Hanna Instruments PL 100 User Manual

PL 100 / PL 101PL 100 / PL 101
PL 100 / PL 101
PL 100 / PL 101PL 100 / PL 101
Amperometric Chlorine Titrators
Instruction Manual
PL 100 Free Chlorine T itrator PL 101 T otal Chlorine T itrator
Both the instruments have features such as recorder outputs, 12VDC power supply for safety and reduction of EMI and built-in stirrer .
Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for the correct use of the instrument, as well as a precise idea of its versatility .
These instruments are in compliance with directives EN 50081-1, EN 50082-1 and EN 61010-1.
ISO 9000 Certified Company since 1992ISO 9000 Certified Company since 1992
ISO 9000 Certified Company since 1992
ISO 9000 Certified Company since 1992ISO 9000 Certified Company since 1992
2
TABLE OF CONTENTS
PRELIMINARY EXAMINATION . . . . . . . . . . . . . . 4
GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . 5
FUNCTIONAL DESCRIPTION . . . . . . . . . . . . . . 6
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . 9
METHOD OF ANALYSIS. . . . . . . . . . . . . . . . . 10
Forward titration . . . . . . . . . . . . . . . . . . . 10
Back titration . . . . . . . . . . . . . . . . . . . . . 11
End Point determination . . . . . . . . . . . . . . 11
HOW TO SELECT THE CORRECT AMPEROMETRIC TITRATION PROCEDURE . . . 13
HOW TO COLLECT THE SAMPLE . . . . . . . . . . 14
OPERATIONAL GUIDE. . . . . . . . . . . . . . . . . . 15
Initial preparation . . . . . . . . . . . . . . . . . . 15
Method A: Free Chlorine forward titration. . 15 Method B: T otal Chlorine forward titration. . 1 6
Method C: T otal Chlorine back titration . . . 1 8 INTERFERENCES AND SOURCES OF ERROR. . 2 1
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . 23
Calibration requirements . . . . . . . . . . . . . 23
Probe conditioning . . . . . . . . . . . . . . . . . 23
Electrode cleaning procedure . . . . . . . . . . 23
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . 24
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . 25
OTHER PRODUCTS FROM HANNA . . . . . . . . 26
CE DECLARATION OF CONFORMITY. . . . . . . 27
3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam­ine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer immediately .
Each titrator is supplied complete with HI 710005 or HI
710006 12VDC power adapter . PL 100C and PL 101C are also supplied with:
HI 3500A-1 10mL burette
HI 3500B reagent container
HI 3500C rubber bulb
• 50mm (2”) long (dia. 7 mm/0.3”) magnetic stir bar
HI 3132B glass-body platinum-platinum electrode with 1 m (3.3”) cable and BNC connector
HI 76405 electrode holder
• small spoon
• graph paper
• recorder plug
Note Save all packing materials until you are sure that the instru-
ment functions correctly . Any damaged or defective items must be returned in their original packing materials together with the supplied accessories.
Safety Precautions Please take the time to read the safety precautions carefully
wherever they appear in this manual. They are provided to prevent personal injury and damage to the instrument. This safety information applies to the operators and service per­sonnel and the following two captions are used:
CAUTION: identifies conditions or practices that could re­sult in damage to the instrument or persons;
WARNING: identifies conditions or practices that could re­sult in personal injury or loss of life.
Note Because of the inherent dangers in handling chemical
samples, standards and reagents, HANNA Instruments strongly recommends the user of this product to review the Material Safety Data Sheets and become familiar with safe handling procedures and proper usage prior to handling any chemicals.
4
GENERAL DESCRIPTION
The Hanna PL 100 and PL 101 Chlorine Analyzers are amperometric titrators which allow the determination of the chlorine content in a known quantity of sample . The PL 100 measures Free Chlorine and the PL 101 T otal Chlorine.
The Hanna PL 100 and PL 101 Amperometric Titrators are laboratory instruments containing a precision adjustable volt­age source, a microammeter with LCD display , and a speed regulated magnetic stirrer .
They operate providing a constant voltage to a dual elec­trode platinum probe and indicating the resulting probe current. As titration proceeds, the change of the electrode current is noted. At the point known as the End P oint, abrupt change in the slope of the current curve occurs and the titra­tion is complete. Sample concentration may be derived from the End Point value.
The magnetic stirrer rotates at 300 RPM during the titration to ensure proper mixing of the sample and titrant, yet slowly enough to avoid volatilization of the measured species. The stirrer motor is turned on and off with a rocker switch on the front panel.
The instrument case features a stainless steel top and easy visible LCD display for viewing the measured probe current. Adjustment of applied probe voltage is provided by a digital potentiometer . The potentiometer setting is controlled by two
keys in the front panel: ” ” and “ ”. A static digital memory retains the last value set when the instrument power is re-
moved. The PL 100 and PL 101 also feature a recorder output which
provides a low impedance output voltage from 0.00 to 2.00V : This voltage corresponds to the probe current displayed on the front panel LCD display . P robe current range is 0.00 to
2.00 µA. A 10 mL glass automatic burette system is available together
with standard titration solutions 0.00564N Phenylarsine Ox­ide (P AO) and 0.000564N PA O, to cover the 0-1.5 ppm of Cl2 or 1-15ppm of Cl2.
In addition to these a 0.00188N Iodine standard titration solution is available in the range 0-4ppm of T otal Chlorine.
5
FUNCTIONAL DESCRIPTION
FRONT PANEL
1. UP key
2. DOWN key
3. Power LED
4. Stirrer ON/OFF switch
5. LCD Display
REAR PANEL
6. ON/OFF switch
7. Fuse holder
8. 12 VDC socket (for HI710005 or HI 710006)
9. BNC Electrode socket
10. Recorder output
6
FRONT PANEL Display
The digital redout indicates the probe current in microamps. An overcurrent situation (greater than 1.99 µA) results in a “1.” display (no tenths or hundredths digit). T o reduce probe current, merely reduce probe voltage by pressing the front
panel “ ” key .
Power on indicator
During normal operation the red front panel indicator LED should be on continuously , indicating that the instrument is turned on.
Stirrer motor switch
This switch, located on the right hand side of the front panel, operates the stirrer . Instrument power must be on (LED indi­cator on) for the stirrer to operate.
“ ” and “ ” bias control keys
Bias voltage applied to the probe is controlled by a digital potentiometer . The potentiometer has sixty-four steps which
are selected with the front panel “ ” and “ ” keys. A single depression causes the potentiometer setting to increase or
decrease by one step. Pressing and holding one of the keys for one second causes the potentiometer to increment or dec­rement at ten steps per second until the key is released or until the end of the taper is reached.
A static digital memory retains the last setting when the power is turned off or removed from the system.
REAR PANEL Power on switch
T o power on and off the instrument.
Probe input connector
The probe is connected to the rear of the instrument with a BNC connector .
Recorder output connector
The recorder output female connector is located on the rear panel of the instrument. The recorder connecting cable is terminated with a male banana type plug provided with the instrument.
7
Recorder output
Power
Fuse
The recommended recorder hookup uses a shielded, twisted­pair cable. The shield should be connected to (earth) ground at the recorder end and left open at the instrument end.
The output is 0.00 to 2.00 V corresponding to 0.00 to 2.00 µA probe current as indicated on the front panel display .
Power is provided from the mains through a HI 710005 or HI 710006 power adapter . Be sure the mains voltage matches the input voltage specified on the power adapter . The power adapter output connector plugs into a socket on the back panel of the instrument.
Caution: The power adapter may be damaged if not operated at the
correct voltage.
The instrument power conditioning circuitry is protected with a 200 mA, 5 x 20mm tubular fuse located on the rear panel. To replace the fuse simply twist off the fuse holder cap and replace the fuse.
Unplug the meter before replacing the fuse.
8
SPECIFICATIONS
Range 0 - 750mVDC probe voltage
Resolution 0.01 µA Accuracy ±0.01 µA Probe HI 3132B glass-body platinum
Recorder output 0.00 to 2.00V
T ypical EMC ±1% f.s. Deviation
Stirrer Motor Speed 300 ±10 RPM (constant) Power Source 12VDC through HI 710005 or HI 710006
Environment 0 to 50°C (32 to 122°F);
Dimensions 180x180x70mm (7.1x7.1x2.8")
PL 100 PL 101
0.00 - 2.00µA probe current
electrode with 1 m (3.3”) cable
corresponding to 0.00 to 2.00 µA
(included)
0 to 95%RH (non condensing)
Weight 1.6 Kg (3.6 lb.)
9
METHOD OF ANALYSIS
Amperometric titration involves measuring the electrical cur­rent flow between two electrodes, usually platinum, immersed in a known quantity of a sample solution which contains an unknown concentration of the chemical to be measured. The titration of the chlorine with the reducing compound Phenylarsine Oxide (PA O) is an application of this technique.
When a small potential is applied across the two platinum electrodes of the titrator probe immersed in the solution con­taining Free Chlorine, a small electrical current will flow . The
reversible reaction Cl2 + 2e- 2 Cl- occurs at both elec­trodes as the reducible form is oxidized at the anode and the
oxidized form is reduced at the cathode.
FORWARD TITRA TION
The gradual addition of the reductant PAO (titrant), in an environment buffered at pH 7, irreversibly reduces the oxi­dized form of the Chlorine present. The reaction it undergoes is:
PhAsO + Cl2 + 2H2O PhAsO(OH)2 + 2Cl- + 2H (Ph =phenyl) (a) The final removal of all oxidized Chlorine terminates the re-
versible reaction and the probe current goes to zero. In the case of Chloramine determination, the pH is lowered
to 4 and potassium iodide is added to convert the chloram­ine species to an equivalent amount of triiodide.
NH2Cl + 3I- +H2O + H+ NH4OH + Cl- + I (monochloramine). NHCl2 + 6I- +H2O + 2H+ NH4OH + 2Cl- + 2I (dichloroamine). The triiodide is titrated with PAO with the current change
measured amperometrically . PhAsO + I
By knowing the exact amount of the reductant added which just extinguishes the probe current, the original concentra­tion of Chlorine present in the sample may be calculated. Required data for the calculation are: sample volume, re-
-
+ 2H2O 3I- + PhAsO(OH)2 + 2H
3
+
-
3
-
3
+
(b)
10
ductant concentration and the activity ratio of the reductant to the measured substance.
BACK TITRA TION
For waters which contain potential chemical interferences or low concentration of Total Chlorine, a back -titration is rec­ommended. In the back-titration procedure, a known excess amount of PAO is added to the sample at pH 4 with an excess of iodide. The PA O reacts with the free chlorine and chloramines present. The amount of unreacted PAO is ti­trated with an iodine solution. A blank back-titration is also required. The total chlorine is then calculated, based on the P AO left in the sample.
The back amperometric End Point is signaled when free io­dine (triiodide ion) is present, which is indicated by a current flow between the electrodes (see chemical reaction (a) and (b)).
The back-titration method is popular in wastewater laborato­ries because:
• the sample chlorine can be “fixed” at the sampling site with the addition of excess reductant.
• Since the End P oint is reversed, there is less interference from iodine-demand substances in the sample.
END POINT DETERMINATION
At the point known as the End Point, abrupt change in the slope of the current curve occurs and the titration is com­plete. Typical titration plots for the forward and back amperometric titration are shown in pictures below.
As the End Point is approached titrant has to be delivered in small amounts, while microampere readings have to be re-
11
corded after each addition (for best results at least 3 points before and 3 points after the End Point). The End Point is determined by the intersection of the two best lines through the points. The titrant volume is multiplied by a factor to ob­tain the sample chlorine concentration or can be read (only in case of forward titration) straight from the graph if the PL 100/PL 101 graph-paper is used.
12
HOW TO SELECT THE CORRECT AMPEROMETRIC TITRATION PROCEDURE
Select the procedure that best fits to your need as suggested in the block diagram below .
Forward titration can be performed in two ranges: LOW RANGE when using 0.000564N PAO HIGH RANGE when using 0.00564N PAO Do not perform measurements using the forward titration
method when chlorine concentration is under 0.05 mg/L. Use back titration for low concentrations of T otal chlorine.
DRINKING AND WASTEWA TER
FREE CHLORINE
(PL 100 only)
FORWARD TITRATION
Low range
• 0-1500 µg/L (detection limit 0.05 mg/L)
High range
• 1000-15000 µg/L
METHOD A
TOT AL CHL ORINE
(PL 101 only)
FORWARD TITRA TION
Low range
• 0-1500 µg/L (detection limit 0.05 mg/L)
High range
• 1000-15000 µg/L
METHOD B METHOD C
BACK TITRA TION
Range
• 0-4000 µg/l
13
HOW TO COLLECT THE SAMPLE
F ree chlorine is a strong oxidizing agent and in natural wa­ters reacts with various inorganic and organic compounds, its decomposition being influenced by parameters like reac­tant concentrations, pH, temperature, salinity and sunlight.
Combined chlorine (chloroamines) is more stable and per­sistent in the environment.
For best results, the delay between sample collection and analysis should be minimized.
Plastic sample containers have a high chlorine demand, thus collect sample in glass bottles. If possible rinse the container with a portion of the sample otherwise rinse with deionized water.
Fill the bottle up to the rim and keep it tightly closed. Avoid excess agitation and exposure to sunlight when sam-
pling. If the back titration method is used for total chlorine determi-
nation, preserve the sample on site. Add 2.00 mL of 0.00564N standard PAO solution and 1.0 mL pH 4 Acetate Buffer to a clean dry glass container with at least 150 mL capacity. At the sampling site, measure 100 mL of sample and carefully transfer it to the sample container . Swirl to mix.
It is important that the entire contents of the sample container be transferred to the beaker used in the titration. Rinse the bottle a few times with a small amount of chlorine free water .
14
OPERATIONAL GUIDE
INITIAL PREPARA TION
• Connect the power supply adapter to the DC input
• Connect the probe to the BNC connector .
• Be sure the front stirring switch is in the OFF position and turn the instrument on by the ON/OFF switch on the rear panel.
PROCEDURES Method A: free chlorine forward titration (PL100 only)
1. Fill the bottle of the automatic burette with
0.000564N P AO solution (HI 70471) for titrations up to 1500 µg/L Cl2 or use the
0.00564N PA O solution (HI 70470) for titrations up to 15000 µg/L Cl2. Fill the 10 mL automatic burette to the zero mark.
2. Use a 100 mL volumetric pipet to trans­fer 100 mL of sample to a 250 mL beaker and add about 100 mL chlorine free wa­ter.
3. Place the stirbar into the beaker .
4. Add 1 mL of pH 7 phosphate buffer solu­tion (HI 70472) to the beaker .
Note: If the pH of the sample is between 6.0 and 7.5
it is not necessary to add the buffer .
5. Turn on the speed controlled stirrer and place the beaker on the top of the PL 100.
6. Immerse the probe tip into the sample, make sure the platinum electrodes are submerged.
7. Adjust the potentiometric setting, using the “ ” and “ ” keys on the front panel until the dis-
play reads about 1.00.
15
8. Dispense the titrant into the beaker in small increments. Note the downward reading on the amperometric titrator. Record the display reading that corre­sponds exactly to the mL of the titrant added. Record at least 3 points before and 3 points after the End P oint.
9. Construct a titration graph using the PL 100 graph paper .
10.Draw the best-fit line through each set of points. The end point is determined by the intersection of the two best lines through the points.
11 .Read directly the free chlorine concentration on the
top of the graph by drawing a straight vertical line through the End Point or read the volume of titrant used until the End Point and multiply by 2 when titrant (a) 0.00564N PA O is used or multiply by 0.2 when titrant (b) 0.000564N P AO is used.
mL
(till End Point)
x 2
= mg/L Free Cl2.
(or 0.2)
Method B: Total Chlorine Forward Titration (PL 101 only)
1. Fill the bottle of the automatic burette with
0.000564N PAO solution (HI 70471) for titrations up to 1500µg/L Cl2 or use the 0.00564N PA O solution (HI 70466) for titrations up to 15000µg/L Cl2. Fill the 10 mL automatic burette to the zero mark.
2. Use a 100 mL volumetric pipet to transfer 100mL of sample to a 250 mL beaker and add about 100 mL chlorine free water.
3. Add one spoon of potassium iodide from the bottle (HI
70468) and swirl to dissolve. Place the stirbar into the beaker.
4. Add 1 mL of pH 4 acetate buffer solution (HI 70467) into the beaker .
5. Turn on the speed controlled stirrer and place the beaker on the top of the PL 101.
16
6. Immerse the probe tip into the sample, make sure the platinum electrodes are submerged.
7. Adjust the potentiometric setting, us­ing the “ ” and “ ” keys on the front panel until the display reads about 1.00.
8. Dispense the titrant into the beaker in small increments. Note the downward reading on the amperometric titrator. Record the display reading that corre­sponds exactly to the mL of titrant added. Record at least 3 points before and 3 points after the End P oint.
10.Construct a titration graph using the PL 101 graph-paper .
11 .Draw the best-fit line through each set of points. The End
Point is determined by the intersection of the two best lines through the points.
12 . Read directly the total chlorine concentration on the top
of the graph by drawing a straight vertical line through the End Point or read the volume of titrant used until the End Point and multiply by 2 when titrant (a) 0.00564N P A O is used or multiply by 0.2 when titrant (b) 0.000564N P AO is used.
mL
(till End Point)
x 2
= mg/L Total Cl2.
(or 0.2)
17
Method C: Total Chlorine back titration (PL 101 only)
Instrument setting
1. Fill the bottle of the automatic burette with
0.00188N I2 solution (HI 70469) and fill the 10 ml automatic burette to the zero mark.
2. Place the stirbar into a clean 250 mL beaker and add about 200 mL deionized water .
3. Add 1 mL of pH 4 acetate buffer (HI
70467) and one spoon of potassium io­dide from the bottle (HI 70468).
4. Turn on the speed controlled stirrer and place the beaker on the top of the PL 101.
5. Immerse the probe tip into the solution, make sure the platinum electrodes are submerged.
6. Add 1.5 mL of the iodine solution into the bea­ker and adjust the potentiometric setting using
the ” ” and “ ” keys on the front panel until the display reads about 0.50 -0.70. Switch off the
stirring action.
7. Remove the probe from the beaker and rinse the plati­num electrodes with deionized water . The probe response slope is adjusted. Don’t change the setting from this point.
18
Blank Titration
8. Refill the automatic burette to the zero mark.
9. Place the stirbar into a clean 250 mL beaker and add about 200 mL of deionized water .
10. Add exactly 2.00 mL of the standard 0.00564N PAO solution (HI 70466) to the beaker and swirl to mix.
11. Add 1 mL of pH 4 acetate buffer solution (HI 70467) and one spoon of potassium iodide from the bottle (HI 70468)
12. Turn on the speed controlled stirrer and place the beaker on the top of the PL 101.
13.Immerse the probe tip into the sample and make sure the platinum electrodes are submerged.
14. Dispense 5 mL of titrant into the bea­ker.
15.Continue dispensing titrant into the beaker in small increments. Record the display reading that corresponds exactly to the mL of titrant added. Record at least 3 points before and 3 points after the End P oint.
16. Construct a titration graph.
17 .Draw a best-fit line through each set of points. The End
Point is determined by the intersection of the two best lines through the points.
18. Read the volume of titrant used until the End Point, this is mL zero.
Note: Standard Iodine is subjected to a normal degradation and
this could lead to an increasing of the End P oint under the same measurement conditions. Blank titration compensates for standard iodine degradation. Discard a standard iodine solution if the End P oint is grater than 8 mL and repeat the procedure with new standard iodine.
Sample titration
19 .Refill the automatic burette to the zero mark.
20. Use a 100 mL volumetric pipet to transfer 100 mL of sample to a 250 mL beaker and add about 100 mL chlo­rine free water .
21 .Add exactly 2 mL of the standard 0.00564N P A O solu-
tion (HI 70466) to the beaker and swirl to mix. Add 1 mL of pH 4 acetate buffer (HI 70467) and one spoon of potassium iodide from the bottle (HI 70468).
19
Note: If the sample is pretreated at the sampling site with the PAO
and the acetate buffer as described before, transfer the sample quantitatively to the beaker and add one spoon of potassium iodide from the bottle (HI 70468)
22. T urn on the speed controlled stirrer and place the beaker on the top of the PL 101.
23.Immerse the probe tip into the sample, make sure the platinum electrodes are sub­merged.
24 .Dispense the titrant into the beaker in
small increments. Note the reading that corresponds exactly to the mL of titrant added. Record at least 3 points before and 3 points after the End Point.
25 .Construct a titration graph. 26 .Draw a best-fit line through each set of points. The End
Point is determined by the intersection of the two best lines through the points.
27. Read the volume of titrant used until the End Point. This is mL sample.
28 .Calculate the total chlorine concentration using the for-
mula:
4.00 - 4.00 x (mL sample)/(mL zero) = mg/L T otal Cl2.
Note: If a negative value is found, the sample contains an excess of
de-chlorinating agent, such as sulfur dioxide, sulfite or bisulfite.
20
INTERFERENCES AND SOURCES OF ERRORS
Despite Standard Methods section 4500 Cl-A.3.b. states that “the amperometric method is the method of choice because it is not subject to interferences from color, turbidity, iron, manganese, or nitrite nitrogen ”, the amperometric method will detect (as all of the common chlorine methods) disinfec­tants such as bromine (Br2), Ozone (O3), Chlorine dioxide (ClO2), and hydrogen peroxide (H2O2).
In general all oxidants which can be reduced by the strong reducing agent P AO will interfere with the free chlorine deter­mination. For the total chlorine determination, interference can be caused by compounds that oxidize iodide to iodine and those that can be reduced by PA O. For example, man­ganese in the lower oxidation states +2, +3, or +4 can be oxidized by the free chlorine. The oxidized formes of manga­nese (+4 to +7) can be reduced by PAO in free chlorine titration or manganese (+4 or +7) can oxidize iodide to iodine during the total chlorine titration.
Hanna Instruments researchers found that nitrite interference can cause either a positive or negative interference depend­ing on the order of reagent addition.
Therefore the preferred procedure in the back titration for T otal Chlorine determination is buffering the solution to pH 4 before adding KI in order to minimize nitrite, manganese and iron interference.
For both free and total chlorine determination, Hanna in­struments has selected PAO as reducing agent because it gives a sharper end point.
The potassium iodide used for the total chlorine determina­tion can be oxidized with enough exposure to oxygen and ultraviolet light. Therefore keep the bottle of HI 70468 tightly closed and out of direct sunlight. Another possible error dur­ing total chlorine determination is volatilization of free iodine. Volatilization from the reaction mixture during the forward titration is minimized because excess iodide is present, but after adding the potassium iodide, start the titration as soon as possible. Keep the standard iodine solution in a closed, dark bottle to avoid volatilization of iodine.
21
Iodine demand of certain samples can cause a shift of the end point as shown in the following graph:
The iodine, formed in case of forward total chlorine determi­nation or added as titrant during the backward titration, can be absorbed by suspended particles or can react with or­ganic matter . This type of interference is common in the case of muddy or highly organic-rich samples.
Another source of error is due to the tendency of some metal to poison the electrodes of the titrator. Iron, copper, silver and some other species can plate or coat the platinum probe electrodes and diminish the probe response. Therefore the dual platinum electrode (DPE) has to be cleaned regularly (see electrode cleaning procedure on page 23).
22
MAINTENANCE
CALIBRATION REQUIREMENT
Calibration of the PL 100 and PL 101 Chlorine Titrators is not required.
If, for any reason, the measurements are inaccurate, contact your dealer or the nearest Hanna Customer Service Center for recalibration.
PROBE CONDITIONING
When the probe has not been used for some time (one week) or it is new , it is recommended that it be conditioned as fol­lows:
1. Add a few drops of bleach to tap water in a 250 mL beaker and place the stir bar in the beaker .
2. Place the beaker on the PL 100/PL 101, turn on the instrument and the stir motor .
3. Immerse the probe in the solution, adjust the digital po­tentiometer for a current of 0.5 to 1.5 µA, and leave for ten minutes.
PROBE CLEANING PROCEDURE
Cleaning involves soaking the probe in a 1:1 nitric acid solu­tion for two hours and than rinsing with deionized water. To stabilize the cleaned probe soak it in chlorinated tap water .
23
ACCESSORIES
HI 3132B Glass-body platinum-platinum electrode with 1 m (3.3”) cable
and BNC connector
HI 3500A-1 10mL glass burette HI 3500B Reagent container HI 3500C Rubber bulb HI 70466 PA O Standard solution 0.00564N (for PL 101 only) HI 70467 Acetate buffer HI 70468 Potassium iodide HI 70469 Iodine standard solution HI 70470 PA O Standard solution 0.00564N (for PL 100 only) HI 70471 PA O Standard solution 0.000564N HI 70472 Phosphate buffer HI 710005 115 VAC to 12 VDC power adapter HI 710006 230VAC to 12 VDC power adapter HI 731320 50mm (2”) long, dia. 7 mm(0.3”) magnetic stirbar (10 pcs) HI 76405 Electrode holder .
24
WARRANTY
All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when
used for their intended purpose and maintained accord­ing to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge.
Damage due to accident, misuse, tampering or lack of pre­scribed maintenance are not covered.
If service is required, contact the dealer from whom you pur­chased the instrument. If under warranty , report the model number , date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty , you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service de­partment and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
25
OTHER PRODUCTS FROM HANNA
• CALIBRATION AND MAINTENANCE SOLUTIONS
• CHEMICAL TEST KITS
• CHLORINE METERS
• CONDUCTIVITY/TDS METERS
• DISSOLVED O XYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS (Colorimeters)
• MAGNETIC STIRRERS
• Na/NaCl METERS
• pH/ORP/Na ELECTRODES
• pH METERS
• PROBES (DO, µS/cm, RH , T , TDS)
• PUMPS
• REAGENTS
• SOFTWARE
• THERMOMETERS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of Accessories
26
Most Hanna meters are available in the following formats:
• BENCH- T OP METERS
• POCKET -SIZED METERS
• PORT ABLE METERS
• PRINTING/LOGGING METERS
• PROCESS METERS (Panel and W all-mounted)
• WATERPROOF METERS
• METERS FOR FOOD INDUSTRY
For additional information, contact your dealer or the near­est Hanna Customer Service Center . Y ou can also e-mail us at tech@hannainst.com.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interference to radio and TV equipment.
The metal band at the end of the sensor is sensitive to electrostatic discharges. Avoid touching this metal band at all times.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
Unplug the instruments from power supply before opening the front cover.
27
HANNA LITERATURE
Lab Recording Water Analysis Handbook
PRINTED IN PORTUGAL
Envirocare General Catalog
These and many others catalogs, handbooks and leaf­lets are available from Hanna. To receive your free copy, contact your dealer or the nearest Hanna Cus­tomer Service Center.
http://www.hannainst.com
MANPL100R1
07/98
Loading...