Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility
in a wide range of applications.
These instruments are in compliance with CSA,
CE Declaration of Conformity ...................... 109
ISO 9000 Certified
Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no
damage has occurred during shipping. If there
is any damage, notify your Dealer.
Each pH meter comes supplied complete with:
• HI 1131B glass-body combination pH electrode
• HI 7669/2W temperature probe
• HI 3131B glass-body platinum ORP electrode
• pH 4.01 & 7.01 buffer solutions (20 mL each)
• 12VDC adapter (HI710005 or HI710006)
• 5 paper rolls (pH 302 only)
The KIT version includes also:
• HI76405 electrode holder
• HI 92000 Windows® compatible applica-
tion software
• HI 920010 PC connection cable
• Storage, cleaning and electolyte solutions
Note: Save all packing material until you are
sure that the instrument functions correctly. All
defective items must be returned in the original
packing with the supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments pH300, pH301 and pH302
are professional bench-top pH meters for pH,
ORP (Oxidation Reduction Potential), ISE and
temperature measurements. Their built-in microprocessor ensures accurate and userfriendly operation.
pH measurements are automatically compensated for the temperature effect (ATC).
These meters come equipped with a large,
easy-to-read LCD which shows the pH or
mV or ION (for pH301 only) with temperature simultaneously. The display has graphic
symbols to make the calibration procedure
easy to follow. Calibration is not allowed until
the electrode has stabilized to ensure accurate measurements.
Through the RS 232 serial port you can
transfer sample measurements to a PC.
32
FUNCTIONAL DESCRIPTION pH 300
THE FRONT PANEL
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
OFF
COMM
CFM
WRONG
CAL
LOG
ONONOFF
COMM
ppmpHpH
INTV
mV°C
m
1
2
BUF
ppm
Rel mV
MEM
BUF
3
1
2
3
mV°C
RANGE CAL DATA
CALMS°C
CFM
DATE TIME
ppmpHpH
RANGE
CALMS°C
CFM
s
pH METER
°C
INTV
m
s
CAL DATA
pH METER
°C
pH300
MR
pH300
MR
Keyboard on the Left
• COMM to display the current baud rate
for RS232.
Keyboard on the Right
!!
•
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not
connected) and to select the calibration
buffer, the baud rate, the command prefix.
• CAL to enter and exit the calibration mode.
After the pH calibration, to enter the calibration date, use this key to toggle between month, day and year.
• CAL DATA to display the date of last pH
calibration. This also displays the offset
and the slope values expressed in mV.
• CFM to confirm calibration values and to
confirm baud rate and command prefix
during communication setting mode.
• MR to recall the stored value.
• MS to hold and memorize the measured
value. The last reading will be frozen on
the display until the key is released. At
the same time this reading is stored in the
memory.
• RANGE to select the measurement range
(pH or mV).
• ON to turn the meter on
• OFF to turn the meter off
• Power LED to confirm the meter is on
54
SPECIFICATIONS pH 300
FUNCTIONAL DESCRIPTION pH 301
THE FRONT PANEL
RangepH0.00 to 14.00
ISE±399.9 mV
ORP±1999 mV
°C-9.9 to 120.0
ResolutionpH0.01
ISE0.1 mV
ORP1 mV
°C0.1
AccuracypH±0.01
(@20°C/68°F) ISE±0.2 mV
ORP±1mV
°C±0.5
Typical EMC pH±0.01
Deviation ISE±0.6 mV
ORP±1 mV
°C±0.5
pH CalibrationAutomatic 1 or 2 points with
3 memorized buffers
(4.01, 7.01, 10.01)
pH OffsetMax ±60mV or ±1pH
pH Slope70-108%
TemperatureAutomatic 1 or 2 points with
Calibration2 memorized values
(0 and 50°C)
TemperatureAutomatic or manual
Compensationfrom -9.9 to 120°C (14 to 248°F)
ElectrodeHI1131B glass body combination
refillable pH electrode (included)
TemperatureHI 76 6 9/ 2 W with 1 m
probe(3.3') cable (included)
Interface
Power supply12 VDC (CSA, UL & CE approved)
Environment0 to 50°C (32 to 122°F); 95% RH
Dimensions280x200x75mm (11x7.9x3")
Shipping Weight1.3 kg (2.9 lb.)
ppm
pH
DATE
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
Rel mV
CFM
MEM
WRONG
CAL
LOG
OFF
CAL
DATE COMM
RANGE
CFM
TIME
TIME
ppm
pH
INTV
mVmV°C
m
1
2
BUF
BUF
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
ppm
pH
DATE
TIME
ppm
pH
INTV
°C
1
m
2
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
• ON to turn the meter on
• OFF to turn the meter off
• Power LED to confirm the meter is on
pH301
pH / ION
METER
°C
°C
pH301
pH / ION
METER
°C
°C
76
Keyboard on the Left
• CAL to start the calibration mode for pH,
mV and ION. To set or quit the date or
time setting mode. Used also during logging (to start logging press pH, mV or ION
plus LOG plus CAL key; to stop logging
press LOG plus CAL key).
• CFM to confirm calibration values.
• COMM to display the current baud rate
and command prefix for RS232.
• DATE to display the date.
• RANGE to display the calibration buffer
or the temperature on the secondary LCD
during calibration. During setting of the
date, to toggle between month, day and
year. During setting of the time, to toggle
between minutes, seconds and logging interval.
• TIME to display the time.
Keyboard on the Right
!!
•
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not
connected) and to select the calibration
buffer, the date and time, the baud rate,
the command prefix.
• CAL DATA to display the date of last pH
calibration. This also displays the offset
and the slope 1 or slope 2 values expressed in mV.
pH OffsetMax ±60mV or ±1pH
pH Slope70-108%
ppm CalibrationAutomatic 1 or 2 points with
5 memorized buffers
(0.1, 1, 10, 100, 1000)
ppm OffsetMax ±1999.9mV
ppm SlopeFrom ±20mV to ±100 mV
[mV/log(0.001xppm)]
TemperatureAutomatic 1 or 2 points with
Calibration2 memorized values (0 and 50°C)
TemperatureAutomatic or manual (for pH only)
Compensationfrom -9.9 to 120°C (14 to 248°F)
ElectrodeHI1131B glass body combination
refillable pH electrode (included)
TemperatureHI 7669/2W with 1 m
probe(3.3') cable (included)
Input impedance1012 Ω
Logging Interval1, 15, 30 seconds or
1, 5, 30, 60, 120, 180 minutes
PC InterfaceRS 232 (optoisolated)
Power supply12 VDC (CSA, UL & CE approved)
Environment0 to 50°C (32 to 122°F); 95% RH
Dimensions280x200x75mm (11x7.9x3")
Shipping Weight1.3 kg (2.9 lb.)
98
FUNCTIONAL DESCRIPTION pH 302
THE FRONT PANEL
Keyboard on the Left
• 0 to 9 for numeric and decimal data in-
put.
• CLEAR to cancel wrong data entry during
programming.
Rel mV
ppm
DATE TIME
Keyboard on the Right
• CAL to perform the pH, mV or tempera-
MEM
CFM
WRONG
CAL
LOG
OFF
ppmpHpH
INTV
mV°C
m
1
2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
ture calibration.
• CFM/
!!
! to confirm pH/mV/temperature
!!
calibration values or to manually set temperature (when the temperature probe is
not connected).
• ENTER/
""
"°C to terminate the numeric data
""
entry or to manually set temperature (when
the temperature probe is not connected)
or to select the buffer value during pH
calibration.
• PAPER to pull the paper.
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
8
9
7
6
5
4
1
3
2
0
CLEAR
.
ppmpHpH
INTV
mV°C
m
1
2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
• PGM to select a program.
• PRINT to obtain a printout (prints time,
date, sample number and pH or mV values).
• RANGE to select the measurement range
(pH or mV) or to display the calibration
buffer or the temperature on the secondary LCD during calibration.
• TIME to display the time and the date.
• ON to turn the meter on
• OFF to turn the meter off
• Power LED to confirm the meter is on
1110
SPECIFICATIONS pH 302
RangepH0.00 to 14.00
ISE±399.9 mV
ORP±1999 mV
°C-9.9 to 120.0
ResolutionpH0.01
ISE0.1 mV
ORP1 mV
°C0.1
AccuracypH±0.01
(@20°C/68°F) ISE±0.2 mV
ORP±1mV
°C±0.5
Typical EMC p H±0.01
Deviation ISE±0.3 mV
ORP±1 mV
°C±0.5
pH CalibrationAutomatic 1, or 2 points with
3 memorized buffers
(4.01,7.01,10.01)
TemperatureAutomatic 1 or 2 points with
Calibration2 memorized values
(0 and 50°C)
TemperatureAutomatic or manual
Compensationfrom -9.9 to 120°C (14 to 248°F)
(for pH only)
ElectrodeHI1131B glass body combination
refillable pH electrode (included)
TemperatureHI 76 6 9/ 2 W with 1 m
probe(3.3') cable (included)
PC InterfaceRS 232 (optoisolated)
Power supply12 VDC (CSA, UL & CE approved)
Environment0 to 50°C (32 to 122°F); 95% RH
Dimensions280x200x75mm (11x7.9x3")
Shipping Weight1.3 kg (2.9 lb.)
THE REAR PANEL
MEASUREMENT
REFERENCE
ELECTRODE
TEMPERATURE
ELECTRODE
123456
RESET 12VDCRS 232
PROBE
ooooo
oooo
1. Reference Electrode Connector (Ø 4 mm)
2. BNC socket for measurement electrode
3. Socket for Temperature Probe
4. Reset Button
5. DC Power Socket (for HI710005 or
HI710006)
6. RS 232 Connector
LCD FUNCTIONAL DESCRIPTION
Primary Display
CFM
CAL
LOG
Rel mV
MEM
WRONG
BUF
ppm
1
2
3
mV°C
ppmpHpH
DATE TIME
INTV
m
s
Secondary Display
1312
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ISE or ORP electrodes (with internal reference) plug the
electrode's BNC to the socket provided (see
page 11).
For electrode with separate reference, connect the measuring electrode's BNC to the
BNC socket and the reference electrode's
plug to the socket provided (see page 11).
For temperature measurements and automatic temperature compensation connect the
temperature probe to the °C socket (see
page 11).
If the temperature probe
is not connected, when
turned on, the meter will
display a default tem-
°C
perature of 25°C or the
last recorded temperature (if the temperature probe is disconnected
when the meter is on) on the secondary
LCD. The "°C" indicator will be flashing.
POWER CONNECTION
Plug the 12VDC adapter (HI710005 or
HI710006) into the DC socket (see page11).
Plug the adapter to the mains.
Note: The instrument uses the following con-
figuration.
+
-
It is recommendable to use the Hanna
HI710005 or HI710006 voltage adapters (sup-
plied with the meters) that use the proper
polarity configuration.
The meters can also be used with other adapters. In this case, remember to check the
correct polarity of your adapter before connecting it to the meter.
Note: Make sure the main line is protected
by a fuse.
Press ON to turn the meter on. The power
LED will remain lit until OFF is pressed to
turn the meter off.
ON
OFF
When the meters are turned on, the microprocessor checks if any calibration data is in
memory (i.e. if the offset is within ±60mV
and the slope between 70-108%).
If the calibration data is
valid, the meter will op-
pH
erate in the normal measurement mode and default to pH measurement
°C
mode with the temperature on the secondary LCD.
If the calibration data is
not valid or the calibra-
pH
tion data is lost, the
"CAL" indicator will be
displayed to prompt the
CAL
°C
user to perform the pH
calibration. If the user chooses not to perform the calibration and presses CAL to quit
from the calibration mode, the offset and the
slope of the instrument would be set to 0 mV
and 100% respectively.
1514
pH300, pH301 and pH302 use an EEPROM
to retain the pH calibration, mV calibration,
temperature calibration as well as the serial
communication setting. The instrument will
store the respective data after a calibration
or serial communication setting, even when it
is not plugged-in.
Press CAL to enter the setting mode and the
month will be flashing.
CAL
DATE
For pH 302 only
pH302 uses a rechargeable battery to retain
the pH calibration data. It lasts approximately
one month when fully charged. The battery is
automatically charged any time the meter is
plugged-in (either turned on or off).
If you need to unplug the
instrument, press OFF before
OFF
disconnecting the meter from
the mains. Only in this case
will the instrument retain the
memorized data. Otherwise,
the instrument needs to be
re-calibrated. In the case of a power failure
while the instrument is operational, re-calibration is required.
When the instrument is initially switched on,
the microprocessor checks if any calibration
data is in memory (i.e. if the offset is within
±1 pH and the slope between 70 - 108%).
If the calibration data are not memorized, the
offset and slope of the instrument will be set
to 0 mV and 100% respectively.
Use UP or DOWN to select the month.
°C
°C
Press RANGE and the day will start to flash.
RANGE
DATE
Use UP or DOWN to select the day.
°C
°C
Press RANGE and the year on the secondary LCD will flash.
RANGE
DATE
SETTING THE DATE AND THE TIME AND
THE LOGGING INTERVAL (for pH 301 only)
Press DATE to display the date. The month
and the day will be displayed on the primary
LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the year.
°C°C
1716
Press CAL to exit the date
setting mode.
CAL
Press RANGE and the logging interval on the
secondary LCD will come up.
RANGE
TIME
Press TIME to display the time. The hour and
the minutes will be displayed on the primary
LCD, the logging interval on the secondary
one.
TIME
TIME
INTV
Press CAL to enter the setting mode and the
hour will flash.
CAL
TIME
INTV
Use UP or DOWN to select the hour.
°C°C
Press RANGE and the minutes will flash.
RANGE
TIME
INTV
Use UP or DOWN to select the logging interval.
°C
°C
CAL
Press CAL to exit the time setting mode.
VIEWING THE DATE AND THE TIME (for
pH 301 only)
Press DATE to display the date. The month
and the day will be displayed on the primary
LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the minutes.
°C
°C
INTV
Press TIME to display the time. The hour and
the minutes will be displayed on the primary
LCD, the logging interval on the secondary
one.
TIME
1918
TIME
INTV
s
SETTING THE DATE AND THE TIME AND
THE LOGGING INTERVAL (for pH 302 only)
See program No. 1 (page 60).
VIEWING THE DATE AND THE TIME (for
pH 302 only)
Press TIME to display the time (HH/MM).
TIME
Press TIME again to display the date (MM/
DD).
TIME
DATE
Allow 1 or 2 minutes for
pH
the electrode to stabilize.
°C
pH is displayed on the
primary display and temperature on the second-
pH
ary one.
If readings are out of
range "---" will appear
on the LCD display.
For pH 301 only:
Press the .00/.000 key to change the resolution from two to three decimal points.
0.00/0.000
pH
°C
TAKING pH MEASUREMENTS
Make sure that the instrument has been calibrated for pH before taking pH measurements (see page 29).
For pH 300 and pH 302 only
:
RANGE
Select the pH measurement mode
by pressing RANGE.
For pH 301 only
:
pH
Select the pH measurement
mode by pressing pH.
To take a pH measurement
simply submerge the tip (4cm/
1½") of the pH electrode and
the temperature probe into the
sample to be tested.
4 cm
(1½")
It is recommended that the electrode is rinsed
thoroughly for better conditioning. For this
rinsing process, it is recommended to use a
liberal amount of the sample to be measured.
If measurements are taken in different samples
successively, it is recommended that the electrode and the temperature probe are rinsed
thoroughly with deionized water or distilled
water (if not available, tap water may also be
used) and then blot dry between each measurement.
The pH reading is directly effected by temperature. In order for the meter to measure
the pH accurately, temperature must be taken
into consideration.
A perfect equilibrium between the pH electrode and the sample is reached in approximately 15 minutes.
To use the meter's Automatic Temperature
Compensation feature, submerge the
HI7669/2W temperature probe into the sample
2120
as close to the electrode as possible and wait
for 1-2 minutes.
If you know the temperature of the sample to
be tested you can compensate for it manually.
If manual temperature compensation is desired the temperature probe must be disconnected from the instrument.
The display will then
show the default temperature of 25°C or the
last recorded tempera-
°C
ture reading. "°C" is
blinking.
Note the temperature of the
C
°
sample using a ChecktempC
or a glass thermometer.
The temperature can now be
adjusted with UP and DOWN.
If the temperature
°C°C
probe is connected to
the meter, the secondary LCD figure displays the measured
temperature reading
and the "°C" indicator
CFM/
ENTER/
does not blink. Pressing UP or DOWN will
not alter the temperature reading.
VIEWING THE pH CALIBRATION DATA
For pH 300:
Press CAL DATA and the meter will display
the last pH calibration data.
CAL DATA
Press UP or DOWN to get this information in
the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month,
day and year)
b.Offset calibration
CAL
value expressed in
mV and displayed on
mV
the primary LCD with
the buffer used to
calibrate the offset on
the secondary LCD.
CAL
pH
1
The offset pH value should be between
±30 mV. For values between ±30 and
±60 mV change the reference electrolyte
and repeat the calibration procedure. For
values higher than +60 mV or lower than
-60mV the electrode is no longer reliable.
See the electrode maintenance section
on page 84.
c.Slope calibration
mV
value directly expressed in mV/pH is
displayed on the primary LCD with the
CAL
pH
2
2nd buffer used to
calibrate the offset on the secondary LCD.
For pH 301:
Press pH followed by
pH
CAL DATA
CAL DATA and the
meter will display the
last pH calibration
data.
2322
Press UP or DOWN to get this information in
the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month,
day and year)
CAL
b. Time when the last
TIME
pH calibration was
completed (hour and
minutes).
CAL
c.Offset calibration
mV
value expressed in
mV and displayed on
the primary LCD
with the buffer used
CAL
pH
1
to calibrate the offset on the secondary LCD.
The offset pH value should be between
±30 mV. For values between ±30 and
±60 mV change the reference electrolyte
and repeat the calibration procedure. For
values higher than +60 mV or lower than
-60mV the electrode is no longer reliable.
See the electrode maintenance section
on page 84.
d. Acid slope calibra-
mV
tion value expressed
in mV/pH and displayed on the primary LCD with the
CAL
pH
2
2nd buffer used to
calibrate the offset on the secondary LCD.
e. Alkaline slope calibra-
tion value expressed
mV
in mV/pH and displayed on the primary
LCD with the 3
buffer used to cali-
rd
CAL
pH
3
brate the offset on the secondary LCD.
For pH 302:
See program #3 (page 66).
TAKING REDOX MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ORP measurements. An optional
ORP electrode must be used to perform these
measurements (see page 105).
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP),
press RANGE.
For pH 301:
mV
To get into the mV mode, press
mV.
To measure the mV of a solution simply submerge the
ORP electrode tip (4 cm/1½")
into the sample to be tested.
Allow a few minutes for the
(1
½
")
reading to stabilize.
The display will indicate the absolute mV value
(positive or negative).
For pH 300 and pH 302 only
When values are less
:
Rel mV
than ±400 mV, tenths of
mV are displayed; for
values greater than
°C
±400mV, the decimal
digit disappears.
4 cm
2524
Note that the change in scale (from 0.1 mV
to 1 mV as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value
exceeds ±1999 mV (out
of range).
Oxidation-reduction potential (REDOX) measurements provide the quantification of the
oxidizing or reducing power of the sample
tested.
To correctly perform a redox measurement,
the surface of the ORP electrode must be
clean and smooth.
When not in use, the tip of the electrode
should be kept moist and safe from any mechanical stress which might cause damage
to the gold/platinum junction.
TAKING ION MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ION measurements. An optional
ISE electrode must be used to perform these
measurements.
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP
or ISE), press RANGE.
For pH 300 and pH 302:
The display will indicate the absolute mV value
(positive or negative).
When values are less
Rel mV
than ±400 mV, tenths of
mV are displayed; for
°C
values greater than
±400mV, the decimal
digit disappears. Note that the change in
scale (from 0.1 mV to 1 mV as resolution) is
automatic.
For pH 301:
The display will indicate
ppm
the ppm value expressed
in ppm or mV (depend-
°C
ing on the measured
sample). Note that the
change in scale (from 0.001ppm to 0.01ppm or
0.1 ppm as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value
exceeds ±1999 mV (out
of range).
When not in use, the tip
of the electrode should be kept moist and
safe from any mechanical stress which might
cause damage to the junction.
For pH 301:
To get into the ION mode, press
ION.
Simply submerge the ISE
electrode tip (4 cm/1½") into
the sample to be tested. Allow a few minutes for the reading to stabilize.
ION
VIEWING THE ION CALIBRATION FACTOR ON THE LCD (for pH 301 only)
Press ION followed
ION
CAL DATA
by CAL DATA and
the meter will display
the last ION cali-bration data.
°C°C
(1
4 cm
½
")
Press UP or DOWN to get
this information in the following sequence:
2726
a. Date of the last ION
DATE
calibration (month,
day and year)
b. Time when the last
CAL
ION calibration was
completed (hour and
TIME
minutes)
c.Offset calibration
value expressed in
CAL
INTV
mV and displayed on
the primary LCD
mV
with the value used
to calibrate the offset on the secondary LCD.
d. Slope calibration
value expressed in
CAL
mV
ppm
1
mV/[log (0.0001xppm)]
and displayed on the
primary LCD with
CAL
ppm
2
the 2nd value used to
calibrate the slope on the secondary LCD.
TAKING TEMPERATURE MEASUREMENTS
Taking a temperature measurement is very
easy.
ON
Connect the temperature probe
to the instrument and turn the
instrument ON.
Dip the liquid/general purpose
temperature probe HI7669/2W
into the sample
Allow the reading on the secondary display to stabilize (1
or 2 minutes).
pH
°C
PRINTING FUNCTION (for pH 302 only)
Press PRINT to get a printout at
PRINT
any time.
All the printouts will be complete
with date, time, sample No., temperature and measurement value
(pH or mV depending on the selected measurement range).
DATE 1/02
TIME 12:21
SAMPLE 1
°C 17.9
pH 7.18
DATE 1/02
TIME 12:40
SAMPLE 5
°C 20.5
mV -350.0
HOLD FUNCTION (for pH 300 only)
To hold and memorize the measured value,
press MS. The last reading will be frozen on
the display until the key is released. At the
same time, this reading is stored in the
memory. During this operation, the "MEM"
indicator is lit.
MS
MEM
pH
°C
2928
MR
To recall the stored memory
value, press MR and the memorized reading will be displayed
with the MEM indicator lit.
RESET BUTTON
The RESET button (see page 11) is used
when the instrument displays erroneous messages due to strong electrical interference or
when the instrument's power supply was disconnected before the meter was switched
off.
It is necessary to press the reset button and
restart the entire operation.
Calibration points should remain memorized.
It is recommended to verify calibration before proceeding.
pH CALIBRATION
For greatest accuracy, it is recommended
that the instrument is calibrated frequently.
For a faster operation, it is possible to standardize the electrode at a value close to the
expected sample value or at pH 7.01 only
(one point calibration), but it is always good
practice to calibrate at least 2 points.
For pH 301 only
:
Due to electrode conditioning time the electrode must be kept immersed for a few seconds to stabilize. The meters are equipped
with a stability indicator and the user will be
guided step by step with easy indications on
the display during the pH calibration. This will
make the calibration a simple and error-free
procedure.
The standard calibration program of the meter
is prepared for 3 (maximum) buffers. The
selectable pH values are: 1.68, 4.01, 6.86,
7.01, 9.18, 10.01.
INITIAL PREPARATION
Pour small quantities of
HI 7004
pH7.01 (HI7007/HI8007) and
pH4.01 (HI7004/HI8004) and/
HI 7004
HI 7007
or pH10.01 (HI7010/HI8010)
solution into individual beakers. If possible,
use plastic beakers to minimize any EMC
interferences.
For accurate calibration use two beakers for
each buffer solution, the first one for rinsing
the electrode, the second one for calibration.
By doing this,
RINSE
CALIBRATION
contamination
between the
buffers is mini-
HI 7007
HI 7007
mized.
To get accurate readings, use pH 7.01 and
3130
pH 4.01 if you are going to measure acid
samples, or pH 7.01 and pH 10.01 for alkaline measurements or, for pH301 only, perform a 3-point calibration for the entire range.
PROCEDURE
RANGE
• Press RANGE (for pH300
and pH 302) to select the
pH measurement mode.
• Press pH (for pH301) to
select the pH measure-
pH
ment mode.
• Remove the protective cap
from the pH electrode.
ONE POINT CALIBRATION (OFFSET)
• Immerse the pH electrode
into the selected buffer (e.g.
pH 7.01) buffer solution and
shake briefly.
Note: the electrode should
HI 7007
be submerged approximately 4 cm
(1½") into the solution. The tempera-
(1
ture probe should be
located as close as
4 cm
½
")
possible to the pH
electrode.
• Press CAL. The primary LCD will display
the pH value using the current offset and
slope, with the "CAL" and "
CAL
1
BUF
" indicators lit.
CAL
pH
pH
1
BUF
For pH 302 only: the printer will also print
step by step instructions for the calibration procedure.
Dip the electrode in the pH 7
solution; wait
30 sec and then
press CFM
• Select the buffer value by pressing UP or
DOWN or just DOWN (for pH 302) until
the secondary LCD displays the value of
the first solution (e.g. “7.01”).
°C
ENTER/
°C
pH
CAL
1
BUF
• Only when the reading is stable (less
" in-
CFM
CFM/
CFM
than ±0.03 variation
in a 10 seconds time
interval) the "
1
BUF
dicator will stop flashing (after about 30
seconds) and the
"CFM" indicator will
start blinking.
• Press CFM to confirm the calibration.
If the reading is not
close to the selected
buffer, "WRONG
WRONG
1
BUF
"
and "WRONG " will
blink alternately. If
the reading is close
to the selected buffer
(±1pH), the meter
stores the reading
WRONG
BUF
3332
CFM/
(and adjusts the offset point).
The buffer value is
pH
then displayed on
the primary LCD and
the secondary LCD
will display the ex-
CAL
2
BUF
pected second buffer value.
For pH 302 only: the printer will print
further instructions for the calibration procedure.
Now dip the electrode in 2nd
solution, wait
30 sec and then
press CFM
• Press CAL and the calibration process is ended with
just the offset of the meter
calibrated. For best accuracy
however, it is recommended
that a two-point calibration is
performed.
For pH 300 only
:
The calibration process
is complete and the display will ask for the date
CAL
entry.
°C
Use UP or DOWN to
select the month, the
day and the year.
Use CAL to toggle between
month, day and year.
CAL
CAL
°C
DATE
After entering the calibration date,
press CFM and the meter will
return to the normal operating
pH
mode.
Note: during calibration, the
RANGE
RANGE
secondary LCD displays
the selected buffer
value. By pressing
RANGE the temperature value can be also
displayed. This will allow you to check
the buffer temperature during calibration.
TWO-POINT CALIBRATION
A two-point calibration is recommended for
best accuracy. As a first point use pH 7.01
(or pH 6.86 can also be used for pH 301). As
a second point use pH 4.01 (or pH 1.68 for
pH 301 only) if you are going to measure acid
samples (pH 7 or less), use pH10.01 (or
pH 9.18 for pH301 only) if you are going to
measure alkaline samples.
CAL
• Proceed as described above
for one point calibration but
do not quit calibration by
pressing CAL at the end.
• After the first calibration
point is confirmed, immerse
the pH electrode into the
second buffer solution (pH
4.01 or pH10.01; for pH301
HI 7004
only, the second buffer can
also be pH 1.68 or pH 9.18) and shake
briefly.
Note: the electrode should
be submerged approximately 4 cm
(1½") into the solu-
(1
4 cm
½
")
tion. The temperature
probe should be located as close
as possible to the pH electrode.
3534
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.