Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instruments.
This manual will provide you with the necessary information for correct use of
the instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or turn the back cover for our worldwide contact list.
These instruments are in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
pH 210, pH 211, pH 212 and pH 213 are guaranteed for two years against
defects in workmanship and materials when used for their intended purpose and
maintained according to instructions. Electrodes and probes are guaranteed for
six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance
is not covered.
If service is required, contact the dealer from whom you purchased the instrument.
If under warranty, report the model number, date of purchase, serial number and
the nature of the problem. If the repair is not covered by the warranty, you will be
notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical
Service department and then send it with shipping costs prepaid. When shipping
any instrument, make sure it is properly packed for complete protection.
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your Dealer or the nearest Hanna
Customer Service Center.
Each instrument is supplied with:
• HI 1131B Glass-body Combination pH Electrode
• HI 7669/2W Temperature Probe
• HI 76404 Electrode Holder
• pH 4.01 & 7.01 Buffer Solutions, 20 mL each
• HI 7071S Electrolyte Solution
• 12VDC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The Hanna pH 210, pH 211, pH 212 and pH 213 are microprocessor-
based pH and Temperature bench meters.
pH 211 and pH 213 can also measure ion concentration (ISE) and
Oxidation Reduction Potential (ORP) in the mV range.
pH measurements are compensated for temperature effect manually or
automatically with the HI 7669/2W temperature probe.
The instrument is equipped with a large easy-to-read LCD which shows
the pH (or mV) and temperature simultaneously, together with graphic
symbols.
The calibration process is guided step by step through clear indications on
the LCD. A stability indicator makes the calibration procedure error-free.
Through the RS232 serial port data can be transferred to a PC
(pH 212 and pH 213 only).
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
23
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
pp
HH
210 AND 210 AND
p
H
210 AND
pp
HH
210 AND 210 AND
pp
HH
211 211
p
H
211
pp
HH
211 211
pp
H 210 AND H 210 AND
p
H 210 AND
pp
H 210 AND H 210 AND
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
pp
H 211H 211
p
H 211
pp
H 211H 211
CAL
/
CFM
/
CFM
Primary LCD
StartCalibration
st
Select1 Buffer
st
Confirm1 Buffer
nd
Select2 Buffer
nd
Confirm2 Buffer
Secondary LCD
C
RANGE
9
8
Front Panel
1
2
3
4
5
6
7
Rear Panel
pH
FunctionspH Calibration
MEM
MemorizeReading
CFM
CAL
MEM
12VDC
POWER
10
MR
MemoryRecall
RANGE
SelectpH / ORP
CC
MR
pH 211
Microprocessor pH Meter
INPUTTEMPREF
121311
1) Liquid Crystal Display (LCD).
2) CFM key, to confirm different values.
3) CAL key, to enter or exit/escape calibration mode.
4)°C key, to manually decrease temperature or select pH buffer.
5)°C key, to manually increase temperature or select pH buffer.
6) MR key, to recall the stored value.
7) MEM key, to store a value into memory.
8) RANGE key, to select measurement range (pH 211 only).
9) ON/OFF switch.
10) Power adapter socket.
11) BNC electrode connector.
12) Electrode reference socket.
13) Temperature probe socket.
EGNAR
±)ylno112Hp(Vm9.993
±)ylno112Hp(Vm0002
NOITULOSER
YCARUCCA
F°86/C°02@
±(Vm2.0±)Vm9.993
±(Vm1±)Vm0002
±)C°0.001–0.0(C°5.0
noitarbilaCHp
noitasnepmocerutarepmeT
edortcelEHp)dedulcni(B1311IH
eborperutarepmeT)dedulcni(W2/9667IH
ecnadepmitupnI01
ylppusrewoP)dedulcni(retpadaCDV21
snoisnemiD)”9.2x1.7x4.9(mm47x281x042
thgieW
tnemnorivnE
ytnarraWsraey2
1±)edistuo(C°
±Hp10.0
Hp00.61ot00.2–
C°0.021ot9.9–
Hp10.0
)ylno112Hp(Vm1.0
)ylno112Hp(Vm1
C°1.0
)rorreeborpgnidulcxe(
sreffub5,noitarbilactniop2ro1
elbaliava
)10.01,81.9,10.7,68.6,10.4(
:morfcitamotuArolaunaM
)F°0.842ot2.41(C°0.021ot9.9–
21
smho
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
)F°221ot23(C°05ot0
gnisnednoc-nonHR%59.xam
45
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
pp
HH
212 AND 212 AND
p
H
212 AND
pp
HH
212 AND 212 AND
pp
HH
213213
p
H
213
pp
HH
213213
pp
HH
212 AND 212 AND
p
H
212 AND
pp
HH
212 AND 212 AND
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
pp
HH
213213
p
H
213
pp
HH
213213
Front Panel
1
2
3
4
5
6
7
Rear Panel
CAL
MEM
12VDC
POWER
11
MR
MemoryRecall
RANGE
SelectpH / ORP
COMM
RS232Communication
CC
MR
pH 213
Microprocessor pH Meter
REF
INPUTTEMP
13
12
FunctionspH Calibration
MEM
MemorizeReading
CFM
Primary LCD
CAL
StartCalibration
/
Select1 Buffer
CFM
Confirm1 Buffer
/
Select2 Buffer
CFM
Confirm2 Buffer
1415
st
st
nd
nd
RS232
Secondary LCD
RANGE
COMM
10
9
8
1) Liquid Crystal Display (LCD).
2) CFM key, to confirm different values.
3) CAL key, to enter or exit/escape calibration mode.
4)°C key, to manually decrease temperature or select pH buffer.
5)°C key, to manually increase temperature or select pH buffer.
6) MR key, to recall the stored value.
7) MEM key, to store a value into memory.
8) COMM key, to set the baud rate and the command prefix.
9) RANGE key, to select measurement range.
10) ON/OFF switch.
11) Power adapter socket.
12) BNC electrode connector.
13) Electrode reference socket.
14) Temperature probe socket.
15) RS232 serial communication connector.
EGNAR
±)ylno312Hp(Vm9.999
±)ylno312Hp(Vm0002
NOITULOSER
YCARUCCA
F°86/C°02@
±(Vm5.0±)Vm9.999
±(Vm1±)Vm0002
±)C°0.001–0.0(C°5.0
±)edistuo(C°1
noitarbilaCHp
erutarepmeT
noitasnepmoc
edortcelEHp)dedulcni(B1311IH
eborperutarepmeT)dedulcni(W2/9667IH
ecnadepmitupnI01
ecafretniretupmoCdetalosi-otpo232SR
ylppusrewoP)dedulcni(retpadaCDV21
snoisnemiD)”9.2x1.7x4.9(mm47x281x042
thgieW
tnemnorivnE
ytnarraWsraey2
±Hp200.0
±Hp10.0
Hp00.61ot00.2–
Hp000.61ot000.2–
C°0.021ot9.9–
Hp10.0
Hp100.0
)ylno312Hp(Vm1.0
)ylno312Hp(Vm1
C°1.0
)rorreeborpgnidulcxe(
sreffub5,noitarbilactniop2ro1
elbaliava
)10.01,81.9,10.7,68.6,10.4(
:morfcitamotuArolaunaM
)F°842ot2.41(C°021ot9.9–
21
smho
;)bl5.2(gK1.1
)bl5.5(gK5.2redlohhtiwtik
)F°221ot23(C°05ot0
gnisnednoc-nonHR%59.xam
67
OPERATIONAL GUIDEOPERATIONAL GUIDE
C
C
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
POWER CONNECTIONPOWER CONNECTION
POWER CONNECTION
POWER CONNECTIONPOWER CONNECTION
Plug the 12 VDC adapter into the power supply socket.
Notes: • These instruments use non volatile memory to retain the
pH, mV, temperature calibrations and all other settings,
even when unplugged.
• Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONS
ELECTRODE AND PROBE CONNECTIONSELECTRODE AND PROBE CONNECTIONS
For pH or ORP measurements connect an electrode with internal
reference to the BNC connector on the back of the instrument.
For electrodes with a separate reference connect the electrode’s BNC to the
BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
INSTRUMENT START-UPINSTRUMENT START-UP
INSTRUMENT START-UP
INSTRUMENT START-UPINSTRUMENT START-UP
• Turn the instrument on by pressing the ON/OFF
switch.
• All LCD tags are displayed and a beep is heard
while the instruments perform a self test (pH 212
and pH 213 only).
pH 210 & pH 211
• The pH is displayed on the primary LCD and the temperature on
the secondary LCD.
pH
C
• If the pH reading is out of range, “----” will be displayed on the
LCD.
If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized water
or tap water and then with some of the next sample in order to
prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the
pH accurately, the temperature effect must be compensated for.
To use the Automatic Temperature Compensation feature, connect
and submerge the HI 7669/2W temperature probe into the sample
as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known,simple manual temperaturecompensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature
of 25 °C or the last recorded temperature reading
C
with the “°C” tag blinking. The temperature can now
be adjusted with the ARROW keys (from -9.9 °C to
120.0 °C).
pH 212 & pH 213
pH MEASUREMENTSpH MEASUREMENTS
pH MEASUREMENTS
pH MEASUREMENTSpH MEASUREMENTS
Make sure the instrument has been calibrated before
taking pH measurements.
• Submerge the electrode tip and the temperature
probe approximately 4 cm (1½”) into the sample
to be tested and stir gently. Allow time for the
electrode to stabilize.
89
ORP MEASUREMENTS (ORP MEASUREMENTS (
ORP MEASUREMENTS (
ORP MEASUREMENTS (ORP MEASUREMENTS (
pp
H 211H 211
p
H 211
pp
H 211H 211
and p and p
and p
and p and p
H 213 H 213
H 213
H 213 H 213
onlyonly
only
onlyonly
An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-reduction potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface
of the ORP electrode must be clean and smooth.
• Press RANGE to enter mV range.
• Submerge the tip of the ORP electrode (4 cm/
1½") into the sample to be tested and allow a
few seconds for the reading to stabilize.
))
)
))
• The instrument displays the mV reading on the primary LCD and
C
C
the temperature on the secondary LCD.
mV
C
• If the reading is out of range, “----” will be displayed on the
LCD.
TEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTS
TEMPERATURE MEASUREMENTS
TEMPERATURE MEASUREMENTSTEMPERATURE MEASUREMENTS
Connect the HI 7669/2W
temperature probe to the
appropriate socket and turn the instrument on.
Immerse the temperature
probe
into the sample and
allow the reading on the secondary LCD to stabilize.
MEMORY FUNCTIONMEMORY FUNCTION
MEMORY FUNCTION
MEMORY FUNCTIONMEMORY FUNCTION
Press and hold down MEM. The last reading will appear on the LCD
along with “MEM” tag until MEM is released.
MEM
MR
Press MR and the previously memorized reading will be displayed
with “MEM” tag on the LCD.
BEEP SETTINGBEEP SETTING
BEEP SETTING
BEEP SETTINGBEEP SETTING
((
pp
(
p
((
pp
H 212 H 212
H 212
H 212 H 212
andand
and
andand
pp
H 213 H 213
p
H 213
pp
H 213 H 213
onlyonly
only
onlyonly
))
)
))
Press and hold down °C and MEM keys and then power on the
instrument. After about 5 seconds the meter will display BEEP setting
menu. Use the ARROW keys to change the current setting. PressCFM to confirm or CAL to escape without saving.
pp
H CALIBRATIONH CALIBRATION
p
H CALIBRATION
pp
H CALIBRATIONH CALIBRATION
Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
PREPARATIONPREPARATION
PREPARATION
PREPARATIONPREPARATION
Pour small quantities of the buffer solutions into clean beakers. If
possible use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two
beakers for each buffer solution: one for rinsing the electrode and one for
calibration.
If you are measuring in the acidic range, use pH 7.01 as first buffer
and pH 4.01 as second buffer. If you are measuring in the alkaline
range, use pH 7.01 as first buffer and pH 10.01 or pH 9.18 as
second buffer.
PROCEDUREPROCEDURE
PROCEDURE
PROCEDUREPROCEDURE
Calibration has a choice of five memorized buffers: pH 4.01, 6.86,
7.01, 9.18 and 10.01.
It is recommended to perform a two-point calibration. However, one-
point calibration is also provided by the instruments.
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
TWO-POINT CALIBRATION
TWO-POINT CALIBRATIONTWO-POINT CALIBRATION
• Submerge the pH electrode and the temperature probe approximately 4 cm (1½") into a
buffer solution and stir gently. The temperature
probe should be close to the pH electrode.
• Press CAL. The “CAL” and “
” tags will appear and “7.01”
buffer will be displayed on the secondary LCD.
pH
CAL
CAL
1
BUF
BUFFER pH
• If necessary, press the ARROW keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
1011
• When the reading is stable and close to the selected buffer,
“READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm calibration.
• The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
CFM
the second expected buffer value.
pH
CAL
2
BUF
BUFFER pH
Note: The instruments will automatically skip the buffer used for the
first point. It also skips 6.86 if 7.01 buffer was used and
viceversa. Likewise, it will skip 9.18 if 10.01 buffer was used and
viceversa.
• After the first calibration point is confirmed,
immerse the pH electrode and the temperature
probe approximately 4 cm (1½”) into the
second buffer solution and stir gently. The
temperature probe should be close to the pH
electrode.
• If necessary, press the ARROW keys to select a different buffer value.
• Press RANGE to display the temperature reading on the LCD during
calibration (pH 211, pH 212, pH213 only).
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
ONE-POINT CALIBRATION
ONE-POINT CALIBRATIONONE-POINT CALIBRATION
• Proceed as described in “Two-point calibration” section.
• Press CAL after the first calibration point
was confirmed.
The instrument will return to measurement mode
CAL
and will memorize the one-point calibration data.
Note: Press and hold down CFM and then press CAL. The instrument
sets default calibration parameters, displays “CLR” message for a
few seconds and then returns to normal measurement mode.
C
C
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable, “READY” tag will appear and “CFM”
tag will blink.
• Press CFM to confirm calibration. The instrument
will return to measurement mode.
CFM
Notes: • If the value measured by the meter is not close to the
selected buffer, “WRONG ” and “WRONG ” tags will
blink alternately. In this case check if the correct buffer
has been used, or regenerate the electrode by following
the cleaning procedure (see page 22). If necessary,
change the buffer or the electrode.
• The “WRONG” tag and temperature value are displayed
blinking if the temperature reading is out of the defined
temperature range of the buffer. Calibration cannot be
confirmed in this situation.
1213
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
C
C
TEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
((
for technical personnel onlyfor technical personnel only
(
for technical personnel only
((
for technical personnel onlyfor technical personnel only
))
)
))
All the instruments are factory calibrated for temperature.
Hanna’s temperature probes are interchangeable and no tempera-
ture calibration is needed when they are replaced.
If the temperature measurements are inaccurate, temperature
recalibration should be performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions bellow.
• Prepare a vessel containing ice and water and another one
containing hot water (at a temperature of around 50 ºC). Place
insulation material around the vessels to minimize temperature
changes.
• Use a calibrated thermometer with a resolution of 0.1 ºC as a
reference thermometer.
• With the instrument off, press and hold down the CAL&MEM
keys, then power on the instrument. The “CAL” tag will appear
and the secondary LCD will show 0.0 ºC.
• Immerse the temperature probe in the second vessel as near as
possible to the reference thermometer. Allow a few seconds for the
probe to stabilize.
• Use the ARROW keys to set the reading on the secondary LCD to
that of the hot water.
• When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The instrument returns
to measurement mode.
Note: If the reading is not close to the selected
CFM
calibration point, “WRONG” tag will blink.
Change the temperature probe and restart
calibration.
CAL
C
• Immerse the temperature probe in the vessel with ice and water
as near as possible to the reference thermometer. Allow a few
seconds for the probe to stabilize.
• Use the ARROW keys to set the reading on the secondary LCD to
that of ice and water, measured by the reference thermometer.
C
C
• When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The secondary LCD will show 50.0 ºC.
CAL
CFM
C
1415
C
C
mm
V CALIBRATION (V CALIBRATION (
m
V CALIBRATION (
mm
V CALIBRATION (V CALIBRATION (
pp
H 213 H 213
p
H 213
pp
H 213 H 213
(for technical personnel only)
onlyonly
only
onlyonly
))
)
))
(p(p
(p
(p(p
PC INTERFACEPC INTERFACE
PC INTERFACE
PC INTERFACEPC INTERFACE
H 212 & H 212 &
H 212 &
H 212 & H 212 &
pp
H 213 H 213
p
H 213
pp
H 213 H 213
onlyonly
only
onlyonly
))
)
))
All the instruments are factory calibrated for mV.
Hanna’s ORP electrodes are interchangeable and no mV calibration
is needed when they are replaced.
If the mV measurements are inaccurate, mV recalibration should be
performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions below.
A two or three-point calibration can be performed at 0.0 mV, 600.0
mV and 1800.0 mV.
• Attach to the BNC connector a mV simulator with an accuracy of
±0.1 mV.
• With the instrument off, press and hold down the MR&CFM
keys, then power on the instrument. The “CAL” tag will appear
and the secondary LCD will show 0.0 mV.
• Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The secondary LCD will display 600 mV.
• Set 600.0 mV on the simulator.
When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The secondary LCD will display 1800 mV.
• Set 1800.0 mV on the simulator.
When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.
• Press CFM to confirm. The instrument returns to measurement
mode.
Notes: • If the reading is not close to the selected calibration point,
“WRONG” tag will blink. Verify calibration condition or
contact your vendor if you can not calibrate.
• Press CAL in any moment of the calibration process. The
instrument will return to measurement mode. If calibration process is stopped after 600 mV is confirmed, the
600 mV range is calibrated and calibration parameters
are memorized.
Data transmission from the instrument to the PC can be done with
the HI 92000 Windows® compatible software (optional). HI 92000
also offers graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for
further analysis.
To connect your instrument to a PC, use the optional Hanna HI 920010
cable connector. Make sure that your instrument is switched off and
plug one connector to the instrument RS232C socket and the other into
the serial port of your PC.
Notes: • Other cables than HI 920010 may use a different
configuration. In this case communication between instrument and PC may not be possible.
• If you are not using Hanna Instruments HI 92000
software, please see the following instructions.
SETTING THE BAUD RATE AND THE COMMANDSETTING THE BAUD RATE AND THE COMMAND
SETTING THE BAUD RATE AND THE COMMAND
SETTING THE BAUD RATE AND THE COMMANDSETTING THE BAUD RATE AND THE COMMAND
PREFIXPREFIX
PREFIX
PREFIXPREFIX
• To set the instrument baud rate, press COMM
while in measurement mode. The primary LCD
will show the current baud rate.
• Use the ARROW keys to change the displayed value (150; 300;
600; 1200; 2400 – factory setting; 4800; 9600).
• Press CFM to confirm. The primary LCD will show the current
command prefix (factory setting is 16).
• Use the ARROW keys to change the displayed value (between 0
and 47).
• Press CFM to confirm. The meter returns to
measurement mode.
Note: The command prefix must not be changed if
HI 92000 PC software is used.
COMM
CFM
1617
SENDING COMMANDS FROM PCSENDING COMMANDS FROM PC
SENDING COMMANDS FROM PC
SENDING COMMANDS FROM PCSENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any
terminal program. Use HI 920010 cable to connect the instrument to
a PC, start the terminal program and set the communication options
as follows: 8, N, 1, no flow control.
COMMAND TYPESCOMMAND TYPES
COMMAND TYPES
COMMAND TYPESCOMMAND TYPES
To send a command to the instrument, follow the next scheme:
<DLE> <
command>
<CR>
This line makes the computer send a Data Link Escape character, the
command expressed as a 3-character sequence and a CR character.
Note: All the terminal programs that support the ANSI escape
sequence, represent the DLE character by the string '^P' and
the CR character by the string '^M'.
SIMPLE COMMANDSSIMPLE COMMANDS
SIMPLE COMMANDS
SIMPLE COMMANDSSIMPLE COMMANDS
PHRsets the range to pH
MVRsets the range to mV (pH 213 only)
CALis equivalent to pressing CAL
CFMis equivalent to pressing CFM
UPCis equivalent to pressing the °C key
DWCis equivalent to pressing the °C key
MEMis equivalent to pressing MEM
MRRis equivalent to pressing MR
COMis equivalent to pressing COMM
COMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWER
COMMANDS REQUIRING AN ANSWER
COMMANDS REQUIRING AN ANSWERCOMMANDS REQUIRING AN ANSWER
pH?Causes the instrument to send the pH reading ("Err 1" is sent
if out of range). If the range is set to mV, "Err 6" is sent.
MV?Causes the instrument to send the mV reading ("Err 2" is sent
if out of range). If the range is set to pH, "Err 6" is sent.
TM?Causes the instrument to send the temperature reading
("Err 3" is sent if out of range).
Notes: • Either small or capital letters can be used.
• The characters sent by the meter
are always capital letters.
• A “CAN” character (ASCII Code 24) is sent when the
instrument receives an unknown or a corrupted command.
Note: Invalid commands will be ignored.
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
p
H BUFFER TEMPERATURE
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
DEPENDENCEDEPENDENCE
DEPENDENCE
DEPENDENCEDEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions. During calibration the instrument will automatically calibrate
to the pH value corresponding to the measured or set temperature.
PMETSREFFUBHp
CºFº10.468.610.781.910.01
02310.489.631.764.923.01
51400.459.601.793.942.01
010500.429.670.733.981.01
519500.409.650.772.921.01
028600.488.630.722.960.01
527710.468.610.781.910.01
036820.458.600.741.969.9
535930.448.699.611.929.9
0440140.448.689.670.988.9
5431150.438.689.640.958.9
0522160.438.689.610.928.9
5513180.448.689.699.897.9
0604190.448.689.679.877.9
5694111.448.699.659.867.9
0785121.458.699.639.857.9
5776141.468.600.719.847.9
0867161.478.610.798.847.9
5858171.478.620.778.847.9
0949191.488.630.758.857.9
5930202.498.640.738.867.9
During calibration the instrument will display the pH buffer value at
25 ºC.
1819
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
PREPARATION PROCEDUREPREPARATION PROCEDURE
PREPARATION PROCEDURE
PREPARATION PROCEDUREPREPARATION PROCEDURE
Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes. They will disappear when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by “shaking down” the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or
HI 80300
Storage Solution for at least one hour.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1") below the
fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for
double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte
Solution for single junction electrodes.
For faster response, unscrew the fill hole screw during measurements.
®®
®
For AmpHelFor AmpHel
For AmpHel
For AmpHelFor AmpHel
®®
electrodes: electrodes:
electrodes:
electrodes: electrodes:
If the electrode does not respond to pH changes, the battery is run
down and the electrode should be replaced.
MEASUREMENTMEASUREMENT
MEASUREMENT
MEASUREMENTMEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom
4 cm /1½”) in the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.
STORAGE STORAGE
STORAGE
STORAGE STORAGE
PROCEDUREPROCEDURE
PROCEDURE
PROCEDUREPROCEDURE
To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300
or HI 80300 Storage Solution or, in its absence, Filling Solution
(HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for
double junction electrodes). Follow the Preparation Procedure on page
20 before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.
PERIODIC MAINTENANCEPERIODIC MAINTENANCE
PERIODIC MAINTENANCE
PERIODIC MAINTENANCEPERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to
the instrument must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry. If any scratches or cracks
are present, replace the electrode. Rinse off any salt deposits with
water.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071
for single junction or HI 7082 or HI 8082 for double junction
electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
2021
CLEANING PROCEDURECLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURECLEANING PROCEDURE
•
General
Soak in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.
•
Protein
Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.
•
Inorganic
Soak in Hanna HI 7074 Inorganic Cleaning
Solution for 15 minutes.
•
Oil/grease
Rinse with Hanna HI 7077 or HI 8077 Oil and
Fat Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte (not necessary for gel-filled electrodes)
and soak the electrode in HI 70300 or HI 80300 Storage Solution for
at least 1 hour before taking measurements.
The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes more time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 25 °C.
Since the resistance of the pH electrode is in the range of 50-200 Mohms,
the current across the membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and
static discharges are detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature1- 3 years
90 °CLess than 4 months
120 °CLess than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interfe-rence is called
alkaline error and causes the pH to be underestimated. Hanna’s glass
formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 °C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
H SENSITIVE GLASSH SENSITIVE GLASS
p
H SENSITIVE GLASS
pp
H SENSITIVE GLASSH SENSITIVE GLASS
+
+
13.00
13.50
14.00
12.50
13.00
13.50
14.00
2425
0.10
0.14
0.20
0.10
0.18
0.29
0.40
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
pH BUFFER SOLUTIONS
HI 70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcs
HI 70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcs
HI 70010P pH 10.01 Buffer Sachets, 20 mL, 25 pcs
HI 7004L pH 4.01 Buffer Solution, 500 mL bottle
HI 7006L pH 6.86 Buffer Solution, 500 mL bottle
HI 7007L pH 7.01 Buffer Solution, 500 mL bottle
HI 7009L pH 9.18 Buffer Solution, 500 mL bottle
HI 7010L pH 10.01 Buffer Solution, 500 mL bottle
HI 8004L pH 4.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8006L pH 6.86 Buffer Sol. in FDA approved bottle, 500 mL
HI 8007L pH 7.01 Buffer Sol. in FDA approved bottle, 500 mL
HI 8009L pH 9.18 Buffer Sol. in FDA approved bottle, 500 mL
HI 8010L pH 10.01 Buffer Sol. in FDA approved bottle, 500 mL
ELECTRODE STORAGE SOLUTIONS
HI 70300L Storage Solution, 460 mL bottle
HI 80300L Storage Solution in FDA approved bottle, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Rinse Sachets, 20 mL, 25 pcs
HI 7061L General Cleaning Solution, 460 mL bottle
HI 7073L Protein Cleaning Solution, 460 mL bottle
HI 7074L Inorganic Cleaning Solution, 460 mL bottle
HI 7077L Oil & Fat Cleaning Solution, 460 mL bottle
HI 8061L General Cleaning Sol. in FDA approved bottle, 460 mL
HI 8073L Protein Cleaning Solution in FDA approved bottle, 460 mL
HI 8077L Oil & Fat Cleaning Sol. in FDA approved bottle, 460 mL
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI 70713.5M KCl+AgCl Electrolyte, 4x50 mL, for single junction
electrodes
HI 70721M KNO3 Electrolyte, 4x50 mL
HI 70823.5M KCl Electrolyte, 4x50 mL, for double junction
electrodes
HI 80713.5M KCl+AgCl Electrolyte in FDA approved bottle,
4x50 mL, for single junction electrodes
HI 80721M KNO3 Electrolyte in FDA approved bottle, 4x50 mL
HI 80823.5M KCl Electrolyte in FDA approved bottle, 4x50 mL
95
ORP PRETREATMENT SOLUTIONS
HI 7091LReducing Pretreatment Solution, 460 mL bottle
HI 7092LOxidizing Pretreatment Solution, 460 mL bottle
pH ELECTRODES
All electrodes part numbers ending with B are supplied with BNC
connector and 1 m (3.3') cable, as shown below :
Plastic-body (Ultem®), gel-filled, combination gold ORP electrode.
Use: general purpose.
12 mm
0.5"
HI 4430
120 mm
"S"VERSION
4.7"
Consult the Hanna General Catalog for more electrodes with screwtype or BNC connectors.
FC 911
110 mm
4.3"
Ultem® is registered Trademark of "General Electric Co."
Kynar® is registered Trademark of "Pennwalt Corp."
2829
EXTENSION CABLE FOR SCREW-TYPE ELECTRODES
(SCREW TO BNC ADAPTER)
HI 7855/1Extension cable 1 m (3.3') long
HI 7855/3Extension cable 3 m (9.9') long
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
OTHER ACCESSORIES
HI 710005 Voltage adapter from 115 VAC to 12 VDC (USA plug)
HI 710006 Voltage adapter from 230 VAC to 12 VDC (European plug)
HI 710012 Voltage adapter from 240 VAC to 12 VDC (UK plug)
HI 710013 Voltage adapter from 230 VAC to 12 VDC (South Africa plug)
HI 710014 Voltage adapter from 230 VAC to 12 VDC (Australia plug)
HI 76405 Electrode holder
RECOMMENDATIONS FOR USERS
Before using these products, make sure they are entirely suitable for
the environment in which they are used.
Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment, requiring the
operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may
degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltages
at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in
microwave ovens.
HI 8427pH and ORP electrode simulator with 1 m (3.3') coaxial
cable ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 7669/2WTemperature probe with 1 m (3.3') cable
HI 92000 Windows® compatible software application
HI 920010 9 to 9-pin RS232 cable
Windows® is registered Trademark of "Microsoft Co."
3031
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. (662) 619.0708.11 • Fax (662) 619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and
Technical offices, please see www.hannainst.com
32
MAN213
12/06
.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.