Hanna Instruments HI9814 Instruction Manual

INSTRUCTION MANUAL
HI9814
pH, EC, TDS and Temperature
Meter with Quick Calibration
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before
information for correct use of this meter, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e‑mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
Dear Customer,
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
3
TABLE OF CONTENTS
Table of Contents ......................................................
Preliminary Examination ...........................................
General Description and Intended Use ........................
Main Features ..........................................................
Display Description ...................................................
Specifications ...........................................................
Operational Guide ....................................................
Calibrations ..............................................................
Battery Replacement .................................................
Accessories ...............................................................
Electrode Maintenance ..............................................
Certification ..............................................................
Recommendations for Users ......................................
Warranty .................................................................
4
4
5
7
14
5
6
15
11
17
3
18
19
19
4
Remove the meter from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, contact your local Hanna Instruments Office.
Each meter is supplied complete with:
HI1285‑7 pH/EC/TDS probe with built‑in temperature sensor, DIN connector and 1m (3.3’) cable
HI50036 Quick calibration solution sachets (3 pcs.)
HI700661 Electrode cleaning solution sachets for agriculture (3 pcs.)
• Alkaline batteries: 1.5V AAA (3 pcs.)
• Calibration certificate of meter
• Calibration certificate of probe
• Instruction manual
Note: Save all packing material until you are sure that the meter functions correctly. All defective items must be returned in the original packing with the supplied accessories.
PRELIMINARY EXAMINATIONGENERAL DESCRIPTION and INTENDED USE
HI9814 is a durable, portable pH, conductivity, total dissolved
solids and temperature meter for most measurements encountered in hydroponics, aquaponics or general agriculture applications. All operations and settings, are made through only two buttons. The housing is water‑ proof and rated for IP67 conditions. The supplied HI1285‑7 multiparameter probe measures pH, EC/TDS, and temperature in one convenient, rugged probe. A preamplifier is integrated into the probe to protect the pH measurement from transient electrical noise. Sources of electrical noise include ballasts used in lighting and pumps to circulate water and nutrient solutions. Other user‑selectable features include selectable TDS factors of 0.5 and 0.7 as well as auto‑off after 8 minutes or 60 minutes to prolong battery life.
5
MAIN FEATURES
• Simultaneous, pH, EC/TDS and temperature measurements on a large three line LCD display
• Automatic and simultaneous pH and EC calibration
• Selectable temperature unit
• pH electrode condition display
• mV of pH measurement for pH electrode check
• Last calibration points for pH and EC
HI1285‑7 dedicated pH/EC/TDS/temperature probe
• Probe quick connect system
• Battery life indication and low battery detection
• Auto‑off function
• Keystroke confirmation tone
• Waterproof casing IP67
1 Quick calibration tag 2 Stability indicator 3 Primary LCD 4 Battery percentage 5 Secondary LCD 6 Electrode condition indicator 7 TDS conversion factor 8 Low battery indicator 9 pH calibration buffer(s) used 10 Automatic Temperature Compensation indicator 11 Measurement unit 12 Temperature unit 13 Third LCD 14 Meter modes indicator
DISPLAY DESCRIPTION
6
Range*
‑2.00 to 16.00 pH ±825 mV (pH‑mV)
0.00 to 6.00 mS/cm (EC) ** 0 to 3000 ppm (0.5 CF)/0 to 3990 ppm (0.7 CF) ‑5.0 to 105.0 °C / 23.0 to 221.0 °F
Resolution
0.01 pH 1 mV (pH‑mV)
0.01 mS/cm 10 ppm (mg/L)
0.1 °C / 0.1 °F
Accuracy @ 20°C / 68°F
±0.02 pH ±1 mV (pH‑mV) ±2% F.S. ±2% F.S. ±0.5 °C / ±1.0 °F
Temperature Compensation
pH ‑ Automatic EC ‑ Automatic, with β=1.9%/ °C
pH Calibration
Automatic, 1 or 2 point, choose from:
4.01; 7.01; 10.0 pH buffers ‑ one‑point calibration using quick calibration solution
EC Calibration
Automatic, one‑point at 1.41 mS/cm or
5.00 mS/cm ‑ one‑point calibration using quick calibration solution
TDS CF***
0.5 (500 ppm) or 0.7 (700 ppm)
Probe (included)
HI1285‑7 pH/EC/TDS/temperature sensor, DIN
connector and 1 m (3.3’) cable
Battery type / life
1.5V AAA (3 pcs.) approx. 600 hours of continuous use
Auto‑Off
user selectable: after 8 min, 60 min or disabled
Environment
0 to 50 °C (32 to 122 °F) RH max. 100%
Meter Dimensions
154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Meter Mass (with batteries)
196 g (6.91 oz.)
Casing Ingress Protection Rating
IP67
*the pH range is limited from 0 to 13 pH and the temperature range from 0 to 50 °C (32 to 122 °F) using HI1285‑7 probe. **mS/cm is displayed as mS on the display. ***TDS Conversion Factor: 1000 μS/cm = 500 ppm with 0.5 CF.
SPECIFICATIONS
Loading...