Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you with the necessary
information for the correct use of the instrument. Please read it carefully before using the meter. If
you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner,
Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
2
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not damaged. If any
damage occured during shipment, please notify your Dealer.
Each HI 96765 Blood Plasma Analyzer is supplied complete with:
• Light Blocking Cap
• 9V Battery
• Cloth for wiping tubes
• Instrument quality certificate
• Instruction Manual
• Rigid carrying case
Note:Save all packing material until you are sure that the instrument works correctly. Any
defective item must be returned in its original packing.
3
GENERAL DESCRIPTION
The HI 96765 meter measures the light absorbance of animal blood plasma: the absorbance
value allows identification of the plasma quality.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the instrument off after 10 minutes of non-use.
This instrument has been developed to measure centrifuged animal blood, coming from abattoirs.
The quality of the blood is checked by measuring its color, expressed in Quality Units. When the
blood cells are damaged, haemoglobin is found in the plasma, which causes high readings. The
lower are the readings in Quality Units (less color), the higher is the blood plasma quality.
This Hanna specialty meter reads directly the blood plasma quality from 0 to 10 units.
ABBREVIATIONS
degree Celsius
°C:
degree Fahrenheit
°F:
milliliter
mL:
mV:
millivolts
SPECIFICATIONS
Range0.0 to 10.0 units
Resolution0.1 unit
Accuracy±0.2 units or 3% of reading @ 25°C (whichever is greater)
Typical EMC Deviation ±0.1
Light SourceLight emitting diode @ 555 nm
Light DetectorSilicon Photocell
MethodDirect Measurement
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type1 x 9 volt
Auto-Shut offAfter 10' of non-use
Dimensions192 x 102 x 67 mm (7.6 x 4 x 2.6")
Weight290 g (10 oz.).
REQUIRED REAGENTS
DescriptionQuantity/test
Blood Animal1.5 mL
4
PRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the physical-chemical characteristics of the substance according to the
Lambert-Beer Law:
-log I/Io = ελ c d
or
A = ελ c d
Where:
-log I/I
Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's HI 96 series colorimeters is based on special subminiature
tungsten lamps and narrow-band interference filters to guarantee both high performance and
reliable results.
=Absorbance (A)
o
Io=intensity of incident light beam
I=intensity of light beam after absorption
ελ=molar extinction coefficient at wavelength λ
c=molar concentration of the substance
d=optical path through the substance
HI 96 series block diagram (optical layout)
5
A microprocessor controlled special tungsten lamp emits radiation which is first optically
conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the
diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to
obtain a light beam of intensity Io or I.
The photoelectric cell collects the radiation I that is not absorbed by the sample and converts
it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the
actual measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires
particular attention. It is important that both, the measurement and the calibration (zeroing)
cuvettes, are optically identical to provide the same measurement conditions. Whenever possible
use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not
scratched. This to avoid measurement interference due to unwanted reflection and absorption of
light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvette to prevent any contamination.
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION
1) CAL key
2) READ key
3) ON/OFF key
4) Liquid Crystal Display (LCD)
5) Tube alignment indicator
6) Tube holder
6
KEYPAD DESCRIPTION
6
5
2
3
4
1
• ON/OFF: to turn the meter on and off.
• CAL: to calibrate the meter prior to measurement.
• READ: to make a measurement.
OPERATING MODES
•
Measurement mode:
default operation mode, enables both calibration and measurement.
DISPLAY ELEMENTS DESCRIPTION
1) The measuring scheme (lamp, tube, detector), appears during different phases of calibration
or reading measurement
2) Error messages and warning
3) The batery icon shows the charging level of the batery
4) The hourglass appears when an interval checking is in progress
5) Four digit main display
6) Four digit secondary display
7
ERRORS AND WARNINGS
The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.
a) on calibration
Light High: There is too much light to perform a measurement.
Light Low: There is not enough light to perform a measurement.
No Light: The instrument cannot adjust the light level. Please
check that the tube holder does not contain any debris.
b) on sample reading
Cal: Calibration was not performed. Follow the instructions of
the measurement procedure for calibrating the meter.
Under range: A blinking “0.0” indicates that the sample
absorbs less light than the zero reference.
8
c) other errors and warnings
Over Range: A flashing value of the maximum concentration
indicates an over range condition.
Cap error: An excess ambient light is reaching the detector.
Please place the light blocking cap and press CAL again. If the
light blocking cap is properly installed, and a calibration can not
be performed, then contact your dealer or the nearest Hanna
Customer Service Center.
Cooling lamp: The instrument waits for the lamp to cool down.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must
be replaced. Once this indication is displayed, the meter will
lock up. Change the battery and restart the meter.
9
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure best accuracy.
• Color or suspended matter in large amounts may cause
interference, therefore these should be removed by
treatement with active carbon and by prior filtration.
• For a correct filling of the tube: the liquid in the tube
forms a concavity on the top; the bottom of this concavity
must be at the same level of the 1.5 mL mark.
• It is important that the sample does not contain any debris. This would corrupt the reading.
• Whenever the tube is placed into the measurement cell, it must be dry outside, and
completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free
cloth prior to insertion.
• It is possible to take multiple readings in a row, but it is recommended to make a new
calibration for each sample.
STARTUP
Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument tube holder.
• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.
10
MEASUREMENT PROCEDURE
To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is calibrated. After the sample is prepared, the sample is measured.
• Turn the meter on by pressing ON/OFF.
The display briefly shows all tags on.
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “CAL” indicates that the
instrument needs to be calibrated first.
• Fill a standard 10 mm tube with
1.5 ml of the blood sample.
• Use a centrifuge to separate the blood
cells from the plasma and to obtain a
clear supernatant of plasma, without any
turbidity or suspended blood cells.
• Please verify that the height of the
deposits does not exceed 12 mm (from
the bottom of the tube), and that the
tube is filled with plasma, up to, at
least, 19 mm.
11
• Press CAL and the lamp, tube and
CAL
detector icons will appear on the display,
depending on the measurement phase.
• Wait for a few seconds and the display
will show “-0.0-”. The meter is now
calibrated and ready for measurement.
NOTE:If the “CAP” message appears, then there is too much ambient light. Please place
the light blocking cap and press CAL again.
12
• Insert the centrifuge tube into the holder.
READ
• Press READ and the lamp, tube and
detector icons will appear on the display,
depending on the measurement phase.
• The instrument directly displays the Blood
Plasma Quality Units on the Liquid Crystal
Display.
INTERFERENCES
Suspended blood cells and solids: repeat pretreatment.
Strong ambient light: use the light blocking cap for calibrating the instrument, and while taking
readings.
BATTERY MANAGEMENT
To save battery, the instrument shuts down after 10 minutes of non-use.
If a valid measurement was displayed before
auto-shut off, the value is displayed when
the instrument is switched on. The blinking
“CAL” means that a new calibration has to
be performed.
One fresh battery lasts for around 750 measurements, depending on the light level.
The remaining battery capacity is evaluated at the instrument startup and after each measurement.
13
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken anymore, the instrument
dead battdead batt
shows “
dead batt” and turns off.
dead battdead batt
To restart the instrument, the battery must be replaced with a fresh one.
BATTERY REPLACEMENT
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.
• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.
ACCESORIES
HI 7213109V battery (10 pcs)
HI 731318Cloth for wiping cuvettes (4 pcs)
HI 741218Carrying case
14
WARRANTY
HI 96765 is warranted for two years against defects in workmanship and materials when used
for its intended purpose and maintained according to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure. If the repair is not covered by the warranty,
you will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization
Number from the Customer Service Department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the
date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for the environment in
which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator
to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or
store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the design, construction and appearance of its
products without advance notice.
15
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA