Hanna Instruments PH 300, PH 301, PH 302 User Manual

Instruction Manual
pH300 pH301 pH302
Microprocessor
GLP Bench-top
pH/mV/ISE/°C Meters
ppm
DATE TIME
Rel mV
MEM
ON OFF
CFM
WRONG
CAL
LOG
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
http://www.hannainst.com
1
PAPER PRINT
pH302
pH METER
RANGE CAL
CFM /
PGM
ENTER /
These Instruments are in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully be­fore using the instrument. This manual will provide you with all the neces­sary information for the correct use of the instru­ment, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA,
UL and directives.
(EN 50081-1 and EN 50082-1)
TABLE OF CONTENTS
Preliminary Examination................................... 1
General Description .......................................... 1
Functional Description pH300 .......................... 2
Specifications pH300 ........................................ 4
Functional Description pH301 .......................... 5
Specifications pH301 ........................................ 7
Functional Description pH302 .......................... 8
Specifications pH302 ...................................... 10
The Rear Panel ................................................ 11
LCD Functional Description ............................. 11
Operational Guide ........................................... 12
pH Calibration ................................................. 29
pH Values at Various Temperature ................. 39
ION Calibration (for pH301 only) ..................... 4 0
mv Calibration ................................................. 44
Temperature Calibration ................................. 50
Logging with pH301 ........................................ 55
Programming Functions with pH302 .............. 58
Interface with PC ............................................. 82
Troubleshooting Guide ................................... 83
Electrode Conditioning and Maintenance ...... 84
Temperature-Resistance Correlation
for HANNA pH Sensitive Glass ....................... 8 8
Additional Information about
Interface with PC ............................................. 90
Accessories..................................................... 99
Electrode Application Reference Guide ........ 10 7
Warranty......................................................... 108
CE Declaration of Conformity ...................... 109
ISO 9000 Certified Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer. Each pH meter comes supplied complete with:
HI 1131B glass-body combination pH electrode
HI 7669/2W temperature probe
HI 3131B glass-body platinum ORP electrode
pH 4.01 & 7.01 buffer solutions (20 mL each)
12VDC adapter (HI710005 or HI710006)
5 paper rolls (pH 302 only)
The KIT version includes also:
HI76405 electrode holder
HI 92000 Windows® compatible applica-
tion software
HI 920010 PC connection cable
Storage, cleaning and electolyte solutions
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments pH300, pH301 and pH302 are professional bench-top pH meters for pH, ORP (Oxidation Reduction Potential), ISE and temperature measurements. Their built-in mi­croprocessor ensures accurate and user­friendly operation. pH measurements are automatically compen­sated for the temperature effect (ATC). These meters come equipped with a large, easy-to-read LCD which shows the pH or mV or ION (for pH301 only) with tempera­ture simultaneously. The display has graphic symbols to make the calibration procedure easy to follow. Calibration is not allowed until the electrode has stabilized to ensure accu­rate measurements. Through the RS 232 serial port you can transfer sample measurements to a PC.
32
FUNCTIONAL DESCRIPTION pH 300
THE FRONT PANEL
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
OFF
COMM
CFM
WRONG
CAL
LOG
ONONOFF
COMM
ppmpHpH
INTV
mV°C
m
1 2
BUF
ppm
Rel mV
MEM
BUF
3
1 2
3
mV°C
RANGE CAL DATA
CAL MS°C
CFM
DATE TIME
ppmpHpH
RANGE
CAL MS°C
CFM
s
pH METER
°C
INTV
m s
CAL DATA
pH METER
°C
pH300
MR
pH300
MR
Keyboard on the Left
COMM to display the current baud rate for RS232.
Keyboard on the Right
!!
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not connected) and to select the calibration buffer, the baud rate, the command pre­fix.
CAL to enter and exit the calibration mode. After the pH calibration, to enter the cali­bration date, use this key to toggle be­tween month, day and year.
CAL DATA to display the date of last pH calibration. This also displays the offset and the slope values expressed in mV.
CFM to confirm calibration values and to confirm baud rate and command prefix during communication setting mode.
MR to recall the stored value.
MS to hold and memorize the measured
value. The last reading will be frozen on the display until the key is released. At the same time this reading is stored in the memory.
RANGE to select the measurement range (pH or mV).
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
54
SPECIFICATIONS pH 300
FUNCTIONAL DESCRIPTION pH 301
THE FRONT PANEL
Range pH 0.00 to 14.00
ISE ±399.9 mV
ORP ±1999 mV
°C -9.9 to 120.0
Resolution pH 0.01
ISE 0.1 mV
ORP 1 mV
°C 0.1
Accuracy pH ±0.01 (@20°C/68°F) ISE ±0.2 mV
ORP ±1mV
°C ±0.5
Typical EMC pH ±0.01 Deviation ISE ±0.6 mV
ORP ±1 mV
°C ±0.5
pH Calibration Automatic 1 or 2 points with
3 memorized buffers
(4.01, 7.01, 10.01)
pH Offset Max ±60mV or ±1pH pH Slope 70-108% Temperature Automatic 1 or 2 points with
Calibration 2 memorized values
(0 and 50°C)
Temperature Automatic or manual Compensation from -9.9 to 120°C (14 to 248°F)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 76 6 9/ 2 W with 1 m probe (3.3') cable (included)
Input impedance 1012 Computer RS 232 (optoisolated)
Interface Power supply 12 VDC (CSA, UL & CE approved) Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3")
Shipping Weight 1.3 kg (2.9 lb.)
ppm
pH
DATE
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
Rel mV
CFM
MEM
WRONG
CAL
LOG
OFF
CAL
DATE COMM
RANGE
CFM
TIME
TIME
ppm
pH
INTV
mVmV°C
m
1 2
BUF
BUF
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
ppm
pH
DATE
TIME
ppm
pH
INTV
°C
1
m
2
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
pH301
pH / ION METER
°C
°C
pH301
pH / ION
METER
°C
°C
76
Keyboard on the Left
CAL to start the calibration mode for pH, mV and ION. To set or quit the date or time setting mode. Used also during log­ging (to start logging press pH, mV or ION plus LOG plus CAL key; to stop logging press LOG plus CAL key).
CFM to confirm calibration values.
COMM to display the current baud rate
and command prefix for RS232.
DATE to display the date.
RANGE to display the calibration buffer
or the temperature on the secondary LCD during calibration. During setting of the date, to toggle between month, day and year. During setting of the time, to toggle between minutes, seconds and logging in­terval.
TIME to display the time.
Keyboard on the Right
!!
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not connected) and to select the calibration buffer, the date and time, the baud rate, the command prefix.
CAL DATA to display the date of last pH calibration. This also displays the offset and the slope 1 or slope 2 values ex­pressed in mV.
ION to select the ION measurement range.
mV to select the mV measurement range.
pH to select the pH measurement range.
.00/.000 to select two or three decimal
points for pH measurements.
SPECIFICATIONS pH 301
Range pH -1.999 to 19.999
mV ±1999.9
ppm 0.001 to 19999
°C -9.9 to 120.0
Resolution pH 0.01 or 0.001
mV 0.1
ppm 0.001 (from 0.001 to 9.999)
0.01 (from 10.00 to 99.99)
0.1 (from 100.0 to 999.9) 1 (from 1000 to 19999)
°C 0.1
Accuracy pH ±0.002 (@20°C/68°F)m V ±0.1 (±799.9) / ±0.2 (±1999.9)
ppm ±0.5%
°C ±0.5
Typical EMC pH ±0.01 Deviation mV ±0.3
ppm ±1%
°C ±0.5
pH Calibration Automatic 1, 2 or 3 points with
6 memorized buffers
(1.68, 4.01, 6.86, 7.01, 9.18, 10.01)
pH Offset Max ±60mV or ±1pH pH Slope 70-108% ppm Calibration Automatic 1 or 2 points with
5 memorized buffers (0.1, 1, 10, 100, 1000)
ppm Offset Max ±1999.9mV ppm Slope From ±20mV to ±100 mV
[mV/log(0.001xppm)]
Temperature Automatic 1 or 2 points with Calibration 2 memorized values (0 and 50°C)
Temperature Automatic or manual (for pH only) Compensation from -9.9 to 120°C (14 to 248°F)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 7669/2W with 1 m probe (3.3') cable (included)
Input impedance 1012 Logging Interval 1, 15, 30 seconds or
1, 5, 30, 60, 120, 180 minutes
PC Interface RS 232 (optoisolated) Power supply 12 VDC (CSA, UL & CE approved) Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3") Shipping Weight 1.3 kg (2.9 lb.)
98
FUNCTIONAL DESCRIPTION pH 302
THE FRONT PANEL
Keyboard on the Left
0 to 9 for numeric and decimal data in- put.
CLEAR to cancel wrong data entry during programming.
Rel mV
ppm
DATE TIME
Keyboard on the Right
CAL to perform the pH, mV or tempera-
MEM
CFM
WRONG
CAL
LOG
OFF
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
ture calibration.
CFM/
!!
! to confirm pH/mV/temperature
!!
calibration values or to manually set tem­perature (when the temperature probe is not connected).
ENTER/
""
"°C to terminate the numeric data
""
entry or to manually set temperature (when the temperature probe is not connected) or to select the buffer value during pH calibration.
PAPER to pull the paper.
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
8
9
7
6
5
4
1
3
2
0
CLEAR
.
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
PGM to select a program.
PRINT to obtain a printout (prints time,
date, sample number and pH or mV val­ues).
RANGE to select the measurement range (pH or mV) or to display the calibration buffer or the temperature on the second­ary LCD during calibration.
TIME to display the time and the date.
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
1110
SPECIFICATIONS pH 302
Range pH 0.00 to 14.00
ISE ±399.9 mV
ORP ±1999 mV
°C -9.9 to 120.0
Resolution pH 0.01
ISE 0.1 mV
ORP 1 mV
°C 0.1
Accuracy pH ±0.01 (@20°C/68°F) ISE ±0.2 mV
ORP ±1mV
°C ±0.5
Typical EMC p H ±0.01 Deviation ISE ±0.3 mV
ORP ±1 mV
°C ±0.5
pH Calibration Automatic 1, or 2 points with
3 memorized buffers
(4.01,7.01,10.01)
Temperature Automatic 1 or 2 points with Calibration 2 memorized values
(0 and 50°C)
Temperature Automatic or manual Compensation from -9.9 to 120°C (14 to 248°F) (for pH only)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 76 6 9/ 2 W with 1 m probe (3.3') cable (included)
Input impedance 1012 Printer Dot matrix, 44mm wide paper Languages English, German, French,
Spanish, Italian & Swedish
PC Interface RS 232 (optoisolated) Power supply 12 VDC (CSA, UL & CE approved)
Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3") Shipping Weight 1.3 kg (2.9 lb.)
THE REAR PANEL
MEASUREMENT
REFERENCE ELECTRODE
TEMPERATURE
ELECTRODE
12345 6
RESET 12VDC RS 232
PROBE
ooooo
oooo
1. Reference Electrode Connector (Ø 4 mm)
2. BNC socket for measurement electrode
3. Socket for Temperature Probe
4. Reset Button
5. DC Power Socket (for HI710005 or HI710006)
6. RS 232 Connector
LCD FUNCTIONAL DESCRIPTION
Primary Display
CFM
CAL
LOG
Rel mV
MEM
WRONG
BUF
ppm
1 2
3
mV°C
ppmpHpH
DATE TIME
INTV
m s
Secondary Display
1312
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ISE or ORP elec­trodes (with internal reference) plug the electrode's BNC to the socket provided (see page 11).
For electrode with separate reference, con­nect the measuring electrode's BNC to the BNC socket and the reference electrode's plug to the socket provided (see page 11).
For temperature measurements and auto­matic temperature compensation connect the temperature probe to the °C socket (see page 11).
If the temperature probe is not connected, when turned on, the meter will display a default tem-
°C
perature of 25°C or the last recorded tempera­ture (if the temperature probe is disconnected when the meter is on) on the secondary LCD. The "°C" indicator will be flashing.
POWER CONNECTION
Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (see page11).
Plug the adapter to the mains.
Note: The instrument uses the following con-
figuration.
+
-
It is recommendable to use the Hanna HI710005 or HI710006 voltage adapters (sup-
plied with the meters) that use the proper polarity configuration.
The meters can also be used with other adapt­ers. In this case, remember to check the correct polarity of your adapter before con­necting it to the meter.
Note: Make sure the main line is protected
by a fuse.
Press ON to turn the meter on. The power LED will remain lit until OFF is pressed to turn the meter off.
ON
OFF
When the meters are turned on, the micro­processor checks if any calibration data is in memory (i.e. if the offset is within ±60mV and the slope between 70-108%).
If the calibration data is valid, the meter will op-
pH
erate in the normal mea­surement mode and de­fault to pH measurement
°C
mode with the tempera­ture on the secondary LCD.
If the calibration data is not valid or the calibra-
pH
tion data is lost, the "CAL" indicator will be displayed to prompt the
CAL
°C
user to perform the pH calibration. If the user chooses not to per­form the calibration and presses CAL to quit from the calibration mode, the offset and the slope of the instrument would be set to 0 mV and 100% respectively.
1514
pH300, pH301 and pH302 use an EEPROM to retain the pH calibration, mV calibration, temperature calibration as well as the serial communication setting. The instrument will store the respective data after a calibration or serial communication setting, even when it is not plugged-in.
Press CAL to enter the setting mode and the month will be flashing.
CAL
DATE
For pH 302 only
pH302 uses a rechargeable battery to retain
the pH calibration data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off).
If you need to unplug the instrument, press OFF before
OFF
disconnecting the meter from the mains. Only in this case will the instrument retain the memorized data. Otherwise, the instrument needs to be re-calibrated. In the case of a power failure while the instrument is operational, re-cali­bration is required.
When the instrument is initially switched on, the microprocessor checks if any calibration data is in memory (i.e. if the offset is within ±1 pH and the slope between 70 - 108%).
If the calibration data are not memorized, the offset and slope of the instrument will be set to 0 mV and 100% respectively.
Use UP or DOWN to select the month.
°C
°C
Press RANGE and the day will start to flash.
RANGE
DATE
Use UP or DOWN to select the day.
°C
°C
Press RANGE and the year on the second­ary LCD will flash.
RANGE
DATE
SETTING THE DATE AND THE TIME AND THE LOGGING INTERVAL (for pH 301 only)
Press DATE to display the date. The month and the day will be displayed on the primary LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the year.
°C °C
1716
Press CAL to exit the date setting mode.
CAL
Press RANGE and the logging interval on the secondary LCD will come up.
RANGE
TIME
Press TIME to display the time. The hour and the minutes will be displayed on the primary LCD, the logging interval on the secondary one.
TIME
TIME
INTV
Press CAL to enter the setting mode and the hour will flash.
CAL
TIME
INTV
Use UP or DOWN to select the hour.
°C °C
Press RANGE and the minutes will flash.
RANGE
TIME
INTV
Use UP or DOWN to select the logging inter­val.
°C
°C
CAL
Press CAL to exit the time set­ting mode.
VIEWING THE DATE AND THE TIME (for pH 301 only)
Press DATE to display the date. The month and the day will be displayed on the primary LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the minutes.
°C
°C
INTV
Press TIME to display the time. The hour and the minutes will be displayed on the primary LCD, the logging interval on the secondary one.
TIME
1918
TIME
INTV
s
SETTING THE DATE AND THE TIME AND THE LOGGING INTERVAL (for pH 302 only)
See program No. 1 (page 60).
VIEWING THE DATE AND THE TIME (for pH 302 only)
Press TIME to display the time (HH/MM).
TIME
Press TIME again to display the date (MM/ DD).
TIME
DATE
Allow 1 or 2 minutes for
pH
the electrode to stabi­lize.
°C
pH is displayed on the primary display and tem­perature on the second-
pH
ary one. If readings are out of
range "---" will appear on the LCD display.
For pH 301 only:
Press the .00/.000 key to change the resolu­tion from two to three decimal points.
0.00/0.000
pH
°C
TAKING pH MEASUREMENTS
Make sure that the instrument has been cali­brated for pH before taking pH measure­ments (see page 29).
For pH 300 and pH 302 only
:
RANGE
Select the pH measurement mode by pressing RANGE.
For pH 301 only
:
pH
Select the pH measurement mode by pressing pH.
To take a pH measurement simply submerge the tip (4cm/ 1½") of the pH electrode and the temperature probe into the sample to be tested.
4 cm
(1½")
It is recommended that the electrode is rinsed thoroughly for better conditioning. For this rinsing process, it is recommended to use a liberal amount of the sample to be measured.
If measurements are taken in different samples successively, it is recommended that the elec­trode and the temperature probe are rinsed thoroughly with deionized water or distilled water (if not available, tap water may also be used) and then blot dry between each mea­surement.
The pH reading is directly effected by tem­perature. In order for the meter to measure the pH accurately, temperature must be taken into consideration.
A perfect equilibrium between the pH elec­trode and the sample is reached in approxi­mately 15 minutes.
To use the meter's Automatic Temperature
Compensation feature, submerge the HI7669/2W temperature probe into the sample
2120
as close to the electrode as possible and wait for 1-2 minutes.
If you know the temperature of the sample to be tested you can compensate for it manu­ally.
If manual temperature compensation is de­sired the temperature probe must be dis­connected from the instrument.
The display will then show the default tem­perature of 25°C or the last recorded tempera-
°C
ture reading. "°C" is blinking. Note the temperature of the
C
°
sample using a ChecktempC or a glass thermometer. The temperature can now be adjusted with UP and DOWN. If the temperature
°C °C
probe is connected to the meter, the second­ary LCD figure dis­plays the measured temperature reading and the "°C" indicator
CFM/
ENTER/
does not blink. Press­ing UP or DOWN will not alter the tempera­ture reading.
VIEWING THE pH CALIBRATION DATA For pH 300:
Press CAL DATA and the meter will display the last pH calibration data.
CAL DATA
Press UP or DOWN to get this information in the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month, day and year)
b.Offset calibration
CAL
value expressed in mV and displayed on
mV
the primary LCD with the buffer used to calibrate the offset on the secondary LCD.
CAL
pH
1
The offset pH value should be between ±30 mV. For values between ±30 and ±60 mV change the reference electrolyte and repeat the calibration procedure. For values higher than +60 mV or lower than
-60mV the electrode is no longer reliable. See the electrode maintenance section on page 84.
c.Slope calibration
mV
value directly ex­pressed in mV/pH is displayed on the pri­mary LCD with the
CAL
pH
2
2nd buffer used to calibrate the offset on the secondary LCD.
For pH 301:
Press pH followed by
pH
CAL DATA
CAL DATA and the meter will display the
last pH calibration data.
2322
Press UP or DOWN to get this information in the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month, day and year)
CAL
b. Time when the last
TIME
pH calibration was completed (hour and minutes).
CAL
c.Offset calibration
mV
value expressed in mV and displayed on the primary LCD with the buffer used
CAL
pH
1
to calibrate the off­set on the secondary LCD.
The offset pH value should be between ±30 mV. For values between ±30 and ±60 mV change the reference electrolyte and repeat the calibration procedure. For values higher than +60 mV or lower than
-60mV the electrode is no longer reliable. See the electrode maintenance section on page 84.
d. Acid slope calibra-
mV
tion value expressed in mV/pH and dis­played on the pri­mary LCD with the
CAL
pH
2
2nd buffer used to calibrate the offset on the secondary LCD.
e. Alkaline slope calibra-
tion value expressed
mV
in mV/pH and dis­played on the primary LCD with the 3 buffer used to cali-
rd
CAL
pH
3
brate the offset on the secondary LCD.
For pH 302:
See program #3 (page 66).
TAKING REDOX MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ORP measurements. An optional ORP electrode must be used to perform these measurements (see page 105).
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP), press RANGE.
For pH 301:
mV
To get into the mV mode, press mV.
To measure the mV of a so­lution simply submerge the ORP electrode tip (4 cm/1½") into the sample to be tested. Allow a few minutes for the
(1
½
")
reading to stabilize. The display will indicate the absolute mV value
(positive or negative).
For pH 300 and pH 302 only
When values are less
:
Rel mV
than ±400 mV, tenths of mV are displayed; for values greater than
°C
±400mV, the decimal digit disappears.
4 cm
2524
Note that the change in scale (from 0.1 mV to 1 mV as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value exceeds ±1999 mV (out of range).
Oxidation-reduction potential (REDOX) mea­surements provide the quantification of the oxidizing or reducing power of the sample tested.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
When not in use, the tip of the electrode should be kept moist and safe from any me­chanical stress which might cause damage to the gold/platinum junction.
TAKING ION MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ION measurements. An optional ISE electrode must be used to perform these measurements.
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP or ISE), press RANGE.
For pH 300 and pH 302:
The display will indicate the absolute mV value (positive or negative).
When values are less
Rel mV
than ±400 mV, tenths of mV are displayed; for
°C
values greater than ±400mV, the decimal digit disappears. Note that the change in scale (from 0.1 mV to 1 mV as resolution) is automatic.
For pH 301:
The display will indicate
ppm
the ppm value expressed in ppm or mV (depend-
°C
ing on the measured sample). Note that the change in scale (from 0.001ppm to 0.01ppm or
0.1 ppm as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value exceeds ±1999 mV (out of range).
When not in use, the tip of the electrode should be kept moist and safe from any mechanical stress which might cause damage to the junction.
For pH 301:
To get into the ION mode, press ION.
Simply submerge the ISE electrode tip (4 cm/1½") into the sample to be tested. Al­low a few minutes for the read­ing to stabilize.
ION
VIEWING THE ION CALIBRATION FAC­TOR ON THE LCD (for pH 301 only)
Press ION followed
ION
CAL DATA
by CAL DATA and the meter will display the last ION cali- bration data.
°C °C
(1
4 cm
½
")
Press UP or DOWN to get this information in the fol­lowing sequence:
2726
a. Date of the last ION
DATE
calibration (month, day and year)
b. Time when the last
CAL
ION calibration was completed (hour and
TIME
minutes)
c.Offset calibration
value expressed in
CAL
INTV
mV and displayed on the primary LCD
mV
with the value used to calibrate the off­set on the second­ary LCD.
d. Slope calibration
value expressed in
CAL
mV
ppm
1
mV/[log (0.0001xppm)] and displayed on the primary LCD with
CAL
ppm
2
the 2nd value used to calibrate the slope on the secondary LCD.
TAKING TEMPERATURE MEASURE­MENTS
Taking a temperature measurement is very easy.
ON
Connect the temperature probe to the instrument and turn the instrument ON.
Dip the liquid/general purpose temperature probe HI7669/2W into the sample
Allow the reading on the sec­ondary display to stabilize (1 or 2 minutes).
pH
°C
PRINTING FUNCTION (for pH 302 only)
Press PRINT to get a printout at
PRINT
any time. All the printouts will be complete
with date, time, sample No., tem­perature and measurement value (pH or mV depending on the se­lected measurement range).
DATE 1/02 TIME 12:21 SAMPLE 1 °C 17.9 pH 7.18
DATE 1/02 TIME 12:40 SAMPLE 5 °C 20.5 mV -350.0
HOLD FUNCTION (for pH 300 only)
To hold and memorize the measured value, press MS. The last reading will be frozen on the display until the key is released. At the same time, this reading is stored in the memory. During this operation, the "MEM" indicator is lit.
MS
MEM
pH
°C
2928
MR
To recall the stored memory value, press MR and the memo­rized reading will be displayed with the MEM indicator lit.
RESET BUTTON
The RESET button (see page 11) is used when the instrument displays erroneous mes­sages due to strong electrical interference or when the instrument's power supply was dis­connected before the meter was switched off.
It is necessary to press the reset button and restart the entire operation.
Calibration points should remain memorized. It is recommended to verify calibration be­fore proceeding.
pH CALIBRATION
For greatest accuracy, it is recommended that the instrument is calibrated frequently.
For a faster operation, it is possible to stan­dardize the electrode at a value close to the expected sample value or at pH 7.01 only (one point calibration), but it is always good practice to calibrate at least 2 points.
For pH 301 only
:
Due to electrode conditioning time the elec­trode must be kept immersed for a few sec­onds to stabilize. The meters are equipped with a stability indicator and the user will be guided step by step with easy indications on the display during the pH calibration. This will make the calibration a simple and error-free procedure.
The standard calibration program of the meter is prepared for 3 (maximum) buffers. The selectable pH values are: 1.68, 4.01, 6.86,
7.01, 9.18, 10.01.
INITIAL PREPARATION
Pour small quantities of
HI 7004
pH7.01 (HI7007/HI8007) and pH4.01 (HI7004/HI8004) and/
HI 7004
HI 7007
or pH10.01 (HI7010/HI8010) solution into individual beakers. If possible, use plastic beakers to minimize any EMC interferences.
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the electrode, the second one for calibration.
By doing this,
RINSE
CALIBRATION
contamination between the buffers is mini-
HI 7007
HI 7007
mized. To get accurate readings, use pH 7.01 and
3130
pH 4.01 if you are going to measure acid samples, or pH 7.01 and pH 10.01 for alka­line measurements or, for pH301 only, per­form a 3-point calibration for the entire range.
PROCEDURE
RANGE
Press RANGE (for pH300 and pH 302) to select the pH measurement mode.
Press pH (for pH301) to select the pH measure-
pH
ment mode.
Remove the protective cap from the pH electrode.
ONE POINT CALIBRATION (OFFSET)
Immerse the pH electrode into the selected buffer (e.g. pH 7.01) buffer solution and shake briefly.
Note: the electrode should
HI 7007
be submerged ap­proximately 4 cm (1½") into the solu­tion. The tempera-
(1
ture probe should be located as close as
4 cm
½
")
possible to the pH electrode.
Press CAL. The primary LCD will display the pH value using the current offset and
slope, with the "CAL" and "
CAL
1
BUF
" indicators lit.
CAL
pH
pH
1
BUF
For pH 302 only: the printer will also print step by step instructions for the calibra­tion procedure.
Dip the electro­de in the pH 7 solution; wait 30 sec and then press CFM
Select the buffer value by pressing UP or DOWN or just DOWN (for pH 302) until the secondary LCD displays the value of the first solution (e.g. 7.01”).
°C
ENTER/
°C
pH
CAL
1
BUF
Only when the read­ing is stable (less
" in-
CFM
CFM/
CFM
than ±0.03 variation in a 10 seconds time
interval) the "
1
BUF
dicator will stop flash­ing (after about 30 seconds) and the "CFM" indicator will start blinking.
Press CFM to con­firm the calibration.
If the reading is not close to the selected buffer, "WRONG
WRONG
1
BUF
"
and "WRONG " will blink alternately. If the reading is close to the selected buffer (±1pH), the meter stores the reading
WRONG
BUF
3332
CFM/
(and adjusts the offset point). The buffer value is
pH
then displayed on the primary LCD and the secondary LCD will display the ex-
CAL
2
BUF
pected second buffer value. For pH 302 only: the printer will print
further instructions for the calibration pro­cedure.
Now dip the e­lectrode in 2nd solution, wait 30 sec and then press CFM
Press CAL and the calibra­tion process is ended with just the offset of the meter calibrated. For best accuracy however, it is recommended that a two-point calibration is performed.
For pH 300 only
:
The calibration process is complete and the dis­play will ask for the date
CAL
entry.
°C
Use UP or DOWN to select the month, the day and the year.
Use CAL to toggle between month, day and year.
CAL
CAL
°C
DATE
After entering the calibration date, press CFM and the meter will return to the normal operating
pH
mode.
Note: during calibration, the
RANGE
RANGE
secondary LCD displays the selected buffer value. By pressing RANGE the tempera­ture value can be also displayed. This will allow you to check the buffer temperature during calibration.
TWO-POINT CALIBRATION
A two-point calibration is recommended for best accuracy. As a first point use pH 7.01 (or pH 6.86 can also be used for pH 301). As a second point use pH 4.01 (or pH 1.68 for pH 301 only) if you are going to measure acid samples (pH 7 or less), use pH10.01 (or pH 9.18 for pH301 only) if you are going to measure alkaline samples.
CAL
Proceed as described above for one point calibration but do not quit calibration by pressing CAL at the end.
After the first calibration point is confirmed, immerse the pH electrode into the second buffer solution (pH
4.01 or pH10.01; for pH301
HI 7004
only, the second buffer can also be pH 1.68 or pH 9.18) and shake briefly.
Note: the electrode should
be submerged ap­proximately 4 cm (1½") into the solu-
(1
4 cm
½
")
tion. The temperature probe should be located as close as possible to the pH electrode.
3534
Press UP or DOWN or just DOWN (for pH 302) until the secondary display changes the value of the second solution (e.g. 4.01).
°C
ENTER/
°C
pH
CAL
2
BUF
Only when the read­ing is stable (less than ±0.03 variation in a 10 seconds time
interval), the "
BUF
1
" in-
CFM
dicator will stop flash­ing (after about 30
CFM/
CFM
seconds) and the "CFM" indicator will blink.
Press CFM to con­firm the calibration.
If the reading is not within 70% to 108% of the buffer solution, "WRONG
BUF
" and
2
WRONG
"WRONG " will blink alternately. If the reading is close to the selected buffer (between 70 to 108%), the meter
WRONG
BUF
stores the reading (and adjusts the slope point).
For pH 300 only
:
DATE
The calibration process is complete and the dis­play will ask for the date entry.
CAL
°C
°C
Use UP or DOWN to select the month, the day and the year.
Use CAL to toggle be­tween month, day and
CAL
year. After entering the cali-
bration date, press CFM and the meter will re-
CFM/
turn to the normal oper­ating mode.
For pH 302 only
:
The calibration process is complete. The meter will print a confirm message and return to the normal operating mode.
Calibration com­pleted
For pH 301:
The second buffer value is then displayed on the primary LCD and the secondary LCD will display the expected third buffer value.
CAL
Press CAL and the calibra­tion process is ended with the offset and the 1st slope of the meter calibrated.
Note: the meter will automatically skip the
buffer that was used for the first calibra­tion to avoid erroneous calibration.
3736
RANGE
RANGE
Note: during calibration, the
secondary LCD dis­plays the selected buffer value. By press­ing RANGE the tem­perature value can be also displayed. This will allow you to check the buffer temperature during calibration.
THREE-POINT CALIBRATION (for pH301 only)
A three-point calibration (for the entire pH range) can be performed with pH301. These are the selectable buffers:
1st buffer 2nd/3rd buffer 3rd/2nd buffer
(offset) (1st slope) (2nd slope):
pH 6.86 pH 1.68 pH 9.18
or 7.01 or 4.01 or 10.01
or 9.18 or 4.01
Proceed as described for the
CAL
two-point calibration but do not quit calibration by pressing CAL at the end.
After the second calibration point is confirmed, immerse the pH electrode into the third buffer solution (e.g. pH10.01) and shake briefly.
HI 7004
HI 7010
Note: the electrode should
be submerged ap­proximately 4 cm (1½") into the solu-
(1
tion. The temperature probe should be lo-
4 cm
½
")
cated as close as possible to the
°C
°C
pH electrode.
Select the 3rd buffer value on the secondary display by pressing UP or DOWN (e.g.10.01”).
Only when the read­ing is stable (less
1
" in-
CFM
than ±0.03 variation in a 10 seconds time
interval), the "
BUF
dicator will stop flash­ing (after about 30
CFM/
seconds) and the "CFM" indicator will blink.
Press CFM to con­firm the calibration.
If the reading is not within 70% to 108% of the buffer solution, "WRONG
BUF
" and
3
WRONG
"WRONG " will blink alternately. If the reading is close to the selected buffer (between 70 and 108%), the meter
WRONG
BUF
stores the reading (and adjusts the 2 slope point).
The calibration process is ended with the offset and the 1st and 2nd slope of the meter calibrated.
Note: the meter will automatically skip the
buffer that was used for the first and second calibration to avoid erroneous calibration.
Note: during calibration, the sec-
RANGE
ondary LCD displays the selected buffer value. By pressing RANGE the tem­perature value can be also displayed. This will allow you to check the buffer temperature dur­ing calibration.
3938
nd
CALIBRATION WITH MANUAL TEMPERA­TURE COMPENSATION
If for some reason the temperature probe is defective, or it is required to calibrate with manual temperature compensation, follow the procedure below:
Unplug the temperature probe from the meter.
Note the temperature of the
°C
buffer solutions with a ChecktempC or a glass thermometer with a resolu­tion of 0.1°C.
For pH300 and pH301 only
Use UP or DOWN to
°C°C
manually adjust the display reading to the value of the reference thermometer (e.g. 20°C). The value will be displayed on the secondary LCD and "°C" symbol will flash.
For pH302 only
°C
Press the +/- key (for eventual negative figures) followed by the noted temperature on the numeric keypad (e.g. 20°C) and ENTER. The value will be displayed on the secondary LCD and "°C" symbol will be flashing.
2
0
ENTER/
°C
Follow the calibration procedure as for pH calibration with temperature probe con­nected (see page 29).
pH VALUES AT VARIOUS
TEMPERATURE
The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions.
When manually entering the calibration val­ues (see program #2 for pH302) refer to the following chart.
TEMP pH VALUES
°C °F 1.68 4.01 6.86 7.01 9.18 10.01
0
32
5
41
10
50
15
59
20
68
25
77
30
86
35
95
40
104
45
113
50
122
55
131
60
140
65
149
70
158
1.67
1.67
1.67
1.67
1.68
1.68
1.68
1.69
1.69
1.70
1.71
1.72
1.72
1.73
1.74
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75
When using automatic calibration, the appro­priate value will be displayed.
For instance, if the buffer temperature is at 25°C, the display will show "pH 4.01" or "7.01" or "10.01".
If the buffer temperature is at 20°C, the dis­play will show "pH 4.00"/"7.03"/"10.06".
If the buffer temperature is at 50°C, the dis­play will show pH "4.06"/"6.98"/"9.82".
4140
ION CALIBRATION (for pH301 only)
For greatest accuracy, it is recommended that the instrument is calibrated frequently.
The standard calibration program of the meter, however, is prepared for 2 (maximum) buff­ers. The selectable ION values are:
ppm 0.1, 1, 10, 100, 1000.
Due to electrode conditioning time the elec­trode must be kept immersed a few seconds to stabilize. The meters are equipped with a stability indicator and the user will be guided step by step with easy indications on the display during the ION calibration. This will make the calibration a simple and error-free procedure.
INITIAL PREPARATION
Pour small quantities of the chosen ION calibration solu­tion into individual beakers. If possible, use plastic beakers to minimize any EMC inter­ferences.
For accurate calibration use two beakers for each solu­tion, the first one for rinsing the electrode, the second one for calibration. By doing this, contamination between the buff­ers is minimized.
PROCEDURE
ION
Press ION to select the ION measurement mode.
Remove the protective cap from the ISE electrode.
ONE POINT CALIBRATION (OFFSET)
Immerse the ISE electrode into the se­lected buffer (e.g. 10 ppm) solution and shake briefly.
Note: the electrode should
be submerged ap­proximately 4 cm (1½") into the solu­tion.
Press CAL. The primary LCD will display the ION value using the current offset and
slope, with the "CAL" and "
CAL
Select the buffer value
CAL
1
BUF
" indicators.
ppm
ppm
1
BUF
°C
by pressing UP or DOWN until the sec­ondary LCD displays the value of the first solution (e.g. 10”).
Only when the read­ing is stable, will the
1
BUF
"
" indicator stop
CFM
flashing (after about 30 seconds) and the "CFM" the indicator start blinking.
CFM/
Press CFM to confirm the cali­bration. The meter stores the reading (and adjusts the offset point). The calibration solution value is then displayed on the primary LCD
ppm
and the secondary LCD will display the ex­pected 2nd calibration value.
CAL
ppm
2
BUF
(1
4 cm
½
")
°C
4342
Press CAL and the calibra-
CAL
tion process is ended with only the offset of the meter calibrated. For best accuracy however, it is recommended that a two-point calibration is performed.
Note: during calibration, the sec-
ondary LCD displays the
RANGE
selected buffer value. By pressing RANGE the tem­perature value can be also displayed. This will allow you to check the buffer tem­perature during calibration.
TWO-POINT CALIBRATION
A two-point calibration is recommended for best accuracy.
Proceed as described above for one point calibration but do not quit calibration by pressing CAL at the end.
After the first calibration point is confirmed, immerse the ISE electrode into the sec­ond calibration solution (e.g. 100 ppm) and shake briefly.
Note: the electrode should
be submerged ap­proximately 4 cm (1½") into the solu-
(1
½
")
tion.
Press UP or DOWN until the secondary display changes the value of the second solution (e.g. 100”).
4 cm
Only when the read­ing is stable, the "
indicator will stop
1
BUF
"
CFM
flashing (after about 30 seconds) and the "CFM" indicator will blink. The meter will check for the slope reading.
CFM/
Press CFM to confirm the calibration.
The meter stores the read­ing (and adjusts the slope point).
The ION calibration process is complete and the meter will return to the normal operating mode.
Note: the meter will automatically skip the buffer
that was used for the first calibration to avoid erroneous calibration.
Note: during calibration, the sec-
ondary LCD displays the
RANGE
selected calibration value. By pressing RANGE the temperature value can be also displayed. This will al­low you to check the solu­tion temperature during cali­bration.
°C
°C
ppm
CAL
2
BUF
4544
mV CALIBRATION
The meter has been factory calibrated for the mV and is ready for measurements. The cali­bration data are stored in the EEPROM.
The ORP electrodes are interchangeable and no mV calibration is needed when the ORP electrode is replaced.
Set the simulator to "390.0mV".
Wait for 5 seconds and press CFM to confirm the value.
The primary LCD will display "-390.0 mV".
CAL
CFM
mV
°C
If, for any reason, the mV measurements are out of accuracy, re-calibration should be car­ried out as follows.
INITIAL PREPARATION
Prepare a simulator with an accuracy of ±0.1mV that can provide:
±390.0mV and ±1000.0mV for pH300
0.0 mV, ±500.0 mV and ±1500.0 mV for
pH301 and pH302 .
PROCEDURE FOR pH 300
Connect the meter to the simulator through the connection cable (connect it to the BNC electrode socket).
Turn the pH300 on by pressing ON plus CFM simultaneously.
ON
CFM
Note: this calibration can also be per-
formed immediately after the tem­perature calibration (see page 50).
The primary LCD will display
mV
"390.0 mV".
°C
CAL
Set the simulator to
CFM
"-390.0mV". Wait for 5 seconds and press
CFM to confirm the value.
mV
The primary LCD will display "1000mV".
CAL
°C
Set the simulator to
"1000.0mV".
CFM
Wait for 5 seconds and press CFM to confirm the value.
mV
The primary LCD will display "-1000mV".
CAL
°C
Set the simulator to
"-1000.0mV".
CFM
Wait for 5 seconds and press CFM to confirm the value.
The mV calibration process is complete and the meter will return to the normal operating mode.
4746
CFM
CFM
PROCEDURE FOR pH 301
Connect the meter to the simulator through the connection cable (connect it to the BNC electrode socket).
Turn the pH301 on by pressing ON plus CAL simultaneously.
ON
CAL
The primary LCD will
mV
display "0.0 mV".
Set the simulator to "0.0mV".
CAL
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM
The primary LCD will
mV
display "0.0 mV".
Set the simulator to "0.0mV".
CAL
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM
The primary
CFM
mV
LCD will display "500 .0 mV".
Set the simulator to "500.0mV".
CAL
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM
The primary
mV
LCD will display "1500.0 mV".
Set the simulator
CAL
to "1500.0mV".
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM
The mV calibration process is complete and the meter will return to the normal operating mode.
PROCEDURE FOR pH 302
Connect the meter to the simulator through the connection cable (connect it to the BNC electrode socket).
Turn the pH302 on by pressing ON plus CFM simultaneously.
ON
CFM/
CFM
4948
CFM
mV
The primary LCD will display "0.0 mV",
Set the simulator to
CAL
"0.0mV".
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM/
CFM
mV
The primary LCD will display "500.0 mV".
Set the simulator to
CAL
"500.0mV".
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM/
The primary
mV
LCD will display "1500.0 mV".
Set the simulator
CAL
to "1500.0mV".
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM/
The mV calibration process is complete and the meter will return to the normal operating mode.
CFM
The primary LCD will
mV
display "0.0 mV".
Set the simulator to "0.0mV".
CAL
As soon as the reading is stable (after about 10 seconds), the "CFM" indicator will start blinking. Press CFM to confirm the value.
CFM/
CFM
5150
ON
TEMPERATURE CALIBRATION (for Technical Personnel only)
The pH meter has been factory calibrated for the temperature and is ready for measure­ments.
The temperature probes are interchangeable and no temperature calibration is needed when the temperature probe is replaced.
If, for any reason, the temperature measure­ments are inaccurate, re-calibration should be carried out as follows.
Wait for about 30 seconds. Press CFM. The secondary LCD section will show "50.0°C".
CFM
°C
CAL
°
C
Immerse the temperature probe in the vessel with hot water.
50 °C
(122 °F)
INITIAL PREPARATION
Prepare a vessel con-
°C
°C
taining ice (at
0.0°C/32°F) and water and another one
0 °C
(32 °F)
50 ºC
(122 ºF)
containing hot water (at a temperature of 50.0°C/122°F). Place insulation material around the con­tainer to minimize temperature changes.
Use a ChecktempC or a calibrated ther- mometer with a resolution of 0.1 °C as a reference thermometer.
PROCEDURE FOR pH 300
Switch the pH300 on while pressing CAL. The "CAL" indicator will be lit. The sec­ondary LCD section will show "0.0°C".
ON
Immerse the temperature
CAL
CAL
°C
°
C
probe in the vessel with the ice and water.
0 °C
(32 °F)
CFM
Wait for about 30 seconds. Press CFM.
Complete your temperature
CAL
calibration by pressing CAL if you do not want to per­form the mV calibration (see page44).
PROCEDURE FOR pH 301
Switch the pH301 on while pressing CFM. The "CAL" indicator will be lit. The sec­ondary LCD section will show "0.0°C".
CFM
°C
CAL
°
Immerse the temperature
C
probe in the vessel with the ice and water and wait for a few minutes.
5352
0 °C
(32 °F)
ON
The secondary LCD will display "0.0 °C" with the "CAL" indi­cator.
As soon as the reading is stable the "CFM" indicator will start blinking.
Press the CFM key to confirm the cali­bration.
Immerse the tem­perature probe in the vessel with hot water and wait for a few minutes.
The secondary LCD will display "50.0 °C" with the "CAL" indica­tor.
As soon as the read­ing is stable the "CFM" indicator will start blinking.
CAL
CFM
CFM
CAL
50 °C
(122 °F)
°C
CFM
PROCEDURE FOR pH 302
Switch the pH302 on while pressing CAL. The "CAL" indicator will be lit. The sec­ondary LCD section will show "0.0°C".
CAL
°C
CAL
°
Immerse the tem-
C
perature probe in the vessel with the ice and water and wait for a few minutes.
°
C
As soon as the reading is stable the "CFM" indica-
CFM
0 °C
(32 °F)
tor will start blink­ing.
Press CFM to con­firm the calibration.
°C
Immerse the tem-
CFM/
°
C
perature probe in the vessel with hot water and wait for a few minutes.
50 °C
(122 °F)
CFM
Press CFM to con­firm the calibration.
The temperature calibration is complete and the meter will return to the normal operating mode.
The secondary LCD will display "50.0 °C" with the "CAL" indi­cator.
As soon as the read­ing is stable the "CFM" indicator will start blinking.
5554
°C
CAL
CFM
CFM/
Press CFM to confirm the cali­bration.
The temperature calibration is complete and the meter will return to the normal operating mode.
LOGGING WITH pH301
This function permits the automatic logging of pH or mV or ION measurements together with the temperature without needing an op­erator and for a long period of time.
The lot number goes from 1 to 99 and then back to #1.
The maximum capacity per lot is 8000 samples.
Set the appropriate logging interval (see page 14) selectable between 1, 15, 30 sec­onds or 1, 5, 30, 60, 120, 180 minutes.
Select the appropriate measurement mode by pressing the pH or mV or ION key.
pH mV ION
Press LOG and then CAL to enter the log­ging mode.
LOG
CAL
Once in the logging mode, the interval can­not be changed.
Exit the logging mode
LOG
CAL
first (by pressing LOG and then CAL) before setting a new interval.
During logging, you can check some infor­mation about the logged data.
5756
ON
Press LOG and the primary LCD will show the current lot number and the secondary LCD will display the current page number.
LOG
LOG
Press LOG again to display the current sample number (the number of values that have been taken in the current lot).
LOG
LOG
TO STOP LOGGING
To stop logging press
LOG
CAL
LOG and then CAL.
when this happens the LCD will show page "0", indicating the overwrit­ing has happened.
Each time a new logging period starts, it automatically starts a new page.
When the samples collected for a single lot are more than the limit (8000 samples) the meter will automatically stop logging.
TO CLEAR LOGGED DATA
The entire logged data can be cleared by pressing LOG as soon as the meter is turned on (i.e. by pressing ON and LOG simultaneously).
LOG
CFM
BUF
The choice has to be confirmed
CFM
by pressing CFM.
Press LOG to display the next lot number.
LOG
Note: if OFF is pressed while
OFF
logging, the meter will stop the logging and then turn off.
MEMORY ORGANIZATION
The memory used for storing the logged data is divided into 16 pages. The capacity of each page is 500 samples. It starts to log from page 16 downwards until 1 and then 16 again, overwriting the previous data. However,
The next logging will start from page 16.
5958
PROGRAMMING FUNCTIONS
WITH pH302
There are 9 programs all together in the instrument. To select a particular program, press PGM followed by the program number on the numeric keyboard.
E.g. To select program 1
PGM
press PGM
1
press 1
By pressing PGM the secondary LCD will show "P" to confirm the programming mode.
By pressing 1, the secondary LCD will show "P1" to confirm the selected program.
At this stage, the secondary LCD shows "Px - y", where x refers to the program number you pressed and y refers to 1 ENTER or 2 ENTER, i.e.:
1
ENTER/
choice #1 selected by pressing 1 fol­lowed by ENTER
or
2
ENTER/
choice #2 selected by pressing 2 fol­lowed by ENTER.
Whenever a numeric value is entered, the entry appears on the primary display.
There will be a long beep after either 1+ EN­TER or 2 +ENTER is pressed, when the en­tered value is either wrong or not acceptable by the particular program.
The 9 programs stored in the instrument are:
Program #0
To end the program function.
Program #1
To set the sample number, the time and the date (see page 60).
Program #2
To calibrate the meter using a nonstandard calibration buffer value (see page 64).
Program #3
To print the pH calibration data complete with date, time, pH offset and slope (see page66).
Program #4
To log and print the pH/mV/temperature val­ues at the selected interval from the start to the finishing time. The printout interval can be selected from 1 minute to 24 hours. This program can be stopped by selecting any other PGM function (see page 67).
Program #5
To log and print the pH/mV/temperature val­ues at a user-selectable interval from a given time up to a selectable pH or mV value. When this pH/mV value is exceeded, the program stops by printing the endpoint hour and minute. The printout interval can be se­lected from 1 minute to 24 hours. This pro­gram can also be stopped by selecting any other PGM function (see page 71).
Program #6
To set the upper and the lower pH or mV alarm limits. A beeper will sound whenever the measured value exceeds the limit. This program can only be terminated by pressing PGM + 0 (see page 75).
6160
Program #7
To select a different printing language (6 languages are available) (see page 78).
Program #8
To print the ion concentration in M for the measured pH value (see page 79).
Program #9
To set the baud rate and the STRINGII code of Escape character for the RS232 (see page 79).
To stop or exit any program:
press PGM
press 0
PGM
0
PROGRAM 1 1 set sample no. 2 set time 3 set date
To set the sample number (choice #1), press 1 followed by ENTER. "P1.1" will be displayed on the secondary LCD with "0" on the primary one.
1
ENTER/
Enter the desired number (choose any number from 1 to 999) followed by EN­TER. E.g. sample number 200.
Note: the instrument must be in pH mode for
program #2, #3 and #8. Otherwise a long beep will sound.
PROGRAM #1 (pH 302)
This program allows you to set sample num­ber, time and date.
To enter program #1:
Push the PGM key and then 1.
PGM
1
The printer will then define the three func­tions:
1 is the sample number, 2 is time, 3 is the date.
2
0
0
ENTER/
The printer will print the value.
PROGRAM 1 1 set sample no. 2 set time 3 set date
Sample number 200
Samples after 999 will be reset to zero and the sample number will automatically advance from 0 to 999 in increment of 1 each time the instrument logs and/or prints a reading.
6362
"P1." will be displayed on the secondary LCD.
To set the time press 2 followed by EN­TER. "P1.2" will be displayed on the sec­ondary LCD with "0" on the primary one.
2
ENTER/
The time value (choice #2) can be set for a 24 hour clock. Enter the desired num­ber also using the decimal point followed by ENTER.
E.g. 2 P.M. equals 14.00. Press "14.00" followed by ENTER.
1
4
.
0
0
ENTER/
The printer will print the time.
PROGRAM 1 1 set sample no. 2 set time 3 set date
Sample number
200
To set the date press 3 followed by EN­TER. "P1.3" will be displayed on the sec­ondary LCD with "0" on the primary one.
3
ENTER/
When setting the date (choice #3), enter the month first. Separate month and day by a decimal point.
E.g. February 4th is entered as "2.04" followed by ENTER.
2
.
0
4
ENTER/
The printer will print the date.
PROGRAM 1 1 set sample no. 2 set time 3 set date
Sample number
200
Time
14.00
Date
2.04
Time
14.00
"P1." will be displayed on the secondary LCD.
Press PGM and then 0 to exit from the program mode.
PGM
6564
0
PROGRAM #2 (pH 302)
This program lets you calibrate with buffer values other than 4.01, 7.01 and 10.01 pH which are pre-programmed into memory.
To enter program #2, press PGM and then 2.
PGM
2
Check the appropriate value for the buffer at the measured buffer temperature (see page39).
After about 30 seconds enter the 1st buffer value at the working temperature followed by ENTER. E.g. 6.92 (at 10°C).
6
.
9
2
ENTER/
The printer will then define the two func­tions:
1 is the first buffer, 2 is the second buffer.
PROGRAM 2 1 enter 1st buf­ fer 2 enter 2nd buf­ fer
To meter will default to the 1
st
buffer value entry displaying "P2.1" on the secondary LCD with "0" on the primary one.
The buffers can be anywhere between 0 and 14 pH.
Example:
To calibrate with NBS buffers 6.86 pH and
9.18 pH at a temperature of 10°C. The corresponding pH values at this tem-
perature are 6.92 and 9.33 respectively.
Immerse the pH electrode and the temperature probe into the 1st buffer (e.g. pH6.86).
The printer will print the 1st buffer value.
PROGRAM 2 1 enter 1st buf­ fer 2 enter 2nd buf­ fer
1st buffer
6.92
The secondary display will show "P2."
To set the second buffer press 2 followed by ENTER. "P2.2" will be displayed on the secondary LCD with "0" on the pri­mary one.
2
ENTER/
Rinse the electrode and immerse it in the 2
nd
buffer (e.g. pH 9.18).
Check the appropriate value for the buffer at the measured buffer temperature (see page39).
6766
After about 30 seconds enter the 2nd buffer value at the working temperature and then press ENTER. E.g. 9.33 (at 10°C).
9
. 3
3
ENTER/
The printer will print the 2nd buffer value.
PROGRAM 2 1 enter 1st buf­ fer 2 enter 2nd buf­ fer
1st buffer
6.92
2nd buffer
9.33
Press PGM to quit this pro-
PGM
gram.
The calibration is now com­plete.
PROGRAM #3 (pH 302)
This program prints the date, time, the offset and slope characteristics of the pH electrode.
Note: the offset and the slope characteristics
of the electrode are referred to the last
performed pH calibration. No programming is required. Simply press PGM and 3.
PGM
3
The printer will print the pH calibration data.
PROGRAM 3 DATE 2/04 TIME 14:04 OFFSET mV 0.16 SLOPE 95.3%
MM/DD HH:MM in mV in %
The program function ends. Changes in an electrode's performance can
be determined from this information. The offset pH value should be between ±30mV. For values between ±30 and ±60 mV change
the reference electrolyte and repeat the cali­bration procedure. For values higher than +60 mV or lower than -60 mV the electrode is no longer reliable. See the electrode mainte­nance section on page 84.
The slope characteristic is in percentage of the theoretical value of 58.16 mV per pH unit at 20°C.
PROGRAM #4 (pH 302)
This program logs and prints pH/mV and temperature values at a preset interval be­tween 1 minute and 24 hours.
To enter the program press PGM and 4.
PGM
4
The printout defines functions 1, 2 and 3:
1 is a start time, 2 is the time interval, 3 is the finish time.
6968
PROGRAM 4
1 enter start time 2 enter time int­ erval 3 enter end time
To set the start time (choice #1), press 1 followed by ENTER. "P4.1" will be dis­played on the secondary LCD with "0" on the primary one.
1
ENTER/
Enter the desired number followed by EN­TER. E.g. Starting time 10:20.
1
0
.
2
ENTER/
0
"P4." will be displayed on the secondary LCD.
To set the time interval (choice #2), press 2 followed by ENTER.
"P4.2" will be displayed on the second­ary LCD with "0" on the primary one.
2
ENTER/
The minimum time interval is 1 minute. The interval is expressed in HH:MM.
E.g printing interval of 5 minutes.
0
0
. 0
ENTER/
5
The printer will print the value.
Hours and minutes must be separated by a decimal point.
The printer will print the value.
PROGRAM 4
1 enter start time 2 enter time int­ erval 3 enter end time
start time
10:20
PROGRAM 4
1 enter start time 2 enter time int­ erval 3 enter end time
start time
10:20
time interval
00:05
"P4." will be displayed on the secondary LCD.
7170
To set the finishing time (choice #3), press 3 followed by ENTER.
"P4.3" will be displayed on the second­ary LCD with "0" on the primary one.
3
ENTER/
DATE 2/04 TIME 10.20 SAMPLE 202 °C 20.3 pH 6.38
DATE 2/04 TIME 10.25 SAMPLE 203 °C 20.4 pH 6.39
Enter the desired number followed by EN­TER. E.g. Finishing time 11:30.
Hours and minutes must be separated by a decimal point.
1
1
. 3
ENTER/
0
The printer will print the value.
PROGRAM 4
1 enter start time 2 enter time int­ erval 3 enter end time
start time
10:20
time interval
00:05
finishing time
11:30
DATE 2/04 TIME 10.30 SAMPLE 204 °C 20.3 pH 6.38
The "LOG" indicator is displayed when the meter starts to log.
LOG
When the instrument is in this mode and
PGM
0
you want to stop be­fore the end time, press PGM plus 0.
PROGRAM #5 (pH 302)
This is the same as program #4 except that function "3" allows you to select an interrupt value for pH/mV instead of end time.
When this interrupt value is exceeded, the program prints the hour and minute of the endpoint.
To enter the program press PGM plus 5.
PGM
5
7372
2
1
The printout defines functions 1, 2 and 3:
1 is a start time, 2 is the time interval, 3 is the interrupt value of pH/mV.
PROGRAM 5 1 enter start
time 2 enter time in­ terval 3 enter end pH/mV value
"P5." will be displayed on the secondary LCD.
To set the time interval (choice #2), press 2 followed by ENTER.
"P5.2" will be displayed on the second­ary LCD with "0" on the primary one.
ENTER/
To set the start time (choice #1), press 1 followed by ENTER. "P5.1" will be dis­played on the secondary LCD with "0" on the primary one.
1
ENTER/
Enter the desired number followed by EN­TER. E.g. Starting time 10:30.
Hours and minutes must be separated by a decimal point.
ENTER/
1
0
.
3
0
The printer will print the value.
PROGRAM 5 1 enter start
time 2 enter time in­ terval 3 enter end pH/mV value
Start time
10.30
The minimum time interval is 1 minute. The interval is expressed in HH:MM.
E.g printing interval of 1 hour and 30 minutes.
.
3
ENTER/
The printer will print the value.
PROGRAM 5
1 enter start time 2 enter time in­ terval 3 enter end pH/mV value
Start time
10.30
Time Interval
1.30
"P5." will be displayed on the secondary LCD.
7574
To set the end pH/mV value (choice #3), press 3 followed by ENTER. "P5.3" will be displayed on the secondary LCD with "0" on the primary one.
3
ENTER/
Enter the desired number followed by EN­TER. E.g. pH 3.50 (supposed that the meter is in pH mode).
Note: the value will automatically be pH or
mV depending on the selected mea­surement range.
3
.
5
ENTER/
DATE 2/04 TIME 10.30 SAMPLE 210 °C 20.3 pH 4.50
DATE 2/04 TIME 12.00 SAMPLE 211 °C 20.4 pH 4.01
DATE 2/04 TIME 12.25 SAMPLE 212 °C 20.3 pH 3.50
The "LOG" indicator is displayed when the meter starts to log.
When the instrument is in this mode and it is desired
PGM
to stop before the end time, press PGM plus 0.
LOG
0
The printer will print the value.
PROGRAM 5
1 enter start time 2 enter time in­ terval 3 enter end pH/mV value
Start time
10.30
Time Interval
1.30
End point
3.50
PROGRAM #6 (pH 302)
This program allows you to set upper and lower limits for pH/mV values.
A beeper will sound whenever the measured value exceeds the limit.
To enter the program press PGM plus 6.
PGM
6
The printout defines functions 1, 2 and 3:
1 is the lower limit, 2 is the upper limit, 3 clears the alarm.
7776
7
PROGRAM 6 1 set lower limit 2 set upper limit 3 clear alarm
To set the lower limit (choice #1), press 1 followed by ENTER. "P6.1" will be dis­played on the secondary LCD with "0" on the primary one.
"P6." will be displayed on the secondary LCD.
To set the upper limit (choice #2), press 2 followed by ENTER. "P6.2" will be dis­played on the secondary LCD with "0" on the primary one.
2
ENTER/
1
ENTER/
Enter the desired number followed by EN­TER. E.g. lower limit pH 6.50 (supposed that the meter is in pH mode).
Note: the value will automatically be pH or
mV depending on the selected mea­surement range.
6
.
5
ENTER/
The printer will print the value.
PROGRAM 6
1 set lower limit
2 set upper limit
Enter the desired number followed by EN­TER. E.g. upper limit pH 7.50.
.
5
ENTER/
The printer will print the value.
PROGRAM 6
1 set lower limit
2 set upper limit
3 clear alarm
lower limit
6.50
upper limit
7.50
3 clear alarm
To stop the beeper from sound-
3
ing when the limit is exceeded,
lower limit
6.50
press 3.
7978
PGM
0
To definitely exit the program press PGM plus 0.
PROGRAM #7 (pH 302)
This program allows the user to select a different working language.
Press PGM and then 7.
PGM
7
The printer will print out a list of languages available.
PROGRAM 7
1. English
2. Deutsch
3. Français
4. Español
5. Italiano
6. Svenska
Press PRINT and the printout will be in French.
PRINT
DATE (M/J) 2/04 HEURE 10.20 ECHANTILLON 215 °C 20.0 pH 6.52
PROGRAM #8 (pH 302)
This program prints out the ion concentration in M for the pH value measured.
To activate the program, press PGM and then 8.
PGM
PROGRAM 8
3.88 pH--
0.132mM H+
8
The default working language is English. To choose a language, key in the corre-
sponding number then press ENTER. E.g. To select French press 3 and ENTER.
PROGRAM 7
3
ENTER/
1. English
2. Deutsch
3. Français
4. Español
5. Italiano
6. Svenska
Français
Note: Make sure the meter is in pH mode,
otherwise a long beep will be heard.
PROGRAM #9 (pH 302)
This program is to set the baud rate and the STRINGII code of Escape character.
To enter the program press PGM plus 9.
PGM
9
8180
The printout defines functions 1 and 2:
1 is the baud rate, 2 to set the STRINGII code of Escape char-
acter
PROGRAM 9 1 enter baud rate 2 enter STRINGII code
of escape character
"P9." will be displayed on the secondary LCD.
To set the command prefix (choice #2), press 2 followed by ENTER. "P9.2" will be displayed on the secondary LCD with "0" on the primary one.
2
ENTER/
To set the baud rate (choice #1), press 1 followed by ENTER. "P9.1" will be dis­played on the secondary LCD with "0" on the primary one.
1
ENTER/
Enter the desired number followed by EN­TER. E.g. 4800 as baud rate.
4
8
0
0
ENTER/
The printer will print the value.
PROGRAM 9
1 enter baud rate 2 enter STRINGII code of escape character
baud rate
4800
Enter the desired number followed by EN­TER. E.g. STRINGII code 25.
5
ENTER/
The printer will print the value.
PROGRAM 9
1 enter baud rate
2 enter STRINGII code of escape character
baud rate
4800
STRINGII code
25
Press PGM to quit this program.
PGM
8382
INTERFACE WITH PC
TROUBLESHOOTING GUIDE
Data transmission from the instrument to the PC is now much easier with the new HI 92000 Windows® compatible application software of­fered by Hanna Instruments.
User friendly, HI 92000 offers a variety of features and has an on-line help feature to support you throughout all situations.
HI92000 allows you to use the powerful means of the most diffused spread sheet programs (e.g. Excel©, Lotus 1-2-3©). Simply run your favorite spread sheet and open the file down­loaded by HI92000. It is then possible to make any elaboration available with your soft­ware (e.g. graphics, statistical analysis).
To install HI 92000 you need a 3.5" drive and a few minutes to follow the instructions con­veniently printed on the disk label.
Contact your Hanna Dealer to request a copy. To connect your pH meter to the PC use
HI 920010, available through your Hanna Dealer. Make sure that your meter is switched off and plug the connectors, one into the meter RS232 connector, the other into the serial port of your PC.
Note: Cables which are different from the
HI920010 may use a different con-
figuration. In which case any commu­nication between the meter and the PC is not possible.
Excel© Copyright of "Microsoft Co." Lotus 1-2-3© Copyright of "Lotus Co." Windows® and Windows Terminal® are registered Trademark of "Microsoft Co."
Symptoms Problem Solution
The meter does not work with the temperature probe The meter fails to calibrate or gives faulty readings
The meter is slow in responding or gives faulty readouts
The meter does not accept the 2nd buffer solution for calibration
The reading drifts
Display shows:
pH
Display shows:
°C
Display shows:
Rel mV
Display shows: "WRONG
and/or "WRONG "
Display shows: "WRONG
and/or "WRONG "
1
BUF
BUF
2
"
"
Defective ATC probe
Defective pH electrode
The electrode is not working or the reference junction is clogged
Defective pH electrode
Defective pH electrode
Out of range-pH scale
Out of range-°C scale
Out of range-mV scale or ION for
pH301
Erroneous buffer solution used for offset cal.
Defective electrode
Erroneous buffer solution used for slope cal.
Defective electrode
8584
Replace the probe
Replace the electrode
Leave the electrode in a storage solution after cleaning the junction. If problem persists, replace the electrode
Follow the cleaning procedure. If this doesn't work replace the electrode
Replace the electrode
a) Check the calibration b) Assure the pH sample
is in the range 0 to 14
c) Check the level of
electrolyte and the state of the electrode itself
Make sure the °C is within 0-100° and temperature probe is plugged
Electrode not connected
Make sure the buffer solution used corresponds to the selected buffer value and replace if necessary
Replace the electrode
Use fresh buffer solution
Replace the electrode
ELECTRODE CONDITIONING
AND MAINTENANCE
Reference Filling Hole
Fill Hole Screw
Sensitive Wire
Reference Wire
Reference Junction
Glass Bulb
Plastic Body
pH Electrode
Reference Wire
Reference Junction
Platinum or Gold tip
Plastic Body
ORP Electrode
Glass Body
pH Electrode
Reference
Wire
Glass Body
ORP Electrode
Reference Filling Hole
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Reference
Junction
Platinum or
Gold tip
ing down" the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI70300 or HI 80300 Storage Solution for at least one hour.
For refillable electrodes:
If the filling solution (electrolyte) is more than 1 cm (½") below the fill hole, add HI7082
or HI8082 3,5M KCl Electrolyte Solution for double junction or HI7071 or HI8071 3,5M KCl+AgCl Electrolyte Solution for
single junction electrodes. For a faster response, unscrew the fill hole
screw during measurements.
For AmpHel® electrodes:
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (4 cm /1½") in the sample
and stir gently for at least 30 seconds. For a faster response and to avoid cross
contamination of the samples, rinse the elec­trode tip with a few drops of the solution to be tested, before taking measurements.
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT. This is normal with electrodes and they will
disappear when rinsed with water. During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode cannot function properly under these condi­tions. The bubbles can be removed by "shak-
STORAGE
To minimize clogging and assure a quick response time, the glass bulb and the junc­tion should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI70300 or HI80300
Storage Solution or, in its absence, Filling Solution (HI7071 or HI8071 for single junc-
tion or HI7082 or HI8082 for double junction
AmpHel® is registered Trademark of "Hanna Instruments"
8786
electrodes). Follow the Preparation Proce­dure above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN
DISTILLED WATER OR DRY.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for the connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the elec­trode stem or bulb. Connectors must be per­fectly clean and dry.
If any scratches or cracks are present, re­place the electrode.
Rinse off any salt deposits with water.
For refillable electrodes:
Refill the reference chamber with fresh elec­trolyte (HI7071 or HI8071 for single junction or HI7082 or HI8082 for double junction electrodes). Allow the electrode to stand up­right for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna HI7061 or
HI 8061 General Cleaning So­lution for approximately 1 hour.
Removal of film, dirt or deposits on the mem­brane/junction:
-
Protein
-
Inorganic
-
Oil/grease
Soak in Hanna HI7073 or
HI8073 Protein Cleaning Solu­tion for 15 minutes.
Soak in Hanna HI7074 or HI8074 Inorganic Cleaning So­lution for 15 minutes.
Rinse with Hanna HI7077 or HI 8077 Oil and Fat Cleaning Solution.
IMPORTANT: After performing any of the
cleaning procedures rinse the electrode thor­oughly with distilled water, refill the reference chamber with fresh electrolyte, (not neces­sary for GEL filled electrodes) and soak the electrode in HI 70300 or HI80300 Storage Solution for at least 1 hour before taking measurements.
TROUBLESHOOTING
Evaluate your electrode performance based on the following.
Noise (Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: Refer to the
Cleaning Procedure above.
- Loss of shielding due to low electrolyte
level (in refillable electrodes only): refill with fresh HI 7071 or HI 8071 for single junction or HI7082 or HI8082 for double junction electrodes.
Dry Membrane/Junction: Soak in Stor- age Solution HI70300 or HI80300 for at
least 1 hour.
Drifting: Soak the electrode tip in warm Hanna Solution HI7082 or HI8082 for one hour and rinse tip with distilled water. Re­fill with fresh HI7071 or HI8071 for single junction electrodes and HI7082 or HI8082 for double junction electrodes.
Low Slope: Refer to the cleaning proce- dure above.
No Slope: Check the electrode for cracks in glass stem or bulb and replace the electrode.
Slow Response/Excessive Drift: Soak the tip in Hanna Solution HI7061 or HI8061 for 30 minutes, rinse thoroughly in dis­tilled water and then follow the Cleaning Procedure above.
8988
TEMPERATURE-RESISTANCE CORRE-
0
0+10+20+30+40+50+60+70+80+90
C
LATION FOR HANNA pH SENSITIVE
GLASS
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes longer for the reading to stabilize if the resis­tance is higher. In addition, the response time will suffer to a greater degree at tem-
9
2x10
-10
9
1x10
-10
8
-10
2x10
8
1x10
-10
7
2x10
-10
-10
7
1x10
-10
-10
-20-2
peratures below 10°C. Since the resistance of the pH electrode is in
the range of 200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for hours.
Typical Electrode Life Ambient Temperature 1- 3 years 90 °C Less than 4 months 120°C Less than 1 month
High concentrations of sodium ions interfere with readings in alkaline solutions; the pH at which the interference starts to be significant depends upon the composition of the glass. This interference is the alkaline error and causes the pH to be underestimated. Hanna's glass formulations have the indicated char­acteristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
°
Concentration pH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
13.00
13.50
14.00
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
For these reasons high humidity environ-
ments, short circuits and static discharges
are detrimental for a stable pH reading.
The life of a pH electrode also depends on temperature. If constantly used at high tem­peratures, electrode life is drastically reduced.
9190
ADDITIONAL INFORMATION ABOUT
INTERFACE WITH PC
(for technical personnel only)
If you are not using Hanna Instruments HI92000 application software, please find here below some additional information to help your connection to the PC.
SETTING THE BAUD RATE AND THE COMMAND PREFIX
The transmission speed (baud rate) of your pH meter and of the external device must be the same.
Procedure for pH 300:
To set the baud rate of the meter press COMM, the primary LCD shows the current
baud rate.
COMM
Note: the Command Prefix has not to be
changed using HI92000 Hanna Soft­ware.
°C
°C
Select a different com­mand prefix (between 0 to 47 in decimal) by pressing UP and DOWN.
Press CFM to confirm the set-
CFM
ting.
Procedure for pH 301:
To set the baud rate of the meter press COMM, the primary LCD shows the current
baud rate.
COMM
The following baud rate can be selected
°C
°C
through UP and DOWN:
150, 300, 600, 1200 (factory setting) and
2400. Press CFM to confirm the setting and the
primary LCD shows the current command
prefix: 16 is the factory setting.
CFM
The following baud rate
°C
°C
can be selected through UP and DOWN:
150, 300, 600, 1200 (factory setting), 2400, 4800, 9600.
Press COMM to confirm the setting and the primary LCD shows the current command
prefix: 16 is the factory setting.
COMM
Note: the Command Prefix has not to be
changed using HI92000 Hanna Soft­ware.
9392
°C
°C
Select a different com­mand prefix (between 0 to 47 in decimal) by pressing UP and DOWN.
COMM
Press COMM to confirm the set­ting.
Procedure for pH 302:
See program #9 (page 79).
SENDING COMMANDS FROM PC
With terminal programs such as, for example, Telix®, Windows Terminal®, it is possible to remotely control your pH meter. Use HI920010 cable to connect the meter to the PC, start the terminal program and set the communication options as follows: 8, N, 1, no flow control.
Command Types
To send a command to the pH meter the scheme is:
<DLE> <
command>
<CR>
This line makes the computer send a Data Link Escape character, the command ex­pressed as a number or a 3-character se­quence) and a CR character.
Note: Windows Terminal® and all the other
terminal programs that support the ANSI escape sequence, represent the DLE character by the string '^P' and the CR character by the string '^M'. E.g. the line '^PPHR^M' sets the range to pH.
Windows Terminal® are registered Trademark of "Microsoft Co." TELIX® is registered Trademark of "Deltacomm"
For pH 300: Commands not requiring an answer from
the pH meter:
PHR sets the range to pH MVR sets the range to mV CAL is equivalent to pressing the CAL key CON is equivalent to pressing the CFM key UPC is equivalent to pressing the °C key DWC is equivalent to pressing the °C key MEM is equivalent to pressing the MEM key MRR is equivalent to pressing the MR key COM is equivalent to pressing the COMM key
Commands requiring an answer:
pH? Causes the meter to send the pH value.
If the reading is out of range "Err 1" is sent.
MV? Causes the meter to send the mV value.
If the reading is out of range "Err 2" is sent.
TM? Causes the meter to send the tempera-
ture value. If the reading is out of range "Err 3" is sent.
FA? Requests the meter to send the calibra-
tion factor of pH. The order of data is as follows:
1) Month, day and Year (when the calibration was done) E.g. "123195" for 31st of Dec. 95
2) Buffer used when the offset was done E.g. "0701" for 7.01 pH
3) Offset in mV E.g. "_00504" for 50.4 mV
"-00504" for 50.4 mV
4) Buffer used when slope was done E.g. "0401" for 4.01 pH
5) Slope in mV E.g. "000592" for 59.2 mV
6) Temperature E.g. "_01199" for 119.9°C
"-00099" for -9.9°C
BAR Requests the meter to send the RS232
baud rate
9594
PRE Requests the meter to send the RS232
command prefix
Commands setting parameters:
/BR To set the RS232 baud rate. E.g.
send "/BR0" to set the meter to baud rate of 150 send "/BR1" to set the meter to baud rate of 300 send "/BR2" to set the meter to baud rate of 600 send "/BR3" to set the meter to baud rate of 1200 send "/BR4" to set the meter to baud rate of 2400
/PF To set the RS232 command prefix. E.g.
send "/PF05" to set the command prefix to
05.
For pH 301: Commands not requiring an answer from
the pH meter:
PHR sets the range to pH PPM sets the range to ppm MVR sets the range to mV OFF is equivalent to pressing the OFF key
Commands requiring an answer:
pH? Causes the meter to send the pH value.
If the reading is out of range "Err 1" is sent.
pM? Causes the meter to send the ppm value.
If the reading is out of range "Err 5" is sent.
MV? Causes the meter to send the mV value.
If the reading is out of range "Err 2" is sent.
TM? Causes the meter to send the tempera-
ture value. If the reading is out of range "Err 3" is sent.
The meter will send "Err6" if in a different measurement range.
FA? Requests the meter to send the calibra-
tion factor of pH. The order of data is as follows:
1) Month, day and Year (when the calibration was done) E.g. "123195" for 31st of Dec. 95
2) Time (when the calibration was done) E.g. "1259" for 12:59 hh/mm
3) Buffer used when the offset was done E.g. "0701" for 7.01 pH
4) Offset in mV E.g. "_00504" for 50.4 mV
"-00504" for 50.4 mV
5) Buffer used when 1st slope was done E.g. "0401" for 4.01 pH
6) 1st slope in mV E.g. "000592" for 59.2 mV
7) Buffer used when 2nd slope was done E.g. "1001" for 10.01 pH
8) 2nd slope in mV E.g. "000569" for 56.9 mV
9) Temperature E.g. "_01105" for 110.5°C
"-00099" for -9.9°C
FI? Requests the meter to send the calibra-
tion factor of ION. The order of data is as follows:
1) Month, day and Year (when the calibration was done) E.g. "060395" for 6th of March 95
2) Time (when the calibration was done) E.g. "1651" for 16:51 hh/mm
2) Buffer used when the offset was done E.g. "000100" for 10.0ppm
3) Offset in mV E.g. "000234" for 23.4 mV
4) Buffer used when slope was done E.g. "010000" for 1000 mV
5) mV measured when the slope was done E.g. "012345" for 1234.5 mV
6) Temperature E.g. "_00105" for 10.5°C
"-00099" for -9.9°C
DA? Requests the meter to send the date
E.g. "022896" for 28th Feb. 96
9796
TI? Requests the meter to send the time
E.g. "233001"
for 23:30 hr, 1 sec. as interval "233002" for 23:30 hr, 15 sec. as interval "233003" for 23:30 hr, 30 sec. as interval "233004" for 23:30 hr, 1 min. as interval "233005" for 23:30 hr, 5 min. as interval "233006" for 23:30 hr, 30 min. as interval "233007" for 23:30 hr, 60 min. as interval "233008" for 23:30 hr, 120 min. as interval "233009" for 23:30 hr, 180 min. as interval
BAR Requests the meter to send the RS232
baud rate
PRE Requests the meter to send the RS232
command prefix
Commands setting parameters:
/BR To set the RS232 baud rate. E.g.
send "/BR0" to set the meter to baud rate of 150 send "/BR1" to set the meter to baud rate of 300 send "/BR2" to set the meter to baud rate of 600 send "/BR3" to set the meter to baud rate of 1200 send "/BR4" to set the meter to baud rate of 2400 send "/BR5" to set the meter to baud rate of 4800 send "/BR6" to set the meter to baud rate of 9600
/PF To set the RS232 command prefix. E.g.
send "/PF05" to set the command prefix to
05.
For pH 302: Commands not requiring an answer from
the pH meter:
PHR sets the range to pH MVR sets the range to mV CAL is equivalent to pressing the CAL key CFM is equivalent to pressing the CFM key PRT is equivalent to pressing the PRINT key PGM is equivalent to pressing the PGM key CLR is equivalent to pressing the CLEAR key OFF is equivalent to pressing the OFF key
Commands requiring an answer:
pH? Causes the meter to send the pH value.
If the reading is out of range "Err 1" is sent.
MV? Causes the meter to send the mV value.
If the reading is out of range "Err 2" is sent.
TM? Causes the meter to send the tempera-
ture value. If the reading is out of range "Err 3" is sent.
The meter will send "Err6" if in a different measurement range.
FA? Requests the meter to send the calibra-
tion factor of pH. The order of data is as follows:
1) Month, day and Year (when the calibration was done) E.g. "123195" for 31st of Dec. 95
2) Buffer used when the offset was done E.g. "0701" for 7.01 pH
3) Offset in mV E.g. "_00504" for 50.4 mV
"-00504" for 50.4 mV
4) Buffer used when slope was done E.g. "0401" for 4.01 pH
5) Slope in mV E.g. "000592" for 59.2 mV
6) Temperature E.g. "_01199" for 119.9°C
"-00099" for -9.9°C
9998
DA? Requests the meter to send the date
status (MMDD)
TI? Requests the meter to send the time
status (HHMM)
Commands setting parameters:
/BR To set the RS232 baud rate. E.g.
send "/BR0" to set the meter to baud rate of 150 send "/BR1" to set the meter to baud rate of 300 send "/BR2" to set the meter to baud rate of 600 send "/BR3" to set the meter to baud rate of 1200 send "/BR4" to set the meter to baud rate of 2400 send "/BR5" to set the meter to baud rate of 4800 send "/BR6" to set the meter to baud rate of 9600
/PF To set the RS232 command prefix.
E.g. send "/PF05" to set the command prefix to 05.
For all models
Note: <ACK> will be sent by the meter if the
command received is accepted, other-
wise it will send <CAN>. <ACK> equals to ASCII code 06 and <CAN> equals to ASCII code 24.
These commands may be sent with either capital or small letters. Invalid commands will be ignored. The characters sent by the pH meter are always capital letters.
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI774P pH 4.01 & 7.01 Buffer Solution,
30mL each
HI7004M pH 4.01 Buffer Solution, 230 mL HI7004L pH 4.01 Buffer Solution, 460 mL HI7006M pH 6.86 Buffer Solution, 230 mL HI7006L pH 6.86 Buffer Solution, 460 mL HI777P pH 7.01 Buffer Sol. 2x30 mL HI7007M pH 7.01 Buffer Solution, 230 mL HI7007L pH 7.01 Buffer Solution, 460 mL HI7009M pH 9.18 Buffer Solution, 230 mL HI7009L pH 9.18 Buffer Solution, 460 mL HI7710P pH 7.01 & 10.01 Buffer Solution,
30 mL each
HI7010M pH 10.01 Buffer Solution, 230 mL HI7010L pH 10.01 Buffer Sol., 460 mL
pH CALIBRATION SOLUTIONS IN FDA APPROVED BOTTLE
HI8004L pH 4.01 Buffer Solution, 460 mL HI8006L pH 6.86 Buffer Solution, 460 mL HI8007L pH 7.01 Buffer Solution, 460 mL HI8009L pH 9.18 Buffer Solution, 460 mL HI8010L pH 1 0.0 1 B uf fe r S ol ut io n, 460 mL
ELECTRODE STORAGE SOLUTIONS
HI70300M Storage Solution, 230 mL HI70300L Storage Solution, 460 mL
ELECTRODE STORAGE SOLUTIONS IN FDA APPROVED BOTTLE
HI80300M Storage Solution, 230 mL HI80300L Storage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI7061M General Cleaning Sol., 230 mL
101100
HI7061L General Cleaning Sol., 460 mL HI7073M Protein Cleaning Sol., 230 mL HI7073L Protein Cleaning Sol., 460 mL HI7074M Inorganic Cleaning Sol., 230 mL HI7074L Inorganic Cleaning Sol., 460 mL HI7077M Oil & Fat Cleaning Sol., 230 mL HI7077L Oil & Fat Cleaning Sol.,460 mL
pH & ORP ELECTRODES
B = BNC CONNECTION PLUG + 1 m (3.3) CABLE
S = SCREW-TYPE CONNECTOR
ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE
HI8061M General Cleaning Solution, 230 mL HI8061L General Cleaning Solution, 460 mL HI8073M Protein Cleaning Solution, 230 mL HI8073L Protein Cleaning Solution, 230 mL HI8077M Oil & Fat Cleaning Solution, 230mL HI8077L Oil & Fat Cleaning Solution, 460 mL
REFILLING ELECTROLYTE SOLUTIONS
HI7071 3.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction elec­trodes
HI7072 1M KNO3 Electrolyte, 4x50 mL HI7082 3.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
REFILLING ELECTROLYTE SOLUTIONS IN FDA APPROVED BOTTLE
HI8071 3.5M KCl + AgCl Electrolyte,
4x50 mL, for single junction elec­trodes
HI8072 1M KNO3 Electrolyte, 4x50 mL HI8082 3.5M KCl Electrolyte, 4x50 mL,
for double junction electrodes
ORP PRETREATMENT SOLUTIONS
HI7091M Reducing Pretreat. Sol., 230 mL HI7091L Reducing Pretreat. Sol., 460 mL HI7092M Oxidizing Pretreat. Sol., 230 mL HI7092L Oxidizing Pretreat. Sol., 460 mL
HI1043B / HI1040S
Glass-body, double junction, refillable, combi­nation pH electrode. Use: strong acid/alkali.
12 mm
0.5"
HI 1043
"S" VERSION
9.5mm DIA
120 mm
4.7"
0.37"
HI1053B / HI1050S
Glass-body, triple ceramic, conic shape, refill­able, co mbination pH electrode. Use: emulsions.
12 mm
0.5"
HI 1053
"S" VERSION
120 mm
4.7"
HI1083B
Glass-body, micro, Viscolene, nonrefillable, combination pH electrode. Use: biotechnology, micro titration.
HI 1083
12 mm
0.5"
5 mm
0.2"
120 mm
3 mm
0.12"
"
3.0 mm DIA
0.12"
HI1131B / HI1111S
Glass-body, single junction, refillable, combi­nation pH electrode. Use: general purpose.
12 mm
0.5"
HI 1131
"S" VERSION
9.5mm DIA
120 mm
4.7"
0.37"
103102
HI1330B / HI1310S
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory.
HI 1330
"S" VERSION
120 mm
4.7"
5mm DIA
0.2"
5mm
0.2"
HI1331B / HI1311S
Glass-body, semimicro, single junction, refill­able, combination pH electrode. Use: flasks.
HI 1331
"S" VERSION
8 mm
0.3"
7.5mm DIA
210 mm
8.25"
0.29"
HI1230B / HI1210S
Plastic-body (Ultem®), double junction, gel-filled, combination pH electrode. Use: general purpose.
12 mm
0.5"
HI 1230
120 mm
"
"
4.7"
HI2031B / HI2020S
Glass-body, semimicro, conic, refillable, combina­tion pH electrode. Use: semisolid products.
6 mm
0.25"
HI 2031
75 mm
"
"
2.95"
HI1332B / HI1312S
Plastic-body (Ultem®), double junction, refillable, combination pH electrode. Use: general purpose.
12 mm
0.5"
HI 1332
"S" VERSION
120 mm
4.7"
FC100B
Plastic-body (Kynar®), double junction, refill­able, combination pH electrode. Use: general purpose for food industry.
12 mm
0.5"
FC 100
120 mm
4.7"
FC200B / FC200S Plastic-body (Kynar®), single junction, conic, Viscolene, refillable, combination pH electrode. Use: meat & cheese.
6 mm
0.25"
FC 200
75 mm
2.95"
FC210B
Glass-body, double junction, conic, Viscolene, combination pH electrode. Use: milk, yogurt.
12 mm
0.5"
FC 210
120 mm
4.7"
FC220B
Glass-body, single junction, refillable, combi­nation pH electrode. Use: food & wine process- ing.
12 mm
0.5"
FC 220
9.5mm DIA
120 mm
4.7"
0.37"
Ultem® is registered Trademark of "General Electrics Co."
Kynar® is registered Trademark of "Pennwalt Corp."
105104
120 mm
4.7"
12 mm
0.5"
HI 4430
"S" VERSION
FC911B
Plastic-body (Kynar®), double junction, refill­able with built-in amplifier, combination pH electrode. Use: very high humidity.
12 mm
0.5"
FC 911
110 mm
4.3"
HI1413B / HI1410S
Glass-body, single junction, flat tip, Viscolene, combination pH electrode. Use: surface mea- surement.
12 mm
0.5"
HI 1413
110 mm
4.3"
Half-cell electrodes:
HI2110B
Glass-body, single half-cell pH electrode. Use: general purpose.
12 mm
0.5"
HI 2110
9.5mm DIA
120 mm
0.37"
"
HI5311
Glass-body, Ag/AgCl reference half-cell elec­trode, double junction, refillable with 4mm plug with 1m (3.3') cable. Use: general purpose with wide temperature range.
12 mm
0.5"
HI5412
Glass-body, single Calomel reference half-cell electrode, refillable with 4mm plug with 1m (3.3') cable. Use: general purpose with constant tem­perature range.
12 mm
0.5"
HI 5412
120 mm
4.7"
ORP electrodes:
HI3131B / HI3111S
Glass-body, refillable, combination platinum ORP electrode. Use: titration.
12 mm
0.5"
HI 3131
150 mm
"S" VERSION
5.9"
HI3230B / HI3210S
Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode. Use: general purpose.
12 mm
0.5"
HI 3230
120 mm
"S" VERSION
4.7"
HI4430B / HI4410S
Plastic-body (Ultem®), gel-filled, combination gold ORP electrode. Use: general purpose.
HI 5311
120 mm
4.7"
Kynar® is registered Trademark of "Pennwalt Corp."
Ultem® is registered Trademark of "General Electrics Co."
107106
EXTENSION CABLES FOR SCREW­TYPE ELECTRODES ONLY (SCREW TO BNC CONNECTOR)
HI7855/1 Extension cable 1m (3.3') long HI7855/3 Extension cable 3m (9.9') long HI7855/5 Extension cable 5m (16.5') long HI7855/10 Extension cable 10m (33') long HI7855/15 Extension cable 15m (49.5') long
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO SCREW TYPE ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
POWER UNITS
HI710005 115VAC-12VDC voltage adapter HI710006 230VAC-12VDC voltage adapter
OTHER ACCESSORIES
ChecktempC Thermometer with penetration
probe (range -50.0 to 150.0°C)
HI76405 Electrode holder HI7669/2W Temperature probe with 1 m (3.3')
screened cable
HI710032 Pack of 10 plain paper spare rolls
(for pH 302 only)
HI710033 Replacement ink cartridge (for
pH 302 only)
HI 92000 Windows® compatible application
software
HI 920010 9 to 25-pin PC connection cable HI 920010/9 9 to 9-pin PC connection cable MANpH300R1 Instruction manual
Windows® is registered Trademark of "Microsoft Co."
ELECTRODE APPLICATION
REFERENCE GUIDE
Application Electrodes
1. Aquarium HI1332B, HI1911B, HI1312S
2. Bath-water HI1910B, HI1130B, HI1110S
3. Beer HI1131B, HI1111S
4. Bread HI2031B, FC200B, HI 2020S, FC200S
5. Cheese FC200B, FC200S
6. Dairy products FC911B, FC100B
7. Dirty water HI1910B, HI1912B
8. Emulsions HI1053B, HI1050S
9. Environment HI1230B, HI1210S
10. Flasks HI1331B, HI1310S
11. Food industry general use FC911B, FC 100B 12 . Fruit FC 200B, FC220B, FC200S 13 . Fruit juices, organic FC210B
14. Galvanizing waste solution HI1130B, HI1912B, HI1110S
15. Heavy-duty applications HI1135B, HI 1115S 16 . High purity water H I1053B, HI1050S
17. Horticulture HI1053B, FC200B, HI1050S, FC200S
18. In-line applications HI 1134B, HI1135B, HI2114B, HI 2910B,
19. Laboratory general use HI1131B, HI 1230B, HI1332B, HI1330B
20. Leather HI1413B, HI1410S
21. Lemon juice FC100B
22. Meat FC200B, HI2031B, FC200S, HI2020S 23 . Micro plate sampling of less than HI1083B, HI1080S
100 mL
24 . Milk and Yogurt FC210B
25. Paints HI1053B, HI1050S
26. Paper HI 1413B, HI1410S
27. Photographic chemicals HI1230B, HI1210S 28 . Quality control HI1332B, HI 1312S
29. Sausages FC200B, HI 2031B, FC200S, HI2020S
30. Semisolid products HI 2031B, HI2020S
31. Skin HI1413B, HI1410S 32 . Soil samples HI1230B, HI1210S
33. Solvents HI1043B, HI 1040S
34. Strong acid HI1043B, HI1040S
35. Submersion application HI1130B, HI1110S
36. Surface measurements HI1413B, HI1410S
37. Swimming pool H 1130B, HI 2114B, HI2910B
38. Titrations with constant temperature HI 1131B, HI1111S range
39 . Titrations with wide temperature HI1131B, HI1111S
range
40. Very high humidity FC911B, HI1912B, HI 1911B
41. Vials and test tube HI1330B, HI 1310S
42. Wine processing FC220B
HI1114S, HI1115S
HI1111S, HI1210S, HI 1312S, HI1310S
B = BNC connection plug S = Screw-type connection plug
109108
WARRANTY
All Hanna Instruments are warranted for two years against defects in workmanship and
materials when used for their intended purpose and maintained according to the instructions.
The probes and the electrodes are war­ranted for a period of six months.
CE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge of the meter only, whenever due to defect of manufacturing.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. If the instrument is to be returned to Hanna Instruments, obtain a Return Goods Authorization from the Customer Ser­vice Department first and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written permission of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
We
Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the bench pH meters
have been tested and found to be in compliance with the following regulations:
IEC 801-2 Electrostatic Discharge IEC 801-3 RF Radiated IEC 801-4 Fast Transient EN 55022 Radiated, Class B
Date of Issue: 30-05-1996
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
pH 300 pH 301 pH 302
D.Volpato - Engineering Manager
Hanna Instruments S.r.l.
On behalf of
111110
PRINTED IN
ITALY
MANpH300R1
07/96
112
http://www.hannainst.com
Loading...