Hanna Instruments HI 98111 User Manual

PICCOLO
®
PICCOLO® 2 PICCOLO® plus
ATC pH Testers
SPECIFICATIONS
Range 1.00 to 13.00 pH
Resolution 0.01 pH 0.01 pH 0.01 pH / 0.1°C Accuracy (@20°C) ±0.01 pH ±0.01 pH ±0.01 pH / ±1°C Calibration Manual, 2 point, through Offset & Slope trimmers Temperature Compensation
Probe (included) HI 1280 HI 1290 HI 1295 Battery Type / Life 3 x 1.5V alkaline / approx. 100 hours Environment 0 to 50ºC (32 to 122ºF); RH max 95% Dimensions 194x29x15 265x29x15 265x29x15 mm
(with probe) (7.6x1.1x0.6") (10.4x1.1x0.6") (10.4x1.1x0.6") Weight 70 g (2.5 oz.)
Each Piccolo® model is supplied complete with pH electrode, pH 4.01 and pH 7.01 buffer sachets, calibration screwdriver, rugged carrying case and instructions.
ACCESSORIES HI 1280 Spare pH electrode for Piccolo® (90 mm/3.5") HI 1290 Spare pH electrode for Piccolo®2 (160 mm/6.3") HI 1295 Spare pH electrode for Piccolo®plus (160 mm/6.3") HI 70004P pH 4.01 buffer solution, 20 mL sachet, 25 pcs HI 70007P pH 7.01 buffer solution, 20 mL sachet, 25 pcs HI 70010P pH 10.01 buffer solution, 20 mL sachet, 25 pcs HI 7004M pH 4.01 buffer solution, 230 mL bottle HI 7007M pH 7.01 buffer solution, 230 mL bottle HI 7010M pH 10.01 buffer solution, 230 mL bottle HI 70300M Electrode storage solution, 230 mL bottle HI 7061M Electrode cleaning solution, 230 mL bottle HI 731326 Calibration screwdriver (20 pcs)
®
PICCOLO®2 PICCOLO®+
-- -- 0.0 to 70.0°C
Automatic, 0 to 70ºC (32 to 150°F)
www.hannainst.com
Piccolo is a registered Trademark of
Hanna Instruments
ISTPICCOLO 05/05
OPERATION
• Connect the electrode to the meter as shown in the drawing, and turn the meter on.
Note: The PICCOLO® plus model
is provided with an OFF/pH/°C switch. Select the "pH" position for pH readings, or "°C" for tem­perature measurements.
• Remove the protective cap and dip the electrode in the sample to be measured, without exceed­ing the maximum immersion level. Stir gently and wait for the reading for stabilize.
• The pH value automatically com­pensated for temperature varia­tions will be displayed.
• After use, always switch the meter off.
MAINTENANCE
• To reactivate a dry electrode, immerse it in HI 70300 storage solution for 4 hours.
• After use, rinse the electrode with water and store it with a few drops of storage solution in the protective cap.
CALIBRATION
• To access the calibration trimmers, slide down the probe. Then turn the meter on.
• Dip the probe in HI 7007 (pH 7.01) buffer solution and allow the reading to stabilize. With the supplied screwdriver, adjust the "pH7" trimmer to read "7.01" pH.
• Rinse the electrode with water and dip it in the second buffer. Use HI 7004 (pH 4.01) buffer if the solution to be tested is acidic, or
HI 7010 (pH 10.01) for alkaline samples.
• Allow the reading to stabilize, then adjust the "pH 4/10" trimmer to read "4.01" or "10.01" pH.
• The calibration is now complete and the electrode can be slided back to its normal position.
BATTERY REPLACEMENT
When the meter cannot be switched on or the display fades, batteries should be replaced.
Unscrew the upper screw and slide off the battery compart­ment. Replace all 3 batteries while pay­ing attention to the correct polarity.
Reinsert the cover and tighten the screw. After battery replacement, it is always recommended to recalibrate the instrument.
Batteries should only be replaced in a safe area and using the bat­tery type specified in this instruc­tion manual.
WARRANTY
These meters are guaranteed for one year against de- fects in workmanship and materials when used for their intended purpose and maintained according to instruc­tions. Electrodes are guaranteed for six months. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model num­ber, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
Recommendations for Users
Before using these products, ensure that they are entirely suitable for the environment in which they are used. Op­eration of these instruments in residential areas could cause unacceptable interference to radio and TV equip­ment. The glass bulb at the end of the electrode is sen­sitive to electrostatic discharges. Avoid touching this glass bulb at all times. During operation of the instrument, ESD wrist straps should be worn to avoid possible damage to the pH electrode by electrostatic discharges. Any varia­tion introduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use these instruments when volt­age at the measurement surface exceeds 24Vac or 60Vdc. To avoid damage or burns, do not perform any measure­ment in microwave ovens.