Hanna Instruments HI 98308 User Manual

SPECIFICATIONS Range 0.0 to 99.9 µS/cm Resolution 0.1 µS/cm Accuracy (@20°C/68°F) ±2% f.s. Typical EMC Deviation ±2% f.s. Calibration Manual, 1 point, with trimmer Cal. Solution HI 70033 (84.0 µS/cm) Temperature Compensation
Automatic, 0 to 50°C (32 to 122ºF),
with ß=2%/°C typical
Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Battery Type 4 x 1.5V alkaline Battery Life approx. 250 hours of use Dimensions 175 x 41 x 23 mm
(6.9 x 1.6 x 0.9")
Weight 95 g (3.4 oz.)
ACCESSORIES HI 70033P 84 µS/cm calibration solution,
20 mL sachet, 25 pcs
HI 7033M 84 µS/cm calibration solution,
HI 73308 Replaceable probe for PWT HI 731326 Calibration screwdriver (20 pcs)
PWT
HI 98308
Pure Water Tester
with Automatic Temperature Compensation
WARRANTY
This meter is guaranteed for one year against defects in workmanship and materials when used for its intended pur­pose and maintained according to instructions. Probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, re­port the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with ship­ment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
The PWT is in compliance with the CE directives.
ISTPWTR2 05/05
www.hannainst.com
OPERATION
• Remove the protective cap and turn the PWT on by sliding the switch on the top of the meter.
• Immerse it into the solution to be tested without exceeding the maxi­mum immersion level.
• Stir gently and wait for the reading to stabilize. PWT automatically compensates for temperature variations.
• Measurements are displayed in µS/cm unit.
HOW TO USE THE CHART
• The following diagram shows the relation be­tween conductivity (µS/cm) and NaCl concen­tration (mg/L) of a watery solution.
CALIBRATION
• Immerse the PWT in the HI 70033 calibration solution, without exceeding the maximum im­mersion level.
• Allow the reading to stabilize, and then using the supplied screwdriver adjust the calibration trimmer on the meter side until the display
shows "84.0".
BEPS AND BATTERY REPLACEMENT: PWT is provided with BEPS (Battery Error Pre-
venting System) which avoids any erroneous reading due to low batteries level, by automatically switching the meter off.
To replace the batteries, slide off the battery compartment cover and replace all four 1.5V batteries, while paying attention to the correct polarity.
• These are some typical conversions of measure­ments performed with PWT:
µS/cm reading Resistivity mg/L NaCl
99.9 10 k 48 10 100 k 4.6
11 M 0.4
0.1 10 M 0.02
Batteries should only be replaced in a non­hazardous area and using the battery type specified in this instruction manual.
PROBE REPLACEMENT
For replacing the probe, contact your Dealer or the nearest Hanna Service Center.