Hanna Instruments HI 98509-01 User Manual

Checktemp-Dip
HI 98509-01 (Checktemp-Dip) C: Celsius
Model
HI 98510-01 (Checktemp-Dip) F: Fahrenheit
Model
SPECIFICATIONS:
Range
ºC -20 to 100.0 &
-60.0 to 0.0°C
ºF -4.0 to 199.9ºF &
200 to 212ºF
Resolution ºC 0.1°C ºF 0.1°F (-4.0 to 199.°F)
1°F (200 to 212°F)
Accuracy
ºC ±0.5°C (-20 to 50°C)
±0.9°C (outside)
ºF ±0.9°F (-4 to 122°F)
±2°F (outside)
Environment 0 to 50ºC (32 to 122ºF);
95% RH max.
Battery Type 1 x 1.5V alkaline Life approx. 3,000 hours
of continuous use
Dimensions 106 x 58 x 19 mm
(4.2 x 2.2 x 0.7")
Weight 80 g (2.8 oz.) Probe Stainless steel probe
with 3 meter (10') cable
Recommendations for Users:
Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause interference to radio and TV equipment. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shocks, do not use this instrument when voltage at the measurement surface exceeds 24 VAC or 60 VDC. To avoid damage or burns, do not perform any measurement in microwave ovens.
Visit our Internet Home Page:
www.hannainst.com
Manufacturers since 1978
Checktemp-Dip
OPERATION:
Turn Checktemp-Dip on.
Dip the probe in the
sample. An immer­sion level of at least
RUN
TEST
0°C/32°F
BATTERY1.5V AAA
ON
OFF
4 cm (1 ½") is recom­mended.
• Do NOT wrench the probe. If the cable is damaged, the meter may have to be replaced.
• Wait for a few sec­onds for the display to stabilize and switch the meter off when not in use.
• Variations in readings or fading dis­play can indicate a weak battery.
Checktemp-Dip is not suitable for con-
tinuous measurement (in-line, es­pecially at high temperatures) and should only be used for spot checks.
Even though Checktemp-Dip can
reach 100ºC, prolonged mea­surements at high tempera­tures will reduce the sensors life dramatically. Ideally operate in the -20 to 50ºC (-4 to 120ºF) range.
CALIBRATION CHECK SWITCH:
Move the Cal-check switch down. The display should show
0.0°C ±0.5°C or 32º F±0.9°F. Contact your dealer or your nearest Hanna Service Center
MEASUREMENT
MODE
CALIBRATION
CHECK
MODE
RUN
TEST
0°C/32°F
BATTERY1.5V AAA
for recalibration, if necessary. Move the Cal­check switch back up and the display will return to measure­ment mode.
BATTERY REPLACEMENT:
When Checktemp-Dip cannot be switched on or the display fades, remove the two screws, open the battery cover and change the
ON ON
RUN RUN
TEST
OFF OFF
0°C/32°F
BATTERY1.5V AAA BATTERY1.5V AAA
ON
OFF
1.5V AAA battery, paying atten-
TEST
0°C/32°F
+-
tion to its polarity. Replace the cover and tighten the two screws. Battery should only be replaced in a safe area using an alkaline
ISTRCKTDIPR2 05/02
1.5V AAA type battery.