Hanna Instruments HI 98101 User Manual

pHel
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode
pH electrodepH electrode
Specifications
Range Resolution Accuracy
(@20°C/°68F) Typical EMC
Deviation Calibration
Environment Battery
Life Dimensions Weight
Operating Instructions
pHel is a very easy in­strument to use. Simply switch it on, remove the protective cap and im­merse the tip of the elec­trode (the bottom 4cm
0.0 to 14.0 pH
0.1 pH
±0.2 pH ±0.2 pH
Two points
through trimmers
0 to 50ºC (32 to
122ºF), 95% RH
2 x 1.4V
3000 hours of
continuous use 200x28x20 mm
7.9x1.1x0.8"
46 g (1.6 oz.)
• Do not be alarmed if white crys­tals appear around the cap. This is normal with pH elec­trodes and they dissolve when rinsed with water.
• If the electrode is dry, soak it in tap water for a few minutes.
• After use, rinse the electrode with water to minimize con­tamination.
• Store the elec­trode with a few drops of
Storage Solution
in the protective cap.
DO NOT USE DISTILLED OR
Always replace the protective cap after use.
Calibration
• Dip the tip of the elec­trode (4cm/1½") of pHel in a sample of pH 7.01 buffer (from
Calibration Kit
the reading to stabilize.
HI70300
DEIONIZED WATER FOR
STORAGE PURPOSES.
HI774P
). Allow
• Adjust the pH 7 trimmer until the display reads "7.0".
pHel
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode pHel
pH electrodepH electrode
pH electrodepH electrode
pH electrode
pH electrodepH electrode
Rinse the electrode with water and dip it in a sample of pH 4.01 (or
10.01 buffer from
HI 7710P Calibration Kit
). Allow the reading
to stabilize.
Adjust the pH 4/10 trim­mer until the display reads the chosen buffer value (either "4.0" or "10.0").
Calibration is now complete. FOR BETTER ACCURACY,
ALWAYS USE FRESH
BUFFERS FOR CALIBRATION
& NEVER REUSE THEM.
Battery Replacement
Replace the batteries when the display fades or pHel cannot be switched on or segments are missing.
Lift the battery cover. Check the polarity of the bat­teries before re­moving them. Insert two new 1.4V bat­teries the same way. Battery re­placement must only take place in a non hazardous area using the battery types specified in this in­struction manual.
Maintenance
If the electrode remains dry for a long period, before taking any measurement or calibrating, it is necessary to reactivate it by dipping it into tion for approximately 4 hours. The electrode should be cleaned after use and stored with the cap on and some
HI70300
HI70300
solution inside.
solu-
Solutions available for the cali­bration, the maintenance, the storage and the cleaning of the electrode:
HI70300
HI7061
solution (soak for 30 min.)
HI7073
solution (soak for 15 min.)
HI7074
solution (soak for 15 min.)
HI7077
ing solution (soak several times with clean solution)
HI774P
4.01 & 7.01 @ 25°C/77°F, 30 ml each)
HI777P
7.01 @ 25°C/77°F, 2 x 30 ml)
HI7710P
4.01 & 10.01 @ 25°C/77°F, 30 ml each)
Storage solution
General cleaning
Protein cleaning
Inorganic cleaning
Oil and fat clean-
Calibration kit (pH
Calibration kit (pH
Calibration kit (pH
Recommendations for Users
Before using this product, ensure that it is entirely suitable for the environment in which it is used. Op­eration of this instrument in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all neces­sary steps to correct interferences. The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electro­static discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC. To avoid damages or burns, do not perform any mea­surement in microwave ovens.
Stay in Touch with the Manufacturer!
SPHEL 09/96
Loading...