Hanna Instruments HI98163 Instruction Manual

INSTRUCTION MANUAL
HI98163
Professional pH and Temperature
Meter for Meat
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the
instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e‑mail us at
tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Dear Customer,
3
PRELIMINARY EXAMINATION ..............................................................................................
SAFETY MEASURES ............................................................................................................
GENERAL DESCRIPTION ......................................................................................................
FUNCTIONAL DESCRIPTION ................................................................................................
SPECIFICATIONS ................................................................................................................
OPERATIONAL GUIDE .........................................................................................................
pH CALIBRATION ...............................................................................................................
pH BUFFER TEMPERATURE DEPENDENCE ............................................................................
GOOD LABORATORY PRACTICE (GLP) ...................................................................................
SETUP ..............................................................................................................................
LOGGING ..........................................................................................................................
AUTOEND .........................................................................................................................
mV AND TEMPERATURE CALIBRATION (FOR TECHNICAL PERSONNEL ONLY) ............................
PC INTERFACE ...................................................................................................................
BATTERIES REPLACEMENT ..................................................................................................
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS .......................................................
ELECTRODE CONDITIONING AND MAINTENANCE ..................................................................
TROUBLESHOOTING GUIDE ................................................................................................
ACCESSORIES ...................................................................................................................
TABLE OF CONTENTS
4
4
6
9
21
23
34
38
43
5
8
11
22
33
35
42
44
47
48
4
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, please notify us.
Each instrument is supplied with:
FC2323 PVDF body, pH probe with internal temperature sensor
HI700630 Acid cleaning solution for meat grease and fats (food industry) (2 pcs.)
HI92000 PC Software
HI920015 USB cable
•pH 4.01 & 7.01 Buffer solutions (230 mL each)
•100 mL Beaker (2 pcs.)
•1.5V AA Batteries (4 pcs.)
HI720161 Hard carrying case
•Instruction Manual and Quick Reference Guide
•Instrument Quality Certificate
PRELIMINARY EXAMINATION
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories.
SAFETY MEASURES
Before using this product, make sure that it is entirely suitable for your specific application and for the environment in which it is used.
Operation of this instrument may cause interference to other electronic equipment, requiring the operator to take steps to correct interference. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave ovens. For your and the instrument’s safety, do not use or store the instrument in hazardous environments.
5
GENERAL DESCRIPTION
Other features include:
• Log on demand up to 200 samples (100 pH and 100 mV).
• Auto Hold feature, to freeze first stable reading on the LCD.
• GLP feature, to view last calibration data for pH.
• PC interface.
HI98163 is a portable pH and temperature meter with a special probe, dedicated to the measure‑
ment of pH in meat processing. The HI98163 meter is provided with a series of new diagnostic features which add an entirely new dimension to the measurement of pH, by allowing the user to dramatically improve the reliability of the measurement:
• Seven standard buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45) for calibration.
• pH calibration up to five calibration points (see instrument specifications).
• Custom calibration with up to five custom buffers.
• Messages on the graphic LCD for an easy and accurate calibration.
• Cal Check™ Diagnostic features to alert the user when the electrode needs cleaning.
• Optional user enabled “Outside Calibration Range” warning.
• Monitoring of the electrode aging.
• User selectable “Calibration Time Out” to remind when a new calibration is necessary.
This meter is supplied with a pH electrode specifically designed for use with meat. The FC2323 pH electrode with a built‑in temperature sensor features a rugged easy to clean PVDF body with a stainless steel blade tip for meat penetration. The free diffusion junction helps to avoid a clogged reference, where the external body material is non‑toxic and food compatible.
6
1) Liquid Crystal Display (LCD).
2) Functional keys.
3)
/ keys to scroll through calibration buffers, setup options, manual temperature compensation
and logged data.
4) ON/OFF (
) key, to turn the instrument ON and OFF.
5) LIGHT (
) key to toggle display backlighting.
6) GLP key, to display Good Laboratory Practice information.
7) CAL key, to enter/exit calibration mode.
8) SETUP key, to enter/exit SETUP mode.
9) RCL key, to enter/exit view logged data mode.
10) MODE key to change pH resolution.
11) RANGE key, to switch between pH and mV range.
12) HELP key to enter/exit contextual help.
13) ESC to leave current mode, exit calibration, setup, help. etc.
FRONT VIEW
FUNCTIONAL DESCRIPTION
7
14) Electrode quick connect DIN connector.
15) USB connector.
TOP VIEW
FUNCTIONAL DESCRIPTION
8
SPECIFICATIONS
pH
Range ‑2.0 to 20.0 pH / ‑2.00 to 20.00 pH / ‑2.000 to 20.000 pH
Resolution 0.1 pH / 0.01 pH / 0.001 pH
Accuracy ±0.1 pH / ±0.01 pH / ±0.002 pH
Temperature
Range ‑20.0 to 120.0 °C (‑4.0 to 248.0 °F)
Resolution 0.1 °C (0.1 °F)
Accuracy ±0.4 °C (±0.8 °F) (excluding probe error)
mV
Range
±2000.0 mV
Resolution
0.1 mV
Accuracy
±0.2 mV
pH Calibration
Up to five point calibration, seven standard buffers available
(pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and five custom
buffers
Slope Calibration From 80 to 110%
Temperature Compensation Automatic
pH Electrode FC2323 pH & temperature
LOG On demand, 200 samples (100 pH and 100 mV) Input Impedance 10¹² Ω Battery Type/Life 1.5V AA batteries (4 pcs.) / approx. 200 hours of continuous use
without backlight (50 hours with backlight)
Auto Power Off User selectable: 5, 10, 30, 60 minutes or disabled
PC Interface opto‑isolated USB
Dimensions 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
Weight 400 g (14.2 oz)
Environment 0 to 50 °C (32 to 122 °F) max. RH 100% IP67
9
OPERATIONAL GUIDE
INITIAL PREPARATION
The instrument is supplied complete with 1.5V AA (4 pcs.) batteries. For placing the batteries inside the meter, see page 42. Make sure that the protective micro USB cover is used when not connected to a computer to ensure waterproof protection. For HI98163 connect the pH/temperature electrode to the DIN connector. Turn the instrument ON by pressing ON/OFF key.
At start‑up the display will show the Hanna logo for a few seconds followed by the percentage indication of the remaining battery life, then enters the measurement mode. After measurement switch the instrument off, clean the electrode and store it with a few drops of
HI70300 storage solution in the protective cap (see page 45).
The Auto Power Off feature turns the instrument off after a set period (default 30 min) with no button pressed to save battery life. To set another period or to disable this feature, see SETUP menu on page 23. The Auto Light Off backlight feature turns the backlight off after a set period (default 1 min) with no buttons pressed. To set another period or to disable this feature, see SETUP menu on page 23.
pH MEASUREMENTS
To take a pH measurement remove the electrode protective cap and simply submerge the tip of the electrode (4 cm/1½") into the sample to be tested. Press RANGE key to choose between pH and mV ranges. Use MODE key to select the pH resolution. Allow time for the electrode to adjust and reading to stabilize (hourglass symbol turns off).
10
OPERATIONAL GUIDE
• pH reading with the selected resolution.
• Temperature reading in the selected unit (°C or °F).
• Electrode condition during the calibration day.
• The buffers used in last pH calibration (if feature is enabled in SETUP).
• Battery level indicator.
• Available functional keys. In order to take more accurate pH measurements, make sure that the instrument is calibrated (see page 11 for calibration details). It is recommended that the electrode is always kept moist and rinsed thoroughly with the sample to be measured before use. The pH reading is directly affected by temperature. For accurate pH measurements, temperature must be taken into consideration. If the sample temperature is different from the temperature at which the pH electrode was kept, allow a few minutes to reach thermal equilibrium.
On the pH screen are displayed:
TEMPERATURE MEASUREMENTS
For HI98163 the temperature sensor is connected through DIN socket.
Note: The temperature can be displayed in Celsius degrees (°C) or in Fahrenheit degrees (°F) (see SETUP for details, page 23).
BACKLIGHT FEATURE
The instrument is provided with a Backlight feature, which can be easily toggled on and off through the keyboard by pressing LIGHT.
Note: The backlight automatically shuts off after a set period (see SETUP for details, page 23) with no buttons pressed.
11
pH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high accuracy is required. The pH range should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When calibration alarm time out is expired ‑ “CAL DUE” blinks (if feature is enabled in SETUP).
• If “Outside Cal Range” message blinks during pH measurement (the measurement range is not covered by current calibration, if feature is enabled in SETUP).
PROCEDURE
HI98163 instrument offers a choice of seven standard buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01
and 12.45). The meter allow the user to set up to five custom buffers. The set custom buffers are the buffer values at 25 °C. When a custom buffer is selected during calibration, the Custom functional key is displayed on the LCD. Press Custom key in order to enter custom buffer changing mode. Use / keys to change the value in a ±1.00 pH window, in according with the temperature reading and then Accept. Press ESC to leave custom buffers value unchanged. For accurate pH measurements, it is recommended to perform a calibration with buffers that bracket the expected pH value. At least two point calibration is recommended.
The instrument will automatically skip the buffers used during calibration and the buffers which are in a ±0.2 pH window around one of the calibrated buffers.
• Remove protective cap and rinse the electrode with distilled or deionized water.
• Pour small quantities of selected buffer solutions into clean beakers. For accurate calibration use
two beakers for each buffer solution, the first one for rinsing the electrode and the second one for calibration.
12
pH CALIBRATION
• Press CAL. The instrument will display the measured pH, the LCD first expected buffer and the
temperature reading.
• If necessary, press the / keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and within range of the selected buffer, CFM functional key is displayed.
FIVE POINT CALIBRATION
• Immerse the pH electrode approximately 4 cm (1½”) into a buffer solution of your choice (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 or a custom buffer) and stir gently.
• Press CFM to confirm first point.
• The calibrated value and the second expected buffer value is then displayed on the LCD.
13
pH CALIBRATION
• After the first calibration point is confirmed, immerse the pH electrode probe approximately 4 cm (1½”) into the second buffer solution and stir gently.
• If necessary, press the / keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and within range of the selected buffer, the CFM functional key is displayed.
• Press CFM to confirm calibration.
• The calibrated value and the third expected buffer value will be displayed.
• After the second calibration point is confirmed, immerse the pH electrode approximately 4 cm (1½”) into the third buffer solution and stir gently.
14
pH CALIBRATION
• If necessary, press the / keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and within range of the selected buffer, the CFM functional key is displayed.
• Press CFM to confirm calibration.
• The calibrated value and the fourth expected value will be displayed.
• After the third calibration point is confirmed, immerse the pH electrode approximately 4 cm (1½”) into the fourth buffer solution and stir gently.
• If necessary, press the / keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and within range of the selected buffer, the CFM functional key is displayed.
15
pH CALIBRATION
• Press CFM to confirm calibration.
• The calibrated value and the fifth expected buffer will be displayed.
• After the fourth calibration point is confirmed, immerse the pH electrode approximately 4 cm (1½”) into the fifth buffer solution and stir gently.
• If necessary, press the / keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and within range of the selected buffer, the CFM functional key is displayed.
• Press CFM to confirm calibration.
• The instrument stores the calibration values and returns to normal measurement mode.
16
FOUR, THREE, TWO or ONE POINT CALIBRATION
• Proceed as described in “FIVE POINT CALIBRATION” section.
• Press CAL or ESC after the appropriate accepted calibration point. The instruments will return to
measurement mode and will memorize the calibration data.
FIRST POINT MODE
Two SETUP options are available to determine how the previous calibration will be affected by a subsequent one point calibration. These options are Replace or Offset. If the Replace option is selected, the slopes between current buffer and nearest lower and higher buffers will be recalculated. If the Offset option is selected, an electrode offset correction is performed keeping unchanged the existing slopes.
ERROR SCREENS
Wrong buffer
The calibration cannot be confirmed.
The pH reading is not within range of the selected buffer. Select another buffer using the / keys or change the buffer.
pH CALIBRATION
Electrode Dirty/Broken alternatively with Buffer Contaminated The calibration cannot be confirmed.
The offset of the electrode is not in the accepted range. Check if the electrode is broken or clean it following the Cleaning Procedure (see page 45). Check the quality of the buffer. If necessary, change the buffer.