Hanna Instruments HI 983327-0 User Manual

Instruction Manual
: follow the above diagram and connect the
colored wires of probe cable as indicated. It is recommended
HI 983327-0 HI 983327-1 HI 98332700
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
FUNCTIONAL DESCRIPTION
FRONT PANEL
Keypad
SET To display
setpoint
MEAS To display
measurement
3. Probe Connection to connect the shield (P5) to avoid any interference.
HI983327 LAYOUT
Panel-Mounted
EC Indicators &
Controllers
Compliance with the CE Directives
WARRANTY
against defects in work-
The probe is warranted for a
This warranty is limited to repair or replacement
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has oc­curred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Cen­ter.
The meter is supplied with:
• Instruction manual;
• Mounting brackets. Note: Conserve all packing material until the in-
strument has been observed to function correctly. Any defective item must be returned in its original pack­ing.
GENERAL DESCRIPTION
HI 983327-0, HI 983327-1 and HI 98332700 are conductivity indicators and controllers with a relay output designed for simplicity of use in a wide range of applica­tions.
HI 98332700 is provided with a DC/DC converter which isolates the measure system from the external power supply. The models are panel mounted with membrane keypads on the front panel and an easy-to-read LCD display. All the meters automatically compensate for the temperature variation. The probe is easy to clean and requires little maintenance. Measurements are highly accurate and the meters can be calibrated at one point.
Power supply, wiring and selection are made via the plug­in terminal blocks on the rear panel.
LED indicators on the front panel identify whether the controller is in set or measurement mode, and if alarm is active.
Trimmers
CAL For
calibration
SET To adjust
setpoint
LEDs
SET ON when LCD displays the set value MEAS ON when LCD displays the measured value ALARM ON when alarm contact is activated
REAR PANEL
1. Power Supply:
HI 983327-0 HI 983327-1
HI 98332700 12 VDC 115/230 VAC
L1: Positive 230 VAC L2: Negative 115 VAC L3: ----- Neutral L4: ----- Protection Earth Note: the power input is internally protected by a 400 mA fuse
2. Alarm Contact: A1, A2. This contact acts as a switch for the power. The contact has to be protected outside by the user.
Front view of the panel-mounted unit
Side view of the panel-mounted unit
Supplied adjustable location brackets al­low the controller to slide into the cutout and will hold the unit securely in place. 95 mm (3.74") is the mini­mum amount of space required to install the controller.
Panel-mounted unit assembling view
OPERATIONAL GUIDE
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal band at all times. During op-
eration, ESD wrist straps should be worn to avoid possible damage to the
probe by electrostatic discharges. Any variation introduced by the user to
the supplied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to mini-
To avoid damage or burns, do not perform any measurement in microwave
to temperature compensation. The “ALARM” LED will light up when the alarm contact is
losed, to indicate an EC value higher than selected setpoint.
CALIBRATION
Make sure the meter is in the measurement mode (the “MEAS” LED lights on).
Immerse the probe in HI 7039 (5.00 mS/cm) calibration solution.
Shake briefly and wait for reading to stabilize. Using a small screwdriver adjust the calibration trimmer
until the meter displays “5.00” on the LCD.
SETPOINT
Press the “SET” key. The display will show the default or previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until the required limit value is displayed.
PROBE MAINTENANCE
To improve probe performance and prolong its life, it is recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning Solution at least for one hour.
• If a more thorough cleaning is required, brush the metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and recalibrate the meter.
• When not in use, clean the probe before storing it.
ACCESSORIES
HI 7632-00 EC probe HI 70039P 5.00 mS/cm (5000 µS/cm) calibration solu-
tion, 20mL sachet (25 pcs)
HI 7039M 5.00 mS/cm (5000 µS/cm) calibration solu-
tion, 230mL bottle
HI 7061M Electrode cleaning solution, 230 mL bottle HI 710005 12 VDC power adapter, US plug HI 710006 12 VDC power adapter, European plug HI 710012 12 VDC power adapter, Australian plug HI 710013 12 VDC power adapter, Southern Africa plug HI 710014 12 VDC power adapter, UK plug HI 731326 Calibration Screwdriver (20 pcs) HI 740146 Mounting brackets
SPECIFICATIONS
HI 983327
Range 0.00 to 10.00 mS/cm Resolution 0.01 mS/cm Accuracy (@ 20°C) ±2% f.s. Setpoint Adjustable through multiturn trimmer Alarm Condition LED ON and alarm contact closed when
measured EC value is higher than setpoint
Alarm Output 2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe HI 7632-00 EC probe Temp.Compensation Automatic from 5 to 50°C
(41 to 122°F); ß=2%/ºC
Calibration Manual, at one point through trimmer Calibration Solution HI 7039 (5.00 mS/cm)
Power supply: External HI 983327-0 12VDC HI 98332700 12VDC isolated inside with DC/DC converter HI 983327-1 115/230 VAC; 50/60Hz
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”) Weight:
HI 983327-0, HI 98332700 140 g (4.96 oz) HI 983327-1 250 g (8.85 oz)
CE DECLARATION
OF CONFORMITY
Recommendations for Users
mize any EMC interferences. ovens.