Hanna Instruments HI 96301 User Manual

Dear Customer, Thank you for choosing a Hanna product. If you need additional technical informa­tion, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
SPECIFICATIONS
Range/Resolution DiST1 (HI96301) 1990ppm / 10ppm DiST2 (HI96302) 10000ppm / 100ppm DiST3 (HI96303) 1990µS/cm / 10µS/cm DiST4 (HI96304)19900µS/cm / 100µS/cm
Accuracy (@20°C) ±2% Full Scale Calibration manual, 1 point Calibration solution (optional)
DiST1 (HI96301) HI70032P DiST2 (HI96302) HI70038P DiST3 (HI96303) HI70031P DiST4 (HI96304) HI70039P
Temperature Compensation
Automatic, 5 to 50°C (41 to 122°F)
Environment
0 to 50°C (32 to 122°F); RH 95% max
Battery type 4 x 1.5V alkaline Battery life approx. 150 hours of use Dimensions
150x30x21 mm (5.9x1.2x0.8")
Weight 70 g (2.5 oz.)
DiST® 1 • DiST® 2 DiST® 3 • DiST® 4
HI 96301 • HI 96302 HI 96303 • HI 96304
Pocket-Sized EC/TDS Meters with ATC
ACCESSORIES
HI 70031P 1413 µS/cm calibration sa- HI 70032P 1382 ppm calibration sa- HI 70038P 6.44 ppt calibration sachet HI 70039P 5000 µS/cm calibration sa-
chet (20 ml, 25 pcs) chet (20 ml, 25 pcs) (20 ml, 25 pcs) chet (20 ml, 25 pcs)
DiST is a registered Trademark of
"Hanna Instruments"
IST96301 12/04
OPERATION
• Remove the protective cap.
• Turn the meter ON sliding the switch on the top.
• Immerse into solution up to the max immersion level with­out touching the bottom of the beaker.
• Stir gently and wait until the display stabilizes. Readings are automatically compensated for temperature variations.
• The value on the display has to be multiplied by a factor of 10 for DiST1 & DiST3 and by a factor of 100 for DiST2 & DiST4 (as shown on the upper left corner of the LCD) to give direct reading in ppm or µS/cm.
CALIBRATION
• Immerse the tester up to the maximum immersion level in the appropriate cali­bration solution (see Specifications).
• Allow the reading to stabilize and using a small screwdriver adjust the calibra­tion trimmer to match the solution value. E.g. calibrating DiST1 with HI70032P solution, turn the trimmer to read "138" (x10 = 1380 ppm).
BATTERY REPLACEMENT
When the meter cannot be switched on or the display fades, pull out the battery compartment and replace all four 1.5V batter­ies, while paying attention to their polarity.
Batteries should only be re­placed in a safe area using the battery type specified in this instruction manual.
Recommendations for users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause interference to radio and TV equipment. The pins at the end of the sensor are sensitive to electrostatic discharges. Avoid touching this metal band at all times. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid damage or burns, do not perform any measurement in microwave ovens.
WARRANTY
These meters are guaranteed for six months against de­fects in workmanship and materials when used for their intended purpose and maintained according to instruc­tions. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protec­tion.