Hanna Instruments HI 983321-0 User Manual

These Instruments are in
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electro-
static discharges. Avoid touching this metal band at all times. During operation,
ESD wrist straps should be worn to avoid possible damage to the probe by
electrostatic discharges. Any variation introduced by the user to the sup-
plied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to
minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in microwave
http://www.hannainst.com
CE DECLARATION
OF CONFORMITY
Compliance with the CE Directives
against defects in work-
Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center. The meter is supplied with:
• Instruction manual;
• Mounting brackets. Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
HI 983321-0 and HI 983321-1 are conductivity indicators and controllers with a relay output designed for simplicity of use in a wide range of applications. The models are panel mounted with membrane keypads on the front panel and an easy-to-read LCD display.
All the meters automatically compensate for the temperature variation. The probe is easy to clean and requires little maintenance.
Measurements are highly accurate and the meters can be calibrated at one point.
Power supply, wiring and selection are made via the plug­in terminal blocks on the rear panel.
LED indicators on the front panel identify whether the controller is in set or measurement mode, and if alarm is active.
SPECIFICATIONS
HI 983321
Range 0.00 to 19.99 ppm Resolution 0.01 ppm Accuracy (@ 20°C) ±2% f.s.
Setpoint Adjustable through multiturn trimmer Alarm Condition LED ON and alarm contact closed when
measured TDS value is higher than setpoint
Alarm Output 2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe HI 7634-00 EC/TDS probe TDS conversion factor 0.5 Temp.Compensation Automatic from 5 to 50°C
(41 to 122°F); ß=2%/ºC
Calibration Factory calibrated
and manual, at one point through trimmer
Calibration Solution recommended >13 ppm
(should be prepared by using HI 9034 TDS meter)
Power supply: External HI 983321-0 12 VDC HI 983321-1 115/230 VAC; 50/60Hz
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”) Weight:
HI 983321-0 140 g (4.96 oz) HI 983321-1 250 g (8.85 oz)
ovens.
ACCESSORIES
EC/TDS probe
Electrode cleaning solution, 230mL bottle
12 VDC power adapter, US plug
12 VDC power adapter, European plug
12 VDC power adapter, Australian plug
12 VDC power adapter, Southern Africa plug
12 VDC power adapter, UK plug
Calibration Screwdriver (20 pcs)
Mounting brackets
Supplied adjustable lo-
mm (3.74") is the mini-
HI 983321 LAYOUT
3. Probe Connection: follow the above diagram and connect the colored wires of probe cable as indicated. It is recommended to connect the shield (P5) to avoid any interference.
OPERATIONAL GUIDE
POWER CONNECTION
Press the “MEAS” key. The LCD will show the TDS value of the solution in ppm unit.
Any initial variation on readings may be due to tempera­ture compensation.
The “ALARM” LED will light up when the alarm contact is losed, to indicate an TDS value higher than selected setpoint.
Front view of the panel-mounted unit
HI 983321-0
Connect a 2-wire power cable to the terminal strip while paying attention to the correct positive and negative con­tacts (12 VDC).
HI 983321-1
Connect a 3-wire power cable to the terminal strip while paying attention to the correct earth, neutral and line contacts (115/230 VAC).
Notes:
• All external cables connected to the rear panel should end with wire lugs.
• It is recommended to cover the unused terminals with insulating tape.
• At start up the meter needs a few seconds to stabilize. Wait until a stable reading is displayed.
ALARM CONTACT
Use this contact (maximum 5A, 240VAC, 30VDC) for connec­tion to an alarm or dosing system. The unit acts as a switch to control an external device.
CALIBRATION
This meter is factory calibrated. If re-calibration becomes necessary, a calibration solution
has to be prepared with the help of HI 9034 TDS-meter (or similar). Recommended conductivity value of calibration solu­tion should be around 13 ppm.
Make sure the meter is in the measurement mode (the “MEAS” LED lights on).
Immerse the probe in the prepared solution. Shake briefly and wait for reading to stabilize. Using a small screwdriver adjust the calibration trimmer
until the meter displays the required TDS value on the LCD.
SETPOINT
Press the “SET” key. The display will show the default or previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until the required limit value is displayed.
PROBE MAINTENANCE
Side view of the panel-mounted unit
cation brackets allow the controller to slide into the cutout and will hold the unit securely in place. 95
mum amount of space required to install the controller.
Panel-
mounted unit
assembling
view
OPERATING THE METER
The contact
All operations can be controlled via front panel keys and trimmers.
“SET” and “MEAS” LEDs light up to indicate which is the operating function.
Make sure that the meter is calibrated and the Setpoint is properly selected before performing any measurement.
Attach the probe to the meter. Install the probe in the fittings or immerse it in the solution to be monitored, while making sure that metal pins are completely submerged.
To improve probe performance and prolong its life, it is recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning Solution at least for one hour.
• If a more thorough cleaning is required, brush the metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and recalibrate the meter.
• When not in use, clean the probe before storing it.