INSTRUCTION MANUAL
HI 95747HI 95747
HI 95747
HI 95747HI 95747
Copper Low Range ISM
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with
directives.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 95747 Ion Selective Meter is supplied complete with:
• Two Sample Cuvets and Caps
• 9V Battery
• Instruction Manual
Note: save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and
accessories see “Accessories”
Technical specifications:
Range 0.000 to 1.500 mg/L
Resolution 0.001 mg/L for
0.01 mg/L for
Precision ±0.015 mg/L @ 0.750 mg/L
Typical EMC Dev. ±0.001 mg/L
Light Source Tungsten lamp with narrow band interfer-
ence filter @ 560 nm
Light Detector Silicon Photocell
Method Adaptation of the USEPA approved
bicinchoninate method.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8”)
Weight 290 g (10 oz.).
measurement
calibration
and
validation
Functional description:
1. Liquid Crystal Display.
2. Cuvet Holder with transparent protective
cup and alignment indicator.
3. ON/OFF key: to turn the meter on and
off.
4. ZERO key: press to zero the meter prior to
measurement.
5. READ/TIMER key: press for making a
measurement, or hold the key for 3
seconds to start a pre-programmed
countdown prior to measurement.
6. CAL CHECK key allows direct validation
with the exclusive Hanna CAL CHECK™
NIST traceable standards.
Guide to display codes:
This prompt appears for 1 second each time the
instrument is turned on.
The parameter code “747” indicates that the
meter is in a ready state and zeroing can be
performed.
Sampling In Progress. Flashing “SIP” prompt
appears each time the meter is performing a
measurement.
“-0.0-”, the meter is in a zeroed state and
measurement can be performed.
“CAP”, Light over range: the cuvet is not inserted
correctly and an excess ambient light is reaching
the detector. If the cover is properly installed,
then contact your dealer or the nearest Hanna
Customer Service Center.
The blinking “BAT” indicates that the battery
voltage is getting low and the battery needs to
be replaced.
“-bA-”, the battery is dead and must be replaced. Once this indication is displayed, the
meter will lock up. Change the battery and
restart.
“Conf”, the meter has lost its configuration. Contact
your dealer or the nearest Hanna Customer Service
Center.
CALIBRATION MODE MESSAGES
The date of the last calibration appears on the
LCD each time the
calibration is performed for the first time, “F.CAL”
appears.
“F.CAL” indicates that the factory calibration is
selected.
The flashing digits indicate that:
1. the month can be set
2. the day can be set
“Stor”, appears for 1 second at the end of the
calibration procedure, to indicate that the
calibration data has been stored.
“Error”, the concentration of the calibration solution
used is not correct. Repeat the calibration procedure
with the right standard solution, and verify it is
not expired.
If the calibration procedure fails again, contact
your dealer or the nearest Hanna Customer Service
Center.
ERROR MESSAGES
On zero reading
Blinking “-0.0-” indicates that the zeroing
procedure failed due to a low signal-to-noise
ratio. In this case press ZERO again.
“no L”, the instrument cannot adjust the light
level. Please check that the sample does not
contain any debris.
calibration mode
is entered. If
“L Lo”, there is not enough light to perform a
measurement. Please check the preparation of the
zero cuvet.
“L Hi”, there is too much light to perform a
measurement. Please check the preparation of the
zero cuvet.
On sample reading
“-SA-”, there is too much light for the sample
measurement. Please check if the right sample
cuvet is inserted.
“Inv”, the sample and the zero cuvet are inverted.
“ZErO”, a zero reading was not taken. Follow the
instruction in the measurement procedure for zeroing the meter.
Under range. A blinking “0.000” indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure you
use the same cuvet for reference (zero) and
measurement.
A flashing value of the maximum concentration
indicates an over range condition. The concentration of the sample is beyond the programmed
range: dilute the sample and re-run the test.
A flashing value lower than the maximum concentration indicates a low signal-to-noise ratio
condition. In this case accuracy of the result is not
guaranteed. Repeat the measurement procedure.
Measurement procedure:
Measurement
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays “747”, it is ready.
3• Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
4• Place the cuvet into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
5• Press ZERO and “SIP” will blink on the
display.
6• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready
for measurement.
7• Remove the cuvet and add the content of
one packet of HI 95747-0 Copper Low Range
reagent.
8• Replace the cap and shake gently for 15
seconds.
9• Replace the cuvet into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
10•Hold READ/TIMER for three seconds. The
display will show the countdown prior to
measurement.
Alternatively, wait for 45 seconds and just
press READ/TIMER.
In both cases “SIP” will blink during
measurement.
11•The instrument directly displays concentration
in mg/L of copper on the Liquid Crystal
Display.
INTERFERENCES
• Cyanide
• Silver
• For strongly buffered alkaline or acidic samples,
pH should be adjusted between 6 and 8 before
addition of reagent.
• To avoid interferences due to fingerprints, oil or
dirt it is very important that the cuvet is wiped
clean prior to insertion in the cuvet holder.
Replacement of scratched cuvets is strongly
recommended.
Validation and Calibration procedures
Warning: do not validate or calibrate the instrument with standard
solutions other than the Hanna CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).
Use the optional CAL CHECK™ cuvets (see
“Accessories”) to validate or calibrate
instruments.
VALIDATION
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays “747”, it is ready.
3• Place the CAL CHECK™ Standard Cuvet A
into the holder and ensure that the notch on
the cap is positioned securely into the groove.
4• Press ZERO and “SIP” will blink on the
display.
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready
for validation.
6• Remove the cuvet.
7• Place the CAL CHECK™ Standard
HI 95747-11 Cuvet B into the holder and
ensure that the notch on the cap is positioned
securely into the groove.
8• Press CAL CHECK key and “SIP” will blink
during measurement.
9• Wait for a few seconds and the display will
show the validation standard value.
Validation
The reading should be within specifications as
reported on the CAL CHECK™ Standard
Certificate. If the value is found out of
specifications, please check that the cuvets are
free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.
CALIBRATION
Note: It is possible to interrupt the calibration
procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument
with standard solutions other than the Hanna
CAL CHECK™ Standards, otherwise erroneous results
will be obtained.
For accurate validation and calibration results, please
perform tests at room temperature (18 to 25°C;
64.5 to 77.0°F).
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays “747”, it is ready.
3• Enter the calibration mode by holding CAL
CHECK key for three seconds.
4• The date of the last calibration appears (e.g.:
month “01”, day “08”). “F.CAL” means that
the factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION
5• Place the CAL CHECK™ Standard Cuvet A
into the holder and ensure that the notch on
the cap is positioned securely into the groove.
6• Press ZERO and “SIP” will blink on the
display.
7• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready
for calibration.
8• Remove the cuvet.
9• Place the CAL CHECK™ Standard
HI 95747-11 Cuvet B into the holder and
ensure that the notch on the cap is positioned
securely into the groove.
10•Press READ/TIMER and “SIP” will blink on the
display.
11•The instrument will show for three seconds the
CAL CHECK™ standard value.
Then the date of last calibration (e.g.:
“01.08”) appears on the display, or “01.01”
if the factory calibration was selected. In both
cases the number of the month is blinking,
ready for date input.
Note: if display will show “ERR” the calibration
procedure failed. Verify that the right CAL
CHECK™ Standard Cuvet B is inserted, that
both A and B cuvets are free from fingerprints
or dirt and that they are inserted correctly.
Calibration
12•Keep READ/TIMER pressed to scroll to the
desired month number (01-12).
13•When the correct month has been set, press
ZERO to confirm. Now the display will show
the day number blinking.
Keep READ/TIMER pressed to scroll to the
desired day number (01-31).
Note: it is possible to change from
day
to
month
and vice versa by pressing ZERO.
14•When both the day and month have been
selected, hold CAL CHECK key for three seconds
to store date and calibration values. The
instrument will show for one second “Stor”, to
confirm that the new calibration data has
been accepted.
15•The instrument will return automatically to
the measurement mode by displaying the
parameter code (“747”) on the LCD.
Factory calibration reset
It is possible to restore factory calibration:
1• Enter the
2• Hold ZERO for 10 seconds. The display will
calibration mode
CHECK key for three seconds.
show for 2 seconds “F.CAL” and the parameter
code “747” appears. The factory calibration
is automatically restored and the instrument
is ready for measurement.
by holding CAL
Factory
Calibration
Reset
Battery replacement
Battery replacement must only take place in a non-hazardous environment.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh 9V battery
while paying attention to the correct polarity. Insert the battery and
replace the cover.
Accessories:
REAGENT SETS
HI 95747-01 Reagents for 100 Copper LR tests
HI 95747-03 Reagents for 300 Copper LR tests
OTHER ACCESSORIES
HI 95747-11 CAL CHECK™ kit
Low Range (1 set)
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731331 Glass cuvets (4 pcs)
HI 731335 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
for Calibration & Validation of Copper
Warranty
HI 95747 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the
design, construction and appearance of its products
without advance notice.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find the Hanna Office in your area,
IST95747R1 12/03
www.hannainst.com
visit our web site