Hanna Instruments HI 98111 User Manual

PICCOLO
ATC pH Meter
OPERATIONOPERATION
OPERATION
OPERATIONOPERATION
• Connect the electrode to the meter with both writings (meter and elec­trode) facing upwards and turn the meter on.
•Dip the probe in the sample to be measured. Under no circum­stances the immersion shall be above the max. level indicated above. Stir gently until the dis­play shows a stable reading. PICCOLO will automatically compen­sate for the temperature variation.
•When testing a new sample, it is recommended to rinse the probe with some of the sample itself prior to measurement. This prevents like­lihood of cross-contamination.
•When not in use, switch the PIC­COLO off.
CALIBRATIONCALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
•Slide down the probe to access the calibration trimmers and turn the meter on.
•Dip the probe in a pH 7.01 buffer solution
HI 7007HI 7007
(
HI 7007). Allow the
HI 7007HI 7007 display to stabilize and using a small screwdriver slowly turn the pH 7 Cali­bration Trimmer to read “7.01”.
• To perform a two point calibration, if you are measuring acidic liquids use a pH 4.01 buffer (
HI 7004HI 7004
HI 7004),
HI 7004HI 7004 and for alkaline mea­surements use a pH 10.01 buffer
HI 7010HI 7010
(
HI 7010).
HI 7010HI 7010
• Rinse the electrode in water and dip it in the second buffer. Allow the reading to stabi­lize and using a small screwdriver turn the pH 4/10 Calibration Trimmer to read “4.01” or “10.01” re­spectively.
• The calibration is now complete. Return the probe to its normal position.
MAINTENANCEMAINTENANCE
MAINTENANCE
MAINTENANCEMAINTENANCE
If the probe remains dry for a long time, it is necessary to reactivate it by immersing it in a storage solution
HI 70300HI 70300
(
HI 70300) for 4 hours before use.
HI 70300HI 70300 The probe should be rinsed with plenty of water after use.
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
When PICCOLO cannot be switched on or the display segments fade, it is necessary to replace all 3 x 1.4V batteries.
Batteries should only be replaced in a safe area using the battery type specified in this instruction manual.
Loosen the upper screw and slide off the cover on the back of the meter. Replace all three batteries while paying attention to their polarity.
Reinsert the cover and tighten the screw.
It is recommended to recalibrate the instrument whenever the batteries are replaced.
Recommendations for UsersRecommendations for Users
Recommendations for Users
Recommendations for UsersRecommendations for Users
Before using this product, ensure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause unacceptable interference to radio and TV equipment. The glass bulb at the end of the elec­trode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times. During opera­tion of the instrument, ESD wrist straps should be worn to avoid possible damage to the pH elec­trode by electrostatic discharges. Any variation in­troduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid electrical shock, do not use these instru­ments when voltage at the measurement surface exceeds 24VAC or 60VDC. To avoid damage or burns, do not perform any measurement in micro­wave ovens.
Loading...
+ 1 hidden pages