Hanna Instruments HI 943500A User Manual

Instruction Manual
HI 8931 • HI 8936 Series
HI943500 Series
Conductivity Process
Controllers and Transmitters
www.hannainst.com
Dear Customer, Thank you for choosing a HANNA instruments® product. Please read this instruction manual carefully before using the instru­ment. This manual will provide you with all the necessary information for correct use of the instruments, as well as a precise idea of thier versatility in a wide range of applications. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com These instruments are in compliance with the directives.
TABLE OF CONTENTS
Preliminary Examination ............................................................ 4
General Description .................................................................... 4
Functional Description HI 8931 & HI 943500 ............................ 6
Functional Description HI 8936 ................................................ 10
Conductivity Probes .................................................................. 13
Specifications HI 8931 & HI 943500 ....................................... 15
Specifications HI 8936 ............................................................. 16
Connections .............................................................................. 17
Operational Guide ................................................................... 25
Calibration Procedure of HI 8931 & HI 8936 with HI 7635 .... 31
Calibration Procedure of HI 8931 & HI 8936 with HI 7638 .... 35
Calibration Procedure of HI 943500 with HI 7638................... 39
Conductivity Versus Temperature Chart ..................................... 41
Diagnostic Tests ........................................................................ 42
Installation Examples ............................................................... 44
Probe Maintenance and Cleaning ............................................ 46
Accessories ................................................................................ 47
Warranty................................................................................. 49
CE Declaration of Conformity ..................................................... 50
32
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your dealer.
Each model is supplied complete with:
• Mounting brackets (not for HI 8936 series)
• Transparent splsh-proof cover (not for HI 8936 series)
• Instruction manual
Note: Save all packing materials until you are sure that the
instrument functions correctly. All defective items must be returned in the original packing materials together with the supplied accessories.
GENERAL DESCRIPTION
The HI 8931 and HI 943500 are panel-mounted conductivity control­lers designed for simplicity of use in a wide range of industrial process applications.
The instruments are designed with a standard DIN panel mount with membrane keypads and large LCD on the front, and provide a series of auto-diagnostic functions.
Four models with different measurement ranges are available to suit any application needs:
HI8931A / HI 943500A / HI8936A / HI 8936AL
from 0.0 to 199.9 mS/cm
HI8931B / HI 943500B / HI 8936B / HI 8936BL
from 0.00 to 19.99 mS/cm
HI 8931C / HI 943500C / HI8936C / HI8936CL
from 0 to 1999 μS/cm
HI 8931D / HI 943500D / HI8936D / HI8936DL
from 0.0 to 199.9 μS/cm
Other features include: recorder output in 0-20 mA or 4-20 mA configuration; LED indicators (for HI 8931 and HI 943500) which identify whether the controller is in operation mode or selection mode.
Each instrument of the HI 8931 and HI 943500 series, is supplied with a plastic front cover and two mounting brackets. Power cables are not included.
Probes, power supply, contacts and recorders are connected on the rear panel through screw terminals.
Using HI 8931 in conjunction with a 4-20 mA output transmitter (HI 8936 or HI 8936L series) will assure you of a strong, interference­free signal at distances up to 300 meters (1000').
For in-line applications use the HI 7635 probe, while for tanks the HI 7638 with external threads is recommended. These probes are provided with built-in NTC sensor for temperature compensated conductivity measurements. The probe cable length is 3 meters (10').
HI 943500 features a direct connection up to 20 m (67'), without intermediate amplifiers, to the conductivity probe HI 7638 with DIN connector and automatic temperature compensation.
54
FUNCTIONAL DESCRIPTION
HI 8931 & HI943500
KEYPAD MEASURE To read measurements and enable diagnostic tests ΔAL To display set tolerance of the alarm SET To set working point TEST SLOPE Diagnostic function TEST OFFSET Diagnostic function
When a key is pressed, the corresponding LED lights up to indicate that the function is active.
TRIMMERS OFFSET For offset calibration
SLOPE For slope calibration ΔAL To set the alarm tolerance COARSE To coarsely adjust the set point SET FINE To finely adjust the set point
REAR PANEL OF HI 8931 SERIES
1. POWER OUTPUT terminals (+20 V and COM) for connec-
tion to a conductivity transmitter (HI 8936)
2. mA INPUT from a conductivity transmitter
3. mA OUTPUT terminals for connection to a recorder
4. SET SELECT terminals for reverse control operation
5. SET terminals for connection to a dosing pump
6. ALARM terminals for connection to an external alarm device
7. Power supply terminals
8. Fuse holder
LEDS SET ON To indicate that the dosage is active
ALARM To indicate an alarm condition
Unplug the instrument from power supply before replacing the fuse.
76
190mm MIN
7.50"
135mm
5.31"
0.25/4mm
0.01/0.160"
144mm
5.67"
ADJUSTABLE LOCATION BRACKET
m "
2.83"
REAR PANEL OF HI 943500 SERIES
MECHANICAL DIMENSIONS OF HI 8931 AND HI 943500
Front view of the panel-mounted unit
1. DIN connector for conductivity probe
2. mA OUTPUT terminals for connection to a recorder
3. SET SELECT terminals for reverse control operation
4. SET terminals for connection to a dosing pump
5. ALARM terminals for connection to an external alarm device
6. Power supply terminals
7. Fuse holder
69mm
2.71"
72mm
141mm
5.55"
144m
5.67
These dimensions show the cutout size for the installation.
Side view of the panel-mounted unit
Unplug the instrument from power supply before replacing the fuse.
Adjustable location brackets (supplied with the meter) allow the indicator to slide into the cutout and will hold the unit securely in place. 190 mm (7.50") is the minimum amount of room required to install the indicator with the cables connected.
98
FUNCTIONAL DESCRIPTION HI 8936
HI 8936A HI 8936B HI 8936C HI 8936D
SIDE VIEW OF HI 8936 SERIES
1. Back cover
2. Top cover
3. Liquid Crystal Display
4. Screws for fastening the top cover
HI 8936AL HI 8936BL HI 8936CL HI 8936DL
1. Top cover
2. Cable glands for wiring
1110
T
142mm
MECHANICAL DIMENSIONS OF
m "
HI 8614
165mm
m "
HI 8614L
3.54"
HI 8936A, HI8936B, HI 8936C, HI8936D
6.50"
100mm
3.94"
4 x 4.3mm
90mm
3.54"
4 x 0.17"
155mm
6.10"
MECHANICAL DIMENSIONS OF
HI 8936AL, HI8936BL, HI8936CL, HI 8936DL
110m
CONDUCTIVITY PROBES
HI 7635 In-line Conductivity Probe
HI 7635 is a one piece, molded conductivity probe with pipe threads
(1" NPT) at both ends.
4.33
This allows the probe to attach to an in-line system, and to be used in conjunction with the HI 8936 conductivity transmitter.
The HI 7635 uses a 4-ring potentiometric measuring method. This method is highly accurate and requires very little maintenance.
The construction of the housing is rugged, fiber-reinforced polypropy­lene.
The maximum working pressure of this unit is 5 bar (72.5 psi). Do not use in systems where the temperature exceeds 80°C (176°F).
90mm
24mm
0.94"
45mm
1.77"
4 x 4.3mm
4 x 0.17"
58mm
2.28"
165mm
6.50"
155mm
6.10"
126mm
5.00"
100mm
3.94"
110m
HI7635
4.33
2.87"
52mm
2.04"
1" NP
5.60"
73mm
1312
HI7638 Tank Conductivity Probe
HI7638 conductivity probe combines the proven 4-ring potentiometric
method of measuring conductivity with the platinum sensor and stainless steel external thread.
This method incorporates a series of four platinum rings into the probe shaft and is highly accurate requiring very little maintenance.
The removable plastic cover resists the harmful effect of most chemicals and can be unscrewed for quick and simple maintenance.
This probe can withstand temperatures of up to 120°C (248°F) and pressure of up to 5 bar (72.5 psi).
This probe is supplied complete with a 7-pin DIN connector.
HI 7638
SPECIFICATIONS HI 8931 & HI 943500
Range HI 8931A - HI 943500A HI 8931B - HI 943500B HI 8931C - HI 943500C HI 8931D - HI 943500D Accuracy (@20°C/68°F) Typical EMC
Deviation 4-20mA INPUT from Transmitter HI 8931A HI 8931B HI 8931C HI 8931D
HI 943500 Conductivity Probe
Calibration
Temp. Compensation HI 8931
HI 943500 Recorder Output
Set Point Relay and Alarm Relay
Power Supply
Environment Panel Cutout Weight Enclosure
0.0 to 199.9 mS/cm
0.00 to 19.99 mS/cm 0 to 1999 μS/cm
0.0 to 199.9 μS/cm ±2% of Full Scale
excluding probe error ±2.5% of Full Scale
±0.4 mA
HI 8936 A or AL (not included) HI 8936 B or BL (not included)
HI 8936 C or CL (not included)
HI 8936 D or DL (not included)
------
HI 7635 for in-line applications or
HI 7638 for tank (not included)
Manual, 2 point, through
offset and slope trimmers
See transmitter HI 8936
Automatic, 0 to 60°C with ß=2%
4 to 20 mA (isolated)
Isolated, 2A, Max. 240V, resistive
load, 1,000,000 strokes
115 or 230 Vac ±10%
(user selectable); 50/60 Hz
-10 to 50°C (14 to 122°F); RH 95% 141 x 69 mm (5.6 x 2.7")
1 kg (2.2 lb.)
DIN 43 700, 144x72 mm (5.7x2.8"),
black anodized aluminum;
front and back with shockproof ABS
plastic, and transparent cover
1514
SPECIFICATIONS HI 8936
Range HI 8936 A/AL HI 8936 B/BL HI 8936 C/CL
HI 8936 D/DL Accuracy (@20°C/68°F)
Typical EMC Deviation Conductivity Probe
Calibration
Temperature Compensation
Output
Power Supply HI 8936 A/B/C/D HI 8936 AL/BL/CL/DL
Protection Environment
Dimensions
Weight
0.0 to 199.9 mS/cm
0.00 to 19.99 mS/cm 0 to 1999 μS/cm
0.0 to 199.9 μS/cm ±2% of Full Scale
excluding probe error
±2% of Full Scale
±0.4 mA
HI 7635 for in-line applications or
HI 7638 for tank (not included)
Manual, 2 point, through
offset and slope trimmers
Fixed or automatic
from 0 to 50°C (32 to 122°F)
with ß=2%
4 to 20 mA not-isolated
max. 500 Ohm
12 to 30 Vdc 17 to 36 Vdc
IP 65
0 to 50°C (32 to 122°F);
RH max 95%
165 x 110 x 90 mm
(6.5 x 4.3 x 3.5")
1 kg (2.2 lb.)
CONNECTIONS
REAR CONNECTIONS FOR HI 8931
Power Connection Terminals
4-screw-terminal-strip for connection to a 3-wire power cable according to the indicated voltage (115 or 230V).
IN/OUT Transmitter
2 wires of the 4-core signal cable from the conductivity transmitter (HI8936) have to be connected to the mA input terminals and the other 2 wires to the "+20 V" and "COM" while paying careful atten­tion to the polarity.
+20 V supply is the regulated DC supply required for the operation of the external conductivity transmitter HI 8936.
Set Contacts
Dosing pumps or other control equip­ment may be connected to the "SET" (Max. 2A, 240 V) terminals. These contacts act only as a "dry" switch allowing electrical continuity, not as a power supply.
1716
TERMINAL BOARD CONNECTIONS FOR H8936
Set Select
These contacts permit the activation of a Set Contact relay when the measured value is lower (connected terminals) or higher (open termi­nals) than the user's set value. See also page 28.
+4-20 mA
These are the output terminals for connection to a recorder or other control equipment.
The output current varies from 4 to 20 mA and is proportional to the measured conductivity value.
Alarm Contacts
During normal operation these termi­nals remain closed.
If the measured conductivity level is not within the tolerance of the set value, the alarm contact is open. These contacts act only as a switch. See also page 29.
• Remove the 4 screws and the top cover of the HI 8936 conductivity
transmitter to obtain access to the terminal board connections.
HI 8936 used in conjunction with HI 8931 controller
Use a PVC insulated 4-core cable to connect the transmitter to the HI 8931 conductivity controller (see also page 25).
1918
The 4-core cable has to be con­nected to the transmitter accord­ing to the label instructions on the 4-terminal strip.
The regulated D.C. supply required for the proper operation of the transmitter is "+20 V", labeled "+20 V" and "COM". The current (mA) output terminals are labeled "4-20 mA" and "COM".
The transmitter is protected against inversion of supply voltage.
• HI 8936 used in conjunction with an external power supply
Use 2 PVC insulated 2-core cables.
Connect a +20 Vdc power supply directly to the terminals labeled "+20 V" and "COM", paying care­ful attention to polarity (see also page 24) or if necessary in series with the receiving device.
The regulated DC supply required for the proper functioning of the circuit is marked "+20 V" and "COM", and the transmitter current output is indicated "4-20 mA" and "COM".
• Probe Connection
The conductivity probe is supplied with a 3 m (10'), 6 core cable. The cable is to be connected to the terminals provided (see also page 24 for proper connection scheme of HI 7635 and HI7638).
The transmitter is protected against inversion of supply voltage.
Max. current required: 40 mA.
• Use of an amplifier
The maximum permissible distance between the power supply unit and the amplifier is 300 m (1000'). It is not necessary to use a shielded cable.
2120
REAR CONNECTIONS FOR HI 943500
• Set
• Power Connection Terminals
4-screw-terminal-strip for connec­tion to a 3-wire power cable ac­cording to the indicated voltage (115 or 230 V).
• DIN connector socket
For connection to the HI 7638 conductivity probe.
• + mA output -
The first and the second terminals are the output terminals for connection to a recorder or other control equipment. The output cur­rent varies from 4 to 20 mA and is proportional to the measured conductivity value.
Dosing pumps or other control equipment may be connected to the "SET" (Max. 2A, 240 V) terminals. These contacts act only as a "dry" switch allowing electrical continuity, not as a power supply.
• Alarm Contacts
During normal operation these ter­minals remain closed.
If the measured conductivity level is not within the tolerance of the set value, the alarm contact is open. These contacts act only as a switch. See also page 29.
• Set Select
These contacts permit the activation of the Set Contact relay when the measured value is lower (connected terminals) or higher (open terminals) than the user's set value. See also page 28.
2322
CONDUCTIVITY PROBE CONNECTIONS
The connections for HI 7635 are color coded for easy in­stallation and are as fol­lows:
HI 7635 cable HI 8936 transmitter
Black or Grey NTC
Red or Pink SENSOR
Brown or Orange probe pin 1
Blue probe pin 2
White probe pin 3
Green or Yellow probe pin 4
OPERATIONAL GUIDE
INITIAL PREPARATION & INSTALLATION
Material needed:
• a 3-wire power cable (to connec the HI 8931 or HI 943500 to
the mains)
• a PVC insulated 4-core cable (to connect the HI 8931 to HI8936)
• rubber seals and a pipe sealant (for installation of HI 7635)
FOR HI 8931 AND HI 8936 ONLY
The connections for HI 7638 are as follows:
HI 7638 HI8936 transmitter
#1 probe pin 1 #2 probe pin 2 #3 probe pin 3 #4 probe pin 4 #5 NTC #6 SENSOR
Note: NTC & SENSOR are equivalent, and are both labeled
"SENSOR" on the HI 8936 conductivity transmitter.
connector
inside view
• Remove the 4 screws and the
top of the HI 8936 conductivity transmitter.
• Connect the 2 wires connected to the "4-20 mA" and "COM" ter­minals of the 4-core signal cable from the HI 8936 to the termi­nals marked "mA input" paying careful attention to polarity. Connect the other 2 wires to the "+20 V" and "COM" terminals, while paying attention to the polarity.
• Connect the 3-wire power cable to the 4-screw terminal strip ac­cording to the voltage level as indicated, and pay particular attention to the correct live, earth and neutral terminal connections.
2524
• The HI 8936 transmitter may be
wall-mounted at any convenient loca­tion close to the measurement site. To minimize thermal drifts due to extreme temperature fluctuations, par­ticularly for outdoors measurements, it is recommended to protect the trans­mitter inside a casing.
• For the installation of the HI 7635 conductivity probe, use rubber
seals between the probe and the pipe or elbow joints. A pipe sealant is also recommended to ensure a leak free joint. When screwing the joints, take care not to overtighten as excessive pressures can damage the probe.
• It is recommended to install the HI 7635 vertically. This is to ensure that trapped air bubbles or turbulent flows cause minimal interference to the measurement system. The maximum working pressure of this unit is 5 bar (72.5 psi).
WARNING: Do not use when temperature exceeds 80°C (176°F).
• The HI 7638 process conductivity probe is also supplied with a 3 m (10') cable. The 6-core cable of this probe is connected to the HI 8936 process conductivity transmitter as shown.
The DIN connector has to be removed when this probe is used in conjunction with an HI8936 transmitter. See page 24 for the proper connection scheme.
FOR HI 943500 ONLY
Ensure that the HI 7638 con­ductivity probe is connected to the meter securely by align­ing the pins with the socket, pushing the plug in and tightening the threaded ring.
• The HI 7635 conductivity probe is
supplied with a 3 m (10') cable. The 6-core cable of the probe is connected to the HI 8936 transmitter as shown. The connections are color coded for an easy installation. See page 24 for the proper connec­tion scheme.
2726
OPERATING INFORMATION
All parameters are set through the front panel keys and trimmers. When any key is pressed, the corresponding LED lights up to indicate
that the function is active. Make sure that the conductivity meter, transmitter and probe are
calibrated before taking measurements (see pages 31, 35 and 39 for calibration procedures).
SET POINT
To set the working point of the controller, press the SET key. The display will indicate the current set value.
Use a small screwdriver to adjust the COARSE and FINE trimmers until the desired value is displayed.
Above Setpoint Control Operation
Leave the "SET SELECT" and "COM" connec­tors open. The set contacts relay will close if the measured value is higher than the setpoint value, and the "SET ON" LED will light up.
COM
SET
SELECT
Below Setpoint Control Operation
Short the "SET SELECT" and "COM" connec­tors with a jumper wire. The set contact relay will close if the measured value is lower than the setpoint value, and the "SET ON" LED will light up.
COM
SET
SELECT
ALARM
Press the "ΔAL" key and the display will show the set tolerance for the alarm.
Use a small screwdriver to adjust the "ΔAL" trimmer un­til the desired tolerance is dis­played.
For example, if the set value is 200 μS/cm and a ΔAlarm of 50 μS/cm is set, an alarm will be activated every time the measured value is higher than 250 μS/cm or lower than 150 μS/cm.
When an alarm occurs, the "ALARM" LED lights up.
2928
The alarm contacts of HI 8931 and HI 943500 remain closed during normal operation. If the measured conductivity level is not within the tolerance of the set value, the alarm contact will be open.
TAKING MEASUREMENTS WITH HI8931 AND HI 943500
After setting the working point and alarm value, press the "MEASURE" key. The actual conductivity value of the test solution will be displayed.
TAKING MEASUREMENTS WITH HI8936AL, HI 8936BL, HI 8936CL, HI8936DL
CALIBRATION PROCEDURE OF
HI 8931 & HI 8936 WITH HI 7635
Material needed:
HI 7635 conductivity probe
HI 8931 conductivity controller
HI 8936 conductivity transmitter
• A 20 mA f.s. ammeter (for transmitters without LCD)
• A reference conductivity meter with automatic temperature com­pensation accurately calibrated (e.g. HI 8733).
PROCEDURE
• Connect the HI 7635 probe to the H I8936 transmitter (see page 24 for details).
• Connect the HI 8936 transmitter to the H I8931 controller (see page 25 for details).
• Connect the HI 8931 controller to the mains (see page 25 for details).
Before proceeding with the calibration, make sure that the meter is in measure­ment mode ("MEASURE" LED is on) and not in set mode.
MEA
SURE
The conductivity transmitters with LCD will always display the measurement value when connected to the HI8931 controllers or a power supply.
• Connect the ammeter to the HI 8936 transmitter to monitor the
signal current (see below picture).
3130
• Ensure that the HI 7635 conductivity probe is dry.
• When the power is on, the ammeter should read "4.0 mA".
The HI 8936 transmitter with LCD should display "0".
• If not, adjust the transmitter OFF­SET trimmer to read "4 mA" or "0" on the HI 8936 LCD.
• The HI 8931 controller should dis­play "0" value.
• If not, adjust the HI 8931 OFF­SET trimmer to display a zero reading.
• The reading will be converted to mA by the following formula: mA = K (measured value x 16/2000) + 4 K = conversion factor depending on the model
Model Conversion factor K HI 8936 A/AL 10
HI 8936 B/BL 100 HI 8936 C/CL 1 HI 8936 D/DL 10
For example, using an HI 8936A, if the measured value is 123.4 mS, then output current = 10 x (123.4 x 16/2000)+4 = 13.9 mA
Adjust the HI 8936 SLOPE trim­mer to read "13.9 mA" on the ammeter, or the reading of the HI 8733 (e.g. 123.4 mS) on the HI 8936L display. .
• Switch the flow on and allow the sample solution to flow through the HI 7635 conductivity probe. Collect a sample of this solution in a beaker.
• Measure the conductivity of the solution with a reference conduc­tivity meter with ATC (HI 8733). The value obtained will be used for the transmitter and controller calibration.
• Adjust the HI 8931 SLOPE trimmer until the controller reading is the same as the one displayed by the HI 8733 (e.g. 123.4 mS).
3332
• The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F).
• If the instrument will not calibrate, refer to the "Probe Mainte­nance and Cleaning" section on page 46.
• If the HI 8936 is not used in conjunction with the HI 8931 controller, connect the transmitter to an external power supply (see page 20), to the HI 7635 conductivity probe (see page 24) and to an ammeter (see page 20).
The calibration of the transmitters with LCD (HI 8936L series) does not require an ammeter.
Follow the above calibration procedure performing the operations referred to the HI 8936 transmitter only.
CALIBRATION PROCEDURE OF
HI 8931 & HI 8936 WITH HI 7638
Material needed:
HI 7638 conductivity probe
HI 8931 conductivity controller
HI 8936 conductivity transmitter
• An ammeter (for transmitters without LCD)
• Calibration solutions, according to the different models:
HI 7034 80 mS/cm @25°C
for HI 8931A, HI 8936 A/AL
HI 7030 12.88 mS/cm @25°C
for HI 8931B, HI 8936 B/BL
HI 7031 1413 μS/cm @25°C
for HI 8931C, HI 8936 C/CL
HI 7033 84 μS/cm @25°C
for HI 8931D, HI 8936 D/DL
PROCEDURE
• Connect the HI 7638 to the HI 8936 transmitter (see page 24).
• Connect the HI 8936 transmitter to the HI 8931 controller (see
page 25).
• Connect the HI 8931 controller to the mains (see page 25).
Before proceeding with the calibration, make sure that the meter is in measure­ment mode ("MEASURE" LED is on) and not in set mode.
• Connect the ammeter to the HI 8936 transmitter to monitor the
signal current (see below picture).
MEA
SURE
3534
• Leave the HI 7638 conductivity probe in air (dry probe).
• Pour at least 8 cm (3¼") of conductiv­ity solution into a plastic beaker.
(3¼")
HI 7030
8 cm
• When the power is on, the ammeter should read "4.0 mA" or the
HI 8936 transmitter with LCD should display "0".
• If not, adjust the HI 8936 OFF-
SET trimmer to obtain "4 mA" or "0" on the HI 8936L.
• The HI 8931 controller should dis-
play "0".
• Immerse the probe into the conductivity solution. The holes on the sleeve must be completely submerged in the solu­tion.
• Tap the probe repeatedly on the bottom of the beaker and stir it to ensure that no air bubbles remain trapped inside the sleeve.
• If the temperature of the probe is close to that of the solution, the display of HI 8931 will stabilize quickly and provide a temperature compensated conductivity measurement. Allow a few minutes if there is a temperature difference of about 5°C (9°F) or more for the ATC circuitry to compensate completely.
For HI 8936 Adjust the SLOPE trimmer of the HI 8936 transmitter to read on
the ammeter: "10.4 mA" using HI 7034 and HI 8936 A/AL "14.304 mA" using HI 7030 and HI 8936 B/BL "15.304 mA" using HI 7031 and HI 8936 C/CL "10.72 mA" using HI 7033 and HI 8936 D/DL
• If not, adjust the HI 8931 OFF-
SET trimmer to display a zero reading.
3736
FOR HI8931
When the reading stabilizes, turn the SLOPE trimmer on the front of the HI 8931 until the LCD reading is the same as the calibration solution at 25°C (77°F), i.e.
"80.0 mS" using HI 7034 with HI 8931A "12.88 mS" using HI 7030 with HI 8931B "1413 μS" using HI 7031 with HI 8931C "84.0 μS" using HI 7033 with HI 8931D
• The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F).
• If the instrument will not calibrate, refer to the "Probe Mainte­nance and Cleaning" section on page 46.
• If the HI 8936 is not used in conjunction with the HI 8931 controller, connect the transmitter to an external power supply (see page 20), to the HI 7638 conductivity probe (see page 24) and to an ammeter (see page 20).
The calibration of the transmitters with LCD (HI 8936L series) does not require any ammeter.
Follow the above calibration procedure performing the operations referred to the HI 8936 transmitter only.
CALIBRATION PROCEDURE OF
HI 943500 WITH HI 7638
Material needed
HI 7638 conductivity probe
HI 943500 conductivity controller
• Calibration solutions, according to the different models:
HI 7034 80mS/cm @25°C for HI 943500A HI 7030 12.88mS/cm@25°C for HI 943500B HI 7031 1413 μS/cm @25°C for HI 943500C HI 7033 84 μS/cm @25°C for HI 943500D
PROCEDURE
• Ensure that the probe is con­nected to the meter securely by aligning the pins with the socket, pushing the plug in and tightening the threaded ring.
• Insert the probe into the sleeve.
• With the conductivity probe in air, press the "MEASURE" key to set the meter to measure­ment mode.
3938
• If the LCD does not show "0", adjust the OFFSET trimmer.
HI 7030
• Pour at least 8 cm (3¼") of conductiv­ity solution into a plastic beaker.
(3¼")
8 cm
• Immerse the probe into the conductivity solution, while paying attention that the holes on the sleeve are completely submerged.
• Tap the probe repeatedly on the bot­tom of the beaker and stir it to ensure that no air bubbles remain trapped inside the sleeve.
• If the temperature of the probe is close to that of the solution, the display will stabilize quickly and provide a temperature compen­sated conductivity measurement.
Allow a few minutes if there is a temperature difference of about 5°C (9°F) or more for the ATC circuitry to compensate completely.
• When the reading stabilizes, turn the SLOPE trimmer on the
HI943500 front panel until the LCD reads the same value of the
calibration solution at 25°C (77°F), i.e.:
80.0 mS using HI 7034 with HI 943500A
12.88 mS using HI 7030 with HI 943500B 1413 μS using HI 7031 with HI 943500C
84.0 μS using HI 7033 with HI 943500D
• The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F).
• If the instrument will not calibrate refer to the "Probe Maintenance and Cleaning" section on page 46.
CONDUCTIVITY VERSUS
TEMPERATURE CHART
°C °F HI7030 HI7031 HI 7033 HI7034 HI7035 HI7039
(mS/cm) (mS/cm) (mS/cm) (mS/cm) (mS/cm) (mS/cm)
0 32 7150 776 64 48300 65400 2760
5 41 8220 896 65 53500 74100 3180
10 50 9330 1020 67 59600 83200 3615
15 59 10480 1147 68 65400 92500 4063
16 60.8 10720 1173 70 67200 94400 4 155
17 62.6 10950 1199 71 68500 96300 4 245
18 64.4 11190 1225 73 69800 98200 4 337
19 66.2 11430 1251 74 71300 100200 4429
20 68 11670 1278 76 72400 102100 4523
21 69.8 11910 1305 78 74000 104000 4517
22 71.6 12150 1332 79 75200 105900 4711
23 73.4 12390 1359 81 76500 107900 4805
24 75.2 12640 1386 82 78300 109800 4902
25 77 12880 1413 84 80000 111800 5000
26 78.8 13130 1440 86 81300 113800 5096
27 80.6 13370 1467 87 83000 115700 5190
28 82.4 13620 1494 89 84900 117700 5286
29 84.2 13870 1521 90 86300 119700 5383
30 86 14120 1548 92 88200 121800 5479
31 87.8 14370 1575 94 90000 123900 5575
If you are calibrating HI 8931A or HI 943500A using HI 7030 (12.88 mS/cm @25°C) standard solution, turn the SLOPE trimmer to display "12.88 mS", with a reference temperature of 25°C (77°F), while for a reference temperature of 20°C (68°F), the SLOPE trimmer should be adjusted to display "11.67 mS".
4140
DIAGNOSTIC TESTS
The HI 8931 and HI 943500 controllers are designed with built-in diagnostic functions to enable the user to check and troubleshoot the instrument. The checks performed are through the front panel keys and can be used to isolate the cause of malfunction.
Press the "MEASURE" key before pro­ceeding the following tests.
A) Test Offset
Press the "TEST OFFSET" key and the display should indicate the following values:
HI8931/HI 943500A 0.0 mS ±1.0 mS HI8931/HI 943500B 0.00 mS ±0.10 mS HI8931/HI 943500C 000 μS ±100 μS HI 8931/HI943500D 0.0 μS ±1.0 μS
B) Test Slope
Press the "TEST SLOPE" key and the display should indicate the following values:
HI8931/HI 943500A 100.0 mS ±35.0 mS HI 8931/HI943500B 10.00 mS ±3.50 mS HI 8931/HI943500C 1000 μS ±350 μS HI 8931/HI943500D 100.0 μS ±35.0 μS
Note: The reading obtained by these functions will vary if the OFFSET
and SLOPE trimmers on the front panel are adjusted.
4342
INSTALLATION EXAMPLES
Some typical installation setups are shown in the following examples:
Example #1
Example #2
Example #3
Example #4
Example #5
4544
PROBE MAINTENANCE & CLEANING
ACCESSORIES
The probe can be compensated for normal contamination by a process of re-calibration. It is recommended to remove the probe from its installation regularly for maintenance.
For HI7635 only: Deposits on the conductivity probe can be removed by immersing the probe in 0.1 N hydrochloric acid for about 30 minutes. Heavier deposits may demand longer immersion periods. Rinse the probe thoroughly with water prior to the reinstallation. On reinstalling, check the seals carefully to ensure that a leak-free connection is obtained.
For HI7638 only: Rinse the probe with tap water. If a more thorough cleaning is desired, unscrew the plastic sleeve and clean the sensors with a nonabrasive cloth or alcohol. After cleaning the probe, re-calibrate the instrument. If the instrument will not calibrate with a clean probe, you must replace the probe.
Note: Always re-calibrate the meter when attaching a new probe.
CONDUCTIVITY CALIBRATION SOLUTIONS
HI 7030L 12880 μS/cm calibration solution, 500 mL HI 7030M 12880 μS/cm calibration solution, 230 mL HI 7031L 1413 μS/cm calibration solution, 500 mL HI 7031M 1413 μS/cm calibration solution, 230 mL HI 7033L 84 μS/cm calibration solution, 500 mL HI 7033M 84 μS/cm calibration solution, 230 mL HI 7034L 80000 μS/cm calibration solution, 500 mL HI 7034M 80000 μS/cm calibration solution, 230 mL HI 7035L 111800 μS/cm calibration solution, 500mL HI 7035M 111800 μS/cm calibration solution, 230 mL HI 7039L 5000 μS/cm calibration solution, 500 mL HI 7039M 5000 μS/cm calibration solution, 230 mL
OTHER ACCESSORIES
HI 731326 Calibration screwdrivers (20 pcs) HI 779/P 6-wire cable (100 m/330' roll) HI 8733 Portable conductivity meter with automatic tem-
perature compensation
4746
WARRANTY
T
142mm
CONDUCTIVITY PROBES
HI 7635 In-line conductivity probe with 3 m (10') cable
(for HI 8936 only)
2.87"
52mm
2.04"
1" NP
5.60"
73mm
HI 7638 Conductivity probe for tank installation, with 3 m
(10') cable (for HI 8936 or HI 943500)
All HANNA instruments® meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
The probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred.
If the instrument is to be returned to HANNA instruments®, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without
advance notice.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interfer­ences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. The trimmers are sensitive to electrostatic discharges. It is recommended to use anti­static screwdrivers. Unplug the instrument from the power supply before replacing the fuse. External cables to be connected to the rear panel should end with cable lugs. To maintain the EMC performance of this equipment, use the recommended cables mentioned on this instruction manual. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc. To avoid damages or burns, do not perform any measurement in microwave ovens.
SALES & TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MANCDPRCR5 09/05
Loading...