Dear Customer,
Thank you for choosing a Hanna Product.
This instruction manual has been written for the following products:
PCA 300Free Chlorine Analyzer with features such as alarm functions, user-se-
lectable sampling periods, LED warning indications, recorder outputs,
RS232 port and a Nema 4X, 12 and 13 enclosure.
PCA 301Tot a l Chlorine Analyzer with features such as alarm functions, user-
selectable sampling periods, LED warning indications, recorder outputs,
RS232 port and a Nema 4X, 12 and 13 enclosure.
PCA 300ASame as PCA 300 except for the RS232 port which is not supplied.
PCA 301ASame as PCA 301 except for the RS232 port which is not supplied.
PCA 300AC This OEM version is exactly the same as PCA 300 except for the
RS232 port and the enclosure.
PCA 301AC This OEM version is exactly the same as PCA 301 except for the
RS232 port and the enclosure.
Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for the correct use of the instrument, as well as a
more precise idea of its versatility.
These instruments are in compliance with directives. The PCA 300 and PCA 301 are
also registered by Underwriters Laboratories Inc. (UL approved).
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive,
Woonsocket, Rhode Island, 02895 , USA.
3
PRELIMINARY EXAMINATION
Remove the analyzer from the packing material and examine
it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your
Dealer immediately.
Each analyzer is supplied complete with:
• 2 reagent bottles (1 indicator and 1 buffer solution)
(PCA 300 and PCA 301 only)
NoteSave all packing materials until you are sure that the instru-
ment functions correctly. Any damaged or defective items must
be returned in their original packing materials together with
the supplied accessories.
WARNING The PCA300 and PCA 301 series of Chlorine Analyzers are
not designed for use with samples that are inflammable or
explosive in nature. If any sample solution other than water is
used with these products, test the sample/product compatibility to assure user safety and proper product performance.
Safety PrecautionsPlease take the time to read the safety precautions carefully
wherever they appear in this manual. They are provided to
prevent personal injury and damage to the instrument. This
safety information applies to the operators and service personnel and the following two captions are used:
CAUTION: identifies conditions or practices that could result in damage to the instrument or persons;
WARNING: identifies conditions or practices that could
result in personal injury or loss of life.
NoteBecause of the inherent dangers in handling chemical
samples, standards and reagents, HANNA Instruments
strongly recommends the users of this product review the
Material Safety Data Sheets and become familiar with safe
handling procedures and proper usage prior to handling
any chemicals.
4
GENERAL DESCRIPTION
The Hanna PCA 300 and PCA 301 series of Chlorine Analyzers are microprocessor-controlled, process analyzers which
continuously monitor a sample stream for Chlorine content.
The PCA 300 series monitor Free Chlorine and the PCA 301
series Total Chlorine in the 0 to 5 mg/L (ppm) range.
In the DPD Colorimetric method, N, N-Diethyl-p-phenylenediamine indicator and a buffer are mixed with the sample.
The resulting chemical reaction causes a magenta color to
form. The color intensity is proportional to the concentration
of Chlorine. The color intensity is measured photometrically
(with a light beam and a photodetector) and converted to
Chlorine concentration, in mg/L, which is displayed on the
front panel, four-digit, LED display.
Indicator and buffer reagent bottles are placed directly into
the instrument case. With a sampling period of 5 minutes,
reagents need to be replenished about once a month. The
reagent bottles are easily visible through the transparent window to allow the operator to check the reagent levels.
The cases of PCA 300 and PCA 301 meet NEMA 4X, 12
and 13 standards. Molded fiberglass polyester has outstanding chemical and temperature resistance.
External mounting feet provide wall mounting capability and
a seamless gasket assures a watertight and dust-tight seal.
The electrical and hydraulic connections (except for
PCA 300AC and PCA 301AC OEM models) are made
through the side of the enclosure.
The front cover is secured with two lockable
latches.
5
Two selectable chlorine level setpoints can be set by the operator: a proportional dosing setpoint and an alarm setpoint.
Both setpoints control a SPDT relay.
The proportional dosing setpoint analyzers is user-selectable
with a delta from 0.1 to 2.0 mg/L (ppm).
The alarm level can be set by the user to be an “activateif-lower-than” setpoint or an “activate-if-greater-than” setpoint.
A system alarm feature provides relay activation to signal
need for operator intervention.
The PCA 300 and PCA 301 analyzers can supply data to
an external computer through an RS232C output connection.
Voltage output levels of 0-10mV, 0-100mV, 0-1V or a current output of 4-20 mA are selectable to drive an external
device such as a chart recorder or an external regulator.
Recorder span minimum and maximum values in mg/L
are programmable by the operator through the keyboard.
MECHANICAL DIMENSIONS *
NoteThe enclosure is not supplied with PCA 300AC or
PCA 301AC
* Wall Mount Dimensions in mm & inches
6
FUNCTIONAL DESCRIPTION
1. Alarms Indicators
2. LED Display
3. Keyboard
4. Alarms Output
5. Recorder Output
6. Peristaltic Pump
7. Access Point to Cell
8. Measuring Cell
9. Output Port
10. Drain Tube
Unplug the meter before any electrical connection.
11. Drain Port of Measuring Cell
12. Drain Port Valve
13. Sample Tubing
14. Buffer Bottle
15. Indicator Bottle
16. Sample Port
17. Incoming Pressure Regulator
18. Electrovalve
19. Reagent Tubing
20. Line Input
21. Pressure Regulator Output Port
7
LED DISPLAY, INDICATORS AND KEYBOARD
DIGITAL DISPLAY
The digital readout will indicate the Chlorine concentration of
the last sample measured in milligrams per liter (mg/L) during normal operation.
Only when a key is pressed to execute some other function,
will the concentration reading be interrupted. In this case, the
Chlorine mg/L indicator is turned off.
If the Chlorine concentration is above 5 mg/L, the display will
blink.
mg/L CHLORINE INDICATOR
During normal operation, this green indicator will be on continuously, indicating that the
LED is displaying Chlorine concentration. During the diagnostic or other modes this indicator
will be off.
DOSE INDICATOR
ALARM INDICATOR
8
When proportional dosing is active, the dose
LED will be on and the corresponding relay
closes. This LED turns off automatically when
the proportional dosing stops.
This LED will be on when the programmed
alarm setpoints are exceeded. The alarm
LED turns off automatically when the alarm
condition disappears.
SYSTEM ERROR INDICATOR
This LED is lit when a system error has occurred. If the situation persists for more than
a few samples, the operator should notify
maintenance personnel for investigation of the
problem. When the meter is in system error
mode, the user can directly access the diagnostic code that indicates the source of error. The analyzer
continues to perform the sampling operations during an alarm
condition.
KEYBOARD DESCRIPTION
RESET
MAX
MIN
DIAG
REC.
MAX
REC.
MIN
The Reset key clears system alarm and restarts the analyzer.
The Max/Min key alternately displays maximum and minimum Chlorine concentration levels since last reset.
The RIGHT arrow key is used to shift digit position within a
displayed number.
The UP arrow key is used to scroll up a displayed digit: 0, 1,
2 ..... 9, 0, 1, 2, etc.
The DIAG key is used to select programming and self-diag-
nostic modes.
The Rec.Max key is used to enter and recall maximum value
of the recorder output in mg/L.
The Rec.Min key is used to enter and recall minimum value
of recorder output in mg/L.
The Set Point key is used to enter and recall concentration
setpoint.
SET
TIME
The Set Alarm key is used to enter and recall concentration
alarm setpoint.
The Set Time key is used to enter and recall time intervals
between two consecutive samples.
9
SPECIFICATIONS
Range0.00 to 5.00 mg/L0.00 to 5.00 mg/L
Resolution0.01 mg/L
Accuracy±8% of reading or ±0.05 mg/L
Typical EMC±0.05 mg/L
Deviation
Minimum0.05 mg/L
detectable level
Repeatability±0.05 mg/L
Response timeRelated to the sampling timeselected.
Sampling rateAdjustable from 3 to 102 minutes
PCA 300PCA 301
Free Cl
2
Total Residual Cl
2
whichever is greater
Typical for a full scale step change and
5 minutes between two consecutive
samples: one sampling cycle for 90%
response and two sampling cycles for
100% response
10
Sample inlet0.07 bar (1psig) min., 4 bar (57.2 psig)
pressuremax. An internal regulator reduces
pressure from 4 bar (57.2 psig) to
1 bar (14.3 psig)
Sample flowFlow rate of 300 mL/min is recom-
rangemended. Minimum and maximum
allowed are 100 mL/min and
500 mL/min, respectively
Sample5 to 40°C (41 to 104°F)
temperature range
InterferencesOxidizing agents such as: Iodine,
Bromine, Ozone, Chlorine Dioxide,
Permanganate, Hexavalent Chromium.
Hardness must not exceed 1000 mg/L
as CaCO3. Alkalinity must not exceed
400mg/L for Free Chlorine analysis (PCA
300) or 700mg/L for Total Chlorine
analysis (PCA 301)
Operating5 to 40°C (41 to 104°F)
temperature range
Recorder output Selectable 0-10mV, 0-100mV, 0-1V or
4-20 mA. Output span is settable
anywhere in the 0-5 mg/L range
DosageProportional on one point with a delta
adjustable at 0.1, 0.2, 0.3, 0.4, 0.5,
0.6, 0.7, 0.8, 0.9, 1.0, 1.5, 2.0 mg/L,
equipped with a SPDT relay with
contacts rated for resistive load: 5 A at
250VAC or 5 A at 30VDC; inductive
load: 2 A at 250VAC or 3 A at 30VDC.
AlarmsOne sample concentration alarm
adjustable as minimum or maximum
acceptable value, equipped with a
SPDT relay with contacts rated for
resistive load: 5 A at 250VAC or 5 A at
30VDC; inductive load: 2 A at 250VAC
or 3 A at 30VDC.
One system error alarm
Power20VA at 115VAC/230VAC;
requirements50/60 Hz
Sample inlet12 mm (1/2”) male NPT fitting
connection
instrument enclosure with transparent
GE Lexan window (not PCA 300AC and
PCA 301AC)
Dimensions318 x 267 x 159 mm
(12.5 x 10.5 x 6.25")
Weight5 Kg (11 lb.)
11
OPERATING DESCRIPTION
Referring to the drawing on page 7 and the Fluidic Diagram on page 13, the Sample Line is connected to the
instrument at the Sample Port (#16); an internal Regulator
(#17) reduces the inlet pressure from a maximum of 4 bar
(57.2 psig) down to 1 bar (14.3 psig); from the Regulator
a nylon tube is connected to the input of the Electrovalve
(#18). The output of the valve goes to the Drain Port (#11
in PCA 300 and PCA 301 only) and then to the Measur-
ing Cell (#8). An optional Filter can be installed to the sample
port if the stream is excessively turbid.
The sample coming from the line normally flows through the
Measuring Cell. It goes out from the Measuring Cell through
the Drain Tube (#10) and the Output Port (#9).
The Measuring Cell is accessible from the port placed on the
top (#7) for speedy cleaning and maintenance.
During the time between two successive sampling, the analyzer solenoid input valve is open to allow sample flow to
flush the colorimeter cell. Every 3 to 102 minutes (user selectable), the valve closes stopping the sample flow and
leaving the sample cell full of fresh sample. Cell volume is
controlled by an overflow gateway.
As the sample inlet valve closes, a series of measurements
(with LED on and off) of the unreacted sample is taken to
determine an average Blank level prior to reagent addition. The measurement of sample blank signal permits
compensation for any turbidity or natural color, and provides the zero reference point for the measurement.
The two channel Peristaltic Pump (#6) starts rotating causing a precise quantity of buffer and indicator (#14 and
#15) to enter the colorimeter sample cell where a magnetically coupled stirrer mixes the reagents with the sample.
After a delay for the development of color, a series of measurements (with LED on and off) are taken (Sample level)
to determine an average Chlorine concentration measurement. The reacted sample signal is then measured and
displayed.
This sequence is repeated every 3 to 102 minutes (user-selectable).
12
METHOD OF ANALYSIS
Fluid Diagram of the Analyzers
Free available Chlorine (monitored by the PCA300 series)
oxidizes the DPD indicator reagent at a pH between 5.5 and
6.0 to form a magenta-colored compound. The intensity of
the resulting color is proportional to the concentration of Chlorine in the sample. The purpose of the buffer solution is to
maintain the proper pH.
To measure Total Residual Chlorine (Free available Chlorine
plus combined Chloramines) the PCA301 series adds Potassium Iodide. The Chloramines in the sample cause
iodide ions to become iodine which then act with free Chlorine to oxidize the DPD indicator. After the chemical reaction
is complete, the optical signal at 555 nm is compared to
the signal measured through the sample (before the reagents were added). From these measurements Chlorine
concentration is calculated.
13
INITIAL PREPARATION AND INSTALLATION
INSTALLATION PERSONNEL
Installation of the PCA 300 and PCA 301 family of Chlorine Analyzers should be undertaken by persons with technical
knowledge of the dangers associated with chemical exposure
and electrical shock.
Hanna Instruments assumes persons performing the installation tasks are aware of the appropriate safety procedures.
Review the Material Safety Data Sheets (MSDS) before handling the supplied chemical reagents.
LOCATION OF THE INSTRUMENT
Analyzer Location
Locate the analyzer as close as is reasonably possible to
the point where the sample is withdrawn from the product
stream (referred to as the “sampling point”). This practice
will minimize the amount of time it takes for the sample to
flow to the analyzer from the sampling point. The result is
improved response time of the analyzer to changes in Chlorine contents.
The instrument should be mounted indoors, out of direct
sunlight. Instrument operating temperature is 5 to 40°C
(41 to 104°F).
Sampling Point Location
Locate the sampling point to obtain a truly representative
sample from the product stream. For example, be sure the
sampling point is well downstream from a Chlorine feed.
This assures that adequate mixing and reaction of the
Chlorine before a sample is extracted.
HYDRAULIC CONNECTIONS
NoteHydraulic connections should be installed only by qualified
personnel to assure conformity to applicable plumbing codes.
Sample Line Installation
Sample lines of 3 mm (1/8") ID tubing are recommended to
keep the sample volume in the tubing to a minimum. This
reduces lag time for the sample flow from the sampling point
to the instrument. In any case note that sample should reach
the measuring cell in less then 100 seconds.
14
Drain Line Installation
Direct routing of sample lines is also recommended.
If the large process pipes are horizontal, taps should be in-
serted vertically in the middle of the pipe to avoid pulling
sediment from the bottom or air bubbles from the top of
the pipe into the sample line.
A 1/2” BSP sample input fitting allows direct connection to
the optional input filter.
Sample line pressure should be between 0.07 and 4 bar (1
and 57.2 psig) with an ideal pressure of 0.7 bar (10 psig).
The drain hose fitting is a 20 mm (3/4”) hose barb on the
bottom of the instrument enclosure. An air gap between the
end of the drain hose and the drain is recommended to prevent
any back flow into the instrument in the event of drain blockage.
15
Return Line Installation
The return hose fitting is a 12 mm (1/2”) hose barb on the
bottom of the regulator output port and should always be
connected even when pressure is below 1 bar.
INSTALLING THE INPUT FILTER
In order to ensure maximum accuracy of measurements, it is
recommended to have always clear sample, with suspended
particles smaller than 0.5 µm. This can be achieved by installing two filters before the sample input.
The type of filters depends on the quality of the water: the first
filter should have 50-100 µm pore size, whereas in any case
the second filter, the one closer to the analyzer, has to be
0.5 µm.
For correct installing procedure and maintenance, see the
instructions of filters.
INSTALLING THE PUMP TUBES
Locate the analyzer reagent tubes in the accessory kit. Each
tube is composed of three sections. The sections are joined
together by plastic connectors with plastic collars at the ends
of the center section.
Locate the peristaltic pump.
Feed one tube from the shorter end section behind the pump
rollers from the right side of the pump. Seat the plastic collar
at the right end of the center section of tubing into the lower right
indentation hole of the pump face.
16
Grasp the other plastic collar and pull, stretching the center
section, and place the grommet in the lower left indentation
hole.
Repeat this process with the second pump tube, placing it in
the upper indentation holes.
Separate reagent caps are provided in the accessory kit. Put
the supplied caps onto each reagent bottle prior to installing
them. Place the indicator bottle (HI 70450 for PCA 300
and HI 70460 for PCA 301) on the right and the buffer
bottle (HI 70451 for PCA 300 and HI 70461 for PCA
301) on the left.
NoteAdd the content of 5 HI 70452 sachets, DPD Compound,
to the Indicator Solution prior to installing it.
Connect the longer tube
ends on the left side of
the pump to the reagent
bottle cap insert fitting.
Connect the short ends
on the right side of the
pump to the measuring
cell reagent input port
through the "Y" connector.
17
ELECTRICAL CONNECTIONS
A power cable (3 mt.) is provided with your analyzer. However, if access to the terminal block is required, see below.
WarningElectrical connections should be installed only by qualified
personnel to assure conformity to applicable electrical codes.
Unplug the meter before any electrical connection.
Power
Power connections are made at a terminal block located in
the center of the electrical compartment to the right of the
fuses.
Hard wiring with 13 mm (½") conduit is recommended and
usually required by most municipal electrical codes.
WarningBefore connecting the instrument to the line:
1) Check the sticker near the fuses for proper voltage.
2) Be sure the power cord is not connected to the line.
3) Open front panel.
18
4) Remove the cover screws (Allen head).
5) Do not remove peristaltic pump or motor.
6) Unplug all alarms and recorder jacks.
Feed the power cord through the watertight grommet and
tighten the grommet nut. See the picture below for proper
wire connections.
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.