Hanna Instruments HI98194, HI98195, HI98196 Instructions Manual

Page 1
HI98194,
HI98195, HI98196
Multiparameter Meters
pH/mV, ORP, EC, TDS, Resistivity,
Salinity, Seawater σ, Dissolved
Oxygen, Atmospheric Pressure
and Temperature
Page 2
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct
use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e‑mail
us at tech@hannainst.com or view our worldwide contact list at
www.hannainst.com.
Dear Customer,
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Page 3
3
TABLE OF CONTENTS
5
5
7
8
10
15
17
19
23
6
10
11
16
18
22
24
25
26
26
28
CHAPTER 1 - INTRODUCTION
Preliminary Examination .....................................................................................................
Model Identification ...........................................................................................................
General Description ............................................................................................................
Display and Keypad Description ...........................................................................................
CHAPTER 2 - QUICK START
Sensor and Probe Installation ..............................................................................................
Basic Operation .................................................................................................................
Help Function .....................................................................................................................
CHAPTER 3 - SPECIFICATIONS
System Specifications ..........................................................................................................
Probe Specifications ............................................................................................................
Sensor Specifications ...........................................................................................................
CHAPTER 4 - PROBE INSTALLATION
Sensor Descriptions .............................................................................................................
Sensor Preparation/Activation ..............................................................................................
Sensor Installation ..............................................................................................................
CHAPTER 5 - INITIALIZATION AND MEASUREMENT
Battery Installation .............................................................................................................
Meter Initialization .............................................................................................................
Measurement Mode ............................................................................................................
Setup Menu Structure .........................................................................................................
CHAPTER 6 - PARAMETER SETUP MENU
Select Parameters ...............................................................................................................
Parameter Units ..................................................................................................................
Parameter Coefficients .........................................................................................................
Page 4
4
29
31
35
41
30
34
38
42
46
50
47
52
54
45
46
49
50
53
56
57
TABLE OF CONTENTS
CHAPTER 7 - CALIBRATION MODE
Calibration Mode.................................................................................................................
Quick Calibration ................................................................................................................
pH Calibration ....................................................................................................................
Relative mV Calibration .......................................................................................................
Dissolved Oxygen Calibration (HI98194, HI98196 only) ........................................................
Conductivity Calibration (HI98194, HI98195 only) ................................................................
Temperature Calibration ......................................................................................................
Atmospheric Pressure Calibration ..........................................................................................
CHAPTER 8 - SYSTEM SETUP
Meter Setup .......................................................................................................................
Probe Setup .......................................................................................................................
CHAPTER 9 - STATUS
Meter Status ......................................................................................................................
Probe Status ......................................................................................................................
GLP Data ...........................................................................................................................
CHAPTER 10 - LOGGING MODE
Logging Mode ....................................................................................................................
Logging Menu Structure ......................................................................................................
Logging On Meter ...............................................................................................................
Log Recall ..........................................................................................................................
Log Notes ..........................................................................................................................
CHAPTER 11 - PC CONNECTION
PC Connection ....................................................................................................................
CHAPTER 12 - TROUBLESHOOTING / ERROR MESSAGES
Troubleshooting/Error Messages............................................................................................
APPENDIX
A ‑ PROBE MAINTENANCE....................................................................................................
B ‑ PROBE DEPLOYMENT ....................................................................................................
C ‑ ACCESSORIES ...............................................................................................................
42
59
61
Page 5
5
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer or the nearest HANNA Customer Service Center immediately.
METERS WITH PROBES ‑ packaged together in a sturdy carrying case with:
HI9298194 PC Application Software
HI7698290 calibration beaker
HI9828‑20 calibration solution (230 mL)
• USB cable
• 1.5V AA batteries (4 pcs.)
• probe maintenance kit
• appropriate probe shield
• specified sensors
• Instruction Manual and Quick Reference Guide
• Certificate
Note: Save all packing materials until you are sure that the instrument functions correctly. Any damaged or defective items must be returned in their original packing material with the supplied accessories.
Chapter 1 - INTRODUCTION
PRELIMINARY EXAMINATION
MODEL IDENTIFICATION
HI98194 pH/mV, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Dissolved Oxygen,
Atmospheric Pressure and Temperature Multiparameter meter with HI7698194 probe.
HI98195 pH/mV, ORP, EC, TDS, Resistivity, Salinity, Seawater σ, Atmospheric Pressure and
Temperature Multiparameter meter with HI7698195 probe.
HI98196 pH/mV, ORP, Dissolved Oxygen, Atmospheric Pressure and Temperature
Multiparameter meter with HI7698196 probe.
Page 6
6
GENERAL DESCRIPTION
HI9819X is a portable logging multiparameter system (instrument, probe) that monitors up to 14
different water quality parameters (7 measured, 7 calculated). The microprocessor‑based intelligent multisensor probes allows measurement of many water quality parameters such as pH, ORP, dissolved oxygen, conductivity and temperature with data logging. The system is easy to setup and easy to use. The HI9819X features a graphic, backlit display that automatically sizes the digits to fit the screen with on‑screen graphing capability. Each parameter is fully configurable.
HI9819X was designed to withstand harsh environments and is the ideal solution for field measure‑
ments of lakes, rivers and sea. The meter meets IP67 standards (30 minute immersion at a depth of 1 m) and the multisensor probe meets IP68 standards (continuous immersion in water).
Main features of the HI9819X systems:
• Rugged meter and probe
• Easy to use
• Measure up to 14 parameters and display of up to 12 parameters
• Waterproof protection (IP67 for the meter and IP68 for the probe)
• Graphic LCD with backlight
• Built‑in barometer for D.O. concentration compensation (HI98194, HI98196 only)
• Quick calibration feature
• Measurement check to eliminate any erroneous readings
• Auto recognition of probe and sensors
• Log‑on‑demand and automatic logging (up to 45,000 samples) on meter for all parameters
• Graphical display of logged data
• USB interface for PC communication
• Auto‑ranging for EC readings (HI98194, HI98195 only)
• Good Laboratory Practice feature, the last 5 calibrations are automatically stored
• Field‑replaceable sensors with color coded caps
• Meter is powered with alkaline batteries
Page 7
7
DISPLAY AND KEYBOARD DESCRIPTION
1. Graphic LCD
2. Battery level indicator
3. Softkeys
4. On/Off key: turn the meter on and off
5. Lamp key: turn the backlight on and off
6. Alphanumeric keyboard: insert alphanumeric codes
7. / Arrow keys: scroll the displayed options/message
8. HELP key: obtain information about the displayed screen
9. ESC key: return to the previous screen
10. Softkey functions defined on display
Page 8
8
Chapter 2 - QUICK START
• Sensor O‑Rings must be lubricated with the supplied grease prior to installation.
HI769819X probes have 2 or 3 sensor connectors sockets identified with color‑coded triangles:
• Connector 1 (red): For either pH/ORP, pH sensor
• Connector 2 (white): For dissolved oxygen sensor (HI98194, HI98196 only)
• Connector 3 (blue): For EC sensor (HI98194, HI98195 only)
• Position the connector key towards the center of the probe, make sure the connector is seated correctly (the sensor will no longer move freely) before tightening the locking threads.
• To protect the sensors, screw the protective shield onto the probe body.
• With the meter off, connect the probe to the instrument input on the top of the meter. Align the
pins and key then push the plug into the socket and tighten the thread.
• Turn the meter on by pressing the On/Off key. The meter will automatically recognize the probe
and the installed sensors and identify them on the probe status screen.
• Press Measure to view the measurement screen.
SENSOR AND PROBE INSTALLATION
Page 9
9
SENSOR AND PROBE INSTALLATION
Page 10
10
BASIC OPERATION
The main operating modes for HI9819X are measurement, logging and setup. The measurement screen can be configured to display a single measurement or up to 12 simultane‑ ous measurements by using the numbers 1‑7 on the keypad. Use the / keys to scroll through the measurements not being displayed. See page 24 for more details. The measurement units will blink if the system has not been calibrated and the measurement number will blink when the reading is out of range. Press Log to display the logging menu. You can either log a single sample on the meter or start an interval log on the meter. See Chapter 10 for more details. Press Menu to enter setup mode. You can configure which parameters you want to measure, calibrate the sensors, change system settings and view the meter and probe status.
HELP FUNCTION
HI9819X features context sensitive HELP, which provides useful information regarding the displayed
screen. Simply press the HELP key to access this function, then use the / keys to scroll through the message. To escape from the HELP window, press the HELP key again or ESC.
Page 11
11
Temperature
Range
‑5.00 to 55.00 °C;
23.00 to 131.00 °F;
268.15 to 328.15 K
Resolution 0.01 °C; 0.01 °F; 0.01 K Accuracy ± 0.15 °C; ± 0.27 °F; ±0.15 K Calibration Automatic at 1 custom point
pH/mV
Range 0.00 to 14.00 pH; ± 600.0 mV Resolution 0.01 pH; 0.1 mV Accuracy ± 0.02 pH; ± 0.5 mV
Calibration
Automatic at 1, 2 or 3 points with automatic recognition of 5 standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) and 1 custom buffer
ORP
Range ± 2000.0 mV Resolution 0.1 mV Accuracy ± 1.0 mV Calibration Automatic at 1 custom point (relative mV)
DISSOLVED OXYGEN (HI98194, HI98196 only)
Range 0.0 to 500.0 %; 0.00 to 50.00 ppm (mg/L) Resolution 0.1 %; 0.01 ppm (mg/L)
Accuracy
0.0 to 300.0 %: ± 1.5 % of reading or ± 1.0 % whichever is greater;
300.0 to 500.0 %: ± 3 % of reading
0.00 to 30.00 ppm (mg/L): ± 1.5 % of reading or ±0.10 ppm (mg/L)
whichever is greater;
30.00 ppm (mg/L) to 50.00 ppm (mg/L): ± 3 % of reading
Calibration Automatic 1 or 2 points at 0, 100 % or 1 custom point
Chapter 3 - SPECIFICATIONS
SYSTEM SPECIFICATIONS
Page 12
12
CONDUCTIVITY (HI98194, HI98195 only)
Range 0 to 200 mS/cm (absolute EC up to 400 mS/cm)
Resolution
Manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm
Automatic: 1 µS/cm from 0 to 9999 µS/cm;
0.01 mS/cm from 10.00 to 99.99 mS/cm;
0.1 mS/cm from 100.0 to 400.0 mS/cm
Automatic (mS/cm): 0.001 mS/cm from 0.000 to 9.999 mS/cm;
0.01 mS/cm from 10.00 to 99.99 mS/cm;
0.1 mS/cm from 100.0 to 400.0 mS/cm
Accuracy ±1 % of reading or ±1 µS/cm whichever is greater
Calibration
Automatic single point, with 6 standard solutions (84 µS/cm, 1413 µS/cm,
5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
RESISTIVITY (HI98194, HI98195 only)
Range 0 to 999999 Ω·cm; 0 to 1000.0 kΩ·cm; 0 to 1.0000 MΩ·cm Resolution Depending on resistivity reading Calibration Based on conductivity or salinity calibration
TDS (Total Dissolved Solids) (HI98194, HI98195 only)
Range 0 to 400000 ppm (mg/L); (the maximum value depends on the TDS factor)
Resolution
Manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L);
1 ppt (g/L)
Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L);
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L);
Automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L);
0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
Accuracy ±1 % of reading or ±1 ppm (mg/L) whichever is greater Calibration Based on conductivity or salinity calibration
SYSTEM SPECIFICATIONS
Page 13
13
SALINITY (HI98194, HI98195 only)
Range 0.00 to 70.00 PSU Resolution 0.01 PSU Accuracy ±2% of reading or ±0.01 PSU whichever is greater Calibration Based on conductivity calibration
Note: For HI98196 Salinity can be set from 0.00 to 70.00 PSU in setup menu.
SEAWATER SIGMA (HI98194, HI98195 only)
Range 0.0 to 50.0 σ
t, σ0, σ15
Resolution 0.1 σ
t, σ0, σ15
Accuracy ±1.0 σ
t, σ0, σ15
Calibration Based on conductivity or salinity calibration
ATMOSPHERIC PRESSURE
Range
450.0 to 850.0 mmHg; 17.72 to 33.46 inHg;
600.0 to 1133.2 mbar; 8.702 to 16.436 psi;
0.5921 to 1.1184 atm;
60.00 to 113.32 kPa Resolution 0.1 mmHg; 0.01 in Hg; 0.1 mbar 0.001 psi; 0.0001 atm; 0.01 kPa Accuracy ±3.0 mmHg within ±15°C from calibration temperature
Calibration Automatic at 1 custom point
SYSTEM SPECIFICATIONS
Page 14
14
METER SPECIFICATIONS
Temperature Compensation Automatic from ‑5 to 55 °C (23 to 131 °F)
Logging Memory
45,000 records (continuous logging or log‑on‑demand of all parameters)
Logging Interval 1 second to 3 hours PC Interface USB (with HI9298194 software) Waterproof Protection IP67
Environment 0 to 50 °C (32 to 122 °F); RH 100 %
Battery Type 1.5V, AA alkaline batteries (4 pcs.)
Battery Life
360 hours of continuous use without backlight / 50 hours with backlight
Dimensions/Weight 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”) / 400 g (14.2 oz.)
METER BATTERY LIFE
The power consumption of the HI9819X multiparameter systems are dependent on two things:
1. The measurement system configuration (sensor configuration)
2. The meter configuration (logging interval and backlight use) The following table estimates the meter’s battery life connected to a HI769819X probe
with backlight off.
Note: Backlighting use consume the most power. The table variables, battery selection and parameter selection.
pH/ ORP, D.O., EC
Backlight OFF, 1 s log 280 hours Backlight OFF, 4 min log 360 hours Backlight OFF, 10 min log 400 hours Backlight ON, 4 min log 50 minutes Backlight ON, 10 min log 50 minutes
SYSTEM SPECIFICATIONS
Page 15
15
Sensor Inputs
3 for HI7698194 2 for HI7698195 and HI7698196
Sample Environment Fresh, brackish, seawater Waterproof protection IP68 Operating Temperature ‑5 to 55 °C Storage Temperature ‑20 to 70 °C Maximum Depth 20 m (66’)
Dimensions (without cable)
342 mm (13.5”), dia=46 mm (1.8”)
Weight (with batteries and sensors)
570 g (20.1 oz.)
Cable Specification
Multistrand‑multiconductor shielded cable with internal strength member rated for 68 kg (150 lb) intermittent use
Wetted Materials
Body: ABS Threads: Nylon Shield: ABS/316 SS Temp probe: 316 SS O‑Rings: EPDM
PROBE SPECIFICATIONS
Page 16
16
HI7698194‑0 HI7698194‑1 HI7698194‑2 HI7698194‑3
Description pH pH/ORP Dissolved Oxygen EC Measure Type
Primary Unit
pH, mV (pH) pH, mV (pH/ORP)
D.O. (% sat. & conc.)
EC
Measure Range
0.00 to 13.00 pH ±600.0 mV
0.00 to 13.00 pH ±600.0 mV ±2000.0 mV
0.0 to 500.0 %
0.00 to 50.00 mg/L
0.0 to 200.0 mS/cm
0.0 to 400 mS/cm (absolute)
Temperature Range
‑5 to 55°C ‑5 to 55°C ‑5 to 55°C ‑5 to 55°C
Color Code Red Red White Blue
Materials
Tip: glass (pH) Junction: ceramic Body: PEI Electrolyte: gel Reference: double
Tip: glass (pH); Pt (ORP) Junction: ceramic Body: PEI Electrolyte: gel Reference: double
Cat/An: Ag/Zn Membrane: HDPE Body: white top ABS CAP
Stainless steel electrodes AISI 316 Body: ABS/EPOXY
Maintenance Solution
HI70300
(storage solution)
HI70300
(storage solution)
HI7042S
(D.O. electrolyte)
none
Dimensions 118 x 15 mm 118 x 15 mm 99 x 17 mm 111 x 17 mm Depth 20 m (65’) 20 m (65’) 20 m (65’) 20 m (65’)
SENSOR SPECIFICATIONS
Page 17
17
Chapter 4 - PROBE INSTALLATION
SENSOR DESCRIPTIONS
HI7698194‑0 Combination pH sensor features a glass pH sensitive
bulb and a silver/silver chloride double junction reference with gelled electrolyte.
HI7698194‑1 Combination pH/ORP sensor features a glass sensitive
bulb for pH readings, a platinum sensor for redox measurements and a silver/silver chloride double junction reference with gelled electrolyte.
Note: See page 18 for pH preparation. See page 18 for ORP activation.
HI7698194‑2 Galvanic dissolved oxygen (D.O.) sensor. The thin gas
permeable membrane isolates the sensor elements from the testing solution but allows oxygen to pass through. The oxygen that passes through the membrane is reduced at the cathode and causes a current, from which the oxygen concentration is determined. The D.O. sensor conforms to Standard Methods 4500‑AG, EPA 360.1.
Note: The D.O. sensor needs to be activated before installation. See page 18 for details.
HI7698194‑3 four ring‑electrode conductivity (EC/TDS/Resistivity/Sa‑
linity) sensor. The sensor is immune to polarization or surface coatings.
Page 18
18
SENSOR PREPARATION / ACTIVATION
pH Preparation
Remove the shipping cap from the pH sensor. If the shipping cap does not contain any liquid, pour
HI70300 into shipping cap, place it back on the sensor and soak for at least 1/2 hour before use. If HI70300 is not available, pH 4.01 buffer may be substituted.
ORP Activation
For improved redox measurements, the surface of the sensor must be clean and smooth. A pretreatment procedure should be performed to ensure quick response. The pretreatment of the sensor is determined by the pH and the ORP potential values of the sample. Use the table below to determine the treatment required. First locate the typical sample pH. If the corresponding ORP value (mV) is higher than the values in the table below, an oxidizing pretreatment is necessary. If the value is lower, a reducing pretreatment is necessary.
pH mV pH mV pH mV pH mV pH mV
0 990 1 920 2 860 3 800 4 740 5 680 6 640 7 580 8 520 9 460
10 400 11 340 12 280 13 220 14 160
For reducing pretreatment: immerse the electrode for at least five minutes in HI7091. For oxidizing pretreatment: immerse the electrode for at least five minutes in HI7092.
D.O. Sensor Activation
The D.O. probe is shipped dry. To prepare the sensor for use:
• Remove the black & red plastic cap. This cap is used for shipping purposes only and can be thrown away.
• Insert the supplied O‑Ring in to the membrane cap.
• Rinse the membrane with some electrolyte solution. Refill with clean electrolyte.
Gently tap the membrane cap to dislodge air bubbles. To avoid damaging the membrane, do not
touch it with your fingers or directly tap the membrane.
• With the sensor facing down screw the membrane cap counterclockwise to the end of the threads. Some electrolyte will overflow.
• Rinse outside of sensor with deionized water.
• Invert sensor and inspect. There should be no bubbles or debris between the membrane and sensor body.
Page 19
19
EC Sensor Preparation
The EC sensor does not need to be soaked or hydrated before use. Use the small brush included in the probe maintenance kit to clean and loosen any debris before using.
SENSOR INSTALLATION
The HI7698194 can support three different sensors: Connector 1: pH, pH/ORP Connector 2: D.O. Connector 3: EC.
To make installation easier, the sensors have color‑coded caps and the sockets are identified with colored triangles, corresponding to the colors of the sensors (pH ‑ red; EC ‑ blue; D.O. ‑ white).
Page 20
20
The HI7698195 support two different sensors:
Connector 1: pH, pH/ORP Connector 2: EC
The HI7698196 support two different sensors:
Connector 1: pH, pH/ORP Connector 2: D.O.
SENSOR INSTALLATION
Page 21
21
For a correct installation:
• Grease the sensor O‑Ring with the lubricant found in the probe maintenance kit.
DO NOT SUBSTITUTE other grease/lubricants as it may cause the O‑Ring to swell.
• Insert the sensor into the correctly color coded opening while positioning the connector key toward the center of the probe. Make sure the connector is seated correctly (the sensor will no longer move freely) before tightening the locking threads with your fingers.
• Continue to tighten the locking threads with the tool supplied in the maintenance kit until the sensor is secured tightly against the probe body.
• To protect the sensors, screw the protective shield onto the probe body.
• With the meter off, connect the probe to the DIN socket on the bottom of the meter. Align the pins and key then push the plug into the socket. Secure the probe by attaching the carabinier of the probe to the meter piston.
• Turn on the meter by pressing the On/Off key. The meter should automatically recognize the installed sensors and identify them on the probe status screen. If you have an error message or the sensor is not recognized, reconnect the sensor(s) or probe and try again.
SENSOR INSTALLATION
Page 22
22
Chapter 5 - INITIALIZATION AND MEASUREMENT
HI9819X are supplied with 4 alkaline, size AA batteries.
The battery symbol on the LCD indicates the remaining battery life. The meter has a low battery warning, and when the symbol starts blinking, batteries should be replaced with new ones. When the batteries are discharged the meter will automatically shut off to avoid erroneous readings.
Replace batteries in safe areas only. Remove the 4 screws on the rear of the instrument and insert the batteries observing polarity.
Note: Do not mix old and new alkaline batteries.
BATTERY INSTALLATION
Page 23
23
METER INITIALIZATION
After connecting the desired sensors to the probe and connecting the probe to the meter (see previous chapter), turn the meter on by pressing On/Off key. After the initialization has been completed if the probe is connected, the meter displays the PROBE STATUS SCREEN. The probe status screen identifies the probe and attached sensors. If an incompatible probe was connected the following message appears and the probe has to be replaced. The compatible meter‑probe pairs are: HI98194 ‑ HI7698194; HI98195 ‑ HI7698195;
HI98196 ‑ HI7698196.
If an incompatible sensor was connected to the probe, “Incompat.sensor!” will be displayed according to the respective connector. If the sensor was placed in a wrong connector, “Wrong input” will be displayed for that connector.
Page 24
24
MEASUREMENT MODE
Two active soft keys are found at the bottom of the status screen.
• Press Measure to access the measurement mode.
• Press Param to access the “Select Parameter” menu.
(This screen can also be accessed from the main menu, see Chapter 6 for a detailed description.).
• Press the DOWN arrow to view additional information about the probe.
Measurement mode is one of the three main operating modes of HI9819X (along with logging mode and setup mode). During measurement mode HI9819X will simultaneously measure data for all enabled parameters.
• Use the numbers on the keyboard to select the number of parameters that are shown on the screen at one time. The display will automatically resize the font.
• Press the / keys to scroll through the enabled parameters if they do not fit on one screen.
Notes: A flashing measurement value indicates that the measurement is out of range. A flashing measurement unit indicates that the user calibration has not been done and is needed for accurate readings.
• Press Log to enter the log menu. See Chapter 11 for details.
• Press Menu to enter the main setup menu. The main menu accesses the parameter setup, calibration, system setup and status options. See the following chapters for details.
Page 25
25
SETUP MENU STRUCTURE
Page 26
26
Chapter 6 - PARAMETER SETUP MENU
From the main menu, use the / keys to highlight Parameter Setup and then press Select.
The following options will be displayed: Warning: Logged data saved on this meter will be changed to selected parameter units or coefficients.
Use the / keys to scroll through the menu. Press the right softkey to enable or disable a single parameter, or the left softkey to enable or disable all parameters. A checked box means that the parameter is enabled. Only the available parameters are present in the list.
Note: If the password protection is enabled, you will be required to enter the password before any parameters can be modified.
Temperature Unit
The user can select the measurement unit: °C, °F or K. The default value is °C.
TDS Unit (HI98194, HI98195 only)
The user can select ppm ‑ ppt or mg/L ‑ g/L measurement unit. The default value is ppm - ppt.
D.O. Concentration Unit (HI98194, HI98196 only)
The user can select ppm or mg/L. Dissolved Oxygen concentration is calculated using % saturation, conductivity (HI98194) or salinity constant (HI98196) and atmospheric pressure. The default value is
ppm.
Pressure Unit
The user can select one the following measurement units: psi, mmHg, inHg, mbar, atm, kPa. The default value is psi.
SELECT PARAMETERS
PARAMETER UNITS
Page 27
27
PARAMETER UNITS
Resistivity Unit (HI98194, HI98195 only)
The user can select resistivity from one of the following measurement
units: Ω·cm, kΩ·cm or MΩ·cm. Resistivity is calculated from the
conductivity measurement. The default unit is MΩ·cm.
Seawater Sigma Unit (HI98194, HI98195 only)
This parameter is used for seawater analysis. It is calculated from the conductivity measurement and depends on water pressure, temperature and salinity. The default value is σt. Users can select the reference temperature: σt, σ
0,
and σ15 (i.e. current temperature,
0 °C or 15 °C).
EC Resolution Unit (HI98194, HI98195 only)
The user can configure the conductivity resolution with one of the following options: Auto: the meter automatically chooses the range to optimize the measurement. Readings can be in µS/cm or mS/cm. Auto mS/cm: the meter automatically chooses the range to optimize the measurement, readings will be in mS/cm only. 1μS/cm, 0.001 mS/cm, 0.01mS/cm, 0.1mS/cm or 1mS/cm: the meter will not autorange, the measurement will be displayed with the selected resolution. The default value is Auto.
Absolute EC Resolution Unit (HI98194, HI98195 only)
Absolute conductivity displays the conductivity without temperature compensation. See EC resolution for resolution details.
Note: A small letter “A” added to the μS/cm or mS/cm unit refers to an absolute conductivity value (i.e. a conductivity reading with no temperature compensation).
TDS Resolution Unit (HI98194, HI98195 only)
The user can configure the TDS resolution with one of the following options: Auto: the meter automatically chooses the range to optimize the measurement, readings can be in ppt or ppm. Auto ppt: the meter automatically chooses the range to optimize the measurement, readings will be in ppt only. 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt or 1 ppt: the meter will display the measurement with selected resolution. The default value is Auto.
Page 28
28
PARAMETER COEFFICIENTS
EC Reference Temperature (HI98194, HI98195 only)
This value is used for temperature compensated conductivity. All EC measurements will be referenced to the conductivity of a sample at this temperature. Press the softkey to select the desired option; 20 °C or at 25 °C. The default value is 25 °C.
EC Temperature Coefficient (HI98194, HI98195 only)
The temperature coefficient Beta (ß) is defined by the following equation (using 25 °C as an example): EC25=ECx /(1+ß(Tx‑25)) Beta is a function of the solution being measured. For freshwater samples Beta is approximately 1.90%/°C. If the actual temperature coefficient of your sample is known, press Modify to enter the value. To confirm press Accept. The value can be within 0.00 and 6.00%/°C. The default value is
1.90%/°C.
TDS Factor (HI98194, HI98195 only)
TDS stands for total dissolved solids, and it is a calculated value based on the conductivity of the solution (TDS = factor x EC25). The TDS conversion factor can be set from 0.00 to 1.00. A typical TDS factor for strong ionic solutions is 0.50, while for weak ionic solutions (e.g. fertilizers) is 0.70. Press Modify to enter the value, press Accept to confirm. The default
value is 0.50.
Salinity (HI98196 only)
This value is used to calculate the D.O. concentration where the EC sensor is not available for the meter. To set the value, press Modify and enter the desired value between 00.00 to 70.00 PSU. Press Accept to save the modification.
Averaging
Averaging is a software filter to minimize sensor noise and provide more stable readings. Averaging is particularly useful to get a rep‑ resentative reading of the “average” value from flowing water. Averaging will affect all measurements.
Page 29
29
This value should be kept low if you want a fast response. Press Modify to select the desired number of samples to average. This value can be set from 1 to 20 samples. The default value is 1.
Note: Each reading takes 1 second, so when logging the first sample will be delayed by a few seconds if averaging is used.
HI9819X’s calibration routines are accessed by highlighting
“Calibration” and pressing Select from the main menu. Calibration is the process that standardizes the electrical signal from the sensors to reagent standards of known value. Calibrations are intuitive and menu driven. All calibration data is stored in the non volatile probe memory, allowing probes to be connected to different meters without recalibration. There are two types of calibrations available: the “Quick calibration”, which is used for a single point calibration of pH, conductivity, and/or dissolved oxygen and is handy for field work; and the Single param. calibration that allows each parameter to be calibrated individually. The user may also restore each parameter to a factory default calibration.
Note: The password will be required if password protection is enabled.
To optimize measurements, it is advisable to establish the optimum calibration period required for the measurement environment. Calibration requirements vary with deployment conditions, for example very turbid biologically‑active waters may require more frequent cleanings and calibrations than cleaner waters. General calibration guidelines are listed below:
• Set up a routine service schedule where measurement integrity is validated. This is especially important for new installation sites or long deployments.
• Inspect sensor connectors for corrosion and replace damaged sensors.
• Inspect sensor O‑Rings for damage and if necessary replace and lubricate with the grease found in the probe maintenance kit.
• Do not handle the sensing surfaces of the sensors.
• Avoid rough handling and abrasive environments that can scratch the reactive surfaces of the sensors.
• Avoid long‑term exposure of sensors to bright sunlight. If possible, calibrate in a shaded area.
• Discard standards after use. Do not return the used standards to the bottles of “fresh” solution.
• For measurements across a temperature gradient (when water temperature is drastically dif‑ ferent from the standards), permit the sensors to reach thermal equilibrium before conducting calibrations or making measurements. The heat capacity of the probe is much greater than the air and the small beakers of calibration standards.
Chapter 7 - CALIBRATION MODE
CALIBRATION MODE
Page 30
30
The quick calibration method provides a quick single point calibration for pH, conductivity and dissolved oxygen sensors. HI9828‑0 calibration solution is used for both pH and conductivity.
• Fill the calibration beaker 2/3 full with HI9828‑0 calibration solution.
• Slowly place the sensors into the solution and dislodge bubbles that may adhere to the sensors.
• Screw the calibration beaker completely on the probe body. Some
solution may overflow.
• Wait a few minutes for the system to stabilize.
• From the “Calibration” menu select “Quick calibration”.
• A three item calibration menu will appear (pH, conductivity and
dissolved oxygen) and “pH” will start to blink along with the “Not ready” message.
• When the pH signal is stable, the “Ready” message appears. Press
Confirm to store the calibration data.
• The “Storing” message will appear as the calibration proceeds to
the next sensor. A checkmark will appear in the box next to “pH” to indicate a successful calibration.
Notes: To bypass any of the calibrations press Skip to move to the next sensor in the quick calibration menu. If the pH sensor is not installed the message “pH sensor not installed! Skip to conductivity calibration” will appear.
• Following the pH calibration, “Conductivity” will start to blink
along with the “Not ready” message.
• When the measurement is stable, “Ready” appears. Press
Confirm to store the calibration data and the “Storing” message will appear.
Note: If EC calibration is not required, skip to the D.O. quick calibration by pressing the Skip
softkey.
• The message “Empty the beaker.” will appear.
• Unscrew the calibration beaker and empty the solution.
• Shake any remaining liquid off the probe and beaker. No droplets should remain on the D.O.
sensor membrane.
QUCIK CALIBRATION
Page 31
31
QUCIK CALIBRATION
Note: Do not attempt to dry wipe the D.O. sensor as damage to the membrane may occur.
• Screw the empty calibration beaker on the probe body. The beaker should not be dry.
• Press Accept to close the displayed message.
• When the measurement is stable, “Ready” appears. Press
Confirm to store the calibration data and the “Storing” message will appear.
• Press OK to return to “Calibration” menu.
Note: To quit the quick calibration procedure, press ESC at any time. After every calibration the quick calibration window will show a check mark in the box next to the calibrated parameter.
To optimize the pH measurement follow the general guidelines mentioned in the Chapter 7 introduction. From the “Calibration” menu select “Single param. calibration” and then “pH calibration”. The display shows two options: “Calibrate pH” and “Restore factory calib.”. If a new pH sensor has been installed use “Restore factory calib.” before performing a user calibration as some warning messages are based on changes from previous calibrations. If “Restore factory calib.” is selected, all user calibration data will be deleted and the default calibration is restored. A user calibration should follow immediately. If “Calibrate pH” is selected, the user can perform a new calibration using up to 3 buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01 or one custom buffer). When a 3‑point calibration is performed, all old data are overwritten, while with a single or 2‑point calibration the meter will also use information from the previous calibration, if exists.
Preparation
Pour small quantities of the selected buffer solutions into clean beakers. To minimize cross‑contamination, use two beakers for each buffer solution: the first one for rinsing the sensor and the second one for calibration.
pH CALIBRATION
Page 32
32
Procedure
The measured pH value is displayed, along with the temperature and the buffer value on the second level. If necessary, press the Cal point softkey and use the / keys to select the correct buffer.
• Immerse the sensors in the first buffer rinse solution and stir gently.
• Immerse the pH sensor and temperature probe into the selected buffer and stir gently. The temperature, pH buffer value and the “Not ready” message are displayed.
• Once the reading has stabilized the countdown timer will count down until the display shows the “Ready” message.
• Press Confirm to accept the calibration point.
• After the calibration point is confirmed, to avoid
cross‑contamination immerse the sensors in the next
calibration buffer rinse solution and stir gently.
• Press Cal Point to select the next buffer (if necessary), and repeat the calibration procedure outlined above with the second and third buffers.
Note: The calibration procedure can be terminated after a single or 2 point calibration by pressing ESC. The message “Storing” followed by “Calibration completed” will be displayed.
• Press OK to return to the Calibration menu.
• Press Measure to return to the measurement screen.
Custom buffer calibration
• The HI9819X permits a single custom buffer to be used for pH calibration. This can be used along with standard buffers as part of a 2 or 3 point calibration or as a single point.
• To select this option first press Cal. point and then Custom while the meter is waiting for stable reading.
• A text box window will appear. Use the keypad to enter the value of the buffer at the current temperature. The valid range for custom a buffer is from 0.00 to 14.00 pH.
pH CALIBRATION
Page 33
33
pH Calibration Error Messages
The HI9819X displays a series of messages if an error has occurred during calibration. If the meter does not accept a pH calibration point, a short message is displayed to indicate the possible error source. The following screens are examples:
These are the available messages:
• “Input out of scale”: the pH value is out of range. The pH sensor may require replacement.
• “Check sensor”: the electrode may be broken, very dirty or the user has attempted to calibrate the
same buffer value twice.
• “Wrong buffer”: the displayed pH reading is too far from the selected buffer value. This is often
seen immediately after a buffer calibration has been completed but before the pH sensor has been moved to the next buffer. Check if the correct calibration buffer has been selected.
• “Invalid temperature”: the buffer temperature is outside the acceptable range.
• “Wrong buffer” / “Contaminated buffer” / “Check electrode”: the buffer is contaminated or the
sensor is broken or very dirty.
• “Check sensor” / “Clean sensor”: the electrode is broken or very dirty.
• “Wrong” / “Clear old calibration”: erroneous slope condition. These messages appear if the
slope difference between the current and previous calibration exceeds the slope window (80% to 110%). Press the Clear softkey to cancel the old data and continue the calibration procedure, or press ESC to quit the pH calibration mode.
pH CALIBRATION
Page 34
34
RELATIVE mV CALIBRATION
The “ORP calibration” allows the user to perform a single point custom calibration (relative mV) or to restore the factory calibration.
The Oxidation‑Reduction Potential (ORP), displayed in mV, is the voltage that results from the difference in potential between the platinum ORP sensor and the silver/silver chloride reference electrode. ORP values are not temperature compensated, although ORP values can change with temperature (e.g. reference electrode potential changes, sample equilibrium changes). It is important to report ORP values together with the reference electrode used and the temperature.
The inert platinum ORP surface provides an electron exchange site with the sample (or standard) and its surface. The electron exchange is typically very fast in well‑poised solutions (standards for example), but may be more lengthy in natural water samples.
Calibration is typically not required for a new ORP sensor, but the process does establish a baseline that can be used as a comparison for future validations.
Calibration is used to compensate for changes due to contamination of the platinum surface and drift in the reference electrode.
A relative mV calibration can also be made to remove the voltage attributable to the Ag/AgCl reference electrode (to display the ORP versus a SHE (standard hydrogen electrode)). This is really an arithmetic correction and is correct only at the standard temperature. For example, HI7022 reads 470 mV at 25 °C versus the Ag/AgCl reference. The ORP mV versus a SHE would be 675 mV. (add 205 mV to the observed value).
Preparation
APPENDIX C – “Accessories” lists Hanna solutions used for ORP calibrations. The calibration should be conducted at temperatures between 20‑26 °C. The sensor should be clean and oil free.
Page 35
35
Procedure
• From the “Calibration” menu select “Single param. calibration” and then “ORP calibration”. The display shows two options: “Custom ORP” and “Restore factory calib.”.
• For a user calibration select “Custom ORP”.
• Fill a beaker with an ORP test solution (see APPENDIX C “Accessories”).
• Using the keypad, insert the numerical ORP value and then press Accept to confirm.
• The stability counter will count down and the message “Ready” and Confirm will be displayed.
• Press Confirm to accept the calibration point.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press OK to return to the Calibration menu.
• Press Measure to return to the measurement screen.
• To restore the factory calibration data, select the corresponding option in the “ORP calibration” menu and then press Select.
DISSOLVED OXYGEN CALIBRATION (HI98194, HI98196 only)
The accuracy of dissolved oxygen measurements is directly related to membrane cleanliness and calibration technique. Oily coating and biological contaminants are the primary cause of calibration drift in dissolved oxygen sensors. Unfortunately, brushes or other cleaning objects may damage the membrane. Replacing the membrane cap and electrolyte is the best way to perform periodic maintenance. Although it may be easier to calibrate the D.O. sensor prior to deployment, it is advised to calibrate at the site of deployment. Errors in measurement may result if altitude and barometric pressure differ between the calibration and measurement site.
Note: Perform either the % D.O. saturation or D.O. concentration calibration.
If the % D.O. saturation range is calibrated, the D.O. concentration range will also be calibrated, and vice versa. Dissolved oxygen concentration values are based on % D.O. saturation, temperature, salinity and atmospheric pres‑ sure. A standard solution or a reference D.O. meter may be used to compare readings during calibration.
RELATIVE mV CALIBRATION
D.O. CALIBRATION (HI98194, HI98196 only)
Page 36
36
The calibration of the D.O. concentration range can only be performed at a single custom point (4 to 50 mg/L). It is recommended to calibrate the D.O. sensor close to the values that will be measured. Choose “DO calibration” from the “Calibration” menu, select the D.O. calibration type using the / keys and press Select to confirm.
% D.O. saturation
The calibration of the % D.O. saturation range can be performed at a single or 2 standard points (0 % and 100 %), or at a single custom point (50 % to 500 %).
Procedure:
• To calibrate at 100 %, fill the calibration beaker with approxi‑ mately 4 mm (5/32") of water and screw it onto the probe. The membrane should not be wet. This condition corresponds to air 100 % saturated with oxygen and water vapor.
• The reading, temperature, calibration point and the “Not ready” message are displayed.
• Once the reading has stabilized the countdown timer will count down until the display shows the “Ready” message.
• Press Confirm to accept the calibration point. After confirmation, put the D.O. and temperature sensors into
HI7040 zero oxygen solution and wait for stability to be reached. The stability timer will count
down and Confirm will appear. Press Confirm to store the calibration.
• The following messages will appear: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press Measure to return to the measurement screen.
Notes: The user can perform a single point calibration by pressing ESC after the first point is accepted. If the D.O. input is not within the acceptable range, the message “Invalid input” is dis­played.
D.O. CALIBRATION (HI98194, HI98196 only)
Page 37
37
D.O. CALIBRATION (HI98194, HI98196 only)
Single point Custom % saturation calibration
• For a calibration at another known value place sensor and temperature probe into the known solu‑ tion and change the calibration value, press the Cal. point softkey and select the desired point.
• To insert a different calibration value, press Cal. point and then Custom. Insert the desired value using the keypad, then press Accept.
• When the reading is stable, the “Ready” message is displayed. Press Confirm to store the calibration point.
• The following messages will appear: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press Measure to return to the measurement screen.
D.O. concentration (HI98194, HI98196 only)
Verify the barometric pressure, conductivity and temperature reading are correct. Calibrate them if necessary. To calibrate the D.O. concentration range, a solution with known dissolved oxygen concentration value is needed. The solutions used to calibrate with, should be determined independently (for instance by Winkler titration). Place the D.O. sensor with temperature sensor into the known solution.
• From the “DO calibration” menu, select the “DO concentration” option, insert the known
concentration. Allow the sensors to reach thermal equilibrium with the solution. Stir or agitate if possible to keep fresh solution in front of the membrane and press OK.
• When the reading is stable, the stability timer will count down and Confirm will appear. Press
Confirm to accept the value.
• When the messages “Storing” and “Calibration completed” appear, the calibration is completed.
To return to the “Calibration” menu, press OK.
• To return to the main menu, press ESC twice.
Page 38
38
CONDUCTIVITY CALIBRATION (HI98194, HI98195 only)
A conductivity calibration is used to adjust for variations in cell factors by using a standard solution of known conductivity. Oily coating and biological contaminants are the primary cause of calibration drift in conductivity sensors. This type of fouling changes the apparent cell geometry, resulting in a shift in cell constant. Before performing a conductivity calibration inspect the EC sensor for debris or blockages. The EC electrodes are situated inside the two small channels found in the bottom of the conductivity sensor. Clean using the small brush from the probe maintenance kit. Flush with water. A mild detergent may be used to remove oily coatings. Always flush with clean water after cleaning.
Note: For a correct conductivity calibration, the probe shield or the calibration beaker must be used.
The conductivity calibration menu includes 3 different types of calibration: Conductivity, Absolute conductivity and Salinity. The “Conductivity” option allows a single point calibra‑ tion with a standard solution selectable by the user. This calibration is temperature compensated. The “Absolute conductivity” option allows a single point calibration with a conductivity solution of known non‑temperature compensated value at the current temperature. The “Salinity” option allows calibration with a standard salinity solution. The 3 calibrations are related, so that each one will calibrate all 3 measurements.
Note: To improve accuracy, choose a calibration standard near the sample conductivity.
Choose “Conductivity calibration” from the “Calibration” menu. Select the calibration type using the / keys and press Select to confirm.
Conductivity
• Select the “Conductivity” option and press Select to confirm.
• Fill the calibration beaker with a conductivity standard (see APPENDIX C ‑ “Accessories” for choos‑ ing the proper HANNA standard solution).
• Pour additional standard into a second beaker to be used to rinse the sensor.
Page 39
39
• Immerse the sensor into the rinse standard by raising and lowering the beaker a few times to ensure that the EC sensor channels are filled with fresh standard.
• Place the calibration beaker over the EC sensor and dislodge any trapped bubbles. Screw the beaker into place. Wait for the reading to stabilize.
• The main display shows the actual reading, while the secondary level displays the current temperature and the standard value.
• To change the standard value, press Cal. point and the list of available standard values is displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and
111.8 mS/cm.
• The third level displays the status message.
• Press Custom to insert a custom value (temperature compensated value). Insert the desired value using the keypad, then press Accept.
• When the reading becomes stable, the stability timer will count down and Confirm will appear. Press Confirm to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to main menu.
• Press Measure to return to the measurement screen.
Absolute Conductivity
• Select “Absolute conductivity” from the “Conductivity calibration” menu.
• Use the keypad to enter the custom value with the desired resolution. Press Accept to confirm.
• Fill the calibration beaker with conductivity standard with known conductivity at the temperature of standardization.
• Pour additional standard into a second beaker to be used to rinse the sensor.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to ensure that the EC sensor channels are filled with fresh standard.
CONDUCTIVITY CALIBRATION (HI98194, HI98195 only)
Page 40
40
• Place the calibration beaker over the EC sensor and dislodge any trapped bubbles. Screw the beaker into place. Wait for the reading to stabilize. The stability timer will count down and Confirm will appear.
• Note the temperature and adjust the conductivity value if needed.
• Press Confirm to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press Measure to return to the measurement screen.
Salinity
The measurement of salinity is based on the Practical Salinity Scale which uses the EC measurement. If the user has a standard with known PSU value it may be used to calibrate the conductivity sensor.
• Select “Salinity” from the “Conductivity calibration” menu.
• Use the keypad to enter the known salinity value of the calibra‑ tion solution. Press Accept to confirm.
• Fill the calibration beaker with salinity standard of known value.
• Pour additional standard into a second beaker to be used to rinse the sensor.
• Immerse the sensor into the rinse beaker and raise and lower the beaker to ensure that the EC sensor channels are filled with fresh standard.
• Place the calibration beaker with standard over the EC sensor and dislodge any trapped gas bubbles. Screw the beaker into place.
• Wait for the reading to stabilize. The stability timer will count down and Confirm will appear.
• Note the temperature and adjust the salinity value if needed.
• Press Confirm to save the calibration.
• After confirmation, the following messages are displayed: “Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press ESC twice to return to the main menu.
• Press Measure to return to the measurement screen.
CONDUCTIVITY CALIBRATION (HI98194, HI98195 only)
Page 41
41
Notes: These procedures calibrate the slope value. To calibrate the offset, set the calibration point at 0 μS/cm and repeat the procedure. If the temperature input is not within the acceptable range (0 to 50 °C), the message “Invalid temperature” is displayed. If the conductivity input is not within the acceptable range, the message “Wrong standard” is displayed.
The probe is factory calibrated for temperature readings. The user can perform a single point tem‑ perature calibration or restore factory calibration. This procedure requires a reference temperature measuring instrument.
• Select “Temperature” from the “Calibration” menu.
• Select “Calibrate temperature”.
• Insert the probe in an isothermal bath with reference instrument and allow the probe to come to
thermal equilibrium.
• Use the keypad to enter the known temperature and then press Accept to confirm.
• The stability timer will count down and the message “Ready”
and “Confirm” will be displayed.
• Press Confirm to store the calibration point.
• After confirmation, the following messages are displayed
“Storing” and “Calibration completed”.
• Press OK to return to the “Calibration” menu.
• Press Measure to return to the measurement screen.
• To restore the factory calibration, select the corresponding
option in the “Temperature calib.” menu and then press
Select.
TEMPERATURE CALIBRATION
Page 42
42
Place HI9819X in a wind‑free area and choose “Custom pressure” to perform a user calibration or “Restore factory calib”.
Note: “Custom pressure” procedure requires a reference barometer.
Select the “Atm. pressure” from the “Calibration” menu.
• Select the “Custom pressure” option.
• Using the keypad, insert the numeric value that agrees with
the reference meter and then press Accept to confirm.
• The stability counter will count down and the message
“Ready” and “Confirm” will be displayed. Press Confirm to store the calibration point.
• After confirmation, the following messages are displayed:
“Storing” and “Calibration completed”.
• Press Measure to return to the measurement screen.
• Press OK to return to the “Calibration” menu.
• To restore the factory calibration, select “Restore factory calib.” in the “Pressure calibration” menu
and press Select.
ATMOSPHERIC PRESSURE CALIBRATION
Chapter 8 - SYSTEM SETUP
From the main menu, select “System setup” and then “Meter setup” or “Probe setup”.
Note: If the password protection is enabled, you will be required to enter the password before any settings can be modified.
Time
The meter uses a real time clock for logging. The time and time format are set in this function. Press Modify and set the time using the keypad. Press Accept to save the time. When using the 12 hour format, press A or P on the keypad for AM or PM after you set the time. Press Format to change between 12 and 24 hour formats.
The default format is 24 hours.
METER SETUP
Page 43
43
Date
The date and date format are set in this function. Press Modify and set the date using the keypad. Press Accept to save the date. Press Format to change between the available date formats: DD/MM/YYYY, MM/DD/YYYY, YYYY/MM/DD, YYYY‑MM‑DD, MM‑DD‑YYYY, and DD‑MM‑YYYY. The default format is YYYY/MM/DD.
Auto Poweroff
The Auto Poweroff function is used to save battery life. After the set time is elapsed, the meter will:
1. Automatically switch off, if in normal measurement mode. Press
On/Off to switch on again.
2. Enter a sleeping mode, if the continuous logging mode is
selected with a logging interval of at least 30 seconds. The “Auto Poweroff” message and the Wake up softkey appear on the LCD; logging is not stopped. Press Wake up to reactivate the display.
Available options are: Not used (disabled), 5, 10, 15, 20, 30 or 60 minutes. Press Modify to select the desired time interval. The default value is “not used”.
Key Beep
If enabled, an acoustic signal sounds every time a key is pressed. A checked box indicates this function has been enabled. The default
setting is disabled.
METER SETUP
Page 44
44
Error Beep
If enabled, an acoustic signal sounds every time an incorrect key is pressed, or when an error occurs. A checked box indicates this function has been enabled. The default setting is disabled.
Decimal Separator
The user can select the type of decimal separator: “dot” or “comma”. Press the softkey to select the desired option. The
default setting is “dot”.
LCD Contrast
The LCD contrast can be adjusted with this function. Press Modify to enter this function. Use the / keys to change the contrast level and press Accept to save the new value. The default value is 8.
LCD Backlight Intensity
The LCD backlight intensity can be adjusted with this function. Use the / keys to change the level and press Accept to save the new value. The default value is 7.
Meter Password
The Meter Password protects against unauthorized configuration changes and log data erasure. When implemented, many setting and functions cannot be modified or viewed. To enable the password proceed as follows:
• Highlight “Meter Password” and press Modify.
• Enter the desired password (of 6 digits) in the text box and press Accept.
Note: While typing, the characters are masked with a “*” (star) symbol.
METER SETUP
Page 45
45
• The meter will require password confirmation. Retype the same password and press Accept to
confirm.
• The meter returns to the “Meter Setup” menu. The checkbox corresponding to the meter password
is checked.
To disable the password protection highlight “Meter Password” and press Modify, enter the password and then press Disable. “No password” appears in the text box. Press Accept to confirm.
Meter ID
The Meter ID may be used to uniquely identify a meter/oper‑ ator. Press Modify and a text box appears. Use the keypad to insert the desired alphanumeric ID and press Accept to store the identification. A maximum of 14 characters can be used.
Language
The language used in the meter user interface can be changed.
The default language is English.
Please contact your local Hanna office for currently available languages.
Restore Factory Settings
This function restores measurement settings to their original factory values. This includes measurement units, coefficients, other measurement configurations and all logged data. The factory calibration for the sensor channels is not affected.
• Select the “Restore factory settings” and press Select.
• The meter will ask to confirm: press Yes to confirm or No to escape.
Probe ID
The probe can be labeled with an identification code: press Modify and a text box will be dis played. Use the keypad to enter the desired alphanumeric code and then press Accept. A maximum of 14 characters can be used.
PROBE SETUP
METER SETUP
Page 46
46
Chapter 9 - STATUS
Useful information regarding the meter, probe (if connected) and GLP calibration data are available for viewing by selecting “Status” from the main menu.
Select “Meter Status” to display information related to the battery, logging, internal temperature, password, Meter ID, serial number and firmware version. Press / to scroll through the status screens. Press ESC to return to the “Status” menu.
Select “Probe Status” to display information related to the probe type, connected sensors, Probe ID, serial number and firmware version.
• Press / to scroll through the status screens.
• Press ESC to return to the “Status” menu.
Note: The probe status screen will automatically be displayed when the probe sensor status has changed. If this occurs, the “Measurement Screen” and “Parameter Selection” softkeys are available (see Meter Initialization).
METER STATUS
PROBE STATUS
Page 47
47
GLP DATA
GLP (Good Laboratory Practice) is a set of functions that allows the user to store or recall data regarding the probe calibration. This feature also allows the user to associate readings with specific calibrations. To view GLP data select “GLP” from the “Status” menu. The complete list of available parameters appears. Select the desired pa‑ rameter to view the stored GLP information.
Notes: If no calibration data is available for the selected parameter, the display shows the message “No GLP data available for this measurement”. Press OK to return to the previous screen. GLP data is stored for the last 5 calibrations. This calibration history allows the user to detect when readings start to change and sensors may require cleaning or replacement.
pH
• From the “GLP” menu, select the “pH” option.
• Data regarding the last pH calibration will be displayed: offset,
acidic slope, basic slope, buffers used, time and date of the
calibration.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
• Press ESC to return to the “GLP” menu.
Notes: A “C” label near the buffer value indicates a custom point, while an “H” indicates a HANNA standard buffer value. If a quick calibration was performed, the buffer values are replaced with the “Quick calibration” indication. If no pH calibration has been performed or if calibration was cleared using the “Restore factory calib.” option the offset and slope values are set to default, and the message “Factory calibration” is displayed. Press ESC to return to the previous screen.
Page 48
48
GLP DATA
ORP
• From the “GLP” menu select the “ORP” option.
• Data regarding the last ORP calibration will be displayed: calibration point, time and date.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
• Press ESC to return to the “GLP” menu.
Notes: If no ORP calibration has been performed or if calibration was cleared using the “Restore factory calib.” option the offset and slope values are set to default, and the message ”Factory calibration” is displayed. Press ESC to return to the previous screen.
Dissolved Oxygen
• From the “GLP” menu select the “Dissolved oxygen” option.
• Data regarding the last D.O. calibration will be displayed: calibration points, % saturation or concentration, time and date.
• Use the / keys to scroll throught the stored data for the last 5 calibrations.
Notes: A “C” label near the calibration point indicates a custom point, while an “H” indicates a HANNA standard value. When the % D.O. range is calibrated, also the D.O. concentration range is calibrated, and vice versa. If no D.O. calibration has been performed or if calibration was cleared using the “Restore factory calib.” option the offset and slope values are set to default, and the message ”Factory calibration” is displayed. Press ESC to return to the previous screen.
Conductivity
• From the “GLP” menu select the “Conductivity” option.
• Data regarding the last conductivity calibration will be displayed:
calibration point, cell constant value, calibration type (conductivity, absolute conductivity or salinity), time and date of the calibration.
• Use the / keys to scroll through the stored data for the last
5 calibrations.
Page 49
49
Temperature
• From the “GLP” menu select the “Temperature”option.
• Data regarding the last temperature calibration will be displayed: calibrated point, time and date.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
Notes: If no user temperature calibration has been performed or if calibration was cleared using the “Restore factory calib.” option the offset value is set to default, and the message ”Factory calibration” is displayed. Press ESC to return to the previous screen.
Atmospheric Pressure
• From the “GLP” menu select “Atm. pressure”.
• Data regarding the last atmospheric pressure calibration will be displayed: custom calibration point, time and date.
• Use the / keys to scroll through the stored data for the last 5 calibrations.
Notes: If no atmospheric pressure calibration has been performed or if calibration was cleared using the “Restore factory calib.” option the offset value is set to default, and the message ”Factory calibration” is displayed. Press ESC to return to the previous screen.
GLP DATA
Chapter 10 - LOGGING MODE
The HI9819X and the corresponding probe offer many logging options that can be combined based on user needs. The following figures describe the available logging options.
LOGGING MODE
Page 50
50
From measurement mode, press Log to access the log menu.
The data logged on the meter are organized by lots. Up to 45,000 complete records can be stored
in up to 100 lots. Each lot can store log‑on‑demand records and/or continuous records with different parameter configurations.
LOGGING MENU STRUCTURE
LOGGING ON METER
Page 51
51
One Sample On Meter
Use this option to log one set of enabled measurement parameters to the meter memory.
• If there are no lots saved on the meter, press New to create a new lot. Use the keypad to enter the desired lot name and press Accept to confirm. Press OK to log the sample in the selected lot.
• If there are existing lots on the meter, the meter will suggest a lot to store the sample. Press OK to use the selected lot or Options to select a different lot. This will add the new sample data to an existing lot. A new lot can also be created by pressing New. Press OK to log the sample in the selected lot.
• On the “Remarks” window, select Yes to go to the Remarks screen. Press No to skip this option.
If Yes is selected, select a remark from the list, or press New to
create a new remark.
• To return to the measurement screen, press ESC.
Continuous Meter Log
• Select “Start meter log” to log the currently enabled parameters at the set logging interval on the meter.
• To set the logging interval, highlight “Start meter log” and press Options. The log interval time
can set from 1 second to 3 hours. Press Modify and use
the / keys and keypad to enter the desired log interval. Press Accept to confirm.
• Press Select to edit the lot, remark or tag.
• To stop the meter log, enter the log menu and select “Stop meter log”.
LOGGING ON METER
Page 52
52
• Select “Meter log recall” to view logs that are stored on the meter. The meter will show the number of available lots. Select “Lots” to view or delete individual lots.
• Use the / keys to select the desired lot and then press View.
• The meter displays a summary of all data related to the selected lot: number of samples, memory space used, time and date of the first and last readings.
• Press View to display the sample details for each point. Use the / keys to change the sample number in the selected lot. The sample number is shown on the bottom right corner of the display.
Notes: The values displayed for D.O. concentration, compensated conductivity and TDS are dependent on the coefficients defined in Parameter Coefficients in Setup menu (EC Reference Temperature, EC Temperature Coefficient, TDS factor and Salinity). If parameter units or coefficients are altered, stored logs on this meter will be altered reflecting these changes. Save logs to a PC before altering parameters or coefficients. Details are available only for the enabled parameters.
• Press Info to see record information for the current sample
(time & date, remark or serial number (if available)).
• Press Data to return to the previous screen or Jump to select
a different sample in the same lot. When Jump is pressed, a text box appears to insert the desired sample number.
• Press ESC to return to the menu.
• Choose “Plot” and the meter will create a list with all available parameters that can be plotted.
• Use the / keys to select the desired parameter. Press Select to view the graph.
LOG RECALL
Page 53
53
• Use the / keys to move the cursor in the graph and
highlight a sample. The sample data are displayed below the graph.
• Press ESC to return to the parameter list.
• Press ESC again to return to the menu.
Note: The number of lot samples that can be plotted is limited by the display resolution. To view a complete graph download data to PC.
Delete all lots
• From ‘’Meter log recall’’ choose “Delete all lots” and the meter
will display the message “Do you want to perform the current operation?”. Press Yes to delete or No to return to the previous
screen.
• To return to the “Log recall” menu, press ESC.
Remarks
A remark can be associated with each sample. The meter can store up to 20 remarks.
• To add a remark, select “Log notes” from the Log menu, and
then select “Remarks”.
• The display shows a list of stored remarks.
• Press New to create a new remark, and use the keypad to enter
the new remark in the text box.
• Press Delete to delete the selected remark from the meter. If the
deleted remark is used in an existing lot, the information will be still available in the lot data.
Delete All Remarks
• Select “Delete all remarks” to delete all remarks. The display
will show the message “Do you want to perform the current operation?”. Press Yes to delete or No to return to the previous screen.
LOG NOTESLOG RECALL
Page 54
54
Chapter 11 - PC CONNECTION MODE
The logged data from a probe or meter can be transferred to a PC using the HI9298194 Windows® compatible application software. HI9298194 offers a variety of features and on‑line‑help is available.
HI9298194 allows data to be imported into most spreadsheet programs (e.g. Excel©,
Lotus 1‑2‑3©). After the data has been imported into a spreadsheet, all features of the spreadsheet program can be used to analyze and graph the data.
SOFTWARE INSTALLATION
• Insert the installation CD into the PC.
• The software menu window should start automatically (if it does not, navigate to the main CD
folder and double‑click “setup.exe”).
Click “Install software” and follow the instructions.
METER TO PC CONNECTION
• With the meter OFF, disconnect the probe.
• Connect the USB cable to the meter and to a USB port on the PC.
• Turn the meter ON and the message “PC connected” will be displayed.
• Run the HI9298194 application software.
• Press Setting button on the top of the screen and select the
measurement units you which your data to appear with.
• To access the meter data select the “Meter” button on the
toolbar at the top of the screen.
The PC‑Meter connection will be established and a new window
will be displayed with meter data: status information (software version and date, SN, ID, battery level and free memory info), as well as a summary of logged data lots.
The lots logged on the meter can be saved to the PC by pressing the “Download lot” button after
the desired lot is selected.
• Once the lot has been downloaded, all the logged samples can be viewed.
Windows® is a registered Trademark of “Microsoft Co.”
PC CONNECTION MODE
Page 55
55
Meter to PC data
a. Select parameter units b. Select Meter from toolbar c. Select Lot
PC CONNECTION MODE
Page 56
56
Chapter 12 - TROUBLESHOOTING / ERROR MESSAGES
HI9819X displays error messages to aid in troubleshooting. Warnings are displayed for most issues,
while Errors are displayed for critical issues. See the calibration chapter for messages that can occur during calibration.
Other messages are listed below.
• “Log space full” appears when the meter memory is full and additional data cannot be logged. Delete one or more lots from the meter (Log / Meter Log).
• “Power fault. Check the probe cable”: this message may appear when powering up the meter with a probe connected. If the meter detects a high load on the probe connection this message is triggered. Check the probe cable. If the problem persists, contact the HANNA service center.
• “Language data not available”: this message appears when powering up the meter if the language file is not seen by the meter. Restart the meter to verify this is a true meter error. If the problem persists, contact the HANNA service center.
• ‘’Dead meter battery!’’: This message appears if the me‑ ter batteries are too low to power the meter and it will automatically turn off. Connect the charger if using rechargeable C batteries or replace the alkaline batteries to continue.
• ‘’User data corrupted!’’: This message appears when
powering up the user data stored on meter are corrupted. Restart the meter. If the problem persists, contact the HANNA service center.
• ‘’Warning x’’: Any other warning that appears at
power‑on is identified using a numeric code. Restart the meter. If the problem persists, contact the HANNA service center. Some meter/probe features can be accessed but with no guarantee.
• ‘’Errors x’’: Any critical errors that appear are identified using a numeric code, and the
meter is automatically switched off. Contact the HANNA service center.
TROUBLESHOOTING / ERROR MESSAGES
Page 57
57
The HI7698292 probe maintenance kit includes HI7042S (electrolyte solution for D.O. sensor), spare membranes with O‑Rings for D.O. sensor, a small brush for cleaning EC , O‑Rings for sensor connectors and a syringe with grease to lubricate these O‑Rings.
General Maintenance
• Inspect all sensor connectors for corrosion and replace sensors if necessary.
• Inspect sensor O‑Rings for nicks or other damage and replace sensor if necessary. Lubricate only with grease from kit.
Use only the supplied grease as some lubricants can cause the O‑Rings to expand or affect the turbidity calibration standards.
• After prolonged storage or cleaning, calibration of the sensors is required.
• After use rinse the probe with tap water and dry it. The pH electrode bulb must be kept moist. Dry
the D.O. and EC sensors.
• Check GLP data under “Status” to ensure the sensor is still functioning properly.
pH and pH/ORP Sensor Maintenance
• Remove the sensor protective cap. Do not be alarmed if any salt deposits are present. This is
normal with pH/ORP electrodes and they will disappear when rinsed with water.
• Shake down the sensor as you would do with a clinical thermometer to eliminate any air bubbles
inside the glass bulb.
• If the bulb and/or junction are dry, soak the electrode in HI70300 storage solution for at least
one hour.
• To ensure a quick response time, the glass bulb and the junction should be kept moist and not
allowed to dry. Store the sensor with a few drops of HI70300 storage solution or pH 4.01 buffer in the protective cap. Tap water may also be used for a very short period (few days).
Never use distilled or deionized water to store pH sensors.
• Inspect the sensor for scratches or cracks. If any are present, replace the sensor.
APPENDIX A - PROBE MAINTENANCE
Page 58
58
• Cleaning procedure: clean the sensor frequently by soaking it for 1 minute in HI70670 or
HI70671 cleaning solution. After cleaning soak the sensor in HI70300 storage solution before
taking measurements.
D.O. Sensor Maintenance (HI98194, HI98196 only)
For a top performance probe, it is recommended to replace the membrane every 2 months and the electrolyte monthly. Proceed as follows:
• Unscrew the membrane by turning it counterclockwise.
• Rinse a spare membrane with some electrolyte while shaking it gently. Refill with clean electrolyte.
• Gently tap the cap over a surface to ensure that no air bubbles remain trapped. Avoid touching the membrane.
• With the sensor facing down, completely screw the cap clockwise. Some electrolyte will overflow.
If any deposit scales the sensor, gently brush the sensor surface with the supplied brush, while
paying attention to not damage the plastic body. Do not use the brush on the membrane.
EC Sensor Maintenance (HI98195, HI98196 only)
• After every series of measurements, rinse the probe with tap water.
• If a more thorough cleaning is required, clean the sensor with the supplied brush or a non‑abrasive
detergent. Ensure that the two cylindrical holes in the sensor are free of foreign material.
APPENDIX A - PROBE MAINTENANCE
Page 59
59
APPENDIX B - PROBE DEPLOYMENT
The Hanna HI769819X have been designed for a variety of water quality measurements both in situ or in active deployments in urban or natural waters.
The HI9819X systems may be used for discrete spot sampling with log on demand function or, continuous monitoring and logging. These data can be downloaded to a PC and plotted with logging software to obtain the graphical log needed for interpretation of the essential physical property of the aqueous body of water.
In all of these deployment situations data quality is dependent upon the site location, service intervals, amount of coatings, sedimentation and vegetation, and the actual installation. The probe may be installed in a horizontal bank (fixed installation) or a vertical suspension. The maximum depth rating of 20 m (65’) for the probe should be adhered to.
Note: actual sensor specifications may be less.
The location must be accessible for the duration of the measurement (consider seasonal flooding, freezing and other acts of nature) when selecting a site. Many conditions may affect the quality of measurements.
Select an installation site that is representative of the water body being monitored. Avoid areas without adequate water circulation. To protect equipment it is best to avoid exposure
to wind, foam, turbulence, air temperature gradients/sun, extended periods of high flow, extended periods of high sediment and floating debris. The standard operating procedures (SOP) for the data gathering must be upheld. This typically includes pre and post deployment checks of the sensors to validate data gathered between calibrations, upholding service intervals, and following any other site‑specific procedures. Grab samples for laboratory analysis or spot sampling with another probe are addition ways to validate the measurements taken by unattended continuous logs.
The probe is suitable for installation in confined locations such as air vaults, river intakes, vertical wells, tanks, etc.. The streamline diameter of the probe permits insertion into 2" pipelines. Unlike probes that require a cable support for active deployments the probe can be manually lowered and raised by the cable due to it’s superior strength member.
It is suitable for installation in open moving waters; rivers, streams, ditches (farmland drainage), conveyance canals, etc.. In these cases protecting the probe from debris is important. If the probe is suspended from a pier or bridge position it behind a support and anchor the cable/probe to a pipe.
It is suitable for deployment in open waters; monitoring lakes, ponds, wetland basin, infiltration basins, bays. Schedule regular service to remove aquatic weed growth that may be interfering with representative water samples.
Page 60
60
The probe is suitable for measurements in a flow cell. Pumping water to a flow‑through monitoring station has obvious pros and cons. Typically a shelter is required to secure a pump, and flow chamber. A power requirement, shelter, pump maintenance and higher installation cost need to be considered. Freeze protection, security, and convenience of calibration and possibility of adding multiple mea‑ surement points and antifouling preconditioning systems are advantages to this type of installation.
General Guidelines for fixed installation:
• Select a water‑sampling site that will allow collection of representative water samples.
• Position the probe so the sensor surfaces face toward the flow. This will minimize air bubble or fluid cavitation. Limit flow rate to moderate.
• Mount Probe 0 to 45° angle from vertical to avoid sensors (pH, pH/ORP) from becoming electrically discontinuous due to internal electrolytes flowing away from their internal cells.
• Install meter and probe where they will be accessible for maintenance as required.
• Regularly visit water sampling sites to: check for damage to sensors, the installation mountings, and the meter battery power.
• Remove aquatic weed growth that may be interfering with water sample collection.
• Set up devices and programs for water monitoring and sampling.
• If the probe is suspended from a pier or bridge ensure that it is protected from debris by positioning behind a support and anchoring the cable / probe to a pipe.
• Have access to spare sensors and proper range standard solutions or buffers.
• Strictly follow the established SOP’s.
• Flow cell installation; Avoid trapped air. Maintain constant flow rate.
APPENDIX B - PROBE DEPLOYMENT
Page 61
61
APPENDIX C - ACCESSORIES
PROBES packed in carton box, without sensors or protective shield.
Code Description
HI7698194/4 HI7698194 probe with 4 m (13.1’) cable.
HI7698194/10 HI7698194 probe with 10 m (33.0’) cable
HI7698194/20 HI7698194 probe with 20 m (65.6’) cable
HI7698194/40 HI7698194 probe with 40 m (131.2’) cable
HI7698195/4 HI7698195 probe with 4 m (13.1’) cable
HI7698195/10
HI7698195 probe with 10 m (33.0’) cable
HI7698195/20 HI7698195 probe with 20 m (65.6’) cable
HI7698195/40 HI7698195 probe with 40 m (131.2’) cable
HI7698196/4 HI7698196 probe with 4 m (13.1’) cable
HI7698196/10 HI7698196 probe with 10 m (33.0’) cable
HI7698196/20 HI7698196 probe with 20 m (65.6’) cable
HI7698196/40 HI7698196 probe with 40 m (131.2’) cable
Note: Probes with different cable length are available upon request. Order protective shields separately.
Page 62
62
APPENDIX C - ACCESSORIES
METERS WITH PROBES
Code Description
HI7698194 HI98194 meter, probe with 4 m (13.1’) cable, with pH/ORP, EC,
D.O. sensors.
HI7698194/10 HI98194 meter, probe with 10 m (33.0’) cable, with pH/ORP, EC,
D.O. sensors.
HI7698194/20 HI98194 meter, probe with 20 m (65.6’) cable, with pH/ORP, EC,
D.O. sensors.
HI7698194/40 HI98194 meter, probe with 40 m (131.2’) cable, with pH/ORP, EC,
D.O. sensors.
HI7698195 HI98195 meter, probe with 4 m (13.1’) cable, with pH/ORP,
EC sensors.
HI7698195/10 HI98195 meter, probe with 10 m (33.0’) cable, with pH/ORP,
EC sensors.
HI7698195/20 HI98195 meter, probe with 20 m (65.6’) cable, with pH/ORP,
EC sensors.
HI7698195/40 HI98195 meter, probe with 40 m (131.2’) cable, with pH/ORP,
EC sensors.
HI7698196 HI98196 meter, probe with 4 m (13.1’) cable, with pH/ORP,
D.O. sensors.
HI7698196/10 HI98196 meter, probe with 10 m (33.0’) cable, with pH/ORP,
D.O. sensors.
HI7698196/20 HI98196 meter, probe with 20 m (65.6’) cable, with pH/ORP,
D.O. sensors.
HI7698196/40 HI98196 meter, probe with 40 m (131.2’) cable, with pH/ORP,
D.O. sensors.
Page 63
63
APPENDIX C - ACCESSORIES
SENSORS
Code Description
HI7698194-0 pH sensor
HI7698194-1 pH/ORP sensor
HI7698194-2 Dissolved Oxygen sensor
HI7698194-3 EC sensor
CABLES, CONNECTORS, ACCESSORIES
Code Description
HI7698290 Short calibration beaker
HI7698295 Short protective shield
HI7698297 Long, quick release flow cell
HI7698292 Probe maintenance kit with HI7042S (electrolyte solution for D.O.
sensor), small brush, small hex key, O‑Rings for D.O. sensor (5 pcs.), O‑Rings for probe (5 pcs.) and syringe with grease to lubricate the O‑Rings
HI76981952 Probe maintenance kit with small brush, small hex key, O‑Rings for
probe (5 pcs.) and syringe with grease to lubricate the O‑Rings.
HI9298194 PC application software
HI920015 Micro USB cable, PC to meter
Page 64
64
APPENDIX C - ACCESSORIES
QUICK CALIBRATION SOLUTIONS
Code Description
HI9828-20 Quick calibration solution, 230 mL
HI9828-25 Quick calibration solution, 500 mL
HI9828-27 Quick calibration solution, 1 gal.
pH BUFFERS
HI5004 pH 4.01 buffer solution, 500 mL
HI5005 pH 5.00 buffer solution, 500 mL
HI5006 pH 6.00 buffer solution, 500 mL
HI5068 pH 6.86 buffer solution, 500 mL
HI5007 pH 7.01 buffer solution, 500 mL
HI5074 pH 7.41 buffer solution, 500 mL
HI5008 pH 8.00 buffer solution, 500 mL
HI5009 pH 9.00 buffer solution, 500 mL
HI5091 pH 9.18 buffer solution, 500 mL
HI5010 pH 10.01 buffer solution, 500 mL
Page 65
65
ORP SOLUTIONS
Code Description
HI7020L ORP test solution, 200/275 mV @ 25 °C, 500 mL
HI7021L ORP test solution, 240 mV @ 25 °C, 500 mL
HI7022L ORP test solution, 470 mV @ 25 °C, 500 mL
HI7091L Reducing pretreatment solution
HI7092L Oxidizing pretreatment solution, 500 mL
APPENDIX C - ACCESSORIES
pH/ORP MAINTENANCE SOLUTIONS
Code Description
HI70670L pH/ORP cleaning solution for salt deposits, 500 mL
HI70671L pH/ORP cleaning and disinfecting solution for algae,
fungi and bacteria, 500 mL
HI70300L pH/ORP electrode storage solution, 500 mL
D.O. SOLUTIONS
Code Description
HI7040L
Zero oxygen solution
HI7042S
Electrolyte solution for D.O. sensor, 30 mL
HI76409A/P
Spare membrane with O‑Ring (5 pcs.)
Page 66
66
CONDUCTIVITY STANDARD SOLUTIONS
Code Description
HI7030L 12880 µS/cm calibration solution, 500 mL
HI7031L 1413 µS/cm calibration solution, 500 mL
HI7033L 84 µS/cm calibration solution, 500 mL
HI7034L 80000 µS/cm calibration solution, 500 mL
HI7035L 111800 µS/cm calibration solution, 500 mL
HI7039L 5000 µS/cm calibration solution, 500 mL
APPENDIX C - ACCESSORIES
Page 67
HI98194, HI98195 and HI98196 meters are warranted for two years
(sensors, electrodes and probes for six months) against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to HANNA Instruments first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
Warranty
Recommendations for Users
Before using Hanna products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipment. Take all necessary steps to correct such interferences. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instruments’ EMC performance. To avoid electrical shock, do not use these instruments when volt‑ ages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid damage or burns, do not perform any measurement in microwave ovens.
Hanna Instruments reserves the right to modify the design, construction or appearance of its products
without advance notice.
Page 68
MAN98194 03/16
Printed in ROMANIA
World Headquarters
Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com
Local Office
Hanna Instruments USA 270 George Washington Highway Smithfield, RI 02917 Phone: 800.426.6287 Fax: 401.765.7575 e‑mail: tech@hannainst.com
Loading...