Hanna Instruments HI 98114 User Manual

Watercheck
pH and Conductivity Meter
Watercheck™1
pH and TDS Meter
OPERATION
• Turn the meter on and select the measurement mode: pH or Conduc­tivity/TDS.
• Dip the probe in the sample to be mea­sured. Under no cir­cumstances the im­mersion shall be above the max. level indicated above. Stir gently until the display shows a stable read­ing. The Conductivity/ TDS displayed value is referenced to 25°C (77°F).
• When testing a new sample, it is recommended to rinse the probe with some of the sample itself prior to measurement. This prevents like­lihood of cross-contamination.
• When not in use, switch the Watercheck™ off.
pH CALIBRATION
• Turn the meter on and set the switch to the pH mode.
• Dip the probe in a pH 7.01 buffer solu­tion (HI 771P or HI 773P). Allow the dis­play to stabilize and using a small screw­driver slowly turn the pH Calibration Trimmer to read “7.0”.
CONDUCTIVITY/TDS CALIBRATION
• Turn the meter on and set the switch to the µS/ppm mode.
• Dip the probe in a conductivity cali­bration solution (HI 771P) for
tion solution (HI 773P) for
, or in a TDS calibra-
1.
Wait for a couple of minutes for the display to stabilize and, if nec­essary, correct the reading to ("1410" for Watercheck™ or "1380" for Watercheck™ 1) by slowly turning the ppm/ mS Calibration Trimmer with a small screw­driver.
MAINTENANCE
If the probe remains dry for a long time, it is necessary to reactivate it by immersing it in an HI 70300 stor­age solution for 4 hours before use. The probe should also be rinsed with plenty of water after use.
If you need to replace the electrode, use an hexagon key to remove the screw on the side of the probe before detaching it from the meter.
BATTERY REPLACEMENT
When the unit cannot be switched on or the display segments fade, it is necessary to replace all 3 x 1.5V batteries. Batteries should only be replaced in a safe area using the battery type specified in this instruc­tion manual. Loosen the upper screw and slide off the cover on the back of the meter. Re­place all three batter­ies while paying atten­tion to their polarity. Reinsert the cover and tighten the screw.
It is recommended to recalibrate the instrument whenever the batteries are replaced.
RECOMMENDATIONS FOR USERS
Before using these products, ensure that they are entirely suitable for the environment in which they are used. Operation of these instruments in resi­dential area could cause unacceptable interference to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC per­formance. To avoid electrical shock, do not use these instruments when voltage at the measure­ment surface exceeds 24VAC or 60VDC. To avoid damage or burns, do not perform any measure­ment in microwave ovens.
Loading...
+ 1 hidden pages