SPECIFICATIONS
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@20°C) ±0.1 pH
Calibration manual, 2 points
Temperature compensation
automatic, 0 to 50°C (32 to 122°F)
Environment RH 95% max;
0 to 50°C (32 to 122°F)
Batteries 3 x 1.5V alkaline
Life approx. 150 hours of use
Dimensions 150x30x21 mm
(5.9x1.2x0.8")
Weight 70 g (2.5 oz.)
ACCESSORIES
HI 7004M pH 4.01 solution (230 ml)
HI 7007M pH 7.01 solution (230 ml)
HI 7010M pH 10.01 solution (230 ml)
HI 70300M Storage solution (230 ml)
HI 7061M Cleaning solution (230 ml)
pHep®+
HI 96108
Pocket-Sized
pH Meter with A TC
WARRANTY
This meter is guaranteed for six months against defects in workmanship and materials when used for
its intended purpose and maintained according to instructions. This warranty is limited to repair or
replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required,
contact the dealer from whom you purchased the instrument. If under warranty, report the model number,
date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number
from the Customer Service department and then send
it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
IST96108 12/04
www.hannainst.com
OPERATION
• Remove the protective cap. Do
not be alarmed if white crystals
appear around the cap. This is
normal with pH electrodes and
they dissolve when rinsed with
water.
• Turn the meter on sliding the
switch on the top.
• Immerse into solution up to
the max immersion level.
• Stir gently and wait until the
display stabilizes.
• After use, rinse the electrode
with water to minimize contamination and store it with a
few drops of storage (HI70300)
or pH 7 (HI 7007) solution in
the protective cap.
DO NOT USE DISTILLED OR DEIONIZED
WATER FOR STORAGE PURPOSES.
• Large fluctuations in readings could be due
to lack of calibration, dry electrode or rundown batteries.
Recommendations for users
Before using this product, make sure that it is entirely
suitable for the environment in which it is used. Operation
of this instrument in residential areas could cause
interference to radio and TV equipment. The glass bulb
at the end of the pH electrode is sensitive to electrostatic
discharge. Avoid touching this glass bulb at all times. Any
variation introduced by the user to the supplied equipment
may degrade the instrument's EMC performance. To avoid
electrical shock, do not use this instrument when voltages
at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any
measurement in microwave ovens.
CALIBRATION
• Immerse the tester up to the maximum
immersion level in pH 7 buffer (HI 7007)
• Allow the reading to
stabilize; using a
small screwdriver
adjust the pH7 calibration trimmer to
read 7.0.
• Rinse the electrode with water and
immerse up to the maximum level in
pH 4 (HI 7004) or pH 10 (HI 7010).
• Allow the reading
to stabilize and
adjust the pH4/10
calibration trimmer to read 4.0
(or 10.0).
Calibration is now complete.
ALWAYS USE FRESH BUFFERS FOR CALIBRATION.
BATTERY REPLACEMENT
When the meter cannot be
switched on or the display
fades, pull out the battery compartment and replace all three
1.5V batteries, while paying
attention to their polarity.
Batteries should only be replaced in a safe area using
the battery type specified in
this instruction manual.
pHep is a registered Trademark of
"Hanna Instruments"