Thank You!
Thank y ou for choosing a Ha nna Instrum ents
produc t. Please re ad this inst ructio n manual
caref ully before usi ng the instr ument. Fo r
addit ional inform ation, visi t www.hannainst.co m
or e-mai l us at tech@ha nnainst.com.
Specications
Range
Resolu tion
Accuracy
Environme nt
Battery Type
Batter y Life
Auto O
Dimens ions
Weight
Certication
-50.0 to 150.0 º C / -58.0 to 302 ºF
0.1 ºC (-50.0 to 150 .0 ºC)
0.1 ºF (-58.0 to 199.9 ºF) / 1 ºF (above 200 ºF)
±0.2 ºC (-30.0 to 120 .0 ºC)
±0.3 ºC (outs ide: -50.0 to -30.0 ºC
and 120.0 to 1 50.0 ºC)
±0.5 ºF (-22.0 to 199.9 º F)
±1 ºF (outsi de: -58.0 to -22.0 ºF
and 200 to 3 02 ºF)
-30 to 50 ºC (-22 to 122 º F)
1 x CR2032
approx. 2000 hour s of continuo us use
8 min (default), 6 0 min or OFF
50 x 185 x 21 mm (2 x 7.3 x 0.9“ )
50 g (1.8 oz.)
EN 13485:2001 - Suitability: Storage;
Environment: E; Accuracy Class: 0.5; IP65
Meter Setup
Enter Se tup to chang e the temper ature unit ( ºC or
Setup MODE , press the S ET button l ocated inside
the ba ttery com partme nt once for tem perature
· Use the O N/OFF but ton to switc h between
“ºC ” or “ºF”. Press S ET once to sav e and continu e
to auto s hut o sett ings.
· Use the O N/OFF but ton to cycle th rough the
follow ing Auto Shu t O options: 8 min utes
(“d08 ”, defau lt value), 60 min utes (“d60 ”),
OFF (“d- -”). Press th e SET butt on once to save
and con tinue to norm al operat ion.
Battery Replacement
Replac e the batte ry when the bat tery sy mbol is
blink ing or when Che cktemp® d oes not swi tch
on. If th e meter has be en submerged in a l iquid,
make sur e it has been co mpletely dr ied. Check t he
polar ity of the b attery, op en the compar tment , and
repla ce the batt ery. Close b attery co ver.
Note: Bat teries sh ould only be r eplaced in a sa fe area using t he batter y
type sp ecied in thi s instruc tion manua l. Old batter ies should b e disposed i n
accordan ce with local regul ations.
Operation
Turn Check temp® ON with ON/OF F button: all LCD
segmen ts are displayed fo r 1 second, and Cal Check®
star ts automat ically to verif y that Check temp®
of “CAL“ and “-0 -“ means that Che cktemp® is
measur ing temperatu re, remember th at:
· The tem perature s ensor is locat ed at the tip
of the s tainless s teel probe.
· It is nec essary to w ait for a few se conds for
the dis play to stab ilize.
Liquid C rystal Di splay
ON/OFF bu tton
Maximum Imme rsion Leve l
Temperature P robe
Batter y
Setup
Butto n
EN 13485
Storage
E / 0.5
4 cm
1.5”
Recommende d
Immersi on Level
Troubleshooting
Error s that may oc cur: A messa ge of “CAL” and
“Err ” means an in ternal error h as occurre d and
Check temp® is no t guaranteed to be w ithin its
and the d isplay show s “---” the temper ature
conne ction to the pro be has been b roken. If a
mess age of “Err ” occurs t hen a calibra tion data
read er ror has occu rred. A mess age of “Er 1” is
cause d by an unexp ected dev ice reset. A m essage
of “Er2” o ccurs if in ternally a temp erature
measu rement malf unction s. After th ese messa ges
Check temp® OFF an d then ON with th e ON/OFF
but ton to reset . If the error m essage co ntinues to
Warra nty
Checkt emp® is guara nteed for a peri od of one year a fter date of p urchase
against defe cts in workma nship and ma terials whe n used for their in tended
purpos e and maintai ned accordi ng to instr uctions. T his warra nty is limit ed
to repair or r eplacemen t free of char ge. Damage du e to accident s, misuse,
tampering o r lack of pres cribed mai ntenance is n ot covered. I f service i s
required , contact yo ur local Hanna O ce. If under wa rranty, re port the mo del
purchas e, serial num ber and the na ture of the prob lem. If
number, date of
the repair i s not covered b y the warra nty, you will be n otied of t he charges
incurre d. If the inst rument is to b e returned to Ha nna Instrum ents, rs t
obtain a Ret urned Goo ds Author ization num ber from th e Technical Ser vice
depart ment and the n send it wit h shipping cos ts prepai d. When shipp ing any
instrum ent, make su re it is properl y packed for co mplete prote ction.
Recomm endatio ns for Users
Before using H anna Produc ts, make sure t hat they are e ntirely sui table for
your speci c applica tion and for t he environm ent in which the y are used.
Operati on of these in strumen ts may cause u naccepta ble interf erences to
other ele ctronic e quipment, t hus requir ing the oper ator to take al l necessar y
steps to cor rect suc h interfer ences. Bat teries sh ould only be r eplaced in a
safe area usi ng the bat tery typ e specied i n this instr uction ma nual. Any
variati on introdu ced by the use r to the suppli ed equipmen t may degrad e
instrum ent’s EMC pe rforma nce. To avoid damages o r burns, do no t put the
instrum ent in microwav e oven. For you rs and the ins trument s afety do n ot
use or store th e instrum ent in hazar dous environm ents.
All rights are reserved. Repr oduction in whole or in par t is
prohibited without the wri tten consent of the copyright owner,
Hanna Instruments Inc., Woonso cket, Rhode Island, 02895, USA.
the
IST98501
09/14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com