Instruction Manual
HI 98127 - HI 98128
Waterproof
pH & Temperature Meters
These Instruments are
www.hannainst.com
WARRANTY
HI 98127 and HI 98128 are warranted for one
year against defects in workmanship and materials
when used for their intended purpose and maintained
according to instructions. The electrode is war-
ranted for a period of six months. This warranty is
limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack
of prescribed maintenance are not covered.
If service is required, contact the dealer from whom
you purchased the instrument. If under warranty,
report the model number, date of purchase, serial
number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of
the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer
Service department and then send it with shipment
costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in
part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket,
Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
in Compliance with
CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This
manual will provide you with the necessary information for correct operation. Please read it carefully
before using the meter.
If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the
directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material
and examine it carefully. If any damage has occurred
during shipment, immediately notify your Dealer or
the nearest Hanna Customer Service Center.
Each meter is supplied with:
• HI 73127 pH electrode
• HI 73128 Electrode removal tool
• 4 x 1.5V batteries
Note: Conserve all packing material until the instru-
ment has been observed to function correctly.
Any defective item must be returned in its
original packing.
US DESIGN PATENT
D462,024
GENERAL DESCRIPTION
HI 98127 and HI 98128 are waterproof pH and
temperature meters. The housing has been completely
sealed against humidity and designed to float.
All pH readings are automatically temperature compensated (ATC), and temperature values can be displayed
in °C or °F units.
The meters can be calibrated at one or two points with
auto-buffer recognition and against five memorized
buffer values.
Measurements are highly accurate with a unique stability indicator right on the LCD.
These meters are also provided with battery level
indication at start-up, and with a low battery symbol
which warns the user when the batteries need to be
replaced. In addition the Battery Error Prevention
System (BEPS) avoids erroneous reading caused by low
voltage level by turning the meter off.
The HI 73127 pH electrode, supplied with the meters,
is interchangeable and can be easily replaced.
The stainless steel encapsulated temperature sensor facilitates faster and more accurate temperature
measurement and compensation.
FUNCTIONAL DESCRIPTION
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. ON/OFF/MODE button
4. HI 73127 pH electrode
5. Temperature sensor
6. SET/HOLD button
1. Automatic temperature compensation
indicator
2. Stability indicator
3. Battery life percentage indicator
4. Low battery indicator
5. Secondary display
6. Primary display
7. Measuring unit for primary display
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is
entirely suitable for the environment in which it is
used.
The glass bulb at the end of the electrode is sensitive
to electrostatic discharges. Avoid touching the glass
bulb at all times.
Any variation introduced by the user to the supplied
equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument
when voltages at the measurement surface exceed
24 VAC or 60 VDC. To avoid damages or burns, do
not perform any measurement in microwave ovens.
ACCESSORIES
HI 73127 Replaceable pH electrode
HI 73128 Electrode removal tool
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70006P pH 6.86 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 70009P pH 9.18 solution, 20 mL sachet (25 pcs)
HI 70010P pH 10.01 solution, 20 mL sachet (25 pcs)
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)
HI 7004M pH 4.01 solution, 230 mL bottle
HI 7006M pH 6.86 solution, 230 mL bottle
HI 7007M pH 7.01 solution, 230 mL bottle
HI 7009M pH 9.18 solution, 230 mL bottle
HI 7010M pH 10.01 solution, 230 mL bottle
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
ISTR98127R4 03/03
SPECIFICATIONS
Range
Resolution
Accuracy
(@20°C/68°F)
Typical
EMC Deviation
Temp. Compensation
Environment
Calibration
Electrode
Battery Type/Life
Auto-off
Dimensions
Weight
-2.0 to 16.0 pH (HI 98127)
pH:
-2.00 to 16.00 pH (HI 98128)
-5.0 to 60.0°C / 23.0 to 140.0°F
Temp:
pH:
Temp:
pH:
Temp:
pH:
Temp:
0.1 pH (HI 98127)
0.01 pH (HI 98128)
0.1°C / 0.1°F
±0.1 pH (HI 98127)
±0.05 pH (HI 98128)
±0.5°C / ±1°F
±0.1 pH (HI 98127)
±0.02 pH (HI 98128)
±0.3°C / ±0.6°F
Automatic
-5 to 50°C (23 to 122°F); RH 100%
1 or 2 points with 2 sets of memorized
buffers (pH 4.01/7.01/10.01 or
4.01/6.86/9.18)
HI 73127 pH electrode (included )
4 x 1.5V with BEPS/typical 300 hours
After 8 minutes
163 x 40 x 26 mm (6.4 x 1.6 x 1.0”)
85 g (3.0 oz)
OPERATIONAL GUIDE
To turn the meter on and check the battery status
Press and hold the /MODE button until the LCD
lights up. All the used segments on the LCD will be
visible for 1 second (or as long as the button is
pressed), followed by the percent indication of the
remaining battery life (E.g. % 100 BATT).
Taking measurements
Submerge the electrode in the solution to be tested
while stirring it gently. The measurements should be
taken when the stability symbol on the top left of
the LCD disappears.
The pH value automatically compensated for temperature is
shown on the primary LCD
while the secondary LCD shows
the temperature of the sample.
To freeze the display
While in measurement mode,
press the SET/HOLD button.
HOLD appears on the secondary display and the reading
will be frozen on the LCD (E.g.
pH 5.78 HOLD).
Press any button to return to normal mode.
To turn the meter off
While in normal mode, press the
OFF will appear on the secondary display. Release
the button.
/MODE button.
Notes:
• Before taking any measurements make sure the
meter has been calibrated (CAL tag present on the
LCD).
• If measurements are taken in different samples
successively, rinse the probe thoroughly to eliminate
cross-contamination; and after cleaning, rinse the
probe with some of the sample to be measured.
CALIBRATION
For better accuracy, frequent calibration of the instrument is recommended. In addition, the instrument
must be recalibrated whenever:
a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.
Calibration procedure
From normal measuring mode, press and hold the
/MODE button until OFF on the secondary LCD is
replaced by CAL. Release the button. The LCD enters
the calibration mode displaying “pH 7.01 USE” (or
“pH 6.86 USE” if the NIST buffer set was selected).
After 1 second the meter activates the automatic buffer
recognition feature. If a valid buffer is detected then its
value is shown on the primary display and REC
appears on the secondary LCD. If no valid buffer is
detected, the meter keeps the USE indication active for
12 seconds, and then it replaces it with WRNG,
indicating the sample being measured is not a valid
buffer.
single-point calibration with buffers pH 4.01,
• For a
9.18 or 10.01, the meter automatically accepts the
calibration when the reading is stable; the meter
displays the accepted buffer, with the message “OK
1”. After 1 second the meter automatically returns to
the normal measuring mode.
If a single-point calibration with buffer pH 7.01 (or
pH 6.86) is desired, then after the calibration point
has been accepted the
pressed in order to return to normal mode. After the
button is pressed, the meter shows “7.01” (or “6.86”)
- “OK 1” and, after 1 second, it automatically returns
to the normal measuring mode.
Note: It is always recommended to carry out a twopoint calibration for better accuracy.
two-point calibration, place the electrode in
• For a
pH 7.01 (or pH 6.86) buffer. After the first calibration
point has been accepted, the “pH 4.01 USE” message
appears. The message is held for 12 seconds, unless a
valid buffer is recognized. If no valid buffer is recognized, then the WRNG message is shown. If a valid
buffer (pH 4.01, pH 10.01, or pH 9.18) is detected,
then the meter completes the calibration procedure.
When the buffer is accepted, the LCD shows the
accepted value with the “OK 2” message, and then
the meter returns to the normal measuring mode.
Note: When the calibration procedure is completed,
the CAL tag is turned on.
/MODE button must be
To quit calibration and to reset to the default
values
• After entering the calibration mode and before the
first point is accepted, it is possible to quit the
procedure and return to the last calibration data by
pressing the
displays “ESC” for 1 second and the meter returns to
the normal measuring mode.
• To reset to the default values and clear a previous
calibration, press the SET/HOLD button after entering
the calibration mode and before the first point is
accepted. The secondary LCD displays “CLR” for 1
second, the meter resets to the default calibration and
the CAL tag on the LCD disappears.
/MODE button. The secondary LCD
SETUP
Setup mode allows the selection of temperature unit
and pH buffer set.
To enter the Setup mode, press the
until CAL on the secondary display is replaced by
TEMP and the current temperature unit (E.g. TEMP
°C). Then:
• for °C/°F selection: Use the SET/HOLD button.
After the temperature unit has been selected, press the
/MODE button
/MODE button to enter the buffer set selection
mode; press the
the normal measuring mode.
to change the calibration buffer set: After setting
•
the temperature unit, the meter will show the current
buffer set: “pH 7.01 BUFF” (for 4.01/7.01/10.01) or
“pH 6.86 BUFF” (for NIST 4.01/6.86/9.18).
/MODE button twice to return to
Change the set with the SET/HOLD button, then press
/MODE to return to normal measuring mode.
pH ELECTRODE MAINTENANCE
• When not in use, rinse the electrode with water to
minimize contamination and store it with a few drops
of HI 70300 storage solution in the protective cap.
DO NOT USE DISTILLED OR DEIONIZED WATER
FOR STORAGE PURPOSES.
• If the electrode has been left dry, soak in storage
solution for at least one hour to reactivate it.
• To prolong the life of the pH electrode, it is
recommended to clean it monthly by immersing it in
the HI 7061 cleaning solution for half an hour.
Afterwards, rinse it thoroughly with tap water and
recalibrate the meter.
• The pH electrode can be easily replaced by using the
supplied tool (HI 73128). Insert the tool into the electrode cavity as shown below.
Rotate the electrode counterclockwise.
Pull the electrode out by using the other side of the tool.
Insert a new pH electrode following the above instructions in reverse order.
BATTERY REPLACEMENT
The meter displays the remaining battery percentage
every time it is switched on. When the battery level is
below 5%, the
lights up to indicate a low battery condition. The batteries should be replaced soon. If the battery level is low
enough to cause erroneous readings, the meter shows
“0%” and the Battery Error Prevention System (BEPS) will
automatically turn the meter off.
To change the batteries, remove the 4 screws located on
the top of the meter.
Once the top has been removed, carefully replace the 4
batteries located in the compartment while paying attention to their polarity.
Replace the top, making sure that the gasket is properly
seated in place, and tighten the screws to ensure a
watertight seal.
- +
symbol on the bottom left of the LCD