Hanna Instruments HI 98129 User Manual

Instruction Manual
HI 98129 - HI 98130
Waterproof pH, EC/TDS &
temperature meters
This Instrument is in
www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 98129 and HI 98130 are warranted for one year against defects in workmanship and materials
when used for their intended purpose and maintained according to instructions. The electrode is war- ranted for a period of six months. This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be re­turned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, con­struction and appearance of its products without advance notice.
Compliance with CE
Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for a correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your Dealer or the nearest Hanna Customer Service Center.
The meters are supplied with:
• 4 x 1.5V batteries
HI 73127 pH electrode
HI 73128 Tool to remove the pH electrode
Note: Conserve all packing material until the instru-
ment has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI 98129 and HI 98130 are waterproof pH, EC/TDS and temperature meters. The housing has been com­pletely sealed against humidity and designed to float.
All pH and EC/TDS readings are automatically tem­perature compensated (ATC), and temperature values can be displayed in °C or °F units.
For EC/TDS readings, the EC/TDS conversion factor (CONV) is selectable by the user, as well as the temperature compensation coefficient β (BETA).
The meters can be calibrated at one or two points for pH (with auto-buffer recognition and against five memorized buffer values), and at one point for EC/ TDS.
Measurements are highly accurate with a unique stability indicator right on the LCD.
A low battery symbol warns the user when the batter­ies are to be replaced. In addition the Battery Error Prevention System (BEPS) avoids erroneous reading caused by low voltage level by turning the meter off.
The HI 73127 pH electrode, supplied with the meters, is interchangeable and can be easily replaced by the user.
The stainless steel encapsulated temperature sensor facilitates faster and more accurate temperature mea­surement and compensation.
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. Stability indicator
4. Low battery indicator
5. ON/OFF/MODE button
6. HI 73127 pH electrode
7. Temperature sensor
8. EC/TDS probe
9. SET/HOLD button
10. Secondary LCD
11. Primary LCD
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
RangeTemperature: 0.0 to 60.0°C or 32.0 to 140.0°F
HI 98129 pH: 0.00 to 14.00
HI 98130 pH: 0.00 to 14.00
Resolution 0.1°C or 0.1°F
HI 98129 0.01 pH; 1 µS/cm ; 1 ppm HI 98130 0.01 pH, 0.01 mS/cm ; 0.01 ppt
Accuracy (@20°C/68°F) Temperature ±0.5°C or ±1°F
Typical EMC Deviation Temperature ±0.5°C or ±1°F
Temp. Compensation pH: Automatic
EC/TDS: with β=0.0 to 2.4%/ºC
Environment 0 to 50°C (32 to 122°F); RH 100% EC/TDS Conversion Factor 0.45 to 1.00 (CONV) Calibration pH: at 1 or 2 points with 2 sets of memo
EC/TDS Calibration Solutions
HI 98129 HI7031 (1413 µS/cm)
HI 98130 HI7030 (12.88 mS/cm)
Electrode HI 73127 Battery Type/Life 4 x 1.5V with BEPS / typical 100 hours Auto-off After 8 min. Dimensions 163 x 40 x 26 mm (6.4 x 1.6 x 1.0”) Weight 85 g (3.0 oz)
Recommendations for Users Before using this product, make sure that it is entirely suit­able for the environment in which it is used. Operation of this instrument in residential areas could cause unaccept­able interferences to radio and TV equipment. The glass bulb at the end of the electrode is sensitive to elec­trostatic discharges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user to the supplied equip­ment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC. T o avoid damages or burns, do not perform any mea­surement in microwave ovens.
EC: 0 to 3999 µS/cm TDS: 0 to 2000 ppm
EC: 0.00 to 20.00 mS/cm TDS: 0.00 to 10.00 ppt
EC/TDS ±2% f.s. pH ±0.01
pH ±0.02 pH EC/TDS ±2% f.s.
rized buffers (pH 4.01/7.01/10.01 or pH 4.01/6.86/9.18)
EC/TDS: automatic, at 1 point
HI7032 (1382 ppm; CONV=0.5) HI70442 (1500 ppm; CONV=0.7)
HI70038 (6.44 ppt; CONV=0.5
or 9.02 ppt; CONV=0.7)
pH electrode (
includedincluded
included)
includedincluded
ISTR98129R2 01/02
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
ToTo
turnturn
thethe
metermeter
onon
andand
toto
checkcheck
batterybattery
To
turn
the
meter
on
and
to
ToTo
turnturn
thethe
metermeter
onon
Press and hold the the used segments on the LCD will be visible for a few seconds, followed by a percent indication of the remaining battery life. Eg. % 100 BATT.
ToTo
change the temperature unitchange the temperature unit
To
change the temperature unit
ToTo
change the temperature unitchange the temperature unit
To change the temperature unit (from °C to °F), from measurement mode, press and hold the MODE but­ton until TEMP and the current temperature unit are displayed on the lower LCD. Eg. TEMP °C.
Use the SET/HOLD button to change the temperature unit, and then press MODE button twice to return to normal measurement mode.
ToTo
freezefreeze
To
freeze
ToTo
freezefreeze
Press the SET/HOLD button for 2-3 seconds until HOLD appears on the secondary display.
Press either button to return to normal mode. ToTo
turnturn
To
turn
ToTo
turnturn
Press the mode. OFF will appear on the lower part of the display. Release the button.
NotesNotes
Notes:
NotesNotes
• Before taking any measurement make sure the meter has been calibrated.
• To clear a previous calibration, press the MODE button after entering the calibration mode. The lower LCD will display ESC for 1 second and the meter will return to normal measurement mode. The CAL symbol on the LCD will disappear. The meter will be reset to the default calibration.
• If measurements are taken in different samples successively, rinse the probe thoroughly to eliminate cross-contamination; and after cleaning, rinse the probe with some of the sample to be measured.
pp
H MEASUREMENTS & CALIBRATIONH MEASUREMENTS & CALIBRATION
p
H MEASUREMENTS & CALIBRATION
pp
H MEASUREMENTS & CALIBRATIONH MEASUREMENTS & CALIBRATION
Taking measurementsTaking measurements
Taking measurements
Taking measurementsTaking measurements Select the pH mode with the SET/HOLD button. Submerge the electrode in the solution to be tested. The measurements should be taken when the stability
symbol on the top left of the LCD disappears.
The pH value automatically com­pensated for temperature is shown on the primary LCD while the secondary LCD shows the temperature of the sample.
Calibration buffer setCalibration buffer set
Calibration buffer set
Calibration buffer setCalibration buffer set
• From measurement mode, press and hold the MODE button until TEMP and the current tempera­ture unit are displayed on the lower LCD. Eg. TEMP °C.
• Press the MODE button again to show the current buffer set: pH 7.01 BUFF (for 4.01/7.01/10.01 cali-
MODE button for 2-3 seconds. All
thethe
displaydisplay
the
display
thethe
displaydisplay
thethe
metermeter
meter
metermeter
offoff
off
offoff
the
thethe MODE button while in normal measurement
andand
check
toto
checkcheck
battery
batterybattery
statusstatus
status
statusstatus
bration) or pH 6.86 BUFF (for NIST 4.01/6.86/9.18 calibration).
• Press the SET/HOLD button to change the buffer value.
• Press the MODE button to return to normal mea­suring mode.
Calibration procedureCalibration procedure
Calibration procedure
Calibration procedureCalibration procedure From measurement mode, press and hold the MODE
button until CAL is displayed on the lower LCD. Release the button. The LCD will display pH 7.01 USE or pH 6.86 USE (if you have selected the NIST buffer set). The CAL tag blinks on the LCD.
single-point pH calibration, place the electrode
• For a in any buffer from the selected buffer set (eg. pH 7.01 or pH 4.01 or pH 10.01). The meter will recognize the buffer value automatically.
If using pH 4.01 or pH 10.01, the meter will display OK for 1 second and then return to measurement mode.
If using pH 7.01, after recognition of the buffer the meter will ask for pH 4.01 as second calibration point. Press the MODE button to return to measure­ment mode or, if desired, proceed with the 2-point calibration as explained below.
NoteNote
Note: It is always recommended to carry out a two-
NoteNote point calibration for better accuracy.
two-point pH calibration, place the electrode
• For a in pH 7.01 (or 6.86 if you have selected the NIST buffer set). The meter will recognize the buffer value and the n di sp lay pH 4.01 USE.
Rinse the electrode thoroughly to eliminate cross­contamination.
Place the electrode in the second buffer value (pH
4.01 or 10.01, or, if using NIST, pH 4.01 or 9.18). When the second buffer is recognized, the LCD will display OK for 1 second and the meter will return to normal measurement mode.
The CAL symbol on the LCD means that the meter is calibrated.
EC/TDSEC/TDS
MEASUREMENTS & CALIBRATION MEASUREMENTS & CALIBRATION
EC/TDS
MEASUREMENTS & CALIBRATION
EC/TDSEC/TDS
MEASUREMENTS & CALIBRATION MEASUREMENTS & CALIBRATION
Taking measurementsTaking measurements
Taking measurements
Taking measurementsTaking measurements Select either EC or TDS mode with the SET/HOLD
button. Submerge the probe in the solution to be tested. Use
plastic beakers to minimize any electromagnetic inter­ferences.
The measurements should be taken when the stability symbol on the top left of the LCD disappears.
The EC (or TDS) value automati­cally compensated for temperature is shown on the primary LCD while the secondary LCD shows the temperature of the sample.
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)
To change the EC/TDS conversion factor (CONV)To change the EC/TDS conversion factor (CONV) and the temperature compensation coefficient and the temperature compensation coefficient
and the temperature compensation coefficient
and the temperature compensation coefficient and the temperature compensation coefficient (BETA)(BETA)
(BETA)
(BETA)(BETA)
• From measurement mode, press and hold the MODE button until TEMP and the current tempera­ture unit are displayed on the lower LCD. Eg. TEMP °C.
• Press the MODE button again to show the current conversion factor. Eg. 0.50 CONV.
• Press the SET/HOLD button to change the conver­sion factor.
• Press the MODE button to show the current tem­perature compensation coefficient β. Eg. 2.1 BETA.
• Press the SET/HOLD button to change the tempera­ture compensation coefficient β.
• Press the MODE button to return to normal mea­suring mode.
Calibration procedureCalibration procedure
Calibration procedure
Calibration procedureCalibration procedure
• From measurement mode, press and hold the MODE button until CAL is displayed on the lower LCD.
• Release the button and immerse the probe in the proper calibration solution: HI7031 (1413 µS/cm) for HI98129 and HI7030 (12.88 mS/cm) for HI98130.
• Once the calibration has been automatically per­formed, the LCD will display OK for 1 second and the meter will return to normal measurement mode.
• Since there is a known relathionship between EC and TDS readings, it is not necessary to calibrate the meter in TDS
The CAL symbol on the LCD means that the meter is calibrated.
pH ELECTRODEpH ELECTRODE
pH ELECTRODE
pH ELECTRODEpH ELECTRODE
• When not in use, rinse the electrode with water to minimize contamination and store it with a few drops of storage (HI 70300) or pH 7 (HI 7007) solution in the protective cap after use. DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
• If the electrode has been left dry, soak in a storage or pH 7 solution for at least one hour to reactivate it.
• To prolong the life of the pH electrode, it is recommended to clean it monthly by immersing it in the HI 7061 cleaning solution for half an hour. Afterwards, rinse it thoroughly with tap water and recalibrate the meter.
• The pH electrode can be easily replaced by using the supplied tool (HI 73128). Insert the tool into the
probe cavity as shown below.
MAINTENANCE MAINTENANCE
MAINTENANCE
MAINTENANCE MAINTENANCE
Remove the probe by rotating it counterclockwise and
ββ
β
then pulling it out.
ββ
Insert a new pH electrode following the above instruc­tions in reverse order .
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
The meter displays the remaining battery percentage every time it is switched on. When the battery level is below 5%, the lights up to indicate a low battery condition. The batter­ies must be immediately replaced. If however the battery level is so low as to cause erroneous reading, the Battery Error Prevention System (BEPS) will automatically turn the meter off. T o change the batteries, remove the 4 screws located on the top of the meter .
Once the top has been removed, carefully replace the 4 batteries located in the compartment while paying atten­tion to their polarity. Replace the top, making sure that the gasket is properly seated in place, and tighten the screws.
HI 73127 Replaceable pH electrode HI 73128 Tool to remove the electrode HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs) HI 70006P pH 6.86 solution, 20 mL sachet (25 pcs) HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs) HI 70009P pH 9.18 solution, 20 mL sachet (25 pcs) HI 70010P pH 10.01 solution, 20 mL sachet (25 pcs) HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each) HI 7004M pH 4.01 solution, 230 mL bottle HI 7006M pH 6.86 solution, 230 mL bottle HI 7007M pH 7.01 solution, 230 mL bottle HI 7009M pH 9.18 solution, 230 mL bottle HI 7010M pH 10.01 solution, 230 mL bottle HI 70030P 12.88 mS/cm @25ºC calibration solu-
HI 70031P 1413 µS/cm @25ºC calibration solu- HI 70032P 1382 ppm @25ºC calibration solution, HI 70038P 6.44 ppt @25ºC calibration solution, 20 HI 70442P 1500 ppm @25ºC calibration solution, HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
- +
symbol on the bottom left of the LCD
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
tion, 20 mL sachet (25 pcs) tion, 20 mL sachet (25 pcs) 20 mL sachet (25 pcs) mL sachet (25 pcs) 20 mL sachet (25 pcs)
Loading...