Instruction Manual
HI 983317-0
HI 983317-1
HI 98331700
Panel-Mounted
EC Indicators &
Controllers
These Instruments are in
Manufacturers since 1978
WARRANTY
These meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and
maintained according to instructions. The probe is warranted for a
period of six months. This warranty is limited to repair or replacement
free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This
manual will provide you with the necessary information for the correct operation of the meter. Please
read it carefully before using the meter. If you need
additional technical information, do not hesitate to
e-mail us at tech@hannainst.com.
These instruments are in compliance with the
directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material
and examine it carefully. If any damage has occurred during shipment, immediately notify your
Dealer or the nearest Hanna Customer Service Center.
The meter is supplied with:
• Instruction manual;
• Mounting brackets.
Note: Conserve all packing material until the in-
strument has been observed to function correctly. Any
defective item must be returned in its original packing.
GENERAL DESCRIPTION
HI 983317-0, HI 983317-1 and HI 98331700 are
conductivity indicators and controllers with a relay output
designed for simplicity of use in a wide range of applications.
HI 98331700 is provided with a DC/DC converter which
isolates the measure system from the external power supply.
The models are panel mounted with membrane keypads on
the front panel and an easy-to-read LCD display.
All the meters automatically compensate for the temperature variation. The probe is easy to clean and requires little
maintenance.
Measurements are highly accurate and the meters can be
calibrated at one point.
Power supply, wiring and selection are made via the plugin terminal blocks on the rear panel.
LED indicators on the front panel identify whether the
controller is in set or measurement mode, and dosing
mode.
FUNCTIONAL DESCRIPTION
FRONT PANEL
Keypad
SET To display
setpoint
MEAS To display
measurement
Trimmers
CAL For calibration
SET To adjust
setpoint
SET
LEDs
SET ON when LCD displays the set value
MEAS ON when LCD displays the measured value
ALARM ON when alarm contact is activated
REAR PANEL
1. Power Supply:
HI 983317-0 HI 983317-1
HI 98331700
12 VDC 115/230 VAC
L1: Positive 230 VAC
L2: Negative 115 VAC
L3: -------- Neutral
L4: --------- Protection Earth
Note: the power input is internally protected by a 400 mA fuse
2. Alarm Contact: A1, A2.
This contact acts as a switch for the power. The contact
has to be protected outside by the user.
CAL
3. Probe Connection: follow the above diagram and connect the
colored wires of probe cable as indicated. It is recommended
to connect the shield (P5) to avoid any interference.
HI 983317 LAYOUT
73mm
79mm
2.87"
ADJUSTABLE
LOCATION
BRACKET
3.11"
Front view of the
panel-mounted unit
Side view of the
panel-mounted unit
Supplied adjustable
location brackets allow the controller to
slide into the cutout
and will hold the unit
securely in place. 95
mm (3.74") is the minimum amount of space
required to install the
controller.
Panel-mounted unit assembling view
79mm
3.11"
42mm
1.65"
49mm
1.93"
0.25/4mm
0.01/0.160"
76mm
3"
95mm MIN
3.74"
HI 740146
OPERATIONAL GUIDE
POWER CONNECTION
HI 983317-0 and HI 98331700
Connect a 2-wire power cable to the terminal strip while
paying attention to the correct positive and negative polarities (12 VDC).
HI 983317-1
Connect a 3-wire power cable to the terminal strip while
paying attention to the correct earth, neutral and line
contacts (115/230 VAC).
Notes:
• All external cables connected to the rear panel should
end with wire lugs.
• It is recommended to cover the unused terminals with
insulating tape.
• At start up the meter needs a few seconds to stabilize.
Wait until a stable reading is displayed.
ALARM CONTACT
Use this contact (maximum 5A, 240VAC, 30VDC) for connection to an alarm or dosing system. The unit acts as a switch
to control an external device.
OPERATING THE METER
All operations can be controlled via front panel keys and
trimmers.
“SET” and “MEAS” LEDs light up to indicate which is the
operating function.
Make sure that the meter is calibrated and the Setpoint is
properly selected before performing any measurement.
Attach the probe to the meter. Install the probe in the
fittings or immerse it in the solution to be monitored, while
making sure that metal pins are completely submerged.
Press the “MEAS” key.
The LCD will show the conductivity value of the solution in
mS/cm unit. Any initial variation on readings may be due
to temperature compensation.
The “ALARM” LED will light up when the alarm contact is
closed, to indicate an EC value lower than selected setpoint.
CALIBRATION
Make sure the meter is in the measurement mode (the
“MEAS” LED lights on).
Immerse the probe in HI 7039 (5.00 mS/cm) calibration
solution.
Shake briefly and wait for reading to stabilize.
Using a small screwdriver adjust the calibration trimmer
until the meter displays “5.00” on the LCD.
SETPOINT
Press the “SET” key. The display will show the default or
previously adjusted value for the setpoint.
Using a small screwdriver, adjust the “SET” trimmer until
the required limit value is displayed.
PROBE MAINTENANCE
To improve probe performance and prolong its life, it is
recommended to clean it regularly.
• Immerse the tip of the probe in HI 7061 Cleaning
Solution at least for one hour.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water and
recalibrate the meter.
• When not in use, clean the probe before storing it.
ACCESSORIES
HI 7632-00 EC probe
HI 70039P 5.00 mS/cm (5000 µS/cm) calibration solu-
tion, 20mL sachet (25 pcs)
HI 7039M 5.00 mS/cm (5000 µS/cm) calibration solu-
tion, 230mL bottle
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 710005 12 VDC power adapter, US plug
HI 710006 12 VDC power adapter, European plug
HI 710012 12 VDC power adapter, Australian plug
HI 710013 12 VDC power adapter, Southern Africa plug
HI 710014 12 VDC power adapter, UK plug
HI 731326 Calibration Screwdriver (20 pcs)
HI 740146 Mounting brackets
SPECIFICATIONS
HI 983317
Range 0.00 to 10.00 mS/cm
Resolution 0.01 mS/cm
Accuracy (@ 20°C) ±2% f.s.
Setpoint Adjustable through multiturn trimmer
Alarm Condition LED ON and alarm contact closed when
measured EC value is lower than setpoint
Alarm Output 2-contact relay, no fuse protected.
5A, 240 VAC, 30 VDC
Probe HI 7632-00 EC probe
Temp.Compensation Automatic from 5 to 50°C
(41 to 122°F); ß=2%/ºC
Calibration Manual, at one point through trimmer
Calibration Solution HI 7039 (5.00 mS/cm)
Power supply: External
HI 983317-0 12VDC
HI 98331700 12VDC isolated inside with DC/DC converter
HI 983317-1 115/230 VAC; 50/60Hz
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7”)
Weight:
HI 983317-0, HI 98331700 140 g (4.96 oz)
HI 983317-1 250 g (8.85 oz)
CE DECLARATION
OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used. Operation of these instruments in
residential areas could cause unacceptable interferences to radio and TV
equipment. The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching this metal band at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the
probe by electrostatic discharges. Any variation introduced by the user to
the supplied equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use these instruments when voltage at the
measurement surface exceeds 24 VAC or 60 VDC. Use plastic beakers to
minimize any EMC interferences.
To avoid damage or burns, do not perform any measurement in microwave
.
ovens
ISTR983317R5 12/2001