HI 98106
Pocket-sized pH Meter
Manufacturers since 1978
Champ is a rugged pH tester with replaceable
electrode and renewable reference junction. The
instrument can be calibrated at one point by the
user, simply acting on the calibration trimmer.
CONDITIONING:CONDITIONING:
CONDITIONING:
CONDITIONING:CONDITIONING:
• Remove the cap.
• To activate the electrode, immerse
ChampChamp
the
Champ in HI70300 storage
ChampChamp
solution for 2 hours.
CALIBRATION:CALIBRATION:
CALIBRATION:
CALIBRATION:CALIBRATION:
ChampChamp
• Immerse the
7.01 (HI7007) buffer solution.
• Allow the reading to stabilize
and with a small screwdriver
turn the Calibration Trimmer
until the display shows "7.0"
The calibration is now complete.
NoteNote
Note: the calibration can be performed
NoteNote
at pH 4.0, pH 7.0 or pH 10.0,
according to the different applications.
OPERATION:OPERATION:
OPERATION:
OPERATION:OPERATION:
• Remove the protective cap.
• Turn the
• Immerse it in the solution to be
• Stir gently and wait for the reading
• After use, rinse the electrode with water to
• Store the electrode with a few
• Always replace the protective cap after use.
ChampChamp
Champ on with the ON/
ChampChamp
OFF switch located on the top of the
meter.
tested without exceeding the maximum immersion level.
to stabilize.
minimize contamination.
drops of storage (HI70300) or
pH7 (HI7007) solution in the
protective cap.
Champ
ChampChamp
in pH
DO NOT USE DISTILLED OR DEIONIZED WATER
FOR STORAGE PURPOSES.
• Large differences in pH readings (±1 pH) could
be due to lack of calibration or dry electrode.
MAINTENANCE:MAINTENANCE:
MAINTENANCE:
MAINTENANCE:MAINTENANCE:
The sensor should be kept moist at all times. For this
purpose, always keep wet the tissue inside the
protective cap. In case of erroneous readings even
after an accurate conditioning and calibration, the
reference junction might be contaminated or clogged.
Pull out 2 mm (1/8'') of
the cloth junction to renew
the electrode reference (it is
recommended to cut the cloth
leaving always at least 2
mm -1/8'' over the reference compartment) and repeat the calibration procedure.
The cloth junction can be pulled out approximately
20 times. After that, the electrode can be replaced
(see the Electrode Replacement section).
BATTERY REPLACEMENT:BATTERY REPLACEMENT:
BATTERY REPLACEMENT:
BATTERY REPLACEMENT:BATTERY REPLACEMENT:
When display fades the batteries must be replaced.
Slide off the battery compartment cover and replace
all four 1.5V batteries while paying attention to
their polarity.
Batteries should only be replaced in a nonhazardous
area using the battery type specified in this instruction
manual.
I 70300
H
®®
®
ChampChamp
Champ
ChampChamp
®®
HI 98106
Pocket-sized pH Meter
SPECIFICATIONS:SPECIFICATIONS:
SPECIFICATIONS:
SPECIFICATIONS:SPECIFICATIONS:
RANGERANGE
RANGE 0.0 to 14.0 pH
RANGERANGE
RESOLUTIONRESOLUTION
RESOLUTION 0.1 pH
RESOLUTIONRESOLUTION
ACCURACY ACCURACY
(@20C/68F)(@20C/68F)
(@20C/68F)
ACCURACY
(@20C/68F)(@20C/68F)
ACCURACY ACCURACY
TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION
TYPICAL EMC DEVIATION ±0.1 pH
TYPICAL EMC DEVIATIONTYPICAL EMC DEVIATION
CALIBRATIONCALIBRATION
CALIBRATION Manual, at 1 point through trimmer
CALIBRATIONCALIBRATION
ENVIRONMENTENVIRONMENT
ENVIRONMENT 0 to 50C
ENVIRONMENTENVIRONMENT
BATTERY TYPEBATTERY TYPE
BATTERY TYPE 4 x 1.5V alkaline (included)
BATTERY TYPEBATTERY TYPE
BATTERY LIFEBATTERY LIFE
BATTERY LIFE approx. 800 hours
BATTERY LIFEBATTERY LIFE
DIMENSIONSDIMENSIONS
DIMENSIONS 175 x 41 x 23 mm
DIMENSIONSDIMENSIONS
WEIGHTWEIGHT
WEIGHT 78 g (2.7 oz.)
WEIGHTWEIGHT
ACCESSORIES:ACCESSORIES:
ACCESSORIES:
ACCESSORIES:ACCESSORIES:
HI 73106HI 73106
HI 73106 Spare electrode
HI 73106HI 73106
HI 77400PHI 77400P
HI 77400P pH 4.01 & 7.01 calibration kit
HI 77400PHI 77400P
(5 x 20 mL, each)
HI 77700PHI 77700P
HI 77700P pH 7.01 calibration kit
HI 77700PHI 77700P
(10 x 20 mL)
HI 770710PHI 770710P
HI 770710P pH 7.01 & 10.01 calibration kit
HI 770710PHI 770710P
(5 x 20 mL, each)
HI 7004MHI 7004M
HI 7004M pH 4.01 buffer solution
HI 7004MHI 7004M
(230 mL bottle)
HI 7007MHI 7007M
HI 7007M pH 7.01 buffer solution
HI 7007MHI 7007M
(230 mL bottle)
HI 7010MHI 7010M
HI 7010M pH 10.01 buffer solution
HI 7010MHI 7010M
(230 mL bottle)
HI 70300MHI 70300M
HI 70300M Storage solution (230mL)
HI 70300MHI 70300M
HI 7061MHI 7061M
HI 7061M Electrode cleaning solution
HI 7061MHI 7061M
(230mL bottle)
HI 731326HI 731326
HI 731326 Calibration screwdriver (20 pcs)
HI 731326HI 731326
Visit our Internet Home Page:
http://www.hannainst.com
±0.2 pH
(32 to 122F); 95% RH
of continuous use
(7.9 x 1.8 x 1")
IS98106-R3 09/01
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:
ELECTRODE REPLACEMENT:ELECTRODE REPLACEMENT:
The electrode can be easily replaced in the following
way:
• Slide off the battery cover.
• Remove the screw on the back of the
located below the battery compartment and slide
off the whole internal part of the tester.
• To remove the electrode proceed as follows:
Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is connected to the electrode through two wires (one
for the glass sensor and one for the reference).
Remove the two small screws (2 & 3) which
fasten the two connecting wires to their sockets.
• Remove the electrode and
replace it with a new
HI73106.
• Connect the new electrode
as follows:
first fasten the screw on the
front (1) to attach the electrode to the circuit. Insert
the two small wires into
their sockets and fasten the
two screws (2 & 3).
• The new electrode is supplied with four new O-
rings (4). Make sure they are placed properly
(as shown on the above drawing) before
reinserting the circuit into the plastic casing.
• Reinsert the internal circuit with the new
electrode into the plastic casing.
ChampChamp
Champ
ChampChamp
HI73106
Internal circuit and electrode:
• Fasten the screw on the back below the
battery compartment.
• Reinsert the battery compartment cover.
• Perform a new calibration before using the
ChampChamp
Champ
again.
ChampChamp
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:
ONE YEAR WARRANTY & SERVICEABLE:ONE YEAR WARRANTY & SERVICEABLE:
Covered by
manship and materials, this tester is now completely
serviceable. Contact your dealer for further information.
The
50082-1, EN 50081-1 and EN 61010-1.