INSTRUCTION MANUAL
HI 96769HI 96769
HI 96769
HI 96769HI 96769
Anionic Surfactants
ISM
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument.
Please read it carefully before using the meter. If you need additional
technical information, do not hesitate to e-mail us at tech@hannainst.com.
Preliminary examination
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96769 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and
accessories see “Accessories”
Technical specifications
Range 0.00 to 3.50 mg/L (as SDBS)
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±3% of reading @ 25°C
Typical EMC Deviation ±0.01 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell with narrow band inter-
Method Adaptation of the USEPA method 425.1 and
Environment 0 to 50°C (32 to 122°F); max 95% RH
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in
Dimensions 192 x 102 x 67 mm
Weight 290 g (10 oz.).
ference filter @ 610 nm
Standard Methods for the Examination of
Water and Wastewater, 20th edition, 5540C,
Anionic Surfactants as MBAS.
non-condensing
after 1 hour of non-use in
with last reading reminder.
(7.6 x 4 x 2.6")
measurement mode
calibration mode
;
;
Functional description
1) GLP/ key: press to enter
edit the date and time.
2) CAL CHECK key: press to perform the validation of the meter, or press
and hold for three seconds to enter the
3) ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4) READ/ /TIMER key: In
measurement, or press and hold for three seconds to start a preprogrammed countdown prior to measurement. In
view the next screen.
5) ON/OFF key: to turn the meter on and off.
6) Liquid Crystal Display (LCD)
7) Cuvette alignment indicator
8) Cuvette holder
GLP mode
. In
calibration mode
calibration mode
measurement mode
press to
.
, press to make a
GLP mode
press to
DISPLAY ELEMENTS DESCRIPTION
1) The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2) Error messages and warnings
3) The battery icon shows the charging level of the battery
4) The hourglass appears when an internal checking is in progress
5) Status messages
6) The chronometer appears when the reaction timer is running
7) The month, day and date icons appear when a date is displayed
8) Four digit main display
9) Measuring units
10) Four digit secondary display
Errors and warnings
ON ZERO READING
ON SAMPLE READING
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light:The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
Inverted cuvettes: The sample and the
zero cuvette are inverted.
Zero: A zero reading was not taken. Follow
the instructions of the measurement
procedure for zeroing the meter.
Under range: A blinking “0.00” indicates that
the sample absorbs less light than the zero reference.
Check the procedure and make sure you use the
same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition. The
concentration of the sample is beyond the programmed
DURING CALIBRATION PROCEDURE
range: dilute the sample and re-run the test.
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
OTHER ERRORS AND WARNINGS
Cap error: Appears when external light enters in
the analysis cell. Assure that the cuvette cap is
present.
Cooling lamp: The instrument waits for the
lamp to cool down.
Battery low: The battery must be replaced
soon.
Dead battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, the meter will lock up.
Change the battery and restart the meter.
Measurement procedure
3
4
5
6
7, 13
8, 14
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
3• Fill the graduated glass vial with 25 mL of
sample.
For most accurate results, use of a class A
laboratory pipette is strongly recommended.
Alternatively, it is possible to use the plastic pipette
to fill the glass vial up to the 25 mL mark.
4• Add 2 drops of HI 95769A-0 Anionic
Surfactants Reagent A and 2 drops of
HI 95769B-0 Anionic Surfactants Reagent B.
5• Close the glass vial with its cap and invert
to mix. The solution will turn blue.
6• Add exactly 10 mL of Chloroform reagent.
For most accurate results, use a class A glass
laboratory pipette is strongly recommended.
Alternatively, it is possible to use another
clean plastic pipette to fill the vial up to the
10 mL mark.
Note: Chloroform is heavier than water
thus it will sink to the bottom of the
graduated glass vial. Do not mix up the
pipettes.
7• Close tightly the glass vial with its cap and
shake it vigorously for 30 seconds.
Note: Block the cap with a finger during shaking.
8• Press and hold
3’’
seconds and the display will show the
countdown. The beeper is playing a beep
READ/READ/
READ/
READ/READ/
/TIMER/TIMER
/TIMER for three
/TIMER/TIMER
at the end of countdown period.
Alternatively, wait for 2 minutes leaving the
vial capped and undisturbed.
or
10
11-12
8
16
Notes: The solution in the cuvette must be limpid.
If the solution is clouded, you can improve separation between
the chloroform and aqueous layers by gently warming the cuvette
(for instance by holding the cuvette in the hand).
If the chloroform layer contains some aqueous drops hanging on
the cuvette wall, gently swirl or invert the cuvette.
It is important to transfer at least 7 mL of chloroform layer into the
measurement cuvette, thus up to 0.5 cm (1/4”) below the 10 mL mark.
If the transferred volume is lower than 7 mL, the accuracy of the test
may be affected. Please repeat the test waiting for longer than 2
minutes, to allow complete separationbetween the two phases.
17
18
19, 23
During this period, the chloroform layer
separates from the aqueous layer: the color
of the aqueous layer will fade slightly if
anionic surfactants are present, while the
chloroform layer will turn blue.
9• Remove the cap.
10•Using the long plastic pipette, remove the
upper aqueous layer and discard. Do not
remove the lower chloroform layer.
11•Add to the vial about 15 mL of deionized
water, up to the 25 mL mark.
12•Add 2 drops of HI 95769A-0 Anionic
Surfactants Reagent A.
13• Close tightly the glass vial with its cap and
shake it vigorously for 30 seconds.
Note: Block the cap with a finger during shaking.
14•Press and hold
seconds and the display will show the
countdown. The beeper is playing a beep
at the end of countdown period.
Alternatively, wait for 2 minutes leaving the
vial capped and undisturbed.
During this period, the chloroform layer
separates from the aqueous layer.
15•Remove the cap.
16•Insert a
upper aqueous layer to transfer only the
lower chloroform layer into a cuvette paying attention
not to transfer the upper aqueous layer too.
17•Cap the cuvette. This is the reacted sample (#2).
18•Fill another cuvette with 10 mL of Chloroform
reagent up to the 10 mL mark and place
the cap. This is the blank (#1).
19•Place the blank (cuvette #1) into the
cuvette holder and ensure that the notch
on the cap is positioned securely into the
groove.
20•Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
21•After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and
READ/READ/
READ/
READ/READ/
clean plastic pipette below the
/TIMER/TIMER
/TIMER for three
/TIMER/TIMER
22•Remove the cuvette.
23•Place the reacted sample (cuvette #2) into
the cuvette holder and ensure that the notch
on the cap is positioned securely into the
groove.
READ/READ/
READ/
READ/READ/
/TIMER/TIMER
. The lamp, cuvette
/TIMER
/TIMER/TIMER
24•Press
and detector icons will appear on the display,
depending on the measurement phase.
25•At the end of measurement, the instrument
directly displays concentration in mg/L (ppm) of
anionic surfactants (MBAS) as SDBD on the LCD.
20-21
24
CONVERSION FACTORS
The anionic detergent content is expressed as
mg/L of SDBS.
Convert the reading to mg/L (ppm) of a different anionic detergent concentration (DX) of known
molecular weight (MWDX), as follows:
DX (mg/L) DX (mg/L)
DX (mg/L)
DX (mg/L) DX (mg/L)
INTERFERENCES
Cationic surfactants negative interference
Absorption particulate matter negative interference
Sulfide negative interference
Organic sulfates, sulfonates positive interference
Strong oxidants (Cl2, H2O2, S2O
Highly buffered samples or with extreme pH may exceed the buffering
capacity of the reagent: pH should be adjusted between 4 and 9 with
diluted NaOH for acidic samples or with diluted HCl for alkaline samples,
prior to addition of the reagent.
Reading (mg/L SDBS) x MW
=
348.48
2-
, etc.) negative interference
8
25
DX
Validation and Calibration procedures
Warning: do not validate or calibrate the instrument with standard
solutions other than the Hanna CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate
instruments.
VALIDATION
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Place the CAL CHECK™ Standard
HI 96769-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
4• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for validation.
6• Remove the cuvette.
7• Place the CAL CHECK™ Standard
Validation
3
4-5
HI 96769-11 Cuvette B into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
8• Press CAL CHECK key and the lamp,
cuvette and detector icons together with
“CAL CHECK” will appear on the display,
depending on the measurement phase.
9• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as
reported on the CAL CHECK™ Standard
Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.
CALIBRATION
Note: It is possible to interrupt the calibration
procedure at any time by pressing CAL
CHECK or ON/OFF keys.
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Press and hold CAL CHECK for three
seconds to enter
display will show “CAL” during calibration
procedure. The blinking “ZERO” asks for
instrument zeroing.
4• Place the CAL CHECK™ Standard HI
96769-11 Cuvette A into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
6• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
7• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for calibration. The blinking “READ”
asks for reading calibration standard.
8• Remove the cuvette.
9• Place the CAL CHECK™ Standard
HI 96769-11 Cuvette B into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
10•Press READ/ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the measurement
phase.
11•The instrument will show for three seconds
the CAL CHECK™ standard value.
Note: If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the standard
value was too low. Verify that both CAL
CHECK™ Standard HI 96769-11 Cu-
calibration mode.
The
vettes, A and B are free from fingerprints
7
8
9
Calibration
3
4
6-7
9
10-11
or dirt and that they are inserted correctly.
Then the date of last calibration (e.g.:
“01.08.2005”) appears on the display, or
“01.01.2005” if the factory calibration
was selected before. In both cases the year
number is blinking, ready for date input.
12•Press GLP/ to edit the desired year
(2000-2099). If the key is kept pressed,
the year number is automatically increased.
13•When the correct year has been set, press
ZERO/CFM or READ/ /TIMER to confirm.
Now the display will show the month blinking.
14•Press GLP/ to edit the desired month
(01-12). If the key is kept pressed, the
month number is automatically increased.
15•When the correct month has been set,
press ZERO/CFM or READ/ /TIMER to
confirm. Now the display will show the
day blinking.
16•Press GLP/ to edit the desired day (01-
31). If the key is kept pressed, the day
number is automatically increased.
Note: It is possible to change the editing from
day to year and to month by pressing
READ/ /TIMER.
17• Press ZERO/CFM to save the calibration date.
18•The instrument displays “Stor” for one
second and the calibration is saved.
19•The instrument will return automatically
to
measurement mode
on the LCD.
by displaying dashes
GLP
In the
GLP mode
factory calibration can be restored.
LAST CALIBRATION DATE
1• Press GLP/ to enter
2• If no calibration was performed, the factory
FACTORY CALIBRATION RESTORE
It is possible to delete the calibration and restore
factory calibration.
1• Press GLP/ to enter
, the last calibration date can be consulted and the
GLP mode
calibration month and day will appear on
the main display and the year on the
secondary display.
calibration message, “F.CAL” will appear
on the main display and the instrument
returns to
measurement mode
. The
after three
seconds.
GLP mode
.
12-13
14-15
16
17-18
1
2
or
or
Last
Calibration
Date
2
3-4
Factory
Calibration
Restore
2• Press READ/ /TIMER to enter in the
factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.
3• Press ZERO/CFM to restore the factory
calibration or press GLP/ again to
abort factory calibration restore.
4• The instrument briefly notifies “done” when
restores factory calibration and returns to
measurement mode
.
Battery management
To save battery, the instrument shuts down after 10 minutes of non-use
in
measurement mode
If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows “
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover by
turning it counterclockwise.
• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.
and after 1 hour of non-use in
dead battdead batt
dead batt” and turns off.
dead battdead batt
calibration mode.
Accessories
REAGENT SETREAGENT SET
REAGENT SET
REAGENT SETREAGENT SET
HI 95769-01 Reagents for 40 Anionic Surfactants tests
OTHER ACCESORIESOTHER ACCESORIES
OTHER ACCESORIES
OTHER ACCESORIESOTHER ACCESORIES
HI 96769-11 CAL CHECK
TM
Standard Cuvettes (1 set)
HI 70000P 20 mL Deionized rinse solution (25 pcs)
HI 740220 25 mL glass test tubes with caps (2 pcs)
HI 721310 9V battery (10 pcs.)
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731331 Glass cuvettes (4 pcs.)
HI 731335 Caps for cuvettes
HI 93703-50 Cuvette cleaning solution (230 mL)
Warranty
HI 96769 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified of
the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the
design, construction and appearance of its products
without advance notice.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find the Hanna Office in your area,
IST96769 01/11
visit our web site
www.hannainst.com