Checktemp-Dip
HI 98509-01 (Checktemp-Dip) C: Celsius
Model
HI 98510-01 (Checktemp-Dip) F: Fahrenheit
Model
SPECIFICATIONS:
Range
ºC -20 to 100.0 &
-60.0 to 0.0°C
ºF -4.0 to 199.9ºF &
200 to 212ºF
Resolution
ºC 0.1°C
ºF 0.1°F (-4.0 to 199.°F)
1°F (200 to 212°F)
Accuracy
ºC ±0.5°C (-20 to 50°C)
±0.9°C (outside)
ºF ±0.9°F (-4 to 122°F)
±2°F (outside)
Environment 0 to 50ºC (32 to 122ºF);
95% RH max.
Battery Type 1 x 1.5V alkaline
Life approx. 3,000 hours
of continuous use
Dimensions 106 x 58 x 19 mm
(4.2 x 2.2 x 0.7")
Weight 80 g (2.8 oz.)
Probe Stainless steel probe
with 3 meter (10') cable
Recommendations for Users:
Before using this product, make sure that it is entirely
suitable for the environment in which it is used.
Operation of this instrument in residential areas could
cause interference to radio and TV equipment. Any
variation introduced by the user to the supplied
equipment may degrade the instrument's EMC
performance. To avoid electrical shocks, do not use
this instrument when voltage at the measurement
surface exceeds 24 VAC or 60 VDC. To avoid damage
or burns, do not perform any measurement in
microwave ovens.
Visit our Internet Home Page:
www.hannainst.com
Manufacturers since 1978
Checktemp-Dip
OPERATION:
• Turn Checktemp-Dip on.
• Dip the probe in the
sample. An immersion level of at least
RUN
TEST
0°C/32°F
BATTERY1.5V AAA
ON
OFF
4 cm (1 ½") is recommended.
• Do NOT wrench the
probe. If the cable is damaged, the
meter may have to be replaced.
• Wait for a
few seconds for the
display to
stabilize and
switch the
meter off
when not in use.
• Variations in readings or fading display can indicate a weak battery.
•Checktemp-Dip is not suitable for con-
tinuous measurement (in-line, especially at high temperatures) and
should only be used for spot
checks.
• Even though Checktemp-Dip can
reach 100ºC, prolonged measurements at high temperatures will reduce the sensors
life dramatically. Ideally operate
in the -20 to 50ºC (-4 to 120ºF)
range.
CALIBRATION CHECK SWITCH:
Move the Cal-check switch down.
The display should show
0.0°C ±0.5°C or
32º F±0.9°F.
Contact your
dealer or your
nearest Hanna
Service Center
MEASUREMENT
MODE
CALIBRATION
CHECK
MODE
RUN
TEST
0°C/32°F
BATTERY1.5V AAA
for recalibration,
if necessary.
Move the Calcheck switch back up and the
display will return to measurement mode.
BATTERY REPLACEMENT:
When Checktemp-Dip cannot be
switched on or the display fades,
remove the two screws, open the
battery cover and change the
ON ON
RUN RUN
TEST
OFF OFF
0°C/32°F
BATTERY1.5V AAA BATTERY1.5V AAA
ON
OFF
1.5V AAA battery, paying atten-
TEST
0°C/32°F
+-
tion to its polarity. Replace the
cover and tighten the two screws.
Battery should only be replaced
in a safe area using an alkaline
ISTRCKTDIPR2 05/02
1.5V AAA type battery.