Hanna Instruments HI 98204 User Manual

ON / OFF SWITCH
RANGE SWITCH
LIQUID
CRYSTAL
DISPLAY
MAXIMUM
IMMERSION
LEVEL
BLACK
COMPARTMENT
FRONT
BATTERY
COVER
OFFSET & SLOPE
CALIBRATION
TRIMMERS
COVER
BACK
Water Test
ON / OFF
4 pH 7
RANGE
IMMERSION LEVEL
MAINTENANCE
• Separate the compartment from the tester.
• Remove deposits formed on the liquid junction and the pH glass bulb by soaking in HI 7061 solution for 1½ hour.
• Soak the platinum wire in acetone for half hour.
• Rinse the bulb, platinum wire and liquid junction thoroughly with distilled water
• Reassemble the compartment and the tester.
• Fill the compartment with tap water.
BATTERY REPLACEMENT
When the Water Test cannot be switched on or the display fades, pull out the battery cover and replace all three 1.5V batteries, while paying attention to their polarity. Battery replacement must only take place in a non hazardous area using the battery types speci­fied in this instruction manual.
ACCESSORIES
HI 7004M Buffer solution pH 4.01 (230 ml) HI 7004L Buffer solution pH 4.01 (460 ml) HI 7007M Buffer solution pH 7.01 (230 ml) HI 7007L Buffer solution pH 7.01 (460 ml) HI 7010M Buffer solution pH 10.01 (230 ml) HI 7010L Buffer solution pH 10.01 (460 ml) HI 7031M Conductivity solution 1413 µS/cm @25°C (230 ml) HI 7031L Conductivity solution 1413 µS/cm @25°C (460 ml) HI 7061M General Cleaning solution (230 ml) HI 7061L General Cleaning solution (460 ml) BATT009/P 1.4V battery (100 pcs) HI 76501/P Calibration screwdriver (20 pcs)
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic discharge.
Avoid touching this glass bulb at all times. For calibration it is recommended to use an antistatic
screwdriver. Any variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance. To avoid electrical shock, do not use this instrument when voltages
at the measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in
microwave ovens.
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the pH / ORP / Conductivity / Temperature meter
has been tested and found to be in compliance with the following regulations:
IEC 801-2 Electrostatic Discharge IEC 801-3 RF Radiated EN 55022 Radiated, Class B
Date of Issue: 12-12-1995
Water Test
Water Test
ON / OFF
RANGE
4 pH 7
IMMERSION LEVEL
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
BATTERY
COVER
WATERTSTR1
03/96
WATER TEST
The "four-in-one" portable meter
GENERAL DESCRIPTION
The HANNA Water Test is a "four-in-one" portable meter designed for fast and accurate analysis and monitoring of water in biological applications, industrial wastewater, rivers, lakes, etc. It measures the main parameters in water analysis, namely, pH, oxidation-reduction potential (REDOX), conductivity and temperature. All is needed for the analysis is a small sample of the water which is poured into the compartment and the 4 measurements are immediately available. The instrument has a complete pH range from 0 to 14, an ORP range from -1000 to +1000 mV, a conductivity range from 0 to 2000 µS/cm and a temperature range from 0 to 60°C. The instrument can be calibrated for pH and conductivity and comes with replaceable batteries.
SPECIFICATIONS
RANGE RESOLUTION
ACCURACY
(@20°C/68°F)
TYPICAL EMC DEVIATION
CALIBRATION
BATTERY
DIMENSIONS
WEIGHT
Stay in Touch with the Manufacturer !
pH ORP Conductivity Temperature
0.0 to 14.0 pH ±1000 mV 0 to 1999 µS/cm 0.0 to 60.0°C
0.1 pH 1 mV 1 µS/cm 0.1°C
±0.2 pH ±5 mV ±2% Full Scale ±1°C
±0.4 pH ±2mV ±2% Full Scale ±1°C
pH: Manual double point through Offset and Slope trimmers µS/cm: Manual single point through trimmer
3 x 1.5V batteries for approx. 200 hours of continuous use
190 x 85 x 85 mm (7.4 x 3.3 x 3.3")
260 g (9.2 oz.)
EN 50081-1 EN 50082-1
OPERATING INFORMATION
• Do not be alarmed if white crystals appear around the cap. This is normal with electrodes and they dissolve when rinsed with water.
Water Test
ON / OFF
RANGE
4 pH 7
• Take a sample of the water to be tested and pour into the black compartment of the Water Test. If you are scooping up the water with the black com­partment, do not immerse above the maximum im­mersion level indicated.
• Turn the tester on by pressing the ON/OFF switch on the front panel.
ON / OFF
°C
• The tester will automatically default the tem­perature measurement indicated by the "°C" symbol on the display.
• Press the RANGE switch once to display
RANGE
µS
the conductivity expressed in µS.
• Press the RANGE switch once more to dis­play pH.
• Press the RANGE switch once again to display ORP expressed in mV.
• If the tester compartment is dry for more than 2
pH
mV
days, you will have to wait for about 5 minutes be­fore taking pH measurements.
• In order to take more accurate pH and conductivity measurements, make sure that Water Test is calibrated for pH and conductivity before use.
• When the measurements are completed, empty the com-
ON / OFF
partment, rinse with tap water to minimize contamination and switch the tester off by pressing the ON/OFF switch again.
• In order to have a fast response time during mea­surement, the compartment of the tester should be filled with tap water when the tester is not in use.
Water Test
ON / OFF
RANGE
4 pH 7
For long term storage, the tester should be kept in the original box.
DO NOT USE DISTILLED OR DEIONIZED
WATER FOR STORAGE PURPOSES.
pH CALIBRATION
• Pour a small quantity of distilled water into the black compartment, rinse by shaking the com­partment and then discard the content.
Water Test
ON / OFF
RANGE
4 pH 7
ON / OFF
07
I 70 H
• Pour a small quantity of pH 7 (HI 7007) buffer solution into the compartment and switch the tester on by pressing the ON/OFF switch.
RANGE
• Press the RANGE switch twice to display pH measure­ments.
HI 7004
• If the display does not show "7.0 pH", adjust the right trimmer located behind the cover until the display shows "7.0 pH".
Water Test
ON / OFF
4 pH 7
RANGE
pH
• Discard the pH 7 solution, pour a small quantity of distilled water into the black compartment, rinse by shaking the com­partment and then discard the content.
Water Test
ON / OFF
RANGE
4 pH 7
4
0 0
I 7 H
• Pour a small quantity of pH 4 (HI 7004) or pH 10 (HI7010) buffer solution into the compartment.
• If the display does not show "4.0 pH" (or "10.0 pH"), adjust the left trimmer lo­cated behind the cover until the display
Water Test
ON / OFF
4 pH 7
RANGE
pH
shows "4.0 pH" (or "10.0 pH").
• Discard the pH 4 (or pH 10) solution, pour a small quantity of distilled water into the black compartment, rinse by shaking the compartment and then discard the content.
• The unit is now calibrated for pH and ready for use.
CONDUCTIVITY CALIBRATION
• Pour a small quantity of distilled water into the black
Water Test
ON / OFF
RANGE
4 pH 7
compartment, rinse by shaking the compartment and then discard the content.
• Fill the compartment with HI 7031 conductivity solu­tion to overflow and switch the tester on by pressing the ON/OFF switch. The tempera­ture of the solution will be displayed.
ON / OFF
°C
• Note the temperature of the solution and read the corresponding conductiv­ity from the table printed on the solution label (e.g. 1413 µS/cm @ 25°C).
• Remove the screw on the right of the casing. Press the RANGE switch to display conductivity measurements and adjust the trimmer on the right of the cas­ing until the display shows the corresponding
°C °F HI7031
0 32 776
5 41 896 10 50 1020 15 59 1147 16 60.8 1173 17 62.6 1199 18 64.4 1225 19 66.2 1251 20 68 1278 21 69.8 1305
RANGE
S/cm)
°C °F HI7031
22 71.6 1332 23 73.4 1359 24 75.2 1386 25 77 1413 26 78.8 1440 27 80.6 1467 28 82.4 1494 29 84.2 1521 30 86 1548 31 87.8 1575
S/cm)
µS
conductivity reading at the noted temperature (e.g. 1278 µS/cm @20°C). Re-
RIGHT
SIDE
place the screw after cali­bration.
ALWAYS USE FRESH BUFFERS
FOR CALIBRATION & NEVER REUSE THEM.
1
3 0
I 7 H