HANNA HI98100 User guide

Page 1
Auto-o
Clear Calibration
Battery Indicator
Care and Maintenance
Accessories
Warranty
INSTRUCTION MANUAL
From measurement mode, press and hold the ON/OFF button. The meter will cycle through “OFF,” “CAL,” then current auto-o setting.
The default setting is 8 minutes (“d08”). Press ON/OFF button to change. “d60” is auto-o after 60 minutes, and “d--” disables the auto-o feature.
Place meter in calibration mode. Press and hold ON/OFF until “CLr” is displayed. The meter will now be at default calibration. No tags will be shown in measurement mode until calibration is performed.
“Err” Message
In calibration mode, if the meter displays an “Err” message when in the correct fresh buer solution then the probe should be cleaned. Place the probe in the HI700601 cleaning solution for 15 minutes. Rinse with puried water and place in storage solution for 1 hour before calibrating. If the “Err” message persists then the HI1271 probe should be replaced.
The Checker®Plus features a low battery indicator. When the battery is running low, the tag will appear on screen. When the battery has been depleted, the tag will blink.
Battery Replacement
To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter counterclockwise to unlock. Remove cover and replace with + side facing up.
Note: Ba tteries s hould only b e replaced in a s afe area using th e batter y type spe cified in t his instru ction manual. O ld batte ries shoul d be dispose d in accordan ce with local regulations.
To obtain the highest accuracy for measurements it is impor tant to follow these tips:
• Calibration is only as good as the buffer being used. The pH buffer values change over time once the sachets are opened. Fresh buffer should be used for each calibration.
• The probe should be rinsed with purified water each time before placing in buffer or sample to be tested.
• When the meter is not in use it is important to add several drops of storage solution to the protective cap to keep the probe hydrated. If storage solution is not available, pH 4 or pH 7 buffer can be used.
• For improved accuracy it is recommended to calibrate a minimum of t wo points. It is important to use buffers that bracket the expected value of the sample to be tested. For example, if the expected value is pH 8, the Checker®Plus should be calibrated using pH 7 and pH 10 buffers.
• It is important to calibrate and measure samples at the same temperature. A dramatic change in temperature between buffer solutions and samples to be tested will give inaccurate readings.
pH Buer Solution
Code Description
HI70004P
HI70007P
HI70010P
HI 7740 0P
HI770710 P
pH 4.01 buf fer solut ion, 20 mL sac hets (25)
pH 7.01 buffer s olution, 20 mL sac hets (25)
pH 10.01 buf fer solut ion, 20 mL sac hets (25)
pH 4.01 & 7.01 buffe r solution , 20 mL sac hets (10 pcs ., 5 ea.)
pH 10.01 & 7.01 buffe r solution , 20 mL sac hets (10 pcs ., 5 ea.)
Electrode Cleaning Solution
Code Description
HI700601P
gener al purpose c leaning solut ion, 20 mL sac hets (25)
Electrode Storage Solution
Code Description
HI70300L
HI70300M
electrode storage solution, 500 mL bo ttle
electrode storage solution, 230 mL bo ttle
Check er®Plus is gua ranteed fo r a period of one yea r for meter and s ix months for p robe afte r date of purch ase agains t defects i n work manship and m aterials wh en used for their in tended pur pose and mai ntained accord ing to instr uctions . This warra nty is limit ed to repair or r eplacement f ree of charg e. Damage due t o accidents , misuse, tamp ering or lack o f prescrib ed maintena nce is not cov ered. If ser vice is requ ired, conta ct your loc al Hanna oce. I f under warr anty, repor t the model nu mber, date of purc hase, seria l number and t he nature of the pr oblem. If th e repair is not co vered by the wa rranty, you w ill be notie d of the charg es incurre d. If the inst rument is to b e retur ned to Hanna In strumen ts, rst ob tain a Retur n Goods Auth orizatio n number fro m the Technic al Servic e departm ent and then send it w ith shippi ng costs pre paid. When shippi ng any instr ument, make s ure it is properly packed for complete protection.
Recommendations for Users
Befor e using Hanna pr oducts , make sure that they ar e entirely su itable for yo ur specic applic ation and fo r the environ ment in which they ar e used. Oper ation of the se instru ments may cau se unaccept able inter ferences to other e lectroni c equipment , thus requi ring the opera tor to take all n ecessar y steps to cor rect such in terferen ces. Any var iation intr oduced by the us er to the suppl ied equipme nt may degra de the inst rument’s E MC perfor mance. To avoid damag es or burns, d o not put the instr ument in micr owave oven. F or yours and the ins trument s afety do no t use or store the ins trument in h azardous en vironmen ts.
All righ ts are rese rved. Rep roducti on in whole or in part i s prohibit ed withou t the writ ten consent of the cop yright own er, Hanna Instr uments Inc ., Woonso cket, Rhod e Island, 0289 5, USA.
HI98100 I NSTMAN 1.0 0 6/15
HI98100
Checker®Plus
Page 2
Thank You Operation
Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument.
For additional information, visit
www.hannainst.com or e-mail
Specications
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accur acy
(@25° C/ 77°F )
Calibration au tomatic, o ne or two-p oint
Electrode HI1271 (included)
Battery Type CR2032 L i-ion
Battery Life
Auto-off
Environment
Dimensions
Weight 50 g (1.8 oz .)
±0.2 pH
approx imately 100 0 hours of continuous use
8 minutes, 60 minutes, or can be d isabled
0 to 50°C ( 32 to 122°F); RH 95% max
50 x 174 x 21 mm (2 x 6.8 x 0.9 “) (excluding probe)
Meter Overview
Preparation:
The pH electrode is shipped dry. Before using the Checker®Plus, remove the protec tive cap and condition the electrode by soaking the tip (bottom 4 cm (1.5”)) in pH 7.01 buer solution for several hours. Then follow the calibration procedure.
• Do not be alarmed if white cr ystals appear around the cap. This is normal with pH electrodes and they dissolve when rinsed with water.
• Connect the electrode to the meter.
• Turn the Checker ®Plus on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the electrode in the sample to be tested.
• Stir gently and wait for a stable reading.
NEVER IMMERSE THE ELEC TRODE OVER THE MAXIMUM IMMERSION LEVEL. THE CONNECTOR MUST ALWAYS BE CLEAN AND DRY.
• After use, rinse the electrode with water and store it with a few drops of HI70300 storage solution in the protective cap.
• Replace protective cap af ter each use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
Press the ON/OFF button to turn the meter on. All tags will be displayed.
The meter will go into measurement mode: current reading and calibrated buers are displayed.
Liquid Cry stal Display
ON/OFF Button
Maximum Immersion Level
pH Probe
Recommended Immersion Level 4 cm ( 1.5”)
Meter Calibration
From measurement mode, press and hold the ON/OFF button until “CAL” is displayed.
When “7.01” blinks on the display, place the tip of the probe into a pH 4, 7, or 10 buer solution.
For one or t wo-point calibration using pH 7
A
buer go to procedure A
For one-point calibration using pH 4 or pH 10 buer
B
go to procedure B
One or Two-Point Calibration with pH 7
A
If pH 7 buer solution is used as the rst point the buer is recognized
with the blinking stability indicator.
When the reading is stable, the stability indicator will disappear and pH 7 will be calibrated. If pH 7 is the only calibration point, nish one-point procedure at right.
If using pH 4 or pH 10 as a second point,
continue two-point procedure at right.
One-Point Calibration with pH 4 or pH 10
B
If pH 4 or 10 buer solution is used as the rst point the value of the
buer is recognized and displayed with the blinking stability indicator.
One-Point
pH 4 will th en blink on the disp lay.
Two-Poi n t
Use pH 4 or 10 to p erform a two-poin t calibr ation. The value is auto matically recogn ized and display ed with the blinking stability indicator.
When the reading is stabl e, the stabili ty indica tor will disapp ear. “Sto” will be displayed when the calibr ation is saved.
Press t he button to e xit and save calibr ation. “Sto” will then be di splayed.
When the reading is stabl e, the stabili ty indica tor will disapp ear. “Sto” will be displayed when the calibr ation is saved.
Meter will exit to m easurement mode and the calibration tag will be displayed.
Meter will exit to m easurement mode and the c alibration t ag will be display ed.
Meter will exit to m easurement mode and the calibration tags will be displayed.
Loading...