo2, Primo3 “5.00” for Primo 4, and “1413” for Primo 5
• Immerse the probe in the proper calibration solution without exceeding the maximum immersion level.
• Stir gently and wait for about 20 seconds.
• Once the display stops blinking, the meter is calibrated and ready for use.
BATTERIES REPLACEMENT
The "Eb" message on the LCD warns the user when the batteries have to be replaced.
Open the bettery compartment and replace the both 1.5 V batteries, while paying attention to the cor­rect polarity.
Batteries should be replaced only in a non hazardous area using the battery type specified in this instruction sheet.
Important note: Once the batteries have been replaced, turn
the meter OFF and then ON again. Wait for about 30 seconds to allow a complete self-test, and then perform a new calibration.
Error in calibration, wrong calibration solu­tion or damaged probe.
Over range.
Low battery indication.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of this instrument in residential areas could cause interference to radio and TV equipment. Any variation intro­duced by the user to the supplied equipment may degrade the instrument EMC performance. To avoid electrical shocks, do not use this instrument when voltage at the measurement surface exceeds 24Vac or 60Vdc. To avoid damage or burns, do not perform any measurement in microwave ovens.