PRIMO 1
Microprocessor
TDS Meter
with Automatic Temperature
Compensation
SPECIFICATIONS:
RANGE 0 to 1999 ppm
RESOLUTION 1 ppm
ACCURACY (@25°C/77°F) ±2% f.s.
TYPICAL EMC DEVIATION ±1% f.s.
TEMPERATURE Automatic from
COMPENSATION
PROBE with 1m cable
ENVIRONMENT 0 to 50°C (32 to 122°F);
BATTERY TYPE 2x1.5V alkaline (included)
BATTERY LIFE approx. 200 hours
DIMENSIONS 66x50x25 mm (2.6x2.2x1")
WEIGHT 80g (2.9oz)
ACCESSORIES:
HI 70032P 1382 ppm calibration kit
This instrument is in compliance with the CE
directives.
0 to 60°C (32 to 140°F)
with ß=2%/°C
95% RH
of continuous use
(meter only)
(25x20mL sachets)
PRIMO 1
Microprocessor
TDS Meter
with Automatic Temperature
Compensation
http://www.hannainst.com
ISPRIMO1 09/00
OPERATION:
• Turn the meter on by pressing the
ON/OFF button.
• Immerse the probe in the sample without
exceeding the maximum immersion level.
• Stir gently and wait for reading to stabilize.
The meter automatically compensates for temperature variations.
• The TDS value of the sample will be shown on
the LCD in ppm unit.
• To extend the batteries life, after 5 minutes of
nonuse the meter provides for auto-off.
CALIBRATION:
BATTERIES REPLACEMENT:
The symbol "Eb" on the LCD warns the user when
the batteries have to be replaced.
Open the compartment and
replace the two 1.5V batteries while paying attention to their polarity.
Batteries should be replaced only in a non hazardous area using the battery type specified in this
instruction sheet.
NoteNote
: :
Note
: Once the batteries have been replaced, turn
NoteNote
: :
the meter OFF and then ON again. Wait for about
30 seconds to allow a complete self-test, and then
perform a new calibration.
ERROR CODES:
Error in calibration, wrong calibration solution or
damaged probe.
Over range.
Low battery indication.
• Switch the meter on.
• Press and hold the ON/OFF
button for about 3 seconds.
"1382" will start blinking on
the display to indicate that
the calibration mode has
been entered.
• Immerse the probe in the
HI 70032 HI 70032
HI 70032 calibration
HI 70032 HI 70032
solution without exceeding the maximum immersion level.
• Stir gently and wait for about 20 seconds to
confirm the reading.
• Once the display stops blinking, the meter is
calibrated and ready for use.
Warranty
The Primo 1 meter is warranted for one year against defects in
workmanship and materials when used for their intended purpose and
maintained according to instructions. This warranty is limited to repair or
replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential areas could cause interference to radio and TV equipment. Any
variation introduced by the user to the supplied equipment may degrade
the instrument EMC performance. To avoid electrical shocks, do not use
this instrument when voltage at the measurement surface exceeds 24 VAC
or 60 VDC. To avoid damage or burns, do not perform any measurement
in microwave ovens.