Hanna Instruments HI 96831 User Manual

Page 1
Instruction Manual
HI 96831
Ethylene Glycol Refractometer
www.hannainst.com
1
Page 2
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary
information for the correct use of this instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ............................................................................................. 2
GENERAL DESCRIPTION .....................................................................................................3
SPECIFICATIONS .................................................................................................................. 3
PRINCIPLE OF OPERATION ................................................................................................4
MEASUREMENT UNITS ........................................................................................................ 5
FUNCTIONAL DESCRIPTION................................................................................................ 5
DISPLAY ELEMENTS............................................................................................................ 6
MEASUREMENT GUIDELINES .............................................................................................. 7
CALIBRATION PROCEDURE ................................................................................................ 7
MEASUREMENT PROCEDURE ............................................................................................. 8
TO CHANGE MEASUREMENT UNIT .....................................................................................9
TO CHANGE TEMPERATURE UNIT...................................................................................... 9
MAKING A STANDARD SOLUTION....................................................................................10
BATTERY REPLACEMENT .................................................................................................10
WARRANTY ........................................................................................................................ 10
ERROR MESSAGES ............................................................................................................11
PRELIMINARY EXAMINATION
Remove the instrument from the packing materials and examine carefully to ensure no damage has occurred during shipping. If any damage has occurred, notify your Dealer or closest Hanna Customer Service Center.
Each instrument is supplied with:
• 9 V battery
• Instruction manual
Note:Save all packing material until you are sure that the instrument functions
correctly. A defective instrument must be returned in its original packing.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
2
Page 3
GENERAL DESCRIPTION
The HI 96831 Digital Refractometer is a rugged portable, water resistant device that utilizes the measurement of refractive index to determine the % volume and freezing point of ethylene glycol based coolants or antifreeze. The HI 96831 device benefits from Hanna’s years of experience as a manufacturer of analytical instruments. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for use in the field to optimize your cooling system. The HI 96831 refractometer is an optical device that is simple and quick to use. Samples are measured after a simple user calibration with distilled or deionized water. Within seconds, the refractive index and temperature are measured and converted into one of two measurement units; % Volume or Freezing Point. The instrument utilizes internationally recognized references for unit conversion and temperature compensation for ethylene glycol solutions (e.g. CRC Handbook of Chemistry and Physics, 87
th
Edition). The temperature (in °C or °F) is also displayed on the large dual level display along with helpful message codes. Key features include:
• IP65 waterproof protection
• Automatic Temperature Compensation (ATC)
• Battery operation with Low Power indicator (BEPS)
• Automatically turns off after 3 minutes of non-use
SPECIFICATIONS
emuloV%tnioPgnizeerF)F°(C°
:egnaR %001ot0)F°85-ot23(C°05-ot0)F°671ot23(C°08ot0
:noituloseR %1.0)F°1.0(C°1.0)F°1
:ycaruccA %2.0±)F°0.1±(C°5.0±)F°5.0±(C°3.0±
.0(C°1.0
Temperature Compensation: Automatic between 0 and 40 °C (32 - 104 °F) Measurement Time: Approximately 1.5 seconds Minimum Sample Volume: 100 μL (cover prism totally) Light Source: Yellow LED Sample Cell: Stainless Steel ring and flint glass prism Case Material: ABS Enclosure Rating: IP 65 Battery Type / Life: 1 X 9V / 5000 readings Auto-Off: After 3 minutes of non-use Dimensions: 19.2(W) x 10.2(D) x 6.7(H) cm Mass: 420 g
3
Page 4
PRINCIPLE OF OPERATION
Ethylene Glycol determinations are made by measuring the refractive index of a solution. Refractive Index is an optical characteristic of a substance and the number of dissolved particles in it. Refractive Index is defined as the ratio of the speed of light in empty space to the speed of light in the substance. A result of this property is that light will “bend”, or change direction, when it travels through a substance of different refractive index. This is called refraction. When passing from a material with a higher to lower refractive index, there is a critical angle at which an incoming beam of light can no longer refract, but will instead be reflected off the interface.
The critical angle can be used to easily calculate the refractive index according to the equation:
sin (θ
) = n2 / n
critical
1
Where n2 is the refractive index of the lower-density medium; n1 is the refractive index of the higher-density medium.
In the HI 96831 refractometer, light from an LED passes through a prism in contact with the sample. An image sensor determines the critical angle at which the light is no longer refracted through the sample. Specialized algorithms then apply temperature compensation to the measurement and convert the refractive index to: % Volume or Freezing Point.
4
Page 5
MEASUREMENT UNITS
Freezing point is displayed as a temperature from 0.0 to -50.0 °C corresponding to 0-58 % by volume. The display blinks the freezing point when the concentration of Ethylene glycol is greater than 78 % concentration by volume. When the display shows “LO”, the freezing point is at a minimum (below -50 °C).
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
TOP VIEW
1. Liquid Crystal Display (LCD)
2. READ Key (User Measurement)
3. ZERO Key (User Calibration)
4. RANGE Key (User Measurement Unit)
5. ON/OFF
6. Stainless Steel Sample Well and Prism
7. Secondary Display
8. Primary Display
5
Page 6
BOTTOM
9. Battery Cover
10.Battery Compartment
DISPLAY ELEMENTSDISPLAY ELEMENTS
DISPLAY ELEMENTS
DISPLAY ELEMENTSDISPLAY ELEMENTS
1. Battery (blinks when low battery condition detected)
2. Measurement in Progress Tag
3. SETUP: Factory Calibration Tag
4. CAL: Calibration Tag
5. Automatic Temperature Compensation (blinks when temperature exceeds 0-40°C / 32-104°F range)
6. Primary Display (displays measurement and error messages)
7. Temperature Units
8. Secondary Display (displays temperature measurements; when blinking, temperature has exceeded operation range: 0-80°C / 32-176°F)
9. Range indicator
6
Page 7
MEASUREMENT GUIDELINESMEASUREMENT GUIDELINES
MEASUREMENT GUIDELINES
MEASUREMENT GUIDELINESMEASUREMENT GUIDELINES
Handle instrument carefully. Do not drop.
Do not immerse instrument under water.
Do not spray water to any part of instrument except the “sample well” located over the prism.
The instrument is intended to measure ethylene glycol solutions. Do not expose instrument or prism to solvents that will damage it. This includes most organic solvents and extremely hot or cold solutions.
Particulate matter in a sample may scratch the prism. Absorb sample with a soft tissue and rinse sample well with deionized or distilled water between samples.
Use plastic pipettes to transfer all solutions. Do not use metallic tools such as needles, spoons or tweezers as these will scratch the prism.
Cover sample well with hand if measuring in direct sun.
CALIBRATION PROCEDURECALIBRATION PROCEDURE
CALIBRATION PROCEDURE
CALIBRATION PROCEDURECALIBRATION PROCEDURE
Calibration should be performed daily, before measurements are made, when the battery has been replaced, between a long series of measurements, or if environmental changes have occurred since the last calibration.
1. Press the ON/OFF key, then release. Two instrument test screens will be displayed briefly;
all LCD segments followed by the percentage of remaining battery life. The meter will then briefly display an indication of the measurement unit set. When the LCD displays dashes, the instrument is ready.
2. Using a plastic pipette, fill the sample well with distilled or deionized water. Make sure the
prism is completely covered.
Note: If the ZERO sample is subject to intense light such as sunlight or another strong source,
cover the sample well with your hand or other shade during the calibration.
7
Page 8
3. Press the ZERO key. If no error messages appear, your unit is calibrated.
(For a description of ERROR MESSAGES see page 11).
Note: The 0.0 screen will remain until a sample is measured or the instrument is turned off.
4. Gently absorb the ZERO water standard with a soft tissue. Use care not to scratch the prism surface. Dry the surface completely. The instrument is ready for sample measurement.
Note: If the instrument is turned off the calibration will not be lost.
MEASUREMENT PROCEDUREMEASUREMENT PROCEDURE
MEASUREMENT PROCEDURE
MEASUREMENT PROCEDUREMEASUREMENT PROCEDURE
Verify the instrument has been calibrated before taking measurements.
1. Wipe off prism surface located at the bottom of the sample well. Make sure the prism and sample well are completely dry.
2. Using a plastic pipette, drip sample onto the prism surface. Fill the well completely.
Note: If the temperature of the sample differs significantly from the temperature of the
instrument, wait approximately 1 minute to allow thermal equilibration.
3. Press the READ key. The results are displayed as % Volume or Freezing Point.
Note: The last measurement value will be displayed until the next sample is measured or the
instrument is turned off. Temperature will be continuously updated.
Note: The ”ATCtag blinks and automatic temperature compensation is disabled if the
temperature exceeds the 0-40 °C / 32-104 °F range.
8
Page 9
4. Remove sample from the sample well by absorbing with a soft tissue.
5. Using a plastic pipette, rinse prism and sample well with distilled or deionized water. Wipe dry. The instrument is ready for the next sample.
TO CHANGE MEASUREMENT UNITTO CHANGE MEASUREMENT UNIT
TO CHANGE MEASUREMENT UNIT
TO CHANGE MEASUREMENT UNITTO CHANGE MEASUREMENT UNIT
Press the RANGE key to select the measurement units. The instrument toggles between the two measurement scales each time the key is pressed and the primary display indicates “Vol” % Volume or “FrPt” for Freezing Point. When the instrument displays the screen with 4 dashes, the instrument is ready for measurement. A number on the display indicates the selected unit: “1” denotes % Volume and “2” denotes Freezing Point (as indicated on the instrument cover).
Note: The temperature measurement is not displayed in Freezing Point mode.
TO CHANGE TEMPERATURE UNITTO CHANGE TEMPERATURE UNIT
TO CHANGE TEMPERATURE UNIT
TO CHANGE TEMPERATURE UNITTO CHANGE TEMPERATURE UNIT
To change the temperature measurement unit from Celsius to Fahrenheit (or vice versa), follow this procedure.
1. Press and hold the ON/OFF key continuously for approximately 8 seconds. The LCD will display the “all segment” screen followed by a screen with the model number on the primary display and the version number on the secondary display. Continue pressing the ON/OFF key.
8 seconds
2. While continuing to hold the ON/OFF key, press the ZERO key. The temperature unit will change from °C to °F or vice versa.
Note: The selected unit will be used in Freezing Point mode.
9
Page 10
MAKING A STANDARD SOLUTIONMAKING A STANDARD SOLUTION
MAKING A STANDARD SOLUTION
MAKING A STANDARD SOLUTIONMAKING A STANDARD SOLUTION
To make a Ethylene Glycol Solution, follow the procedure below:
To make an X % ethylene glycol solution volumetrically add X mL of high purity ethylene
glycol (CAS #: 107-21-1; MW 62.068) to a 100 mL Grade A volumetric flask.
Use distilled or deionized water to bring the flask close to the total volume, mix, allow the
solution to return to room temperature.
Once the solution has returned to room temperature use distilled or deionized water to bring
the total volume to 100 mL. Mix solution well before use.
locylGenelyhtEemuloVlatoTeulaVtnioPgnizeerFdetcepxE
V%01 Lm00.01Lm00.001)2.52(C°8.3-
V%04 Lm00.04Lm00.001)4
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
To replace the instrument’s battery, follow these steps:
• Make sure the instrument is off.
• Turn instrument upside down and remove the battery cover by turning it counterclockwise.
• Extract the battery from its location.
• Replace with a new 9V battery making sure to observe polarity.
• Insert the back battery cover and fasten it by turning clockwise to engage.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
The HI 96831 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
.51-(C°3.62-
10
Page 11
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGESERROR MESSAGES
edoCrorrEDCLnoitpircseD
“rrE“
otrewopelcyC.eruliaflareneG tcatnoc,stsisreprorrefI.tnemurtsni
.annaH
“OL“ yalp
“IH“ yalpsidyramirP
“OL“ yalpsidyramirP
“LAC“ NOtnemges
“IH“ yalpsidyramirP
“LAC“ NOtnemges
“LAC“ NOtnemges
“IHt“ ya
“LAC“ NOtnemges
“riA“ .derevocyltneici
“tLE“
“tLn“
sidyramirP
m
tsid OREZ .
“OLt“ yalpsidyramirP
lpsidyramirP
.dnahhtiw
thgilDEL
.annaHtcatnoC
.egnartnemerusaem
.egnartnemerusae
sserP.retawdellitsid
sserP.retawdelli
ffusniecafrusmsirP
muminimsdeecxeelpmaS
mumixamsdeecxeelpmaS
orezotdesunoitulosgnorW rodezinoiedesU.tnemurtsni
OREZ .
orezotdesunoitulosgnorW rodezinoiedesU.tnemurtsni
timilwolCTAsdeecxeerutarepmeT
.noitarbilacgnirud)C°0(
timilhgihCTAsdeecxeerutarepmeT
.noitarbilacgnirud)C°04(
rofthgillanretxehcumooT
llewelpmasrevoC.tnemerusaem
.detcetedtonsi
0.0 “ro“ C°0.08
CTA “gniknilbtnemges
PUTES “gniknilbtnemges
gniknilbtnemgesyrettaB.gniniamersiefilyrettabfo%5<
gniknilberaseulaverutarepmeT
11
.)C°08ot0(egnar
stuO
.)C°04ot0(egnar
.tsolnoitarbilacyrotcaF
.annaHtcatnoC
fotuotnemerusaemerutarepmeT
noitasnepmocerutarepmetedi
Page 12
Hanna Instruments Inc.
Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA
Local Sales and Customer Service office
Hanna Instruments United States Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
Tel. (800) 426 6287
Fax (401) 765 7575
www.hannainst.com/usa
Technical Support for customers
Telephone (800) 426 6287
Fax (401) 765 7575
E-mail tech@hannainst.com
Printed in Romania, EU
MAN96831 10/09
12