Hanna Instruments HI 98300 User Manual

SPECIFICATIONS Range / Resolution DiST®1 (HI98300/301) 1999 ppm / 1ppm DiST®2 (HI98302) 10.00 ppt / 0.01ppt DiST®3 (HI98303) 1999 µS/cm / 1µS/cm DiST®4 (HI98304) 19.99 mS/cm / 0.01mS/cm Accuracy (@20°C/68°F) ±2% f.s.
Typical EMC Deviation ± 1% f.s. TDS Factor
HI 98300 0.65 HI 98301 / HI 98302 0.5
Calibration Solutions HI 98300 HI 70442 (1500 ppm) HI 98301 HI 70032 (1382 ppm) HI 98302 HI 70038 (6.44 ppt) HI 98303 HI 70031 (1413 µS/cm) HI 98304 HI 70039 (5.00 mS/cm) or
HI 70030 (12.88 mS/cm)
Temperature Compensation
Automatic, 0 to 50°C (32 to 122ºF)
Environment 0 to 50°C (32 to 122°F);
RH 95% non-condensing
Battery Type 4 x 1.5V alkaline Battery Life approx. 200 hours of use Dimensions 175 x 41 x 23 mm
(6.9 x 1.6 x 0.9")
Weight 95 g (3.4 oz.)
DiST
®
HI 98300 / 301 / 302 HI 98303 / 304
EC and TDS Testers with ATC
WARRANTY
These meters are guaranteed for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge. Damages due to acci­dent, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with ship­ment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
The DiST® models are in compliance with the CE directives.
ISTDISTNEW-R5 05/05
www.hannainst.com
OPERATION
• Remove the protective cap, turn the DiST® on, and immerse it into the solution to be tested, without ex­ceeding the maximum immersion level.
• Stir gently and wait for the read­ing to stabilize. DiST® automati­cally compensates for temperature variations.
• The reading on display is directly expressed in: ppm (mg/L) for DiST®1 (HI 98300 / 301),
ppt (g/L) for DiST®2 (HI 98302), µS/cm for DiST®3 (HI 98303) and mS/cm for DiST®4 (HI 98304).
CALIBRATION
• Immerse the DiST® in the proper calibration solution (see Specifications table), without ex­ceeding the maximum immersion level.
• Allow the reading to stabilize and with a small screwdriver adjust the calibration trim­mer until the display shows the required solu­tion value.
For example HI 70032 is the correct calibration so­lution for HI 98301, and the trimmer should be turned until the display shows "1382".
PROBE REPLACEMENT
For replacing the probe, contact your Dealer or the nearest Hanna Service Center.
• Replace the built-in probe with a new:
HI 73301 for DiST®1 HI 73302 for DiST®2 HI 73303 for DiST®3 HI 73304 for DiST®4
BEPS AND BATTERY REPLACEMENT DiS®T is provided with BEPS (Battery Error Pre-
venting System), which avoids any erroneous reading due to low battery level, by switching the meter off.
To replace the batteries, slide off the battery compartment cover and replace all four 1.5V batteries while paying attention to their polarity.
Batteries should only be replaced in a non­hazardous area, and using the battery type specified in this instruction manual.
ACCESSORIES HI 70030P 12880 µS/cm calibration solu-
tion, 20 mL sachet, 25 pcs
HI 70031P 1413 µS/cm calibration solution,
20 mL sachet, 25 pcs
HI 70032P 1382 ppm (mg/L) calibration so-
lution, 20 mL sachet, 25 pcs
HI 70038P 6.44 ppt (g/L) calibration solu-
tion, 20 mL sachet, 25 pcs
HI 70039P 5000 µS/cm calibration solution,
20 mL sachet, 25 pcs
HI 70442P 1500 ppm calibration solution,
20 mL sachet, 25 pcs
HI 73301 Spare electrode for DiST®1 HI 73302 Spare electrode for DiST®2 HI 73303 Spare electrode for DiST®3 HI 73304 Spare electrode for DiST®4 HI 731326 Calibration screwdriver (20 pcs.)