Instruction Manual
HI 98120
Waterproof
ORP & Temperature Meter
This Instrument is
www.hannainst.com
WARRANTY
HI 98120 is warranted for one year against defects
in workmanship and materials when used for its
intended purpose and maintained according to instructions. The electrode is warranted for a period
of six months. This warranty is limited to repair or
replacement free of charge.
Damages due to accidents, misuse, tampering or lack
of prescribed maintenance are not covered.
If service is required, contact the dealer from whom
you purchased the instrument. If under warranty,
report the model number, date of purchase, serial
number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of
the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer
Service department and then send it with shipment
costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in
part is prohibited without the written consent of the
copyright owner, Hanna Instruments Inc., Woonsocket,
Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
in Compliance with
CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This
manual will provide you with the necessary information for correct operation. Please read it carefully
before using the meter.
If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the di rectives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material
and examine it carefully. If any damage has
occurred during shipment, immediately notify your
Dealer or the nearest Hanna Customer Service
Center.
Each meter is supplied with:
• HI 73120 ORP electrode
• HI 73128 Electrode removal tool
• 4 x 1.5V batteries
Note: Conserve all packing material until the in-
strument has been observed to function
correctly. Any defective item must be returned in its original packing.
US DESIGN PATENT
D462,024
GENERAL DESCRIPTION
HI 98120 is a waterproof ORP and temperature
meter. The housing has been completely sealed
against humidity and designed to float.
The temperature values can be displayed in °C or
°F units.
The meter is factory calibrated and measurements
are highly accurate with a unique stability indicator
right on the LCD.
This meter is also provided with battery level
indication at start-up, and with a low battery symbol which warns the user when the batteries need
to be replaced. In addition the Battery Error Prevention System (BEPS) avoids erroneous reading caused
by low voltage level by turning the meter off.
The HI 73120 ORP electrode, supplied with the
meter, is interchangeable and can be easily replaced by the user.
The stainless steel encapsulated temperature sensor facilitates faster and more accurate temperature
measurement.
FUNCTIONAL DESCRIPTION
1. Battery compartment
2. Liquid Crystal Display (LCD)
3. ON/OFF/MODE button
4. HI 73120 ORP electrode
5. Temperature sensor
6. SET/HOLD button
1. Stability indicator
2. Battery life percentage indicator
3. Low battery indicator
4. Secondary display
5. Primary display
6. Measuring unit for primary display
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is
entirely suitable for the environment in which it is
used.
Avoid touching the electrode at all times.
Any variation introduced by the user to the supplied
equipment may degrade the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument
when voltages at the measurement surface exceed
24 VAC or 60 VDC. To avoid damages or burns, do
not perform any measurement in microwave ovens.
ISTR98120 03/03
SPECIFICATIONS
Range
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Environment
Calibration
Electrode
Battery Type
Life
Auto-off
Dimensions
Weight
ORP:
Temp:
ORP:
Temp:
ORP:
Temp:
ORP:
Temp:
± 1000 mV
-5.0 to 60.0°C/23.0 to 140.0°F
1 mV
0.1°C / 0.1°F
±2 mV
±0.5°C / ±1°F
±2 mV
±0.3°C / ±0.6°F
-5 to 50°C (23 to 122°F);
RH 100%
factory calibrated
HI 73120 ORP electrode
(included)
4 x 1.5V with BEPS
typical 300 hours
After 8 minutes
163 x 40 x 26 mm
(6.4 x 1.6 x 1.0”)
85 g (3.0 oz)
OPERATIONAL GUIDE
To turn the meter on and to check battery status
Press and hold the /MODE button until the LCD
lights up. All the used segments on the LCD will
be visible for 1 second (or as
long as the button is pressed),
followed by the percent indication of the remaining battery
life (E.g. % 100 BATT).
Taking measurements
Submerge the electrode in the solution to be
tested while stirring it gently. The measurements
should be taken when the stability symbol on
the top left of the LCD disappears.
The ORP (mV) value is shown
on the primary LCD while the
secondary LCD shows the temperature of the sample.
To freeze the display
While in measurement mode,
press the SET/HOLD button.
HOLD appears on the secondary display and the reading
will be frozen on the LCD (E.g.
mV 825 HOLD).
Press any button to return to normal mode.
To turn the meter off
While in normal measuring mode, press the /
MODE button. OFF will appear on the secondary
display. Release the button.
Note:
If measurements are taken in different samples
successively, rinse the probe thoroughly to eliminate cross-contamination; and after cleaning, rinse
the probe with some of the sample to be measured.
SETUP
Setup mode allows the selection of temperature
unit.
To enter the Setup mode, °C/°F selection, press
the /MODE button until TEMP and the current
temperature unit are displayed on the lower part
of the LCD (E.g. TEMP °C).
Use the SET/HOLD button to change the temperature unit, and then press the /MODE button
to return to the normal measuring mode.
BATTERY REPLACEMENT
The meter displays the remaining battery percentage every time it is switched on. When the battery
level is below 5%, the
left of the LCD lights up to indicate a low battery
condition. The batteries should be replaced soon.
If the battery level is low enough to cause erroneous readings, the meter shows “0%” and the Battery
Error Prevention System (BEPS) will automatically
turn the meter off.
To change the batteries, remove the 4 screws located on the top of the meter.
Once the top has been removed, carefully replace
the 4 batteries located in the compartment while
paying attention to their polarity.
- +
symbol on the bottom
Replace the top, making sure that the gasket is properly seated in place, and tighten the screws to ensure
a watertight seal.
FACTORY RECALIBRATION
The ORP range is factory calibrated.
Contact your nearest Hanna Service Center for
recalibration, if necessary.
ELECTRODE MAINTENANCE
• When not in use, rinse the electrode with water
to minimize contamination and store it with a few
drops of HI 70300 storage solution in the protective cap after use. DO NOT USE DISTILLED
OR DEIONIZED WATER FOR STORAGE PURPOSES.
• If the electrode has been left dry, soak in
storage solution for at least one hour to reactivate it.
• To prolong the life of the electrode, it is recommended to clean it monthly by immersing it in the
HI 7061 cleaning solution for half an hour.
Afterwards, rinse it thoroughly with tap water.
• The ORP electrode can be easily replaced by
using the supplied tool (HI 73128). Insert the tool
into the electrode cavity as shown below.
Rotate the electrode counterclockwise.
Pull the electrode out by using the other side of the
tool.
Insert a new ORP electrode following the above
instructions in reverse order.
ACCESSORIES
HI 73120 Replaceable ORP electrode
HI 73128 Electrode removal tool
HI 7021M ORP test solution (240 mV), 230 mL bottle
HI 7022M ORP test solution (470 mV), 230 mL bottle
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
HI 7091M Pretreatment reducing solution, 230 mL
HI 7092M Pretreatment oxidizing solution, 230 mL
bottle
bottle
HANNA WATERPROOF TESTERS
Hanna offers a complete line of waterproof,
microprocessor-based testers for measurement of
pH, Temperature, EC, TDS and ORP.
Choose the instruments which best meet your
needs:
• HI 98120 ORP (mV) and Temperature (°C
& °F)
• HI 98121 pH (0.01 resolution), ORP (mV)
and Temperature (°C & °F)
• HI 98127 pH (0.1 resolution) and Temperature (°C & °F)
• HI 98128 pH (0.01 resolution) and Temperature (°C & °F)
• HI 98129 pH (0.01 resolution), EC/TDS
Low Range (µS/cm & ppm) and
Temperature (°C & °F)
• HI 98130 pH (0.01 resolution), EC/TDS
High Range (mS/cm & ppt) and
Temperature (°C & °F)
• HI 98311 EC/TDS Low Range (µS/cm &
ppm) and Temperature (°C & °F)
• HI 98312 EC/TDS High Range (mS/cm &
ppt) and Temperature (°C & °F)