Hanna Instruments HI 981401 User Manual

Instruction Manual
(HI 981401N)
On-line, waterproof
pH meter with large LCD
This Instrument is in
http://www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for a correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the directives.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.
The meter is supplied with:
 HI 1286 pH electrode;  HI 1283 grounding probe;  Calibration screwdriver;  pH 4.01 and 7.01 buffer solutions (20 mL each);  12 VDC power adapter.
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range 0.0 to 14.0 pH Resolution 0.1 pH Accuracy (@ 25°C/77°F) ±0.2 pH Typical EMC Deviation ±0.1 pH Calibration Manual with two trimmers
for offset and slope
Probes HI 1286 interchangeable pH electrode
and HI 1283 grounding probe (included)
Casing IP54 Power supply External 12 VDC (included) Dimensions 86 x 94 x 33 mm (3.4 x 3.7 x 1.3) Weight 150 g (5.3 oz)
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
pH ELECTRODE CONNECTION & MAINTENANCE
In order to protect pH against vapors and humidity, the BNC connector is shielded behind a waterproof sheath.
pH is warranted for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrode is warranted for a period of six months. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of pur­chase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service de­partment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protec­tion.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instru­ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
pH is a pH meter specially designed to meet the needs of growers in greenhouses and hydroponic appli­cations. It is equipped with a large LC display which make easy to read measurements from distance at any time.
The housing has been completely sealed against vapors and humidity. The HI 1286 gel-filled pH electrode is inter­changeable and the BNC connector is protected behind a waterproof sheath.
The meter is also supplied with HI 1283 stainless steel probe to prevent potential grounding problems and thus ensuring longer life for your electrode, which has been designed for use in fertilizer solutions with high concentration of phos­phate and nitrate.
You can simply hang the meter right above the sample to be tested for continuous measurement, and it will run without interruption on 12 VDC power supply.
Measurements are highly accurate and the meter can be calibrated at one or two points.
1. Molded eye
2. pH 4.0 calibration trimmer
3. BNC connector
4. HI 1283 stainless steel grounding probe
5. HI 1286 pH electrode
6. 12 VDC power adapter
7. Power supply connector
8. Protective sheath
9. pH 7.0 calibration trimmer
10. Liquid Crystal Display (LCD)
 Slide the protective sheath down. Connect the HI 1286
pH electrode to the BNC connector and then slide the protective sheath back up all the way up to the pH
casing. Since the protective sheath is rub­berized and to ensure maximum waterproof protection, make sure the connector is completely covered.
 Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH elec­trodes and they dissolve when rinsed with water.
 When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops of storage (HI 70300) or pH7 (HI7007) solution in the protective cap. Always replace the protective cap after use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
 If the electrode has been left dry, soak the tip in a
storage (HI 70300) or pH7 (HI7007) solution for at least one hour to reactivate it.
CALIBRATIONCALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
For the greatest accuracy, frequent calibration of the instru­ment is recommended. In addition, the instrument must be recalibrated whenever:
a) The pH electrode is replaced. b) After testing aggressive chemicals. c) Extreme accuracy is required. d) At least once a month.
Note: the electrode should be sub-
merged approximately 4cm (1½) in the solution.
 Adjust the right hand trimmer
with the calibration screwdriver until the LCD shows 7.0.
(1½")
4 cm
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
 To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it monthly. Im­merse the tip of the electrode in HI 7061 for one hour and then rinse it with tap water.
TAKING pH MEASUREMENTS
 Turn the meter on by connecting the
12 VDC power adapter to the meter and to the mains.
 In addition to the pH electrode, the
pH is supplied with a differential input and HI1283 stain­less steel probe to prevent potential grounding problems. Remove the pro­tective cap from the pH electrode and immerse the tips (4cm/1½) of both pH electrode and grounding probe in the sample.
 The LCD will show the pH value. Allow the reading to stabilize
and pH will start continuous monitoring.
Note: to prolong the electrode life, after switching the meter
off, remove the electrode from the solution or, alterna­tively, detach it from the meter.
PREPARATION
Pour small quantities of pH 7.01 (HI7007) and pH 4.01 (HI7004) solution into two clean beakers.
HI 7004
HI 7004 HI 7007HI 7004 HI 7007
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the tip of the electrode and the second one for calibration.
RINSE
HI 7007
CALIBRATION
HI 7007
This way, contamination of the buffers is minimized. pH4.01 (HI7004) is recommended for measuring acidic
samples. Use pH 10.01 (HI7010) if subsequent samples are alkaline.
CALIBRATION PROCEDURE
 Remove the protective cap from the
electrode, rinse and immerse it in a pH 7.01 buffer with the grounding probe. Stir gently and then wait a couple of minutes for the reading to stabilize.
Rinse and immerse the pH electrode
and the grounding probe in pH4.01 (or pH10.01) buffer and stir gently.
 Wait a couple of minutes and
then adjust the left hand trim­mer until the LCD shows the value of the second buffer.
The pH calibration is now complete.
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause unaccept­able interferences to radio and TV equipment. The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC. To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
HI 1283 * Stainless steel grounding probe with 2 m
(6.6) cable
HI 1286 Double junction, plastic body pH electrode
with 2 m (6.6) cable and BNC connector
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs) HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs) HI 7004L pH 4.01 solution, 460 mL bottle HI 7007L pH 7.01 solution, 460 mL bottle HI 7061L Electrode cleaning solution, 460 mL bottle HI 710005 12 VDC power adapter, US plug HI 710006 12 VDC power adapter, European plug HI 710012 12 VDC power adapter, Australian plug HI 710013 12 VDC power adapter, Southern Africa plug HI 710014 12 VDC power adapter, UK plug HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
* To be replaced only by authorized service technicians
IST981401NR1