Hanna publishes a wide range of catalogs and
handbooks for an equally wide range of applications.
The reference literature currently covers areas such
as:
• Water Treatment
Instruction Manual
pH 210 pH 211
pH 212 pH 213
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is constantly
being added to the library.
For these and other catalogs, handbooks and leaflets,
contact your dealer or the Hanna Customer Service
Center nearest to you. To find the Hanna Office in
your vicinity, check our home page at
www.hannainst.com.
Microprocessor-based
Bench pH/mV/°C Meters
MANPH213R2 02/00
http://www.hannainst.com
Manufacturers since 1978
These Instruments are in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna Product. Please read this instruction
manual carefully before using the instrument.
This manual will provide you with the necessary information for a
correct use of the instrument, as well as a precise idea of its versatility.
If you need more technical information, do not hesitate to e-mail us
at tech@hannainst.com.
These instruments are in compliance with directives.
CE DECLARATION OF CONFORMITY ........................................... 31
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is
used.
Operation of this instrument in residential areas could cause unacceptable interferences to radio and
TV equipment, requiring the operator to take all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching
this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC
performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface
exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
2
31
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
All Hanna Instruments meters are guaranteed for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodesand the probes are guaranteed for a period of six months. This
warranty is limited to repair or replacement free of charge.
Damage due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization number from the Customer
Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.
To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner,
Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode
Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your Dealer or the nearest Hanna
Custumer Service Center.
Each meter comes supplied complete with:
• HI 1131B glass-body combination pH Electrode with 1 m
(3.3') cable
• HI 7669/2W Temperature Probe
• HI 76404 Electrode Holder
• pH 4.01 & 7.01 Buffer Solutions (20mL each)
• HI 7071S Electrolyte Solution
• 12VDC Power Adapter
• Instruction Manual
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective item must be
returned in the original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The Hanna pH 210, pH 211, pH 212 and pH 213 are microprocessor-based bench meters for pH and temperature measurements.
pH 211 and pH 213 can also measure ion concentration (ISE) and
ORP (Oxidation Reduction Potential) in the mV range.
The pH measurements are compensated for temperature effect manually or automatically with the HI 7669/2W temperature probe.
The meter comes equipped with a large easy-to-read LCD which shows
the pH (or mV) and temperature simultaneously together with graphic
symbols.
The calibration process is guided step by step through clear indications
on the LCD. A stability indicator make the calibration procedure errorfree.
Through the RS 232C serial port you can transfer the data to a PC
(pH 212 and pH 213 only).
30
3
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
pp
HH
210 AND 210 AND
p
H
210 AND
pp
HH
210 AND 210 AND
pp
HH
211 211
p
H
211
pp
HH
211 211
1) Liquid Crystal Display (LCD)
2)
CFM key, to confirm calibration values
3) CAL key, to enter or exit calibration mode
4) WC key, to manually decrease temperature or select pH buffer
5) VC key, to manually increase temperature or select pH buffer
6) MR key, to recall the stored value
7) MEM key, to store a value in memory
8) RANGE key, to select measurement range (pH 211 only)
9) ON/OFF switch
10) Power adapter socket
11) BNC electrode connector
12) Electrode reference socket
13)
Temperature probe socket
14) RESET button
Extension cables for screw-type electrodes (screw to BNC
adapter)
HI 7855/1Extension cable 1m (3.3') long
HI 7855/3Extension cable 3m (9.9') long
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
OO
THER ACCESSORIESTHER ACCESSORIES
O
THER ACCESSORIES
OO
THER ACCESSORIESTHER ACCESSORIES
HI 710005Voltage adapter from 115 VAC to 12 VDC (USA plug)
HI 710006Voltage adapter from 230 VAC to 12 VDC (European
plug)
HI 710012Voltage adapter from 240 VAC to 12 VDC (UK plug)
HI 710013Voltage adapter from 230 VAC to 12 VDC (South
Africa plug)
HI 710014Voltage adapter from 230 VAC to 12 VDC (Australia
plug)
ChecktempC Pocket-size thermometer (range -50.0 to 150.0C)
HI 76405 Electrode holder
HI 8427 pH and ORP electrode simulator with 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m
(3.3') coaxial cable ending in female BNC connectors
HI 7669/2W Temp. probe with 1 m (3.3') cable
HI 92000 Windows® compatible software
HI 920010 9 to 9-pin RS232 cable
Windows® is registered Trademark of "Microsoft Co."
pH CalibrationAutomatic 1 or 2 points with 5 memorized buffers
TemperatureFrom 0 to 100C (32 to 212F) manual or
Compensationautomatic with HI 7669/2W probe (included)
ElectrodeHI 1131B glass-body combination
Temperature probeHI 7669/2W (included)
Input impedance10
Computer InterfaceOpto-isolated RS232C
Power supply12 VDC adapter (included)
Environment0 to 50C (32 to 122F)
Dimensions240x182x74 mm (9.4x7.1x2.9")
Weight1.1 kg (2.5 lb.); kit with holder: 3.0 kg (6.5 lb.)
Plug the 12VDC adapter into the power supply socket (#10 on page
4; #11 on page 6).
Note:pH 212 and pH 213 use an EEPROM to retain the pH, mV
and temperature calibrations as well as the serial communication setting. The instrument will retain the data after a
calibration or serial communication setting, even when it is
unplugged.
Note:Note:
Note: Make sure the main line is protected by a fuse.
Note:Note:
Electrode and Probe connections
For combination pH or ORP electrodes (with internal reference)
connect the electrode's BNC to the socket on the back of the meter
(#11 on page 4; #12 on page 6).
For electrodes with a separate reference, connect the single cell
electrode's BNC to the BNC socket (#11 on page 4; #12 on page 6)
and the reference electrode plug to the reference socket (#12 on
page 4; #13 on page 6).
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket
(#13 on page 4; #14 on page 6).
Note: to prevent damage to the electrode, remove the pH electrode
from the solution before turning the meter off.
If the meter is OFF, detach the electrode from the meter before
immersing the electrode in the storage solution.
pH MEASUREMENTS
Make sure that the instrument has been calibrated before taking pH
measurements.
• Switch the instrument on by pressing the ON/OFF
switch.The meter automatically defaults to the pH measurement mode.
• Submerge the tip (4cm/1½") of the electrode and the temperature probe into the
sample to be tested. Allow time for the
electrode to stabilize.
8
(1½
")
4 cm
ORP PRETREATMENT SOLUTIONS
HI 7091LReducing Pretreatment Solution, 460 mL
HI 7092LOxidizing Pretreatment Solution, 460 mL
pH ELECTRODES
All electrodes part numbers ending in B are supplied with a BNC
connector and 1 m (3.3') cable, as shown below :
Glass-body, single junction, refillable, combination pH electrode.
Use: general purpose.
12 mm
0.5"
HI 1131
"S"VERSION
25
9.5mm DIA
120 mm
4.7"
0.37"
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 Buffer Sachets, 20mL, 25 pcs
HI 70007P pH 7.01 Buffer Sachets, 20mL, 25 pcs
HI 70010P pH 10.01 Buffer Sachets, 20mL, 25 pcs
HI 7004L pH 4.01 Buffer Solution, 460 mL
HI 7006L pH 6.86 Buffer Solution, 460 mL
HI 7007L pH 7.01 Buffer Solution, 460 mL
HI 7009L pH 9.18 Buffer Solution, 460 mL
HI 7010L pH 10.01 Buffer Sol., 460 mL
HI 8004L pH 4.01 Buffer Solution in FDA approved bottle, 460 mL
HI 8006L pH 6.86 Buffer Solution in FDA approved bottle, 460 mL
HI 8007L pH 7.01 Buffer Solution in FDA approved bottle, 460 mL
HI 8009L pH 9.18 Buffer Solution in FDA approved bottle, 460 mL
HI 8010L pH 10.01 Buffer Solution in FDA approved bottle, 460 mL
ELECTRODE STORAGE SOLUTIONS
HI 70300L Storage Solution, 460 mL
HI 80300L Storage Solution in FDA approved bottle, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Rinse Sachets, 20 mL, 25 pcs
HI 7061L General Cleaning Sol., 460 mL
HI 7073L Protein Cleaning Sol., 460mL
HI 7074L Inorganic Cleaning Sol., 460mL
HI 7077L Oil & Fat Cleaning Sol.,460 mL
HI 8061L General Cleaning Solution in FDA approved bottle, 460 mL
HI 8073L Protein Cleaning Solution in FDA approved bottle, 230 mL
HI 8077L Oil & Fat Cleaning Solution in FDA approved bottle,460mL
REFILL ELECTROLYTE SOLUTIONS
HI 70713.5M KCl + AgCl Electrolyte, 4x50mL, for single junc-
tion electrodes
HI 70721M KNO3 Electrolyte, 4x50 mL
HI 70823.5M KCl Electrolyte, 4x50 mL, for double junction elec-
trodes
HI 8071 3.5M KCl + AgCl Electrolyte in FDA approved bottle,
4x50mL, for single junction electrodes
HI 8072 1M KNO3 Electrolyte in FDA approved bottle, 4x50 mL
HI 8082 3.5M KCl Electrolyte in FDA approved bottle, 4x50 mL, for
double junction electrodes
• pH is displayed on the primary display and temperature on the
secondary one.
• With pH 211 and pH 213 it is also possible to view the mV
reading by pressing the "RANGE" key.
If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized water
or, if not available, tap water first and then with some of the next
sample to condition the electrode before immersing it in the sample.
The pH reading is affected by temperature. In order for the meter to
measure the pH accurately, temperature must be taken into consideration. To use the Automatic Temperature Compensation feature,
connect and submerge the HI 7669/2W temperature probe into the
sample as close to the electrode as possible and wait for a couple of
minutes.
If the temperature of the sample is known or tests are always
performed at the same temperature you can simply manually
compensate for it. For this purpose the temperature
probe must be disconnected.
The display will then show the default temperature
of 25C or the last recorded temperature reading
with the "C" symbol blinking.
Note the temperature of the sample using a
ChecktempC or an accurate reference thermometer.
The temperature can now be adjusted with the VC and WC keys.
24
9
ORP MEASUREMENTS (pH211 and pH213 only)
An optional ORP electrode must be used to perform ORP measurements
(see accessories).
Oxidation-reduction potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the sample tested.
To correctly perform a redox measurement, the surface of the ORP
electrode must be clean and smooth.
• Switch the meter on and press the "RANGE" key to
enter the mV mode.
• Submerge the ORP electrode tip (4cm/1½") into
the sample. Allow a few minutes for the reading
to stabilize.
• The meter displays the mV reading on the primary display.
(1½
")
4 cm
TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDE
TROUBLESHOOTING GUIDE
TROUBLESHOOTING GUIDETROUBLESHOOTING GUIDE
SymptomsProblemSolution
The meter is slow in
responding or gives
faulty readings
The meter does not accept the 2nd buffer
solution for calibration
The reading drifts
Display shows:
The electrode is not working or the reference
junction is clogged
Out of order pH electrode
Defective pH electrode
Out of range pH scale
Leave the electrode in a storage solution after cleaning
the junction. If problem persists, replace the electrode
Follow the cleaning procedure. If this doesn't work
replace the electrode
Replace the electrode
a) Recalibrate
b) Make sure the pH sample
is in the specified range
c) Check the electrolyte level
and the general state of
the pH electrode
• If readings are out of range, "---" will appear on the LCD.
TAKING TEMPERATURE MEASUREMENTS
Taking a
temperature measurement is very easy.
Connect the HI 7669/2W
turn the instrument on. Dip the temperature
temperature probe and
probe
in the sample and allow the reading on the secondary display to stabilize.
HOLD FUNCTION
Press and hold MEM. The last reading will be frozen on the display
until MEM is released. During this operation, the MEM indicator is also
shown on the LCD.
Press MR and the previously memorized reading will be
displayed with the MEM indicator on the LCD.
Display shows:
Display shows:
Display shows:
"WRONG 1 "
and
"WRONG "
Display shows:
"WRONG
and
"WRONG "
The meter does not
work with the temperature probe
The meter fails to calibrate or gives faulty
readings
"
2
Out of range temperature scale
Out of range mV scale
Erroneous buffer solution
used for offset calibration
Defective electrode
Erroneous buffer solution
used for slope calibration
Defective electrode
Out of order temperature
probe
Out of order pH electrode
Make sure the temperature
is in the 0 to 100C range
and the temperature probe is
plugged in
Electrode not connected
Make sure the buffer setting
is correct and the solution is
fresh. Replace the buffer if
necessary
Replace the electrode
Make sure the buffer setting
is correct and the solution is
fresh. Replace the buffer if
necessary
Replace the electrode
Replace the probe
Replace the electrode
10
23
TEMPERATURE CORRELATION FOR TEMPERATURE CORRELATION FOR
20
0+10+20+30+40+50+60+70+80+90
C
TEMPERATURE CORRELATION FOR
TEMPERATURE CORRELATION FOR TEMPERATURE CORRELATION FOR
SENSITIVE GLASSSENSITIVE GLASS
SENSITIVE GLASS
SENSITIVE GLASSSENSITIVE GLASS
pp
HH
p
H
pp
HH
The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes longer time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 10C.
Ω
Ω
9
2x10
-10
9
1x10
-10
8
-10
2x10
8
1x10
-10
7
2x10
-10
-10
7
1x10
-10
-20-
-20 -10 0 +10 +20+30+40 +50 +60 +70 +80 +90 C
°
Since the resistance of the pH electrode is in the range of 200 Mohm,
the current across the membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits andstatic discharges are detrimental to a stable pH reading.
The pH electrode's life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature1- 3 years
90 °CLess than 4 months
120°CLess than 1 month
High concentrations of sodium ions interfere with readings in alkaline
solutions; the pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is the
alkaline error and causes the pH to be underestimated. Hanna's glass
formulations have the indicated characteristics.
Alkaline Error
Sodium Ion Correction for the Glass at 20-25°C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
+
13.00
13.50
14.00
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
RESET
The RESET button (#14 on page 4;#15 on page 6) should only be
used when the instrument displays erroneous messages due to strong
electrical interference or when the instrument's power supply was
disconnected before the meter was switched off.
After pressing RESET always recalibrate the unit before proceeding.
pp
H CALIBRATIONH CALIBRATION
p
H CALIBRATION
pp
H CALIBRATIONH CALIBRATION
Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated :
a) Whenever the pH electrode is replaced.
b) At least once a week.
c) After testing aggressive chemicals.
d) After pressing RESET.
e) If higher accuracy is required.
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If
possible use plastic beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross contamination, use two
HI 7004
HI 7004
beakers for each buffer solution. One for
HI 7007
RINSE
CALIBRATION
rinsing the electrode and the second for
calibration.
HI 7007
HI 7007
If you are measuring in the acid range,
use pH 4.01 as second buffer; if you are measuring in the alkaline
range, use pH 10.01 or pH 9.18 as second buffer.
PROCEDURE
Calibration has a choice of 5 memorized buffers: pH 4.01, 6.86,
7.01, 9.18 and 10.01.
It is always recommended to perform a two-point calibration. The pH
meters however also provide for one-point calibration, as described
below.
22
11
One-point calibration
••
• Immerse the pH electrode and the tem-
••
perature probe approximately 4 cm (1½")
into a buffer solution of your choice (pH
4.01, 6.86, 7.01, 9.18 or 10.01) and
stir gently. The temperature probe should
(1½
")
4 cm
be close to the pH electrode.
• Press CAL. The "CAL" and "" indicators and the most common
"7.01" buffer will be displayed on the secondary LCD.
• If necessary, press "VC" or "WC" to
select a different buffer value.
• The "NOT READY" indication will blink on LCD until
the reading has stabilized.
• When the reading is stable, "READY" and "CFM" will blink. Press CFM
to confirm the calibration.
• If the reading is close to the selected buffer, the meter stores the
reading. The calibrated value is then displayed on the primary LCD
and the secondary LCD will display the second expected buffer value.
• If the value measured by the meter is not
close to the selected buffer, "WRONG " and
"WRONG " will blink alternately. In this case
check if the correct buffer has been used or
refresh the electrode by following the maintenance procedure below.
If necessary change the buffer or the electrode.
• Press CAL to quit calibration. The meter will return to
normal operation and memorize the one-point calibration data.
CLEANING PROCEDURE
GeneralSoak in Hanna HI 7061 or HI 8061 General Cleaning
Solution for approximately ½ hour.
Removal of films, dirt or deposits on the membrane/junction:
-
Protein
-
Inorganic
-
Oil/grease
IMPORTANT: After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte (not necessary for gel-filled electrodes)
and soak the electrode in HI 70300 or HI 80300 Storage Solution
for at least 1 hour before taking measurements.
TROUBLESHOOTING
Evaluate your electrode performance based on the following.
• Noise (Readings fluctuate up and down) could be due to:
– Clogged/Dirty Junction: Refer to the Cleaning Procedure above.
– Loss of shielding due to low electrolyte level (in refillable
• Dry Membrane/Junction: Soak in HI 70300 or HI 80300 Stor-
age Solution for at least 1 hour.
• Drifting: Soak the electrode tip in warm (approx. 50-60C) HannaHI 7082 or HI 8082 Solutions for one hour and rinse the tip
with distilled water. Refill with fresh HI 7071 or HI 8071 for
single junction electrodes and HI 7082 or HI 8082 for double
junction electrodes (refillable electrodes only).
• Low Slope: Refer to the cleaning procedure above.
• No Slope: Check the electrode for cracks in glass stem or bulb
and replace the electrode.
• Slow Response/Excessive Drift: Soak the tip in HI 7061 or
HI 8061 Solutions for 30 minutes, rinse thoroughly in distilled
water and then follow the Cleaning Procedure above.
Soak in Hanna HI 7073 or HI 8073 Protein Cleaning
Solution for 15 minutes.
Soak in Hanna HI 7074 or HI 8074 Inorganic Cleaning Solution for 15 minutes.
Rinse with Hanna HI 7077 or HI 8077 Oil and Fat
Cleaning Solution.
electrodes only): refill with fresh HI 7071 or HI 8071 for
single junction or HI 7082 or HI 8082 for double junction
electrodes.
12
21
For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1") below the
fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for
double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte
Solution for single junction electrodes.
For a faster response, unscrew the fill hole screw during measurements.
For AmpHel® electrodes:
If the electrode does not respond to pH changes, the battery is run
down and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom 4
cm /1½") in the sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.
STORAGE
To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300
or HI 80300 Storage Solution or, in its absence, Filling Solution
(HI 7071 or HI 8071 for single junction or HI 7082 or HI 8082 for
double junction electrodes). Follow the Preparation Procedure above
before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
Two-point calibration
Any two of the 5 memorized buffers can be used for this purpose.
However, it is recommended that pH 6.86 or 7.01 is chosen as the
first point and pH 4.01 (acidic) or pH 9.18/10.01 (alkaline) as the
second point.
• Proceed as described in “One-point calibration” above but do not
quit calibration by pressing CAL at the end.
••
• After the first calibration point is con-
••
firmed, immerse the pH electrode and the
temperature probe approximately 4 cm
(1½") in the second buffer solution and
stir gently. The temperature probe should
(1½
")
be close to the pH electrode.
• If necessary, press "VC" or "WC" to
select a different buffer value.
Note: The meter will automatically skip the buffer used for the first
point. It also skips 6.86 if 7.01 was used, and viceversa.
Likewise, it will skip 9.18 if 10.01 was used, and viceversa.
• The "NOT READY" indication will blink on LCD until
the reading has stabilized.
• When the reading is stable, "READY" and "CFM" will blink. Press
CFM to confirm the calibration.
4 cm
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry. If any scratches or cracks
are present, replace the electrode. Rinse off any salt deposits with
water.
For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071
for single junction or HI 7082 or HI 8082 for double junction electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
AmpHel® is a registered Trademark of "Hanna Instruments"
20
• If the reading is close to the selected buffer, the meter stores the
reading and returns to normal operational mode.
• If the value measured by the meter is not
close to the selected buffer, "WRONG " and
"WRONG " will blink alternately. In this case
check if the correct buffer has been used, or regenarate the
electrode by following the cleaning procedure below. If necessary
change the buffer or the electrode.
Note: Press RANGE to display the temperature reading
on the LCD during calibration (pH211, pH212,
pH213 only).
13
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
pp
H 212 AND H 212 AND
p
H 212 AND
pp
H 212 AND H 212 AND
pp
H 213 ONLYH 213 ONLY
p
H 213 ONLY
pp
H 213 ONLYH 213 ONLY
(for technical personnel only)
All the meters are factory calibrated for temperature.
The temperature probes are interchangeable and no temperature
calibration is needed when they are replaced.
If, for any reason, the temperature measurements are inaccurate,
temperature recalibration may be carried out.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center.
• Prepare a vessel containing ice and water and another one
containing hot water (at a temperature of around 50C). Place
insulation material around the vessels to minimize temperature
changes.
• Use a ChecktempC or a calibrated thermometer with a resolution of 0.1C as a
reference thermometer.
• With the instrument off, press and hold CAL and then the ON
switch. The "CAL" indicator will come on and the secondary LCD
will show 0.0C.
(32 °F)
°C
0 °C
50 ºC
(122 ºF)
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &
ELECTRODE CONDITIONING &ELECTRODE CONDITIONING &
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
Note: To prevent damage to the electrode, remove the pH electrode
from the solution before turning the meter off.
If the meter is OFF, detach the electrode from the meter before
immersing the electrode in the storage solution.
°
C
• Immerse the temperature probe in the vessel with
the ice and water as near to the ChecktempC as
possible. Allow a couple of minutes for the probe to
stabilize.
• Use the "VC" or "WC" key to set the
reading on the secondary LCD to that of
ice and water, measured by checktempC
(for example, 0.1 ºC).
• Press CFM to confirm. The secondary LCD will show
50.0C.
• Immerse the temperature probe in the second
vessel as near to the ChecktempC as possible.
14
0 °C
(32 °F)
50 °C
(122 °F)
°
C
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
°
C
removed by "shaking down" the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or
HI 80300
Storage Solution for at least one hour.
19
pp
H VALUES AT VARIOUSH VALUES AT VARIOUS
p
H VALUES AT VARIOUS
pp
H VALUES AT VARIOUSH VALUES AT VARIOUS
TEMPERATURESTEMPERATURES
TEMPERATURES
TEMPERATURESTEMPERATURES
Temperature has an effect on pH. The calibration buffer solutions are
affected by temperature changes to a lesser degree than normal
solutions. During calibration the meter will automatically calibrate to
the pH value corresponding to the measured or set temperature.
TEMPpH VALUES
°C°F4.016.867.019.1810.01
0
32
4.01
6.98
7.13
9.46
10.32
5
41
4.00
6.95
7.10
9.39
10.24
10
50
4.00
6.92
7.07
9.33
10.18
15
59
4.00
6.90
7.04
9.27
10.12
20
68
4.00
6.88
7.03
9.22
10.06
25
77
4.01
6.86
7.01
9.18
10.01
30
86
4.02
6.85
7.00
9.14
9.96
35
95
4.03
6.84
6.99
9.10
9.92
40
104
4.04
6.84
6.98
9.07
9.88
45
113
4.05
6.83
6.98
9.04
9.85
50
122
4.06
6.83
6.98
9.01
9.82
55
131
4.07
6.84
6.98
8.99
9.79
60
140
4.09
6.84
6.98
8.97
9.77
65
149
4.11
6.85
6.99
8.95
9.76
70
158
4.12
6.85
6.99
8.93
9.75
Allow a couple of minutes for the probe
to stabilize.
• Use the "VC" or "WC" key to set the
reading on the secondary LCD to that of the hot
water
• Press CFM to confirm. The meter returns to normal
operational mode.
mm
V CALIBRATIONV CALIBRATION
m
V CALIBRATION
mm
V CALIBRATIONV CALIBRATION
pp
H 213 ONLYH 213 ONLY
p
H 213 ONLY
pp
H 213 ONLYH 213 ONLY
(for technical personnel only)
All the meters are factory calibrated for mV.
The ORP electrodes are interchangeable and no mV calibration is
needed when they are replaced.
If, for any reason, the mV measurements are inaccurate, mV
recalibration may be carried out.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center.
• Attach to the electrode BNC connector a mV simulator with a
range of at least ±500.0 mV and an accuracy of ±0.1 mV.
• With the instrument off, press and hold CFM and then the ON
switch. The "CAL" indicator will come on and the primary LCD will
show 380.0 mV.
• Set 380.0 mV on the simulator.
For instance, if the buffer temperature is 25C, the display will show
pH 4.01 or 7.01 or 10.01. If the buffer temperature is 20C, it will
show pH 4.00/7.03/10.06 or at 50C, the display will show pH 4.06/
6.98/9.82.
18
• Wait for 3 sec. and press CFM. The LCD will display -380.0 mV
asking for this value to be set.
• Set -380.0 mV on the simulator.
• Wait for 3 sec. and press CFM. The LCD will display 500.0 mV.
• Set 500.0 mV on the simulator.
• Wait for 3 sec. and press CFM. The LCD will display -500.0 mV.
• Set -500.0 mV on the simulator.
• Wait for 3 sec. and press CFM.
• The secondary LCD will display 0.0C. It is now possible
to proceed with the temperature calibration, if necessary, as described above, or return to normal operation
by pressing CAL.
15
INTERFACE WITH INTERFACE WITH
INTERFACE WITH
INTERFACE WITH INTERFACE WITH
(p(p
H 212 AND H 212 AND
(p
H 212 AND
(p(p
H 212 AND H 212 AND
Data transmission from the instrument to the PC is now much easier
with the new HI 92000 Windows® compatible application software
(optional). HI 92000 also offers an on-line help feature.
To install HI 92000 you need a 3.5" drive and a few minutes to
follow the instructions printed on the disk label.
To connect your meter to the PC use the optional Hanna HI 920010
cable connector. Make sure that your meter is switched off and plug
the connectors, one into the meter RS232C socket and the other into
the serial port of your PC.
Note: Cables other than HI 920010 may use a different configura-
tion, in which case, communication between the meter and
the PC may not be possible.
If you are not using Hanna Instruments HI 92000
software, please find below some additional information to help you
link-up to your PC.
Setting the Baud Rate and the Command Prefix
The baud rate of the meter and of the external device must be the
same. To set the baud rate of the meter press COMM, the primary
LCD shows the current baud rate.
The following baud rates can be selected
with the "VC" or "WC" key: 150, 300,
600, 1200 (factory setting) and 2400.
Press CFM to confirm the setting. The primary LCD will show the
current command prefix (factory setting is 16).
Note: The Command Prefix must not be changed when using
Hanna HI 92000 Software.
pp
H 213 ONLY)H 213 ONLY)
p
H 213 ONLY)
pp
H 213 ONLY)H 213 ONLY)
Windows® is registered Trademark of "Microsoft Co."
Select, if necessary, a different command
prefix (between 0 and 47) by pressing the
"VC" or "WC" key.
Press CFM to confirm the setting.
Sending Commands from PC
With any terminal program it is possible to remotely control your
meter. Use HI 920010 cable to connect the meter to a PC, start the
terminal program and set the communication options as follows: 8, N,
1, no flow control.
Command Types
To send a command to the pH meter the scheme is:
<DLE> <
This line makes the computer send a Data Link Escape character, the
command expressed as a 3-character sequence and a CR character.
Note: All the terminal programs that support the ANSI escape se-
quence, represent the DLE character by the string '^P' and
the CR character by the string '^M'.
Commands not requiring an answer from the pH meter:
PHRsets the range to pH
MVRsets the range to mV (pH 213 only)
CALis equivalent to pressing the CAL key
CFMis equivalent to pressing the CFM key
UPCis equivalent to pressing the VC key
DWCis equivalent to pressing the WC key
MEMis equivalent to pressing the MEM key
MRRis equivalent to pressing the MR key
COMis equivalent to pressing the COMM key
OFFsets the meter in standby mode
Commands requiring an answer:
pH?Causes the meter to send the pH reading ("Err 1" is sent if
out of range). If the range is set to mV, "Err 6" is sent.
MV?Causes the meter to send the mV reading ("Err 2" is sent if
out of range). If the range is set to pH, "Err 6" is sent.
TM?Causes the meter to send the temperature reading ("Err 3" is
sent if out of range).
These commands may be sent with either capital or small letters.
Invalid commands will be ignored. The characters sent by the meter
are always capital letters. When the meter receives an unknown or a
corrupted command, it will send a character CAN (ASCII Code 24).
command>
17
<CR>
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.