Hanna Instruments PH 300, PH 301, PH 302 User Manual

Instruction Manual
pH300 pH301 pH302
Microprocessor
GLP Bench-top
pH/mV/ISE/°C Meters
ppm
DATE TIME
Rel mV
MEM
ON OFF
CFM
WRONG
CAL
LOG
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
http://www.hannainst.com
1
PAPER PRINT
pH302
pH METER
RANGE CAL
CFM /
PGM
ENTER /
These Instruments are in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully be­fore using the instrument. This manual will provide you with all the neces­sary information for the correct use of the instru­ment, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA,
UL and directives.
(EN 50081-1 and EN 50082-1)
TABLE OF CONTENTS
Preliminary Examination................................... 1
General Description .......................................... 1
Functional Description pH300 .......................... 2
Specifications pH300 ........................................ 4
Functional Description pH301 .......................... 5
Specifications pH301 ........................................ 7
Functional Description pH302 .......................... 8
Specifications pH302 ...................................... 10
The Rear Panel ................................................ 11
LCD Functional Description ............................. 11
Operational Guide ........................................... 12
pH Calibration ................................................. 29
pH Values at Various Temperature ................. 39
ION Calibration (for pH301 only) ..................... 4 0
mv Calibration ................................................. 44
Temperature Calibration ................................. 50
Logging with pH301 ........................................ 55
Programming Functions with pH302 .............. 58
Interface with PC ............................................. 82
Troubleshooting Guide ................................... 83
Electrode Conditioning and Maintenance ...... 84
Temperature-Resistance Correlation
for HANNA pH Sensitive Glass ....................... 8 8
Additional Information about
Interface with PC ............................................. 90
Accessories..................................................... 99
Electrode Application Reference Guide ........ 10 7
Warranty......................................................... 108
CE Declaration of Conformity ...................... 109
ISO 9000 Certified Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer. Each pH meter comes supplied complete with:
HI 1131B glass-body combination pH electrode
HI 7669/2W temperature probe
HI 3131B glass-body platinum ORP electrode
pH 4.01 & 7.01 buffer solutions (20 mL each)
12VDC adapter (HI710005 or HI710006)
5 paper rolls (pH 302 only)
The KIT version includes also:
HI76405 electrode holder
HI 92000 Windows® compatible applica-
tion software
HI 920010 PC connection cable
Storage, cleaning and electolyte solutions
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments pH300, pH301 and pH302 are professional bench-top pH meters for pH, ORP (Oxidation Reduction Potential), ISE and temperature measurements. Their built-in mi­croprocessor ensures accurate and user­friendly operation. pH measurements are automatically compen­sated for the temperature effect (ATC). These meters come equipped with a large, easy-to-read LCD which shows the pH or mV or ION (for pH301 only) with tempera­ture simultaneously. The display has graphic symbols to make the calibration procedure easy to follow. Calibration is not allowed until the electrode has stabilized to ensure accu­rate measurements. Through the RS 232 serial port you can transfer sample measurements to a PC.
32
FUNCTIONAL DESCRIPTION pH 300
THE FRONT PANEL
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
OFF
COMM
CFM
WRONG
CAL
LOG
ONONOFF
COMM
ppmpHpH
INTV
mV°C
m
1 2
BUF
ppm
Rel mV
MEM
BUF
3
1 2
3
mV°C
RANGE CAL DATA
CAL MS°C
CFM
DATE TIME
ppmpHpH
RANGE
CAL MS°C
CFM
s
pH METER
°C
INTV
m s
CAL DATA
pH METER
°C
pH300
MR
pH300
MR
Keyboard on the Left
COMM to display the current baud rate for RS232.
Keyboard on the Right
!!
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not connected) and to select the calibration buffer, the baud rate, the command pre­fix.
CAL to enter and exit the calibration mode. After the pH calibration, to enter the cali­bration date, use this key to toggle be­tween month, day and year.
CAL DATA to display the date of last pH calibration. This also displays the offset and the slope values expressed in mV.
CFM to confirm calibration values and to confirm baud rate and command prefix during communication setting mode.
MR to recall the stored value.
MS to hold and memorize the measured
value. The last reading will be frozen on the display until the key is released. At the same time this reading is stored in the memory.
RANGE to select the measurement range (pH or mV).
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
54
SPECIFICATIONS pH 300
FUNCTIONAL DESCRIPTION pH 301
THE FRONT PANEL
Range pH 0.00 to 14.00
ISE ±399.9 mV
ORP ±1999 mV
°C -9.9 to 120.0
Resolution pH 0.01
ISE 0.1 mV
ORP 1 mV
°C 0.1
Accuracy pH ±0.01 (@20°C/68°F) ISE ±0.2 mV
ORP ±1mV
°C ±0.5
Typical EMC pH ±0.01 Deviation ISE ±0.6 mV
ORP ±1 mV
°C ±0.5
pH Calibration Automatic 1 or 2 points with
3 memorized buffers
(4.01, 7.01, 10.01)
pH Offset Max ±60mV or ±1pH pH Slope 70-108% Temperature Automatic 1 or 2 points with
Calibration 2 memorized values
(0 and 50°C)
Temperature Automatic or manual Compensation from -9.9 to 120°C (14 to 248°F)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 76 6 9/ 2 W with 1 m probe (3.3') cable (included)
Input impedance 1012 Computer RS 232 (optoisolated)
Interface Power supply 12 VDC (CSA, UL & CE approved) Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3")
Shipping Weight 1.3 kg (2.9 lb.)
ppm
pH
DATE
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
Rel mV
CFM
MEM
WRONG
CAL
LOG
OFF
CAL
DATE COMM
RANGE
CFM
TIME
TIME
ppm
pH
INTV
mVmV°C
m
1 2
BUF
BUF
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
ppm
pH
DATE
TIME
ppm
pH
INTV
°C
1
m
2
s
3
LOG
CAL DATA
pH
ION
0.00/0.000
mV
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
pH301
pH / ION METER
°C
°C
pH301
pH / ION
METER
°C
°C
76
Keyboard on the Left
CAL to start the calibration mode for pH, mV and ION. To set or quit the date or time setting mode. Used also during log­ging (to start logging press pH, mV or ION plus LOG plus CAL key; to stop logging press LOG plus CAL key).
CFM to confirm calibration values.
COMM to display the current baud rate
and command prefix for RS232.
DATE to display the date.
RANGE to display the calibration buffer
or the temperature on the secondary LCD during calibration. During setting of the date, to toggle between month, day and year. During setting of the time, to toggle between minutes, seconds and logging in­terval.
TIME to display the time.
Keyboard on the Right
!!
!°C and
!!
""
"°C to manually set tempera-
""
ture (when the temperature probe is not connected) and to select the calibration buffer, the date and time, the baud rate, the command prefix.
CAL DATA to display the date of last pH calibration. This also displays the offset and the slope 1 or slope 2 values ex­pressed in mV.
ION to select the ION measurement range.
mV to select the mV measurement range.
pH to select the pH measurement range.
.00/.000 to select two or three decimal
points for pH measurements.
SPECIFICATIONS pH 301
Range pH -1.999 to 19.999
mV ±1999.9
ppm 0.001 to 19999
°C -9.9 to 120.0
Resolution pH 0.01 or 0.001
mV 0.1
ppm 0.001 (from 0.001 to 9.999)
0.01 (from 10.00 to 99.99)
0.1 (from 100.0 to 999.9) 1 (from 1000 to 19999)
°C 0.1
Accuracy pH ±0.002 (@20°C/68°F)m V ±0.1 (±799.9) / ±0.2 (±1999.9)
ppm ±0.5%
°C ±0.5
Typical EMC pH ±0.01 Deviation mV ±0.3
ppm ±1%
°C ±0.5
pH Calibration Automatic 1, 2 or 3 points with
6 memorized buffers
(1.68, 4.01, 6.86, 7.01, 9.18, 10.01)
pH Offset Max ±60mV or ±1pH pH Slope 70-108% ppm Calibration Automatic 1 or 2 points with
5 memorized buffers (0.1, 1, 10, 100, 1000)
ppm Offset Max ±1999.9mV ppm Slope From ±20mV to ±100 mV
[mV/log(0.001xppm)]
Temperature Automatic 1 or 2 points with Calibration 2 memorized values (0 and 50°C)
Temperature Automatic or manual (for pH only) Compensation from -9.9 to 120°C (14 to 248°F)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 7669/2W with 1 m probe (3.3') cable (included)
Input impedance 1012 Logging Interval 1, 15, 30 seconds or
1, 5, 30, 60, 120, 180 minutes
PC Interface RS 232 (optoisolated) Power supply 12 VDC (CSA, UL & CE approved) Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3") Shipping Weight 1.3 kg (2.9 lb.)
98
FUNCTIONAL DESCRIPTION pH 302
THE FRONT PANEL
Keyboard on the Left
0 to 9 for numeric and decimal data in- put.
CLEAR to cancel wrong data entry during programming.
Rel mV
ppm
DATE TIME
Keyboard on the Right
CAL to perform the pH, mV or tempera-
MEM
CFM
WRONG
CAL
LOG
OFF
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
ture calibration.
CFM/
!!
! to confirm pH/mV/temperature
!!
calibration values or to manually set tem­perature (when the temperature probe is not connected).
ENTER/
""
"°C to terminate the numeric data
""
entry or to manually set temperature (when the temperature probe is not connected) or to select the buffer value during pH calibration.
PAPER to pull the paper.
ppm
DATE TIME
Rel mV
MEM
CFM
WRONG
CAL
LOG
ONONOFF
8
9
7
6
5
4
1
3
2
0
CLEAR
.
ppmpHpH
INTV
mV°C
m
1 2
BUF
s
3
PAPER
PRINT
pH302
pH METER
RANGE
CFM /
CAL
PGM
TIME
ENTER /
PGM to select a program.
PRINT to obtain a printout (prints time,
date, sample number and pH or mV val­ues).
RANGE to select the measurement range (pH or mV) or to display the calibration buffer or the temperature on the second­ary LCD during calibration.
TIME to display the time and the date.
ON to turn the meter on
OFF to turn the meter off
Power LED to confirm the meter is on
1110
SPECIFICATIONS pH 302
Range pH 0.00 to 14.00
ISE ±399.9 mV
ORP ±1999 mV
°C -9.9 to 120.0
Resolution pH 0.01
ISE 0.1 mV
ORP 1 mV
°C 0.1
Accuracy pH ±0.01 (@20°C/68°F) ISE ±0.2 mV
ORP ±1mV
°C ±0.5
Typical EMC p H ±0.01 Deviation ISE ±0.3 mV
ORP ±1 mV
°C ±0.5
pH Calibration Automatic 1, or 2 points with
3 memorized buffers
(4.01,7.01,10.01)
Temperature Automatic 1 or 2 points with Calibration 2 memorized values
(0 and 50°C)
Temperature Automatic or manual Compensation from -9.9 to 120°C (14 to 248°F) (for pH only)
Electrode HI1131B glass body combination
refillable pH electrode (included)
Temperature HI 76 6 9/ 2 W with 1 m probe (3.3') cable (included)
Input impedance 1012 Printer Dot matrix, 44mm wide paper Languages English, German, French,
Spanish, Italian & Swedish
PC Interface RS 232 (optoisolated) Power supply 12 VDC (CSA, UL & CE approved)
Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 280x200x75mm (11x7.9x3") Shipping Weight 1.3 kg (2.9 lb.)
THE REAR PANEL
MEASUREMENT
REFERENCE ELECTRODE
TEMPERATURE
ELECTRODE
12345 6
RESET 12VDC RS 232
PROBE
ooooo
oooo
1. Reference Electrode Connector (Ø 4 mm)
2. BNC socket for measurement electrode
3. Socket for Temperature Probe
4. Reset Button
5. DC Power Socket (for HI710005 or HI710006)
6. RS 232 Connector
LCD FUNCTIONAL DESCRIPTION
Primary Display
CFM
CAL
LOG
Rel mV
MEM
WRONG
BUF
ppm
1 2
3
mV°C
ppmpHpH
DATE TIME
INTV
m s
Secondary Display
1312
OPERATIONAL GUIDE
ELECTRODE AND PROBE CONNECTIONS
For combination pH or ISE or ORP elec­trodes (with internal reference) plug the electrode's BNC to the socket provided (see page 11).
For electrode with separate reference, con­nect the measuring electrode's BNC to the BNC socket and the reference electrode's plug to the socket provided (see page 11).
For temperature measurements and auto­matic temperature compensation connect the temperature probe to the °C socket (see page 11).
If the temperature probe is not connected, when turned on, the meter will display a default tem-
°C
perature of 25°C or the last recorded tempera­ture (if the temperature probe is disconnected when the meter is on) on the secondary LCD. The "°C" indicator will be flashing.
POWER CONNECTION
Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (see page11).
Plug the adapter to the mains.
Note: The instrument uses the following con-
figuration.
+
-
It is recommendable to use the Hanna HI710005 or HI710006 voltage adapters (sup-
plied with the meters) that use the proper polarity configuration.
The meters can also be used with other adapt­ers. In this case, remember to check the correct polarity of your adapter before con­necting it to the meter.
Note: Make sure the main line is protected
by a fuse.
Press ON to turn the meter on. The power LED will remain lit until OFF is pressed to turn the meter off.
ON
OFF
When the meters are turned on, the micro­processor checks if any calibration data is in memory (i.e. if the offset is within ±60mV and the slope between 70-108%).
If the calibration data is valid, the meter will op-
pH
erate in the normal mea­surement mode and de­fault to pH measurement
°C
mode with the tempera­ture on the secondary LCD.
If the calibration data is not valid or the calibra-
pH
tion data is lost, the "CAL" indicator will be displayed to prompt the
CAL
°C
user to perform the pH calibration. If the user chooses not to per­form the calibration and presses CAL to quit from the calibration mode, the offset and the slope of the instrument would be set to 0 mV and 100% respectively.
1514
pH300, pH301 and pH302 use an EEPROM to retain the pH calibration, mV calibration, temperature calibration as well as the serial communication setting. The instrument will store the respective data after a calibration or serial communication setting, even when it is not plugged-in.
Press CAL to enter the setting mode and the month will be flashing.
CAL
DATE
For pH 302 only
pH302 uses a rechargeable battery to retain
the pH calibration data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off).
If you need to unplug the instrument, press OFF before
OFF
disconnecting the meter from the mains. Only in this case will the instrument retain the memorized data. Otherwise, the instrument needs to be re-calibrated. In the case of a power failure while the instrument is operational, re-cali­bration is required.
When the instrument is initially switched on, the microprocessor checks if any calibration data is in memory (i.e. if the offset is within ±1 pH and the slope between 70 - 108%).
If the calibration data are not memorized, the offset and slope of the instrument will be set to 0 mV and 100% respectively.
Use UP or DOWN to select the month.
°C
°C
Press RANGE and the day will start to flash.
RANGE
DATE
Use UP or DOWN to select the day.
°C
°C
Press RANGE and the year on the second­ary LCD will flash.
RANGE
DATE
SETTING THE DATE AND THE TIME AND THE LOGGING INTERVAL (for pH 301 only)
Press DATE to display the date. The month and the day will be displayed on the primary LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the year.
°C °C
1716
Press CAL to exit the date setting mode.
CAL
Press RANGE and the logging interval on the secondary LCD will come up.
RANGE
TIME
Press TIME to display the time. The hour and the minutes will be displayed on the primary LCD, the logging interval on the secondary one.
TIME
TIME
INTV
Press CAL to enter the setting mode and the hour will flash.
CAL
TIME
INTV
Use UP or DOWN to select the hour.
°C °C
Press RANGE and the minutes will flash.
RANGE
TIME
INTV
Use UP or DOWN to select the logging inter­val.
°C
°C
CAL
Press CAL to exit the time set­ting mode.
VIEWING THE DATE AND THE TIME (for pH 301 only)
Press DATE to display the date. The month and the day will be displayed on the primary LCD, the year on the secondary one.
DATE
DATE
Use UP or DOWN to select the minutes.
°C
°C
INTV
Press TIME to display the time. The hour and the minutes will be displayed on the primary LCD, the logging interval on the secondary one.
TIME
1918
TIME
INTV
s
SETTING THE DATE AND THE TIME AND THE LOGGING INTERVAL (for pH 302 only)
See program No. 1 (page 60).
VIEWING THE DATE AND THE TIME (for pH 302 only)
Press TIME to display the time (HH/MM).
TIME
Press TIME again to display the date (MM/ DD).
TIME
DATE
Allow 1 or 2 minutes for
pH
the electrode to stabi­lize.
°C
pH is displayed on the primary display and tem­perature on the second-
pH
ary one. If readings are out of
range "---" will appear on the LCD display.
For pH 301 only:
Press the .00/.000 key to change the resolu­tion from two to three decimal points.
0.00/0.000
pH
°C
TAKING pH MEASUREMENTS
Make sure that the instrument has been cali­brated for pH before taking pH measure­ments (see page 29).
For pH 300 and pH 302 only
:
RANGE
Select the pH measurement mode by pressing RANGE.
For pH 301 only
:
pH
Select the pH measurement mode by pressing pH.
To take a pH measurement simply submerge the tip (4cm/ 1½") of the pH electrode and the temperature probe into the sample to be tested.
4 cm
(1½")
It is recommended that the electrode is rinsed thoroughly for better conditioning. For this rinsing process, it is recommended to use a liberal amount of the sample to be measured.
If measurements are taken in different samples successively, it is recommended that the elec­trode and the temperature probe are rinsed thoroughly with deionized water or distilled water (if not available, tap water may also be used) and then blot dry between each mea­surement.
The pH reading is directly effected by tem­perature. In order for the meter to measure the pH accurately, temperature must be taken into consideration.
A perfect equilibrium between the pH elec­trode and the sample is reached in approxi­mately 15 minutes.
To use the meter's Automatic Temperature
Compensation feature, submerge the HI7669/2W temperature probe into the sample
2120
as close to the electrode as possible and wait for 1-2 minutes.
If you know the temperature of the sample to be tested you can compensate for it manu­ally.
If manual temperature compensation is de­sired the temperature probe must be dis­connected from the instrument.
The display will then show the default tem­perature of 25°C or the last recorded tempera-
°C
ture reading. "°C" is blinking. Note the temperature of the
C
°
sample using a ChecktempC or a glass thermometer. The temperature can now be adjusted with UP and DOWN. If the temperature
°C °C
probe is connected to the meter, the second­ary LCD figure dis­plays the measured temperature reading and the "°C" indicator
CFM/
ENTER/
does not blink. Press­ing UP or DOWN will not alter the tempera­ture reading.
VIEWING THE pH CALIBRATION DATA For pH 300:
Press CAL DATA and the meter will display the last pH calibration data.
CAL DATA
Press UP or DOWN to get this information in the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month, day and year)
b.Offset calibration
CAL
value expressed in mV and displayed on
mV
the primary LCD with the buffer used to calibrate the offset on the secondary LCD.
CAL
pH
1
The offset pH value should be between ±30 mV. For values between ±30 and ±60 mV change the reference electrolyte and repeat the calibration procedure. For values higher than +60 mV or lower than
-60mV the electrode is no longer reliable. See the electrode maintenance section on page 84.
c.Slope calibration
mV
value directly ex­pressed in mV/pH is displayed on the pri­mary LCD with the
CAL
pH
2
2nd buffer used to calibrate the offset on the secondary LCD.
For pH 301:
Press pH followed by
pH
CAL DATA
CAL DATA and the meter will display the
last pH calibration data.
2322
Press UP or DOWN to get this information in the following sequence:
°C
a. Date of the last pH
°C
DATE
calibration (month, day and year)
CAL
b. Time when the last
TIME
pH calibration was completed (hour and minutes).
CAL
c.Offset calibration
mV
value expressed in mV and displayed on the primary LCD with the buffer used
CAL
pH
1
to calibrate the off­set on the secondary LCD.
The offset pH value should be between ±30 mV. For values between ±30 and ±60 mV change the reference electrolyte and repeat the calibration procedure. For values higher than +60 mV or lower than
-60mV the electrode is no longer reliable. See the electrode maintenance section on page 84.
d. Acid slope calibra-
mV
tion value expressed in mV/pH and dis­played on the pri­mary LCD with the
CAL
pH
2
2nd buffer used to calibrate the offset on the secondary LCD.
e. Alkaline slope calibra-
tion value expressed
mV
in mV/pH and dis­played on the primary LCD with the 3 buffer used to cali-
rd
CAL
pH
3
brate the offset on the secondary LCD.
For pH 302:
See program #3 (page 66).
TAKING REDOX MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ORP measurements. An optional ORP electrode must be used to perform these measurements (see page 105).
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP), press RANGE.
For pH 301:
mV
To get into the mV mode, press mV.
To measure the mV of a so­lution simply submerge the ORP electrode tip (4 cm/1½") into the sample to be tested. Allow a few minutes for the
(1
½
")
reading to stabilize. The display will indicate the absolute mV value
(positive or negative).
For pH 300 and pH 302 only
When values are less
:
Rel mV
than ±400 mV, tenths of mV are displayed; for values greater than
°C
±400mV, the decimal digit disappears.
4 cm
2524
Note that the change in scale (from 0.1 mV to 1 mV as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value exceeds ±1999 mV (out of range).
Oxidation-reduction potential (REDOX) mea­surements provide the quantification of the oxidizing or reducing power of the sample tested.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
When not in use, the tip of the electrode should be kept moist and safe from any me­chanical stress which might cause damage to the gold/platinum junction.
TAKING ION MEASUREMENTS
pH300, pH301 and pH302 have the capabil-
ity to take ION measurements. An optional ISE electrode must be used to perform these measurements.
RANGE
For pH 300 or pH 302:
To get into the mV mode (ORP or ISE), press RANGE.
For pH 300 and pH 302:
The display will indicate the absolute mV value (positive or negative).
When values are less
Rel mV
than ±400 mV, tenths of mV are displayed; for
°C
values greater than ±400mV, the decimal digit disappears. Note that the change in scale (from 0.1 mV to 1 mV as resolution) is automatic.
For pH 301:
The display will indicate
ppm
the ppm value expressed in ppm or mV (depend-
°C
ing on the measured sample). Note that the change in scale (from 0.001ppm to 0.01ppm or
0.1 ppm as resolution) is automatic.
For all models:
Rel mV
"---" appears if the value exceeds ±1999 mV (out of range).
When not in use, the tip of the electrode should be kept moist and safe from any mechanical stress which might cause damage to the junction.
For pH 301:
To get into the ION mode, press ION.
Simply submerge the ISE electrode tip (4 cm/1½") into the sample to be tested. Al­low a few minutes for the read­ing to stabilize.
ION
VIEWING THE ION CALIBRATION FAC­TOR ON THE LCD (for pH 301 only)
Press ION followed
ION
CAL DATA
by CAL DATA and the meter will display the last ION cali- bration data.
°C °C
(1
4 cm
½
")
Press UP or DOWN to get this information in the fol­lowing sequence:
2726
a. Date of the last ION
DATE
calibration (month, day and year)
b. Time when the last
CAL
ION calibration was completed (hour and
TIME
minutes)
c.Offset calibration
value expressed in
CAL
INTV
mV and displayed on the primary LCD
mV
with the value used to calibrate the off­set on the second­ary LCD.
d. Slope calibration
value expressed in
CAL
mV
ppm
1
mV/[log (0.0001xppm)] and displayed on the primary LCD with
CAL
ppm
2
the 2nd value used to calibrate the slope on the secondary LCD.
TAKING TEMPERATURE MEASURE­MENTS
Taking a temperature measurement is very easy.
ON
Connect the temperature probe to the instrument and turn the instrument ON.
Dip the liquid/general purpose temperature probe HI7669/2W into the sample
Allow the reading on the sec­ondary display to stabilize (1 or 2 minutes).
pH
°C
PRINTING FUNCTION (for pH 302 only)
Press PRINT to get a printout at
PRINT
any time. All the printouts will be complete
with date, time, sample No., tem­perature and measurement value (pH or mV depending on the se­lected measurement range).
DATE 1/02 TIME 12:21 SAMPLE 1 °C 17.9 pH 7.18
DATE 1/02 TIME 12:40 SAMPLE 5 °C 20.5 mV -350.0
HOLD FUNCTION (for pH 300 only)
To hold and memorize the measured value, press MS. The last reading will be frozen on the display until the key is released. At the same time, this reading is stored in the memory. During this operation, the "MEM" indicator is lit.
MS
MEM
pH
°C
2928
MR
To recall the stored memory value, press MR and the memo­rized reading will be displayed with the MEM indicator lit.
RESET BUTTON
The RESET button (see page 11) is used when the instrument displays erroneous mes­sages due to strong electrical interference or when the instrument's power supply was dis­connected before the meter was switched off.
It is necessary to press the reset button and restart the entire operation.
Calibration points should remain memorized. It is recommended to verify calibration be­fore proceeding.
pH CALIBRATION
For greatest accuracy, it is recommended that the instrument is calibrated frequently.
For a faster operation, it is possible to stan­dardize the electrode at a value close to the expected sample value or at pH 7.01 only (one point calibration), but it is always good practice to calibrate at least 2 points.
For pH 301 only
:
Due to electrode conditioning time the elec­trode must be kept immersed for a few sec­onds to stabilize. The meters are equipped with a stability indicator and the user will be guided step by step with easy indications on the display during the pH calibration. This will make the calibration a simple and error-free procedure.
The standard calibration program of the meter is prepared for 3 (maximum) buffers. The selectable pH values are: 1.68, 4.01, 6.86,
7.01, 9.18, 10.01.
INITIAL PREPARATION
Pour small quantities of
HI 7004
pH7.01 (HI7007/HI8007) and pH4.01 (HI7004/HI8004) and/
HI 7004
HI 7007
or pH10.01 (HI7010/HI8010) solution into individual beakers. If possible, use plastic beakers to minimize any EMC interferences.
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the electrode, the second one for calibration.
By doing this,
RINSE
CALIBRATION
contamination between the buffers is mini-
HI 7007
HI 7007
mized. To get accurate readings, use pH 7.01 and
3130
pH 4.01 if you are going to measure acid samples, or pH 7.01 and pH 10.01 for alka­line measurements or, for pH301 only, per­form a 3-point calibration for the entire range.
PROCEDURE
RANGE
Press RANGE (for pH300 and pH 302) to select the pH measurement mode.
Press pH (for pH301) to select the pH measure-
pH
ment mode.
Remove the protective cap from the pH electrode.
ONE POINT CALIBRATION (OFFSET)
Immerse the pH electrode into the selected buffer (e.g. pH 7.01) buffer solution and shake briefly.
Note: the electrode should
HI 7007
be submerged ap­proximately 4 cm (1½") into the solu­tion. The tempera-
(1
ture probe should be located as close as
4 cm
½
")
possible to the pH electrode.
Press CAL. The primary LCD will display the pH value using the current offset and
slope, with the "CAL" and "
CAL
1
BUF
" indicators lit.
CAL
pH
pH
1
BUF
For pH 302 only: the printer will also print step by step instructions for the calibra­tion procedure.
Dip the electro­de in the pH 7 solution; wait 30 sec and then press CFM
Select the buffer value by pressing UP or DOWN or just DOWN (for pH 302) until the secondary LCD displays the value of the first solution (e.g. 7.01”).
°C
ENTER/
°C
pH
CAL
1
BUF
Only when the read­ing is stable (less
" in-
CFM
CFM/
CFM
than ±0.03 variation in a 10 seconds time
interval) the "
1
BUF
dicator will stop flash­ing (after about 30 seconds) and the "CFM" indicator will start blinking.
Press CFM to con­firm the calibration.
If the reading is not close to the selected buffer, "WRONG
WRONG
1
BUF
"
and "WRONG " will blink alternately. If the reading is close to the selected buffer (±1pH), the meter stores the reading
WRONG
BUF
3332
CFM/
(and adjusts the offset point). The buffer value is
pH
then displayed on the primary LCD and the secondary LCD will display the ex-
CAL
2
BUF
pected second buffer value. For pH 302 only: the printer will print
further instructions for the calibration pro­cedure.
Now dip the e­lectrode in 2nd solution, wait 30 sec and then press CFM
Press CAL and the calibra­tion process is ended with just the offset of the meter calibrated. For best accuracy however, it is recommended that a two-point calibration is performed.
For pH 300 only
:
The calibration process is complete and the dis­play will ask for the date
CAL
entry.
°C
Use UP or DOWN to select the month, the day and the year.
Use CAL to toggle between month, day and year.
CAL
CAL
°C
DATE
After entering the calibration date, press CFM and the meter will return to the normal operating
pH
mode.
Note: during calibration, the
RANGE
RANGE
secondary LCD displays the selected buffer value. By pressing RANGE the tempera­ture value can be also displayed. This will allow you to check the buffer temperature during calibration.
TWO-POINT CALIBRATION
A two-point calibration is recommended for best accuracy. As a first point use pH 7.01 (or pH 6.86 can also be used for pH 301). As a second point use pH 4.01 (or pH 1.68 for pH 301 only) if you are going to measure acid samples (pH 7 or less), use pH10.01 (or pH 9.18 for pH301 only) if you are going to measure alkaline samples.
CAL
Proceed as described above for one point calibration but do not quit calibration by pressing CAL at the end.
After the first calibration point is confirmed, immerse the pH electrode into the second buffer solution (pH
4.01 or pH10.01; for pH301
HI 7004
only, the second buffer can also be pH 1.68 or pH 9.18) and shake briefly.
Note: the electrode should
be submerged ap­proximately 4 cm (1½") into the solu-
(1
4 cm
½
")
tion. The temperature probe should be located as close as possible to the pH electrode.
3534
Loading...
+ 39 hidden pages