Hanna Instruments HI 98107 User Manual

pHep® is a rugged pH tester with replaceable
pHep® is in compliance with the CE direc-
tives.
SPECIFICATIONS Range 0.0 to 14.0 pH Resolution 0.1 pH Accuracy (@20°C/68°F) ±0.1 pH Typical EMC Deviation ±0.1 pH Calibration Manual, 2 point, through
offset and slope trimmers
Environment 0 to 50°C (32 to 122°F);
RH max 95%
Battery Type 4 x 1.5V alkaline Battery Life approx. 700 hours of use Dimensions 175 x 41 x 23 mm
(6.9 x 1.6 x 0.9")
Weight 95 g (3.4 oz.)
pHep
®
HI 98107
Pocket-sized pH Meter
ACCESSORIES HI 73106 Spare electrode HI 7004M pH 4.01 solution, 230 mL bottle HI 7007M pH 7.01 solution, 230 mL bottle HI 7010M pH 10.01 solution, 230 mL HI 70300M Storage solution, 230 mL bottle HI 7061M Cleaning solution, 230 mL bottle HI 731326 Calibration screwdriver (20 pcs)
pHep is a registered Trademark of
Hanna Instruments
IST98107 R5 05/05
www.hannainst.com
CONDITIONING
• Remove the cap.
• To activate the electrode, immerse it in HI70300 storage solution for 2 hours.
CALIBRATION
• Immerse the electrode in pH 7.01 (HI7007) buffer solution.
• Allow the reading to stabilize and with the supplied screwdriver, ad­just the “pH7” calibration trimmer to read “7.0” pH.
• Rinse the electrode with tap water and immerse it into the second buffer.
• Use pH 4.01 (HI7004) if the solution to be measured is acidic (pH < 7), or pH 10.0 (HI7010) for alka­line samples (pH > 7).
• Allow the reading to stabilize and with the screwdriver adjust the “pH4/10” calibration trimmer to read “4.0” or “10.0” pH.
OPERATION
• Remove the protective cap, turn the
®
pHep
on, by sliding the ON/OFF
switch on the top of the meter.
• Immerse it in the solution to be tested without exceeding the maxi­mum immersion level.
• Stir gently and wait for the reading to stabilize.
• After use, switch the meter off, rinse the electrode with tap water and store it with a few drops of storage (HI70300) or pH7 (HI7007) solution in the protec­tive cap. Always replace the protective cap after use.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STOR­AGE PURPOSES.
• Large differences in pH readings (±1 pH) could be due to lack of calibration or dry electrode.
MAINTENANCE
The sensor should be kept moist at all times. For this purpose, always keep wet the tissue inside the protective cap. In case of erroneous readings even after an accurate conditioning and calibration, the reference junction might be contaminated or clogged.
Pull out 2 mm (1/8'') of the cloth junction to renew the electrode refer­ence (it is recommended to cut the cloth leaving always at least 2 mm ­1/8'' over the reference compartment) and recalibrate the meter.
The cloth junction can be pulled out approximately 20 times. After that, the electrode should be replaced.
ELECTRODE REPLACEMENT
For replacing the electrode, contact your Dealer or the nearest Hanna Service Center.
BATTERY REPLACEMENT
If calibration can not be performed or the display fades, batteries must be replaced. Slide off the battery compart­ment cover and replace all four 1.5V alkaline batteries, while paying attention to the correct polarity.
Batteries should only be replaced in a non-hazardous area and using the battery type specified in this instruction manual.
WARRANTY
This meter is guaranteed for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This war­ranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is re­quired, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the fail­ure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service depart­ment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.