Hanna Instruments HI 98311 User Manual

Instruction Manual
HI 9831 1 - HI 98312
Waterproof
EC/TDS & Temperature Meters
These Instruments are
www.hannainst.com
WARRANTY
HI 98311 and HI 98312 are warranted for one year against defects in workmanship and materials
when used for their intended purpose and maintained according to instructions. The probe is warranted for a period of six months. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be re­turned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, con­struction and appearance of its products without advance notice.
in Compliance
with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for a correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.
Each meter is supplied with:
HI 73311 EC/TDS probe
HI 73128 Probe removal tool
• 4 x 1.5V batteries Note: Conserve all packing material until the instru-
ment has been observed to function correctly. Any defective item must be returned in its original packing.
US DESIGN PATENT
D462,024
GENERAL DESCRIPTION
HI 98311 and HI 98312 are waterproof EC/TDS and temperature meters. The housing has been com­pletely sealed against humidity and designed to float.
All EC/TDS readings are automatically temperature compensated (ATC), and temperature values can be displayed in °C or °F units.
The EC/TDS conversion factor (CONV) is selectable by the user, as well as the temperature compensation coefficient β (BETA).
The meters can be calibrated at one point. Measurements are highly accurate with a unique
stability indicator right on the LCD. A low battery symbol warns the user when the batter-
ies are to be replaced. In addition the Battery Error Prevention System (BEPS) avoids erroneous reading caused by low voltage level by turning the meter off.
The HI 73311 EC/TDS probe, supplied with the meter, is interchangeable and can be easily replaced by the user.
The stainless steel encapsulated temperature sensor facilitates faster and more accurate temperature mea­surement and compensation.
FUNCTIONAL DESCRIPTION
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. ON/OFF/MODE button
4. HI 73311 EC/TDS probe
5. Temperature sensor
6. SET/HOLD button
1. Automatic temperature compensation indicator
2. Stability indicator
3. Battery life percentage indicator
4. Low battery indicator
5. Secondary display
6. Primary display
7. Measuring units for primary display
SPECIFICATIONS
Range
Temperature:
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Temperature
Compensation
Environment
EC/TDS Conversion
Factor
Calibration
EC/TDS Cal.solutions
HI 98311:
HI 98312:
Probe (included)
Battery Type/Life
Auto-off
Dimensions
Weight
Recommendations for Users Before using this product, make sure that it is entirely suit­able for the environment in which it is used. Operation of this instrument in residential areas could cause unaccept­able interferences to radio and TV equipment. Avoid touching the probes at all times. Any variation introduced by the user to the supplied equip­ment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any mea­surement in microwave ovens.
0.0 to 60.0°C / 32.0 to 140.0°F 0 to 3999 µS/cm (HI 98311)
EC:
0.00 to 20.00 mS/cm (HI 98312) 0 to 2000 ppm (HI 98311)
TDS:
0.00 to 10.00 ppt (HI 98312)
Temp:
EC/TDS:
Temp:
EC/TDS:
Temp:
EC/TDS:
0.1°C / 0.1°F
1 µS/cm ; 1 ppm (HI 98311)
0.01 mS/cm ; 0.01 ppt (HI 98312) ±0.5°C / ±1°F
±0.5°C / ±1°F
Automatic, with β=0.0 to 2.4%/ºC
0 to 50°C (32 to 122°F); RH 100%
0.45 to 1.00 (CONV)
Automatic, at 1 point
HI7031 (1413 µS/cm) HI7032 (1382 ppm; CONV=0.5) HI70442 (1500 ppm; CONV=0.7) HI7030 (12.88 mS/cm) HI70038 (6.44 ppt; CONV=0.5
or 9.02 ppt; CONV=0.7)
HI 73311 EC/TDS probe
4 x 1.5V with BEPS/typical 100 hours
After 8 minutes
163 x 40 x 26 mm (6.4 x 1.6 x 1.0”)
85 g (3.0 oz)
ISTR98311R2 03/03
±2% f.s.
±2% f.s.
OPERATIONAL GUIDE
CALIBRATION
BATTERY REPLACEMENTPROBE MAINTENANCE
To turn the meter on and to check battery status
Press and hold the /MODE button for 2-3 sec­onds. All the used segments on the LCD will be visible for a few seconds, followed by a percent indication of the remaining battery life (E.g. % 100 BATT).
Taking measurements
Submerge the probe in the solution to be tested. Use plastic beakers to minimize any electromagnetic inter­ferences.
Select either EC or TDS mode with the SET/HOLD button.
The measurements should be taken when the stability symbol on the top left of the LCD disappears.
The EC (or TDS) value automati­cally compensated for temperature is shown on the primary LCD while the secondary LCD shows the temperature of the sample.
To change the temperature unit
To change the temperature unit (from °C to °F), from measurement mode, press and hold the
button until TEMP and the current temperature unit are displayed on the lower portion of the LCD (E.g. TEMP °C).
Use the SET/HOLD button to change the temperature unit, and then press the
return to the normal measuring mode.
To freeze the display
Press the SET/HOLD button for 2-3 seconds until HOLD appears on the secondary display.
Press either button to return to the normal measuring mode.
To turn the meter off
Press the /MODE button while in normal measur­ing mode. OFF will appear on the lower part of the display. Release the button.
Notes:
• Before taking any measurement make sure the meter has been calibrated.
• If measurements are taken in different samples successively, rinse the probe thoroughly to eliminate cross-contamination; and after cleaning, rinse the probe with some of the sample to be measured.
/MODE button twice to
/MODE
For better accuracy, frequent calibration of the instru­ment is recommended. In addition, the instrument must be recalibrated whenever:
a) The EC/TDS probe is replaced. b) After testing aggressive chemicals. c) Where high accuracy is required. d) At least once a month.
To change the EC/TDS coversion factor (CONV) and the temperature compensation coefficient (BETA)
• From measurement mode, press and hold the / MODE button until TEMP and the current tempera­ture unit are displayed on the lower LCD (E.g. TEMP °C).
• Press the /MODE button again to show the current conversion factor (E.g. 0.50 CONV).
• Press the SET/HOLD button to change the conver­sion factor.
• Press the temperature compensation coefficient β (E.g. 2.1 BETA).
• Press the SET/HOLD button to change the the temperature compensation coefficient β.
• Press the measuring mode.
Calibration procedure
• From measurement mode, press and hold the / MODE button until CAL is displayed on the lower LCD.
• Release the button and immerse the probe in the proper calibration solution: HI7031 (1413 µS/cm) for HI98311 and HI7030 (12.88 mS/cm) for HI98312.
• Once the calibration has been automatically per­formed, the LCD will display OK for 1 second and the meter will return to normal measurement mode.
• Since there is a known relathionship between EC and TDS readings, it is not necessary to calibrate the meter in TDS. If the EC/TDS conversion factor is either
0.5 or 0.7, the meter will allow a direct calibration in ppm by using the Hanna calibration solutions listed below.
The CAL symbol on the LCD means that the meter is calibrated.
To reset to the default calibration
To clear a previous calibration, press the MODE button after entering the calibration mode. The lower LCD will display ESC for 1 second and the meter will return to normal measurement mode. The CAL symbol on the LCD will disappear. The meter will be reset to the default calibration.
/MODE button to show the current
/MODE button to return to the normal
The EC/TDS probe can be easily replaced by using the supplied tool (HI 73128). Insert the tool into the probe cavity as shown below.
ββ
β
ββ
Rotate the probe counterclockwise.
Pull the probe out by using the other side of the tool. Insert a new probe following the above instructions in reverse order.
The meter displays the remaining battery percentage every time it is switched on. When the battery level is below 5%, the lights up to indicate a low battery condition. The batter­ies should be replaced soon. If the battery level is low enough to cause erroneous readings, the meter shows “0%” and the Battery Error Prevention System (BEPS) will automatically turn the meter off. To change the batteries, remove the 4 screws located on the top of the meter.
Once the top has been removed, carefully replace the 4 batteries located in the compartment while paying atten­tion to their polarity.
Replace the top, making sure that the gasket is properly seated in place, and tighten the screws to ensure a
watertight seal.
- +
symbol on the bottom left of the LCD
ACCESSORIES
HI 73311 Replaceable EC/TDS probe HI 73128 Probe removal tool HI 70030P 12.88 mS/cm @25ºC calibration solu-
HI 70031P 1413 µS/cm @25ºC calibration solu-
HI 70032P 1382 ppm @25ºC calibration solution,
HI 70038P 6.44 ppt @25ºC calibration solution, 20
HI 70442P 1500 ppm @25ºC calibration solution,
tion, 20 mL sachet (25 pcs)
tion, 20 mL sachet (25 pcs)
20 mL sachet (25 pcs)
mL sachet (25 pcs)
20 mL sachet (25 pcs)