CE Declaration of Conformity ................... 45
ISO 9000 Certified
Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no
damage has occurred during shipping. If there
is any damage, notify your Dealer.
Each bench D.O. meter HI964400 comes
supplied complete with:
• HI 76407/2 D.O. probe with 2 m (6.7')
cable
• HI 76407A membrane cap (2 pcs)
• HI 7041S electrolyte solution (30 ml)
• 12 VDC power adapter (HI 710005 or
HI 710006)
• dust cover.
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packing with the
supplied accessories.
GENERAL DESCRIPTION
HI964400 is a bench microprocessor-based,
logging meter for Dissolved Oxygen measurements.
It can store in memory up to 99 lots with up
to 8,000 readings. These readings can be
transferred to a computer system for elaboration or permanent storage.
Dissolved oxygen is indicated in ppm (parts
per million) or in %.
Temperature is compensated for by the meter's
ATC circuitry. Salinity compensation in water
allows direct determination of dissolved oxygen in saline waters and the altitude compensation readjusts for the altitude variance.
1
The dissolved oxygen probe has a membrane
covering the polarographic sensors and a builtin thermistor for temperature measurements
and compensation.
This permeable PTFE membrane isolates
the sensor elements from the testing solution, but allows oxygen to pass through. When
a voltage is applied across the sensor, oxygen that has passed through the membrane
reacts causing a current flow, and hence
determining a reading.
2
FUNCTIONAL DESCRIPTION PROBE
3
4
4
1
7
8
10
9
5
6
1. D.O. Probe
2. Protective Cap
3. Watertight Shielded Cable
4. Polypropylene Probe Body
5. Temperature Sensor
6. O-Ring Seal
7. Silver Chloride Anode
8. Platinum Cathode (sensor)
9. Oxygen Permeable PTFE Membrane
10. Membrane Cap
3
2
FUNCTIONAL DESCRIPTION OF HI964400
THE FRONT PANEL
1. Liquid Crystal Display
2. Keyboard:
ALT keyto display altitude settings
CAL keyto enter or exit the calibra-
tion mode; to enable or disable the date or time settings; to start or exit the D.O.
logging mode
CFM keyto confirm the calibration val-
ues
COMM key to display the RS232C baud
rate and command prefix
DATE keyto display the date
LOG keyto display the lot number on
the primary LCD and the page
number on the secondary
one. Press again to display
the sample number of the lot
!!
! keyto select the calibration value;
!!
to set date, time, logging interval; to set the altitude and
the salinity factors; to set
the baud rate and command
4
prefix for communication to
PC
NaCl keyto display salinity settings
OFF keyto turn the meter on off
ON keyto turn the meter on
RANGE key to select dissolved oxygen
measurement mode in ppm
or in %
TEMP keyto select temperature mea-
surement mode
TIME keyto display the time
""
" keyto select the calibration value;
""
to set date, time, logging interval; to set the altitude and
the salinity factors; to set
the baud rate and command
prefix for communication to
PC.
THE REAR PANEL
1. Socket for D.O. Probe
2. Reset Button
3. RS 232C Connector
4. DC Power Socket (for HI710005 or
HI710006)
Unplug the meter from the power supply
before replacing the fuse.
5
SPECIFICATIONS OF HI964400
RangeD.O. ppm0.00 to 45.00
Resolution D.O. ppm0.01
AccuracyD.O.±1.5% of full scale
Typical EMCD.O.±1.5% of full scale
DeviationTemp. °C±0.5
CalibrationD.O. single or double point at 0%
Altitude Compensation0 to 1,900 m (6,230')
Salinity Compensation0 to 40 g/l
Temperature
Compensation0.0 to 50.0°C (32 to 122°F)
ProbeHI76407/2 with
Logging Interval1, 15, 30 seconds
ComputerRS 232C (optoisolated)
Interface
PowerPower socket for 12 VCD
Environment0 to 50°C (32 to 122°F);
Dimensions230x170x70 mm
Weight1 kg (2.2 lb.)
Response Time The response time is approximately
20 seconds for a 95% reading at a constant temperature
of 25°C. The response time for low oxygen readings or
at low temperature is approximately 40 seconds. Allow
more time to obtain more accurate readings.
D.O. %0.0 to 300.0
Temp. °C0.0 to 50.0
D.O. %0.1
Temp. °C0.1
Temp. °C±0.5
(HI 7040) and 100% (in air)
Temp. °C single point or double point
at 0.0°C and/or 50.0°C
Resolution100 m (328')
Resolution1 g/l
2 meters (6.7') cable
or 1, 2, 5, 15, 30, 60, 120,
180 minutes
95% RH
(9.1x6.7x2.7")
6
LCD FUNCTIONAL DESCRIPTION
Pri
mary Display
DATEppmTIME
CFM MEM
WRONG
CAL
LOG
BUF
°C
1
2
INTV
m
S
Secondary Display
OPERATIONAL GUIDE
POWER CONNECTION
Plug the 12VDC adapter to the meter and to
the mains.
HI964400 uses an EEPROM to retain the
D.O. calibration and temperature calibration
as well as the serial communication setting.
The instrument will store the respective data
after a calibration or serial communication
setting, even when it is not plugged-in.
7
PROBE CONNECTION & PREPARATION
S
To prepare the instrument for use, connect
the D.O. probe to the meter securely by
aligning the pins with the socket located on
the back of the meter, pushing the plug in
and tightening the threaded ring.
All probes shipped from Hanna Instruments
are dry. To hydrate the probe and prepare it
for use, connect it to the meter and proceed
as follows.
1. Remove the red and black plas-
hipping
cap
tic cap. This cap is for shipping
purposes and can be thrown
away.
2. Wet the sensor by soaking the
bottom 2½ cm (1") of the probe
black
red
in electrolyte (HI 7041S) for 5 minutes.
3. Rinse the membrane cap (HI 76407A supplied in the kit with the meter) with electrolyte solution while shaking it gently.
Refill with clean electrolyte solution.
4. Gently tap the sides of the membrane cap
with your finger tip to
ensure that no air
bubbles remain trapped.
To avoid damaging the
membrane, do not tap
the membrane directly
on the bottom.
FILL FIRST
THEN TAP
5. Make sure that the rubber O-ring sits properly
inside the membrane
cap.
THEN SCREW
BACK ON
6. With the sensor facing down,
slowly screw the cap clockwise.
Some electrolyte will overflow.
When not in use and during polarization (see below), place the protective transparent cap supplied in
the kit with the meter.
8
TURNING THE METER ON
To switch the meter on, press
ON
the ON key and let the probe
in the auto-conditioning (polarization) mode before proceeding. After approximately
5 minutes, the instrument can
be calibrated (see page 17).
If the probe is disconnected,
the meter will display "----".
ppm
°C
This also indicates the possibility of a broken
probe cable.
PROBE POLARIZATION
The probe is under polarization with a fixed
voltage of approximately 800 mV.
Probe polarization is essential for stable measurements with the same recurring degree of
accuracy.
With the probe properly polarized, oxygen is
continually "consumed" when it passes
through the sensitive diaphragm and dissolves
in the electrolyte solution contained in the
probe.
If polarization is interrupted, the electrolyte
solution continues to be enriched with oxygen until it reaches an equilibrium with the
surrounding solution.
Whenever measurements are taken with a
non-polarized probe, the oxygen level revealed
is both that of the tested solution as well as
that present in the electrolyte solution. This
reading is incorrect.
The calibration of this instrument is very
simple.
Before proceeding with the calibration make
sure the probe is ready for measurements
(see page 8), i.e. the membrane cap is filled
with electrolyte and the probe is connected
9
to the meter and properly polarized.
For an accurate calibration, it is recommended
to wait for 5 or 10 minutes to ensure precise
conditioning of the probe.
Keep the protective cap on during the polarization time and remove it for the calibration
and the measurements.
Follow the calibration procedure on page 17.
D.O. MEASUREMENTS (in ppm or in %)
Make sure the meter has been calibrated
(see page 17) and the protective cap has
been removed.
Salinity and Altitude compensation
If the sample contains significant concentration of salinity or if you are performing
measurements at an altitude different from
sea level, the readout values must be corrected, taking into account the lower degree of oxygen solubility in such occasions
as explained on pages 23-25.
Remember to set the altitude and/or the
salinity before taking any D.O. measurements. The meter will automatically compensate for these factors.
Taking measurements
Immerse the tip of the probe in the sample to
be tested. Make sure the temperature sensor
is also immersed.
Press the RANGE key to display D.O. measurements. Allow approximately one minute
for the meter to stabilize and
read the
ppm
value of dissolved oxygen on
the display.
RANGE
10
Press the RANGE key to change the reading
from ppm to % and vice-versa.
RANGE
For accurate dissolved oxygen measurements
a water movement of 0.3 m/sec is required
at a minimum. This is to ensure that the
oxygen-depleted membrane surface is constantly replenished. A moving stream will
provide adequate circulation.
The use of a magnetic stirrer to ensure a certain velocity in the fluid is recommended. In this way, errors
due to the diffusion of the
oxygen present in the air in
the solution are reduced to
a minimum.
At all times, the time necessary for thermal
equilibrium to occur between the probe and
the measurement sample must be allowed (a
few minutes if the temperature difference is
only several degrees).
TAKING TEMPERATURE MEASUREMENTS
The probe has a built-in temperature sensor.
Press the TEMP key to display the measured
temperature on the secondary display.
TEMP
11
°C
Allow the probe to reach the thermal equilibrium before taking any measurement. Reaching thermal equilibrium can take several minutes. The greater the difference between the
temperature at which the probe was stored
and the temperature of the sample, the longer
the time will be.
If "----" is displayed, it indicates that the D.O. probe is
not properly connected or the
°C
temperature is out of range.
This also indicates the possibility of a broken
probe cable.
SETTING THE DATE AND THE TIME AND
THE LOGGING INTERVAL
Press the DATE key to display the date. The
month and the day will be displayed on the
primary LCD, the year on the secondary one.
DATE
DATE
Press the CAL key to enter the setting mode,
the month starts blinking.
CAL
DATE
Use the UP or DOWN arrow keys to select
the month.
12
Press the RANGE key, the day starts blinking.
DATE
RANGE
Use the UP or DOWN arrow keys to select
the day.
Press the RANGE key and the year on the
secondary LCD will blink.
RANGE
DATE
Use the UP or DOWN arrow keys to select
the year.
Press the CAL key to
CAL
exit the date setting
mode.
13
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.