To obtain the hi ghest accurac y for measureme nts it is import ant to follow the se tips:
• Fresh buf fer should be use d for each ca libratio n. Calibration is only as g ood as the buffe r being use d. The pH buf fer values c hange over ti me once the sa chets are op ened.
• When the me ter is not in use it i s important to a dd several dro ps of storage s olution to the prote ctive cap to ke ep the probe hydr ated. If stora ge solution is no t available, pH
4.01 or pH 7.01 buffer can be u sed.
• For improv ed accuracy it is r ecommended t o calibrate in tw o buffers .
• It is import ant to calibrate and me asure samples at the s ame temperature as t here is no temperature compensation. A dramatic change in temperature between buffer solutio ns and samples to b e tested will gi ve inaccurate re adings.
Refilling the Electrode
• If fouled, remove the s leeve and r inse slee ve and sens or tip wit h purifi ed water. Turn over the probe, remo ve the sle eve by care fully rot ating it and pull strai ght along the axis of the el ectro de. Use care as the pH s tem is made of glass. Rinse of f any traces of ele ctrolyte g el.
• Soak the s ensor tip i n HI70061G, HI700661P, HI700663P o r HI700664P cleaning s olution f or 20 minu tes t hen rinse wit h purified wat er.
• Refill the r eferen ce well wit h HI9071 Gelled Bridge Electrolyte.
• Replace the refere nce sleeve: inser t and push the sl eeve onto the ele ctrode. Make sure the bla ck O-ring i s fixed in side the electr ode body. Any excess o f gel will be expelle d from the end of t he electro de through the o pen junction .
• Rinse any ex cess gel off wi th purified wate r and gently dry of f body wit h a soft cloth o r tissue.
• Soak assembled probe in Electrode storage solution for a minimum of 30 min utes.
• Rinse prob e with purifi ed water.
• Shake the el ectro de down as yo u would do wi th a clinical thermom eter to elimina te any air bubbles in side the glass b ulb.
• Calibrate in fresh buffers before using for measurements.
Care and Maintenance
To change the C R2032 Li-ion ba ttery, tur n the batter y cover locat ed on the back of t he meter counterc lockwise t o unlock. Remov e cover and repl ace the batt ery with + side f acing up.
Note: Batte ries should only be r eplaced i n a safe area u sing the b atter y type spe cified in t his inst ructio n manual. Old bat teries sh ould be di sposed in accordance with local regulations.
Battery Indicator
The meter features a low battery indicator. When the battery is running low, the tag will blink on screen. When the battery has been depleted, “Erb” will appear on screen and the meter will turn off.
Clear Calibration
Place meter in calibration mode. Press and hold ON/OFF until “CLr” is displayed. The meter will now be at default calibration.
“Err” Message
In calibration mode, if the meter displays an “Err” message when in the correct fresh buffer solution then the probe should be cleaned. Place the probe in the cleaning solution for 20 minutes. Rinse with purified water and place in storage solution for a minimum of 30 min utes before calibrating.
Battery Replacement
Auto-off
From measurement mode, press and hold the ON/OFF button. The meter will cycle through “OFF,” “CAL,” then current auto-off setting. The default setting is 8 minutes (“d08”). Press ON/OFF button to change. “d60” is auto-off after 60 minutes, and “d--” disables the auto-off feature. Press and hold the button to exit the menu.
Electrode Cleaning Solutions
Code Description
HI700 61G GroLine general purpose cleaning solution, 20 mL sachets (25 pcs.) HI70 0661P General p urpose cl eaning sol ution for a gricultur e, 20 mL sache ts (25 pcs.) HI700663P Cleaning solu tion for soil d eposits, 2 0 mL sachets (25 p cs.) HI700664P Cleaning solu tion for humus d eposits , 20 mL sachets (2 5 pcs.)
pH Buffe r Solution
Code Description
HI70004G Gro Line pH 4.01 buf fer solut ion, 20 mL sach ets (25 pcs.) HI70007G Gro Line pH 7.01 buffe r solution , 20 mL sachet s (25 pcs.) HI774 00P pH 4.01 & 7.01 buff er solutio n, 20 mL sachet s (10 pcs., 5 ea.)
Accessories
Recommendations for Users
Before us ing this produc t, make sure it is enti rely suitable f or your specif ic application an d for the environm ent in which it i s used. Any var iation intr oduced by th e user to the s upplied equi pment may degrade t he meters’ pe rformance. F or yours and th e meter’s safet y do not use or st ore the meter in hazardous environments.
Warranty
The meter is warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
Certification
All Hanna Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment. The product should not be treated as household
waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to www.hannainst.com.
RoHS
compliant
HI981030
Soil pH Tester
Other Accessories
Code Description
HI721 319 Soil auger HI705 1M Soil preparation solution, 230 mL bottle
Electrode Storage Solutions
Code Description
HI70300G G roLine elec trode sto rage solut ion, 20 mL sachet s (25 pcs.) HI70300-023 GroLine e lectrode s torage sol ution, 230 mL bo ttle HI9072 Electr ode storage s olution, 13 mL dro pper
Electro de Fill Solution
Code Description
HI9071 Gelled Br idge Electr olyte
Specifications
Range 0.0 to 14.0 pH Resolution 0.1 pH Accuracy ±0.2 pH @25 °C/ 77 °F Calibration Automatic, one or two-point Electrode Built-in probe for specific application Batte ry Type CR2032 Li-ion Batter y Life Approximat ely 1000 hours of co ntinuous use Auto-off 8 minutes, 6 0 minutes or can be di sabled Environment 0 to 50 °C (32 to 122 °F); RH 95% max Dimensions 51 x 151 x 21 mm (2 x 5.9 x 0.9“) Weight 44 g (1.6 oz.)
Preliminary Examination
Remove the m eter from th e packing mat erial and examin e it carefully t o make sure that no damage has o ccurred durin g shipment. If not iceable damag e is evident, con tact your local Hanna In struments O ffice. Each meter i s supplied with:
• pH 4.01 buffer s olution liquid sa chet (2 pcs.)
• pH 7.01 buffer solut ion liquid sachet (2 p cs.)
• Cleaning solution for soil deposits
• Cleaning solution for humus deposits
• Electrode storage solution, 13 mL dropper
• Electrode fill solution
• Instruction manual
• Quality certificate
Note: Save all p acking material u ntil you are sure that t he instrument functions correctl y. All defective item s must be returned i n the original pa ckaging with the supplied accessories.
Liquid Cr ystal Display
ON/OFF Button
Maximum Immersio n Level
Recommended Immersion Level
Procedure for Direct Ground Measurement
1) Verify meter is set up correctly and pH probe is calibrated.
2) Dig, discarding 5 cm of topsoil.
3) Perforate the soil (with HI721319 soil auger) to a depth of about 20 cm or more.
4) If the soil is dry, moisten it with a small amount of tap water.
5) Wash the electrode with tap water (not distilled).
6) Insert the electrode pushing it slightly into the soil to ensure proper contact.
7) Observe the measurement.
8) Wash the electrode with tap water (not distilled) and (using a finger) gently remove any soil remaining on the electrode (avoid using a rag or cloth).
9) Repeat the procedure in different locations in the field.
10) Consider the average of the measured data.
For best r esult, it is advisabl e to measure the pH of a s oil solution, using a s ample of soil and soil preparation solution HI7051; it is bette r to use this proce dure if you have to test a st ony field in whic h you risk damaging t he electrode .
Probe Features
Removable PVDF reference sleeve
Polyvi nylidene fluorid e (PVDF) is a durable plas tic that is resist ant to most chemical s and solve nts, including so dium hypochlori te. It has high abrasion r esistance, mec hanical strengt h and resistan ce to ultravio let. PVDF is also re sistant to fun gal growth.
Operation
Press th e ON/OFF but ton to turn th e meter on. All ta gs will be displayed.
The meter w ill go into m easure ment mode : current reading an d calibrated b uffers a re displayed .
Meter Calibration
From measurement mode, press and hold the ON/OFF button until “CAL” is displayed.
When “7.01” blinks on the display, place the tip of the probe into a pH 4.01, 7.01 buffer solution.
When the reading is s table, the stab ility indicator will disappe ar.“Sto” will be disp layed when the calibration is saved.
If pH 4.01 buff er solutio n is used as the fir st point
the value of the buffe r is recogni zed and displayed with the blinking stability indicator.
Meter wil l exit to measurement mode and th e calibration tag will be displayed.
One-Point Calibration with pH 4.01
B
One-Point
Meter will exit to measurement mode and the calibration tag will be displayed.
pH 4.01 will then blink on the display. Ignore it an d press ON/OFF button.
Two- Point
Use pH 4.01 to perform a two­point calibration. The value is automatically recognized and displayed with the blinking stability indicator.
When the reading is stable, the stability indicator will disappear. “Sto” will be displayed when the calibration is saved.
Meter will exit to measurement mode and the calibration tags will be displayed.
If pH 7.01 buffer solution is used as the first point the buffer is recognized
with the blinking stability indicator. When the reading is stable, the stability indicator will disappear and pH 7.01 will be calibrated.
If pH 7.01 is the only calibration point,
finish one-point procedure at right.
If using pH 4.01 as a second point,
continue two-point procedure at right.
Meter Overview
Preparation:
The pH elec trode is shipp ed ready for u se with a prote ctive cap con taining stor age solution . Before using the m eter, remove the pro tective cap and in spect the probe t o verify the bri dge gel has not dried ou t. Refresh the brid ge electroly te if necessar y. Following care pro cedure (ignoring t he cleaning procedures).
Alterna tely, rinse o ff the pro be tip with w ater and con dition the e lectrod e by soakin g the tip (bot tom 4 cm (1.5”)) in pH 7.01 buffer s olution f or sever al minutes . Perform a p robe calib ration be fore usin g.
• Do not be alarme d if white c ryst als appear ar ound the c ap. This is nor mal with p H electr odes. Rinse o ff the probe t ip with water b efore using.
• Tur n the meter on by p ressing ON/ OFF button.
• R emove the prote ctive cap. DO NOT S CRATCH THE GLA SS. Do not use the pr obe to bore a hole thro ugh the soil. Mak e sure an entry ho le is made in the soil a nd water has bee n added. Af ter placing t he probe in the ho le, wait for a sta ble reading.
NEVER IMM ERSE THE ELECTR ODE OVER THE MA XIMUM IMMERSI ON LEVEL.
• For best results, recalibrate periodically.
• A fter use, r inse the elec trode wit h water and sto re it with a few dr ops of storage s olution in the prot ective cap.
• R eattach th e protectiv e cap after ea ch use.
DO NOT USE DIS TILLED OR DEIONIZ ED WATER FOR STOR AGE PURPOSES.
A
One or Two-Point Calibration with pH 7.01
For one-p oint calibrati on using pH 4.01 buff er go to proced ure B
For one or t wo-point calibr ation using pH 7.01 buffe r go to procedure A
A B
Open Reference Junction with removable sleeve design
Suspende d solids c an permane ntly clog t he pores of a ceramic re ference junctio n. The open junction design pro vides a con stant junc tion potential and minimizes b lockage b y providin g an open gel interf ace betwee n the sample and in ternal Ag/A gCl referen ce. Should s oil enter into the junc tion, the junction c an be easily cleaned and refreshed with new bridge electrolyte.
pH Probe
Thank You
Thank you f or choosing a Hanna In struments pro duct. Please re ad this instruc tion manual care fully before u sing this instru ment.
For techni cal support, con tact your local Han na Instrument s Office or em ail us at tech@hannainst.com.
To find your lo cal Hanna Instr uments Of fice or for add itional info rmation on Hann a Instruments products, visit www.hannainst.com
Battery indicator
Unstable tag
Measurement
unit
Digits
Calibration tags
"Sto" will b e displayed when the calibr ation is saved.