pHel
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode
pH electrodepH electrode
Specifications
Range
Resolution
Accuracy
(@20°C/°68F)
Typical EMC
Deviation
Calibration
Environment
Battery
Life
Dimensions
Weight
Operating Instructions
pHel is a very easy instrument to use. Simply
switch it on, remove the
protective cap and immerse the tip of the electrode (the bottom 4cm
0.0 to 14.0 pH
0.1 pH
±0.2 pH
±0.2 pH
Two points
through trimmers
0 to 50ºC (32 to
122ºF), 95% RH
2 x 1.4V
3000 hours of
continuous use
200x28x20 mm
7.9x1.1x0.8"
46 g (1.6 oz.)
or 1½ inch) into the solution. Under no circumstances the immersion
shall be above 8 cm or
3 inches. Stir gently and
wait until the display stabilizes.
If the electrode is dry,
allow a few minutes for
the reading to stabilize.
For best results, recalibrate periodically using fresh calibration
buffers.
• Do not be alarmed if white crystals appear around the cap.
This is normal with pH electrodes and they dissolve when
rinsed with water.
• If the electrode is dry, soak it in
tap water for a few minutes.
• After use, rinse the electrode
with water to minimize contamination.
• Store the electrode with a few
drops of
Storage Solution
in the protective
cap.
DO NOT USE DISTILLED OR
Always replace the protective
cap after use.
Calibration
• Dip the tip of the electrode (4cm/1½") of pHel
in a sample of pH 7.01
buffer (from
Calibration Kit
the reading to stabilize.
HI70300
DEIONIZED WATER FOR
STORAGE PURPOSES.
HI774P
). Allow
• Adjust the pH 7 trimmer
until the display reads
"7.0".
pHel
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode
pH electrodepH electrode
• Rinse the electrode with
water and dip it in a
sample of pH 4.01 (or
10.01 buffer from
HI 7710P Calibration
Kit
). Allow the reading
to stabilize.
• Adjust the pH 4/10 trimmer until the display reads
the chosen buffer value
(either "4.0" or "10.0").
• Calibration is now complete.
FOR BETTER ACCURACY,
ALWAYS USE FRESH
BUFFERS FOR CALIBRATION
& NEVER REUSE THEM.
Battery Replacement
Replace the batteries when the
display fades or pHel cannot be
switched on or segments are
missing.
Lift the battery
cover. Check the
polarity of the batteries before removing them. Insert
two new 1.4V batteries the same
way. Battery replacement must
only take place in a
non hazardous area using the
battery types specified in this instruction manual.
Maintenance
• If the electrode remains dry for
a long period, before taking any
measurement or calibrating, it
is necessary to reactivate it by
dipping it into
tion for approximately 4 hours.
The electrode should be
cleaned after use and stored
with the cap on and some
HI70300
HI70300
solution inside.
solu-
• Solutions available for the calibration, the maintenance, the
storage and the cleaning of the
electrode:
•
HI70300
•
HI7061
solution (soak for 30 min.)
•
HI7073
solution (soak for 15 min.)
•
HI7074
solution (soak for 15 min.)
•
HI7077
ing solution (soak several
times with clean solution)
•
HI774P
4.01 & 7.01 @ 25°C/77°F,
30 ml each)
•
HI777P
7.01 @ 25°C/77°F, 2 x 30 ml)
•
HI7710P
4.01 & 10.01 @ 25°C/77°F,
30 ml each)
Storage solution
General cleaning
Protein cleaning
Inorganic cleaning
Oil and fat clean-
Calibration kit (pH
Calibration kit (pH
Calibration kit (pH
Recommendations for Users
Before using this product, ensure that it is entirely
suitable for the environment in which it is used. Operation of this instrument in residential area could
cause unacceptable interferences to radio and TV
equipments, requiring the operator to take all necessary steps to correct interferences. The glass bulb at
the end of the electrode is sensitive to electrostatic
discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn
to avoid possible damage to the electrode by electrostatic discharges. Any variation introduced by the user
to the supplied equipment may degrade the
instrument's EMC performance. To avoid electrical
shock, do not use these instruments when voltages at
the measurement surface exceed 24VAC or 60VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
Stay in Touch with the Manufacturer!
SPHEL 09/96