DiST
HI98300/1/2/3/4
Conductivity/TDS Meters
with Automatic Temperature
Compensation
OPERATION:
• Remove the protective cap.
• Turn the
• Immerse it in the solution to be tested without
• Stir gently and wait for the reading
• The reading on display is directly expressed in:
DiSTDiST
DiST on.
DiSTDiST
exceeding the maximum immersion level.
to stabilize.
DiSTDiST
DiST automatically compensates for
DiSTDiST
temperature variations.
DiST1DiST1
HI98300 HI98300
ppm (mg/L) unit for
ppt (g/L) unit for
µS/cm for
mS/cm for
DiST3DiST3
DiST3 (
DiST3DiST3
DiST4DiST4
DiST4 (
DiST4DiST4
DiST1 (
DiST1DiST1
DiST2DiST2
HI98302HI98302
DiST2 (
HI98302),
DiST2DiST2
HI98302HI98302
HI98303HI98303
HI98303) and
HI98303HI98303
HI98304HI98304
HI98304).
HI98304HI98304
HI98300 &
HI98300 HI98300
HI98301 HI98301
HI98301),
HI98301 HI98301
BEPS AND BATTERY REPLACEMENT:
DiST DiST
DiST is supplied with BEPS (Battery Error Preventing
DiST DiST
System) which avoids any erroneous reading due to
low batteries level, by switching the meter off.
To replace the batteries, slide off the battery compartment cover and replace all four 1.5V batteries
while paying attention to their polarity.
Batteries should only be replaced in a nonhazardous
area using the battery type specified in this instruction
sheet.
Manufacturers since 1978
CALIBRATION:
• Immerse the
solution (see Specifications), without exceeding
the maximum immersion level.
• Allow the reading to stabilize and with a small
screwdriver turn the calibration trimmer until the
display shows the required solution value.
E.g. for calibrating
tion solution is
be turned until the display shows "1382".
DiST DiST
DiST in the indicated calibration
DiST DiST
HI98301HI98301
HI98301 the correct calibra-
HI98301HI98301
HI70032HI70032
HI70032, and the trimmer should
HI70032HI70032
ACCESSORIES:
HI 70030P 12880 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70031P 1413 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70032P 1382 ppm (mg/L) calibration kit
(25 x 20 mL sachets)
HI 70038P 6.44 ppt (g/L) calibration kit
(25 x 20 mL sachets)
HI 70039P 5000 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70442P 1500 ppm calibration kit
(25 x 20 mL sachets)
HI 73301 Spare electrode for DiST1
HI 73302 Spare electrode for DiST2
HI 73303 Spare electrode for DiST3
HI 73304 Spare electrode for DiST4
HI 731326 Calibration screwdriver (20 pcs)
Visit our Internet Home Page:
http://www.hannainst.com
SPECIFICATIONS:
RANGE / RESOLUTION
DiST1 (HI98300 & HI98301) 0 to 1999 ppm/1ppm
DiST2 (HI98302) 0.00 to 10.00 ppt / 0.01ppt
DiST3 (HI98303) 0 to 1999 µS/cm / 1µS/cm
DiST4 (HI98304) 0.00 to 19.99 mS/cm / 0.01mS/cm
ACCURACY (@20°C/68°F) ±2% f.s.
TYPICAL EMC DEVIATION ± 1% f.s.
TDS FACTOR
HI98300 0.65
HI98301 & HI98302 0.5
CALIBRATION SOLUTIONS
HI98300 HI70442 (1500 ppm)
HI98301 HI70032 (1382 ppm)
HI98302 HI70038 (6.44 ppt)
HI98303 HI70031 (1413 µS/cm)
HI98304 HI70039 (5.00 mS/cm)-HI70030 (12.88 mS/cm)
TEMP.COMPENSATION 0 to 50C (32 to 122ºF)
ENVIRONMENT 0 to 50C (32 to 122F);
95% RH not condensing
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY LIFE approx. 200 hours
of continuous use
DIMENSIONS 175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT 78 g (2.7 oz.)
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:ONE YEAR WARRANTY AND SERVICEABLE:
Covered by
one yearone year
one year warranty against defects in work-
one yearone year
manship and materials, these testers are now completely
serviceable. Contact your dealer for further information.
DiST DiST
The
DiST is in compliance with the CE directives.
DiST DiST
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and
can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the
DiSTDiST
DiST located
DiSTDiST
below the battery compartment and slide off the
whole internal part of the tester.
• Replace the built-in electrode with a new:
HI73301HI73301
HI73301 for
HI73301HI73301
HI73302HI73302
HI73302 for
HI73302HI73302
HI73303HI73303
HI73303 for
HI73303HI73303
HI73304HI73304
HI73304 for
HI73304HI73304
DiST1DiST1
DiST1
DiST1DiST1
DiST2DiST2
DiST2
DiST2DiST2
DiST3DiST3
DiST3
DiST3DiST3
DiST4DiST4
DiST4
DiST4DiST4
• Before reinserting the internal circuit into the
new plastic casing, make sure the O-rings (2)
of the temperature sensor (3) are properly
seated (see drawing).
To avoid bending the connections to the conductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
Internal circuit:
• Fasten the screw on the back, below the
battery compartment.
• Reinsert the battery compartment cover.
• Recalibrate the
measurements.
DiSTDiST
DiST before performing new
DiSTDiST
DISTNEW-R3 07/01