Hanna Instruments HI 98106 User Manual

®®
®
ChampChamp
Champ
ChampChamp
®®
HI 98106
Pocket-sized pH Meter
Manufacturers since 1978
Champ is a rugged pH tester with replaceable electrode and renewable reference junction. The instrument can be calibrated at one point by the user, simply acting on the calibration trimmer.
CONDITIONING:CONDITIONING:
CONDITIONING:
CONDITIONING:CONDITIONING:
• Remove the cap.
• To activate the electrode, immerse
ChampChamp
Champ in HI70300 storage
ChampChamp
solution for 2 hours.
CALIBRATION:CALIBRATION:
CALIBRATION:
CALIBRATION:CALIBRATION:
ChampChamp
• Immerse the
7.01 (HI7007) buffer solution.
• Allow the reading to stabilize and with a small screwdriver turn the Calibration Trimmer until the display shows "7.0"
The calibration is now complete. NoteNote
Note: the calibration can be performed
NoteNote
at pH 4.0, pH 7.0 or pH 10.0, according to the different applications.
OPERATION:OPERATION:
OPERATION:
OPERATION:OPERATION:
• Remove the protective cap.
• Turn the
• Immerse it in the solution to be
• Stir gently and wait for the reading
• After use, rinse the electrode with water to
• Store the electrode with a few
• Always replace the protective cap after use.
ChampChamp
Champ on with the ON/
ChampChamp
OFF switch located on the top of the meter.
tested without exceeding the maxi­mum immersion level.
to stabilize.
minimize contamination.
drops of storage (HI70300) or pH7 (HI7007) solution in the protective cap.
Champ
ChampChamp
in pH
DO NOT USE DISTILLED OR DEIONIZED WATER
FOR STORAGE PURPOSES.
• Large differences in pH readings (±1 pH) could be due to lack of calibration or dry electrode.
MAINTENANCE:MAINTENANCE:
MAINTENANCE:
MAINTENANCE:MAINTENANCE:
The sensor should be kept moist at all times. For this purpose, always keep wet the tissue inside the protective cap. In case of erroneous readings even after an accurate conditioning and calibration, the reference junction might be contaminated or clogged.
Pull out 2 mm (1/8'') of the cloth junction to renew the electrode reference (it is recommended to cut the cloth leaving always at least 2 mm -1/8'' over the refer­ence compartment) and re­peat the calibration proce­dure. The cloth junction can be pulled out approximately 20 times. After that, the electrode can be replaced (see the Electrode Replacement section).
BATTERY REPLACEMENT:BATTERY REPLACEMENT:
BATTERY REPLACEMENT:
BATTERY REPLACEMENT:BATTERY REPLACEMENT:
When display fades the batteries must be replaced. Slide off the battery compartment cover and replace all four 1.5V batteries while paying attention to their polarity.
Batteries should only be replaced in a nonhazardous area using the battery type specified in this instruction manual.
I 70300
H
®®
®
ChampChamp
Champ
ChampChamp
®®
HI 98106
Pocket-sized pH Meter
SPECIFICATIONS:SPECIFICATIONS:
SPECIFICATIONS:
SPECIFICATIONS:SPECIFICATIONS:
RANGERANGE
RANGE 0.0 to 14.0 pH
RANGERANGE RESOLUTIONRESOLUTION
RESOLUTION 0.1 pH
RESOLUTIONRESOLUTION ACCURACY ACCURACY
(@20C/68F)(@20C/68F)
(@20C/68F)
ACCURACY
(@20C/68F)(@20C/68F)
ACCURACY ACCURACY TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION
TYPICAL EMC DEVIATION ±0.1 pH
TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION CALIBRATIONCALIBRATION
CALIBRATION Manual, at 1 point through trimmer
CALIBRATIONCALIBRATION ENVIRONMENTENVIRONMENT
ENVIRONMENT 0 to 50C
ENVIRONMENTENVIRONMENT
BATTERY TYPEBATTERY TYPE
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY TYPEBATTERY TYPE BATTERY LIFEBATTERY LIFE
BATTERY LIFE approx. 800 hours
BATTERY LIFEBATTERY LIFE
DIMENSIONSDIMENSIONS
DIMENSIONS 175 x 41 x 23 mm
DIMENSIONSDIMENSIONS
WEIGHTWEIGHT
WEIGHT 78 g (2.7 oz.)
WEIGHTWEIGHT
ACCESSORIES:ACCESSORIES:
ACCESSORIES:
ACCESSORIES:ACCESSORIES:
HI 73106HI 73106
HI 73106 Spare electrode
HI 73106HI 73106 HI 77400PHI 77400P
HI 77400P pH 4.01 & 7.01 calibration kit
HI 77400PHI 77400P
(5 x 20 mL, each)
HI 77700PHI 77700P
HI 77700P pH 7.01 calibration kit
HI 77700PHI 77700P
(10 x 20 mL)
HI 770710PHI 770710P
HI 770710P pH 7.01 & 10.01 calibration kit
HI 770710PHI 770710P
(5 x 20 mL, each)
HI 7004MHI 7004M
HI 7004M pH 4.01 buffer solution
HI 7004MHI 7004M
(230 mL bottle)
HI 7007MHI 7007M
HI 7007M pH 7.01 buffer solution
HI 7007MHI 7007M
(230 mL bottle)
HI 7010MHI 7010M
HI 7010M pH 10.01 buffer solution
HI 7010MHI 7010M
(230 mL bottle)
HI 70300MHI 70300M
HI 70300M Storage solution (230mL)
HI 70300MHI 70300M HI 7061MHI 7061M
HI 7061M Electrode cleaning solution
HI 7061MHI 7061M
(230mL bottle)
HI 731326HI 731326
HI 731326 Calibration screwdriver (20 pcs)
HI 731326HI 731326
Visit our Internet Home Page:
http://www.hannainst.com
±0.2 pH
(32 to 122F); 95% RH
of continuous use
(7.9 x 1.8 x 1")
IS98106-R3 09/01
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the located below the battery compartment and slide off the whole internal part of the tester.
• To remove the electrode proceed as follows: Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con­nected to the electrode through two wires (one for the glass sensor and one for the reference). Remove the two small screws (2 & 3) which fasten the two connecting wires to their sock­ets.
• Remove the electrode and replace it with a new HI73106.
• Connect the new electrode as follows:
first fasten the screw on the front (1) to attach the elec­trode to the circuit. Insert the two small wires into their sockets and fasten the two screws (2 & 3).
The new electrode is supplied with four new O-
rings (4). Make sure they are placed properly (as shown on the above drawing) before reinserting the circuit into the plastic casing.
• Reinsert the internal circuit with the new electrode into the plastic casing.
ChampChamp
Champ
ChampChamp
HI73106
Internal circuit and electrode:
• Fasten the screw on the back below the battery compartment.
• Reinsert the battery compartment cover.
• Perform a new calibration before using the
ChampChamp
Champ
again.
ChampChamp
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
Covered by manship and materials, this tester is now completely serviceable. Contact your dealer for further information.
The
50082-1, EN 50081-1 and EN 61010-1.
Champ® is a registered Trademark of "Hanna Instruments"