These Instruments are in
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal band at all times. During op-
eration, ESD wrist straps should be worn to avoid possible damage to the
probe by electrostatic discharges. Any variation introduced by the user to
the supplied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to mini-
To avoid damage or burns, do not perform any measurement in microwave
Compliance with the CE Directives
against defects in work-
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:
• Instruction manual;
• Mounting brackets.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
GENERAL DESCRIPTION
HI 983324-0 and HI 983324-1 are conductivity indicators
and controllers with a relay output designed for simplicity of
use in a wide range of industrial applications.
The models are panel mounted with membrane keypads on
the front panel and an easy-to-read LCD display.
All the meters compensates for the temperature variation
automatically. The probe is easy to clean and requires little
maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
Power supply, wiring and selection are made via the plugin terminal blocks on the rear panel.
LED indicators on the front panel identify whether the
controller is in set or reading mode and if alarm is active.
SPECIFICATIONS
HI 983324
Range 0.0 to 49.9 ppm
Resolution 0.1 ppm
Accuracy (@20°C/68ºF) ±2% f.s.
Setpoint Adjustable from 0 to 49.0 ppm
Alarm Condition LED ON and alarm contact closed when
measured TDS value is higher than setpoint
Alarm Output 2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe HI 7634-00 EC/TDS probe
Temp. Compensation Automatic from 0 to 50°C
(32 to 122°F) with β=2%
TDS Conversion Factor 0.5
Calibration Manual, at one point through trimmer
Calibration Solution HI 7033 (84 µS/cm)
Power supply: External
HI983324-0 12 VDC
HI983324-1 115/230 VAC; 50/60Hz
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI983324-0 140 g (4.96 oz)
HI983324-1 250 g (8.85 oz)
mize any EMC interferences.
ovens.
OPERATIONAL GUIDE
Adjustable location brack-
meter) allow the controller
to slide into the cutout and
will hold the unit securely
in place. 95 mm (3.74") is
space required to install
84 µS/cm Calibration solution, 20mL
84 µS/cm Calibration solution, 230mL
84 µS/cm Calibration solution, 460mL
Electrode cleaning solution, 230 mL
12 VDC power adapter, US plug
12 VDC power adapter, European plug
12 VDC power adapter, Australian plug
12 VDC power adapter, South African plug
12 VDC power adapter, UK plug
Calibration Screwdriver (20 pcs)
POWER CONNECTION
HI 983324-0
Connect a 2-wire power cable to the terminal strip while
paying attention to the correct positive and negative contacts (12 VDC).
HI 983324-1
Connect a 3-wire power cable to the terminal strip while
paying attention to the correct earth, neutral and line
contacts (115/230 VAC).
The “ALARM” LED will light up when the alarm contact is
closed, to indicate a TDS value higher than selected setpoint.
CALIBRATION
Make sure the meter is in the measurement mode (the
“MEAS” LED lights on).
Immerse the probe in
solution.
Shake briefly and wait for reading to stabilize.
Using a small screwdriver adjust the calibration trimmer
until the meter displays “42” ppm.
HI 7033 (84 µS/cm) calibration
Front view of the
panel-mounted unit
Side view of the
panel-mounted unit
Notes:
• All external cables connected to the rear panel should
end with wire lugs.
• It is recommended to cover the unused terminals with
insulating tape.
• At start up the meter needs a few seconds to stabilize.
Wait until a stable reading is displayed.
ALARM CONTACT
Use this contact (maximum 2A, 220V) for connection to an
alarm or dosing system. The unit acts as a switch to control
an external device.
OPERATING THE METER
All operations can be controlled via front panel keys and
trimmers.
“SET” and “MEAS” LEDs light up to indicate which is the
operating function.
Make sure that the meter is calibrated and the Setpoint is
properly selected before performing any measurement.
Attach the probe to the meter. Install the probe in the
fittings or immerse it in the solution to be monitored, while
making sure that metal pins are completely submerged.
Press the “MEAS” key.
The LCD will show the TDS value of the solution in ppm unit.
Any initial variation on readings may be due to temperature compensation.
SETPOINT
Press the “SET” key. The display will show the default or
previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until
the required limit value is displayed.
PROBE MAINTENANCE
To improve probe performance and prolong its life, it is
recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning
Solution at least for one hour.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and
recalibrate the meter.
• When not in use, clean the probe before storing it.
the controller.
Panel-
mounted unit
assembling
view
HI 7634-00
HI 70033P
HI 7033M
HI 7033L
HI 7061M
HI 710005
HI 710006
HI 710012
HI 710013
HI 710014
HI 731326