OPERATION:
• Remove the protective cap.
• Turn the
• Immerse it in the solution to be tested without
• Stir gently and wait for the reading
PWTPWT
PWT on.
PWTPWT
exceeding the maximum immersion level.
PWT automatically compensates for
PWTPWT temperature variations.
HOW TO USE THE CHART:
• The following diagram shows the relation be­tween conductivity (µS/cm) and NaCl concentra­tion (mg/L) of a watery solution.
• The reading on display is directly expressed in µS/cm unit.
CALIBRATION:
• Immerse the solution, without exceeding the maximum im­mersion level.
• Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer until the display shows "84.0".
PWT PWT
PWT in the HI70033 calibration
PWT PWT
BEPS AND BATTERY REPLACEMENT:
PWT PWT
PWT is supplied with BEPS (Battery Error Preventing
PWT PWT System) which avoids any erroneous reading due to low batteries level, by switching the meter off. To replace the batteries, slide off the battery com­partment cover and replace all four 1.5V batteries while paying attention to their polarity.
Batteries should only be replaced in a nonhazardous area using the battery type specified in this instruction sheet.
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
Covered by manship and materials, this tester is now completely serviceable. Contact your dealer for further information.
The
one yearone year
one year warranty against defects in work-
one yearone year
PWT PWT
PWT is in compliance with the CE directives.
PWT PWT
Manufacturers since 1978
Visit our Internet Home Page:
http://www.hannainst.com
PWT
HI98308
Pure Water Tester
with Automatic Temperature Compensation
SPECIFICATIONS:
RANGE 0.0  99.9 µS/cm RESOLUTION 0.1 µS/cm ACCURACY (@20°C/68°F) ±2% f.s. TYPICAL EMC DEVIATION ±2% f.s. CALIBRATION Manual, at 1 point through trimmer CAL. SOLUTION HI70033 (84.0 µS/cm) TEMPERATURE 0 to 50C (32 to 122ºF) COMPENSATION with ß=2%/C typical ENVIRONMENT 0 to 50C (32 to 122F);
95% RH not condensing
BATTERY TYPE 4 x 1.5V alkaline (included) BATTERY LIFE approx. 250 hours
of continuous use
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT 78 g (2.7 oz.)
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the below the battery compartment and slide off the whole internal part of the tester.
PWTPWT
PWT located
PWTPWT
Internal circuit:
• Fasten the screw on the back, below the battery compartment.
ACCESSORIES:
HI 70033P 84 µS/cm calibration kit
(25 x 20 mL sachets)
HI 7033M 84 µS/cm calibration solution
(230 mL bottle)
HI 73308 Probe for PWT HI 731326 Calibration screwdriver (20 pcs)
ISTPWTR1 07/01
• Replace the built-in electrode with a new HI73308.
• Before reinserting the internal circuit into the new plastic casing, make sure the O-rings (2) of the temperature sensor (3) are properly seated (see drawing).
To avoid bending the connections to the con­ductivity sensors (1), do not apply excessive force while inserting the circuit. In case of any problem, reslide and reinsert it.
• Reinsert the battery compartment cover.
• Recalibrate the measurements.