OPERATION
• Turn the meter on and
select the measurement
mode: pH or EC/TDS.
• Dip the probe in the
sample to be measured. Under no circumstances the immersion shall be
above the max. level
indicated above. Stir
gently until the display
shows a stable reading. The EC/TDS displayed value is referenced to 25°C (77°F).
• When testing a new sample, it is
recommended to rinse the probe
with some of the sample itself prior
to measurement. This prevents likelihood of cross-contamination.
• When not in use,
switch the
AGRITESTAGRITEST
AGRITEST
AGRITESTAGRITEST
™™
™
™™
off.
pH CALIBRATION
• Turn the meter on and
set the switch to the
pH mode.
• Dip the probe in a
pH 7.01 buffer solution (HI 771P or
HI 772P). Allow the
reading to stabilize
and using a small
screwdriver slowly
turn the pH Calibration Trimmer to read
“7.0”.
EC/TDS CALIBRATION
• Turn the meter on and
set the switch to the
ppm/mS mode.
• Dip the probe in an EC calibration
solution (HI 771P) for
or in a TDS calibration solution
(HI 772P) for
••
• Wait for a couple of
••
AGRITESTAGRITEST
AGRITEST
AGRITESTAGRITEST
AGRITESTAGRITEST
AGRITEST
AGRITESTAGRITEST
™™
™
™™
1. 1.
1.
1. 1.
minutes for the display to stabilize and,
if necessary, correct
the reading (to “1.41”
AGRITESTAGRITEST
for
AGRITEST
AGRITESTAGRITEST
“1500” for
™™
™
™™
TESTTEST
1) 1)
TEST
1) by slowly
TESTTEST
1) 1)
™™
™
™™
or
AGRI-AGRI-
AGRI-
AGRI-AGRI-
turning the ppm/mS
Calibration Trimmer
with a small screwdriver provided.
MAINTENANCE
If the probe remains dry for a long
time, it is necessary to reactivate it
by immersing it in an HI70300 stor-
age solution for 4 hours before use.
The probe should also be rinsed
clean with plenty of
water after use.
If you need to replace
the electrode, use an
hexagon key to remove
the screw on the side
of the probe before
detaching it from the
meter.
BATTERY REPLACEMENT
When the unit cannot be switched
on or the display segments fade, it
is necessary to replace all 3 x 1.4V
batteries.
Batteries should only be replaced in
a safe area using the battery type
specified in this instruction manual.
Loosen the upper
screw and slide off
the cover on the back
of the meter. Replace
all three batteries while
paying attention to their
polarity. Reinsert the
cover and tighten the screw.
It is recommended to recalibrate the
instrument whenever the batteries
are replaced.
RECOMMENDATIONS FOR USERS
Before using these products, ensure that they are
entirely suitable for the environment in which they
are used. Operation of these instruments in residential area could cause unacceptable interference
to radio and TV equipments, requiring the operator
to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied
equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use
these instruments when voltage at the measurement surface exceeds 24VAC or 60VDC. To avoid
damage or burns, do not perform any measurement in microwave ovens.
™™
™
™™
,